S 61 Servant ENT workstation and are therefore a must besides regular cleaning. ® Repair work and period tests may only be carried out by expert personnel authorised by ATMOS. By applying only original spare parts you will have the guarantee that operational safety, readiness for work and the value of your ATMOS S 61 Servant ENT workstation will be preserved.
Single-use product No single use product / reprocessing: 1.3 Function The treatment unit ATMOS S 61 Servant ENT workstation makes modular assembly possible for the following functional ® modules: • Suction system • Compressed-air system • Ear irrigation module (Hygrotherm) •...
Replace Check Please read, Engage, check click important information correct fit Icons for the ATMOS S 61 Servant ENT workstation ® Warning, special diligent notice Mirror quick heater Ground wire connection Fuse according to IEC 417/5016, Adjustment Alternating current DIN 30600/0186 Compressed air...
• The control panel must be clearly visible and accessible Compatibility“ (Medical Electrical Equipment). for the user. • ATMOS is not liable for personal injury and damage to • This product is not re-sterilizable. Repeated reuse of property if components which are marked with a is forbidden.
2.2 Hygrotherm / Variotherm: Important advice for maintaining the hygienic status For loosening cerumen in the auditory meatus and for the stimulation of the labyrinthine system, ATMOS offers the warm water units Hygrotherm plus (37°C) and Variotherm plus (20° - 47°C).
Connecting hoses are not included in delivery. 3.1 Required connections for all units of ATMOS S 61 Servant ® Flexible multi-port distributors may not be used as a power supply for the ATMOS S 61 Servant workstation, ATMOS S 61 ®...
• The connection may be closed without any additional effort. • Connection including pipe aeration is recommended. • Minimum height of the water tap: 35 cm. • Distance of the water tap to the ATMOS S 61 Servant Workstation: < 3 m. ®...
3.0 Setting up and starting up 3.4 Waste water drain Prerequisites: • Permanently installed connection fitting with G3/4" external thread or • Drain outlet, standard HT 40 or HT 50 (connection adapters can be ordered for the adoption of the discharge hose: 510.2130.0 for HT 40; 510.2129 for HT 50) or •...
3.0 Setting up and starting up 3.6 Controls and options when fully equipped Mirror quick heater Heating module for instrument tray Nebulization of medications (compressed air) LED light source max. 2 kg! LED headlight Clipboard Automatic suction equipment Compressed air Ear irrigation / thermal nystagmus...
3.0 Setting up and starting up 3.7 Rear view Plug the power cable into the IEC power socket. Please observe the safety instructions on page 7! 3.8 Assembly of secretion canister Prior to use, check the system for leak tightness, otherwise it is possible that the vacuum cannot built up! Insert splash guard into Insert bacterial filter ( Insert suction hose (...
First start-up Prior to shipment each ATMOS S 61 Servant was checked by the manufacturer for function and safety. In order to ensure the ® safe functioning of the device after transport and installation, the following points should be observed: The user should only operate the device when 1.
4.0 Operation 4.1 On / off switch After switching on the mains switch, all integrated and connected devices are ready for operation. 4.2 Suction system Do not operate the suction system without a bacterial filter. Always use the splash guard, this will prolong the service life of the bacterial filter. The guarantee will expire in the case of improper use! Change the bacterial filter regularly. See details in chapter 5.0. The hose attachment must be exchanged after each patient. 4.2.1 Control and display of the suction performance Display vacuum Adjustment vacuum Remove the handle from the support to activate the suction system.
4.0 Operation Refilling storage canister for hose rinsing A) Fully automatic refilling (530.1060.0) If required, the water storage canister will be automatically refilled. Attention: Make sure that the water supply system is on and the water supply secured! Do not use any foaming cleaning agents or disinfectants. B) Manual refilling (530.1050.0) Fill the canister with water and/or a non-foaming instrument disinfectant.
4.0 Operation 4.3 Ear irrigation module Hygrotherm 4.3.1 Overview Front view Temperature display (double-digit, resolution 1°C), actual value indication Setting mode for the temperature Adjustment button up or down Irrigation handle Handle Connection for double-barrelled hose Three-hole gasket Knurled screw Release lever Valve plunger Jet connection...
4.0 Operation 4.3.3 Activation of the Hygrotherm Remove the handle from the support. The Hygrotherm is automatically activated via the installed light barrier. The adjustment control of the Hygrotherm requires ca. 20 s, until Ear irrigation the desired temperature is reached. Check the temperature bowl indication! As long as the temperature of the actual value deviates from the target value, the LED button „set“ flashes.
