Page 3
Contents 8. Operation ........................................26 General ........................................ 26 Attaching/removing the adaptor ................................ 27 Loading/Removing of the transmission instruments with ISO or Roto Quick connection .............. 28 Loading/Removing of the fixed RM connection..........................29 Maintenance process ..................................30 Oil service of chucking system ................................32 9.
Symbols in the Instructions for Use WARNING! ATTENTION! General explanations, (risk of injury) (to prevent damage occuring) without risk to persons or property CE mark Do not dispose of with domestic waste...
Page 5
Symbols on the device Catalogue number Consult instructions Do not dispose of with domestic for use waste Serial number Date of manufacture Data Matrix code Supply voltage for product information of the unit including UDI (Unique Device Identification) DC – direct current CE mark...
Page 6
Symbols on the packaging/cartridge This way up Permitted temperature Caution: According to Federal range for storage law, this medical device may only be sold by or on the order of a dentist, physician or any Fragile, handle with care Permitted temperature other medical practitioner range for transport: licensed by the law of the...
1. Introduction For your safety and the safety of your patients Prior to initial use please read the Instructions for use. These explain how to use your medical device and guarantee a smooth and efficient operation. Observe the safety notes. Intended use Unit for cleaning the spray channels and lubricating the moving parts of dental and surgical handpieces, turbines, air motors and air-driven dental scalers.
Page 8
Introduction Production according to EU Directive The medical device meets the requirements of Directive 93/42/EEC. Responsibility of the manufacturer The manufacturer can only accept responsibility for the safety, reliability and performance of the medical device if it is used in compliance with the following directions: >...
2. Electromagnetic compatibility (EMC) Medical electrical equipment is subject to particular precautions with regard to EMC and must be installed and put into operation in accordance with the EMC notes included. W&H only guarantees compliance of the medical device with the EMC Directives when it is used with original W&H accessories and spare parts.
3. Unpacking W&H packaging is environmentally friendly and can be Remove the cardboard box disposed of by industrial recycling companies. with the TWIN Care SET. However, we recommend that you keep the original packaging. Lift out the device. e Remove the cardboard box with the: >...
5. Safety notes Please observe the following instructions under all circumstances > Before using the medical device for the first time, store it at room temperature for 24 hours. > Only set up the device in sufficiently well ventilated rooms! >...
Page 13
Safety notes Device The medical device is classed as »conventional equipment« (closed equipment without protection against the ingress of water). Process fluids > If inhaled: Supply fresh air; consult doctor in case of complaints. > Following contact with skin: Wash off with soap and plenty of water. >...
6. Description Front panel Cover cartridge compartment Cartridge compartment for W&H Service Oil F1 MD-302 Operating controls Cartridge compartment for W&H Active Fluid MC-302 Connection port with union nut Sliding window Left chamber Right chamber Air duct...
Description Rear panel USB port (for Service only!) Filter compartment cover Connecting socket for Connecting socket for power supply unit Air connection hose...
7. Start-up Insertion/removal of a cartridge The cartridges must be at room temperature when they are inserted. The cartridge spike is sharp. Be careful not to injure yourself on the projecting spike. Remove the cartridge e Removing a cartridge. compartment cover.
Start-up Insertion/removal of a filter Open the filter e Close the filter compartment cover. compartment cover. Insertion/removal of a filter. > Push the filter in as far as it will go. > Pull out the filter. Always clean and disinfect the filter compartment whenever you replace a filter. Observe the manufacturer’s instructions on use of the disinfectant.
Start-up General Only use clean, filtered compressed air. > Unfiltered or contaminated compressor air is not cleaned. > Replace the filter if contaminated. Connection/disconnection of the air connection hose Connect the air connection hose. Disconnect the air connection hose. Connection of the power supply unit e Connect the power supply unit to the medical device.
Start-up Initial filling > Carry out the initial filling without a transmission instrument. > It does not matter which chamber you fill first. Close one chamber using Press the »Start« button the sliding window. for approx. 5 seconds. The initial filling of the first chamber will take approx.