4.0 Operation 4.4 Thermal nystagmus stimulation, Variotherm 4.4.1 Overview Front view Display of stimulation time (double-digit, resolution 1s) Button for selecting the irrigation level (37°C) Adjustment button up or down (for temperature or Button for selecting the warm stimulation level (e.g. 44°C) seconds setting) Button for selecting the cold stimulation level (e.g.
4.0 Operation 4.4.3 Activation of the Variotherm Ear irrigation bowl Remove the handle from the support. The Variotherm is automatically activated via the installed light barrier. The adjustment control of the Variotherm requires ca. 20 s, until Handle thermal the desired temperature is reached. Check the temperature nystagmus stimulation indication! As long as the temperature of the actual value deviates from the target value, the LED button „set“ flashes.
4.4.8 Information about the service display of the ATMOS S 61 Servant up from software version V 1.2 ® The ATMOS S 61 Servant workstation has a service display on the front panel. ® The display consists of a red and a yellow LED. red LED yellow LED After switching on the module the electronics goes through a test cycle.
4.0 Operation 4.5 Ear irrigation bowl with separate suction channel (530.2070.0) The ear irrigation bowl with optional, separate suction channel may only be used for suction during ear irrigation or stimulation. Only the ear irrigation bowl with integrated sieve (hole diameter 1 mm at a maximum) may be used. The suction channel is not suitable for any other suction procedure in the medical field! The liquid pump is not suitable for sucking off bigger particles, for example, cerumen, etc.
4.0 Operation 4.6.2 Adapt / deposit sprayer Adapt: Deposit: Insert the medication bottle Insert () the handle onto the medication bottle. top down in the support. Engage, then lift medication Press activator () and pull handle () backwards. bottle with the handle from the support ().
4.0 Operation 4.8 Instrument heating 4.8.1 Mirror quick-heater Press A. The mirror quick-heater heats up for 10 seconds and then switches off automatically. Risk of burns! Prior to each use on your patient check the mirror temperature (e.g. on the back of your hand). Only heat up disinfected instruments! 4.8.2 Heating module for instrument tray When the unit is switched on, the heating system for the instrument tray will also be activated and heated to 37°C.
4.0 Operation 4.9 Clipboard endoscopy Integration of 3-channel endoscope management for rigid and flexible endoscopes. 4.9.1 Option LED Light Cube The LED Light Cube is automatically activated when the unit is switched on. With a simple rotary motion the adapters are screwed in up to the stop position and fixed in the Clipboard. The light conductors can be plugged into the connection.
5.0 Cleaning and care 5.1 General information on cleaning and disinfection Prior to cleaning Set main switch of the ATMOS S 61 Servant ENT workstation to OFF prior to cleaning and disinfection! ® Prior to each use: required ) The described measures relating to cleaning and ) Always observe the concentration specifications and disinfection resp.
) When placing the twin tube nozzle again, make sure the and remove filter. Thoroughly rinse all parts under mark (0, X or milling area) on the nozzle shows upwards! running water. A detergent or cleaning agent may also ) Sprayer tubes are available as spare parts at ATMOS. be used if required. • Check the DDS bacterial filter / oversuction stop: - The DDS bacterial filter / oversuction stop are disposable products.
5.0 Cleaning and care 5.2 Recommended instrument disinfectants Manual disinfection of instruments Disinfectant Ingredients in 100 g Manufacturer Korsolex med AF N-dodecylpropane-1,3-diamine 15.6 g Bode Chemie, ® (Application concentrate) N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine 5.1 g Hamburg surfactants, corrosion inhibitors ph-value regulators, foam inhibitors Korsolex basic glutaral...
5.0 Cleaning and care 5.3 Recommended surface disinfectants Coated surfaces Disinfectant Ingredients in 100 g Manufacturer Green & Clean SK Di alkyl dimethyl ammonium chloride < 1 g Metasys, Rum Alkyl dimethyl ethyl benzyl ammonium chloride < 1 g (Austria) Alkyl dimethyl benzyl ammonium chloride <...