Page 23
Start-up Initial filling e Close the other chamber using the sliding window. The initial filling continues automatically. The initial filling of the second chamber will take approx. 1 minute. > The LED for the chamber flashes green. > The W&H Service Oil and W&H Active Fluid LED are green. The initial filling has completed successfully.
Start-up Standby mode The device switches to standby mode: > when the power supply unit is inserted in the socket; or > if it is not operated for 15 minutes. All LED are off. Activating the Assistina TWIN > Press the »Start« button. All of the LED are white for 1 second.
Start-up Test run > Place a pre-disinfected transmission instrument onto the connection port (see pages 27 – 29). > Start the Assistina TWIN (see page 30). The transmission instrument will be slightly moist when you take it out. It then needs to be manually disinfected and sterilized. In the event of malfunctions (e.g., vibrations, unusual noises, overheating, failure or leakage), switch the medical device off immediately and contact an authorized W&H service partner (see page 53).
8. Operation General > Wear protective clothing, safety glasses, face mask and gloves. > Refer to the »hygiene and maintenance« instructions in the instructions for use of the transmission instruments.
Operation Attaching/removing the adaptor Damaged or leaking O-rings should be replaced immediately (see page 37). Connecting Push the adaptor as far as it Firmly tighten the union nut anti-clockwise by hand will go on the connection port. (leak tightness). Make sure that it is correctly positioned.
Operation Loading/removing Loading/Removing of the transmission instruments with ISO or Roto Quick connection Push the transmission instrument with the coolant outlet facing downwards onto the adaptor. For air motors set the direction of rotation via the switching ring to clockwise or anti-clockwise.
Operation Loading/removing Loading/Removing of the fixed RM connection Transmission instruments with a fixed RM connection can be pushed straight onto the connection port. Loading transmission instruments with fixed RM connection Push on the transmission Firmly tighten the union instrument.
Operation Maintenance process > The amount of W&H Active Fluid, MC-302 and W&H Service Oil F1, MD-302 used in the maintenance process is carefully metered. > The maintenance process always runs automatically and identically in both chambers. Close the sliding window. ...
Page 31
Operation Maintenance process The maintenance process is performed. > The LED flashes green. > The maintenance process runs automatically and takes approx. 10 seconds. The maintenance process has completed successfully. > The LED for the chamber is green. > A signal tone sounds.
Operation Oil service of chucking system > Remove the rotary instrument. > Use the HPI adaptors REF 07477400 and REF 02693000. > It is possible to oil the chucking system in both chambers. Close the sliding window. e Press the »Start« button twice.
Page 33
Operation Oil service of chucking system Align the instrument t Press the »Start« button. with the maintenance attachment. The oil service is performed. > The LED flashes purple. > The oil service runs automatically and takes approx. 5 seconds. Wear protective clothing, safety glasses, face mask and gloves.
Page 34
Operation Oil service of chucking system The oil service has completed successfully. > The LED for the chamber and the start button are purple. > A signal tone sounds. Continuing the oil service y Align the instrument u Press the »Start« button. with the maintenance attachment.
9. Cleaning of the Assistina TWIN > Follow your country-specific directives, standards and guidelines in relation to the disposal of process fluids and to cleaning and disinfection. > W&H recommends cleaning the medical device at least once daily or as required. Wear protective clothing, safety glasses, face mask and gloves.
Cleaning of the Assistina TWIN Switching OFF Remove the sliding window. Unplug the device from the mains supply. Device Manual cleaning of the interior and exterior Clean the chambers with a dry cloth. Wipe-disinfect the exterior and the chambers. >...
10. Maintencance Replacing O-rings Damaged or leaking O-rings should be replaced immediately. Do not use sharp tools! Firmly squeeze the O-ring between your thumb and index finger so that it forms a loop. Pull off the O-ring. e Slide on a new O-ring.
Maintencance Replacing the air filter Switch off the central air supply before replacing the air filter. > Replace the air filter whenever it is dirty or at least 1x per year. > Only use clean, filtered compressed air. Removing the air e Replace the air filter and re- connection hose connect the air hose (a, b).