5.0 Cleaning and care 5.4 Recommended endoscope disinfectants Manual disinfection of endoscopes Disinfectant Ingredients in 100 g Manufacturer Helipur H plus N glutaral 12 g, BBraun, Melsungen ® 2-propanol 7.5 g ethyl hexanol 0.5 g surfactants, complexing agents, corrosion inhibitors, colorants, perfumes Helix Ultra peracetic acid...
6.0 Hygiene Plan Cleaning and disinfection plan ATMOS S 61 Servant ® What When Qualifi ed and trained staff who are familiar Notices with reprocessing. (Please fi ll in the Parts to be reprocessed Cleaning Disinfection Sterilization responsible person -> use a water-based...
Page 32
Instru AF (Schülke & Mayr) ® • Dismozon plus (Bode Chemie) ® Gigasept FF neu (Schülke & Mayr) the ATMOS S 61 Servant. ATMOS MedizinTechnik cannot be hold liable for ® ® • Gigazyme (Schülke & Mayr) ® • Kohrsolin FF (Bode Chemie) ®...
ATMOS recommends: Work should be carried out by an authorized ATMOS service partner. This ensures that repairs and testing are carried out professionally, original spare parts are used and warranty claims remain unaffected. - At least every 12 months a repeat test of the electrical safety should be performed according to IEC 62353. ATMOS recommends an inspection according to the manufacturer‘s specifications.
• The device must be well padded and packed in suitable packaging. • Place the form QD 434 „Delivery complaint / return shipment“ and the respective decontamination certificate in an envelope. • Affix the envelope to the outside of the package. • Send the product to ATMOS or to your dealer.
8.0 Troubleshooting Description Cause Measure Electric power supply No function, main switch is active, No voltage to power plug Check house fuse, possibly it burnt through control light does not light up Cable/plug defect Exchange cable/plug Suction Low or no suction rate, but vacuum Suction hose is clogged Clean suction hose gauge indicates vacuum...
Page 36
8.0 Troubleshooting Description Cause Measure Automatic filling of the hose rinsing container Container overflows Charging valve does not close Inform the service department Sensor does not detect filling level Sensor defect - inform the service department Container does not get filled Water supply is interrupted Assure water supply Safety switch in the container is not Place hose rinsing container correctly in the activated holder...
Accessories for light module economy 507.4510.0 Lithium-ionic rechargeable battery for the mobile use For ATMOS LED light source respectively LED headlight with power and eco mode, charging indication and holding clip. ® Universal battery quick charging power pack (100-240 V) 011.1199.0...
9.0 Consumables, Accessories, Spare Parts 9.3 Spare parts Spare parts for ear irrigation / thermal nystagmus stimulation Irrigation handle 530.1190.0 Rubber bush for splash guard 501.0331.1 501.0320.0 3-hole-sealing device side 3-hole-sealing handle side 530.1186.0 Double lumen hose for irrigation handle 530.1181.0 502.0984.0 Jet connection for water irrigation handle 40 mm...
10.0 Technical data at complete equipment Please note the technical specifications of the single devices in the separate operating instructions! Voltage 230 V~ ± 10 %; 50/60 Hz Special voltage: 100 V~ ± 10 %; 50/60 Hz 115 V~ ± 10 %; 50/60 Hz 127 V~ ± 10 %; 50/60 Hz Current consumption Max.
11.0 Disposal • The materials of the housing can be recycled completely. • The ATMOS S 61 Servant ENT workstation does not contain any hazardous goods. ® • The component parts of the ATMOS S 61 Servant ENT workstation must be disposed of correctly ®...
12.2 Guidelines and Manufacturer's Declaration - Immunity The ATMOS S 61 Servant ENT workstation is intended for use in the electromagnetic environment specified below. The ® customer or user of the ATMOS S 61 Servant ENT workstation should ensure that it is used in such an environment. ® Immunity Test IEC 60601- Test Level Compliance Level...
Page 42
12.3 Guidelines and Manufacturer´s Declaration - Immunity The ATMOS S 61 Servant ENT workstation is intended for use in the electromagnetic environment specified below. The ® customer or user of the ATMOS S 61 Servant ENT workstation should ensure that it is used in such an environment. ® IEC 60601- Test Immunity Test Compliance Level...
Page 43
S 61 Servant ENT workstation can help prevent electromagnetic ® interference by maintaining a minimum distance between portable and mobile RF Communications equipment and the ATMOS ® S 61 Servant ENT workstation as recommended below, according to the maximum output power of the communications equipment. Safety distance, depending on transmit-frequency m Nominal output of the...