11. Error messages LED display Description Solution Device in standby mode > Press the »Start« button. Device not connected to the mains supply > Connect the power supply. No LED lit up Electronics error > Unplug device and plug back in again. >...
Page 41
Error messages LED display Description Solution Problem with air supply > Restore the air supply. The start button LED is blue The maintenance process did not > Press the »Start« button. complete successfully. > Repeat the maintenance process. The LED for the right or left chamber is red start button LED is green The maintenance process did not complete...
Page 42
Error messages LED display Description Solution The initial filling did not complete > Check if the cartridges are: successfully. > connected correctly; > empty (replace cartridge). > Press the »Start« button. > Repeat the initial filling. The LED for the right or left chamber, the W&H Service Oil LED and W&H Active Fluid LED are red / start button LED is green...
Page 43
Error messages LED display Description Solution The cartridges are almost empty > Replace the cartridges as soon as possible. The W&H Service Oil and W&H Active Fluid LED are orange The cartridges are empty > Replace the cartridges. > Carry out the initial filling. The W&H Service Oil and W&H Active Fluid LED are red...
Page 44
Error messages LED display Description Solution The oil service of the chucking system > Press the »Start« button. did not complete successfully. > Repeat the oil service. The LED for the right or left chamber is red The start button LED is purple The oil service of the chucking system >...
12. W&H Accessories and spare parts Use only original W&H accessories and spare parts or accessories approved by W&H! 02690400 03104500 02679000 Adaptor for all W&H products Adaptor for W&H LW-style air Adaptor for detachable with Roto Quick system high-speed handpieces contra-angle handpiece heads 03103500 07535400...
Page 46
W&H Accessories and spare parts 05204600 07477400 02693000 Adaptor for Entran HPI adaptor for Adaptor for oil service contra-angle head Assistina TWIN MB-302 chucking system 02675200 02697000 07477500 Air filter Air connection hose (2 m) Power supply unit MB-302...
Page 47
W&H Accessories and spare parts 07484000 02695700 TWIN Care SET O-ring, HPI adaptor, rotation MB-302 02060100 O-ring, Roto Quick adaptor, large 02060200 O-ring, Roto Quick adaptor, small...
13. Servicing Repairs If defects occur, always send in the complete device. Returns > Refer all questions to an authorized W&H service partner (see page 53). > Always return devices in their original packaging.
14. Technical data Assistina TWIN MB-302 Supply voltage: 100 – 240 V AC Permitted voltage fluctuation: ±10% Nominal current 0.4 – 0.2 A Frequency: 50 – 60 Hz Max. power consumption: 18 VA Air consumption: approx. 40 Nl/min. Operating pressure: 73 –...
Page 50
Technical data Ambient conditions Temperature during transport: +32 °F to +122 °F (0 °C to +50 °C) Temperature during storage: +41 °F to +104 °F (+5 °C to +40 °C ) Air humidity during storage and transport: 8% to 80% (relative), non-condensing Ambient temperature during operation: +50 °F to +104 °F (+10 °C to +40 °C) Air humidity during operation:...
15. Disposal Ensure that the parts are not contaminated on disposal. Follow your local and national laws, directives, standards and guidelines for disposal. > Medical device > Waste electrical equipment > Packaging...
E x p la na ti on o f war ranty ter ms This medical device has been manufactured with great care by highly qualified specialists. A wide variety of tests and controls guarantee faultless operation. Please note that claims under warranty can only be validated when all the directions in the instructions for use have been followed.
Authorized W&H service partners Find your nearest W&H service partner at http://wh.com Simply go to the menu option »Service« for full details. Alternatively please contact: A-dec Inc., 2601 Crestview Drive, Newberg, OR 97132 USA t + 1.800.547.1883 Within USA/Canada, t + 1.503.538.7478 Outside USA/Canada f + 1.503.538.0276, www.a-dec.com...
Page 56
Manufacturer W&H Dentalwerk Bürmoos GmbH Ignaz-Glaser-Straße 53, 5111 Bürmoos, Austria Form-Nr. 50927 AAE t + 43 6274 6236-0, f + 43 6274 6236-55 Rev. 002 / 22.08.2019 office@wh.com wh.com Subject to alterations...