Summary of Contents for Accora FloorBed 2

  • Page 2: Table Of Contents

    Any actions that are inconsistent with the Storage manual are performed at your own risk and Accora shall not be liable for any injury or Daily inspection damage. Please ensure that the manual is available to operators throughout the bed’s General maintenance service life.
  • Page 3 Do not move the bed when it is in the Accora shall not be held liable for any Floor-level position. damage, injuries or accidents arising from unauthorised modifications, non-genuine...
  • Page 4: Means Of Delivery

    The bed may only be operated by persons who It is essential to consult Accora in advance if you wish are able to operate it in accordance with the to use the bed for any purpose outside the use manual.
  • Page 5: Accessories

    Do not dispose of in household waste. Headboard Degree of protection against electric shock: Footboard Class II Double Insulated. Mattress platform, head section For indoor use only Mattress platform, foot section Voltage supply Voltage input frequency © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 6: Assembly

    Identify all components: a. Box including mattress guides, handset, mattress platform joining bars and power supply. b. Mattress platform, head section Mattress platform, foot section d. Footboard Headboard © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 7 R-clips as shown in caps as shown below. two positions. Join the two parts of the bed together using the joining bars and secure with locking pins and R- clips as shown in two positions. © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 8: Positioning

    Reverse Trendelenburg and The patient should only operate bed if they Trendelenburg function are competent to do so. © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 9: Functionality Check

    Second swipe – orange light  Third swipe – no light The locked and unlocked options may be chosen at any time depending on the needs of the user. Brake locked Brake unlocked © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 10: Mattresses Selection

    Mattress platform head section 9. Ensure all R-clips and pins are refitted so that the bed on the transport bracket is safe to move. 10. Ensure all wires are secure to prevent damage. © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 11: Moving And Repositioning

    All functions MUST be tested and approved by Accora cannot be liable for any damage or risk of a competent person after moving or damage if inappropriate cleaning or disinfectant repositioning agents are used.
  • Page 12: Storage

    In the event of the disposal of materials from the bed, end-of-life parts must be disposed of in accordance Do not carry out maintenance with service with current environmental regulations. user or patient on the bed. © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 13 Harmonic emissions Class A connected to the public low-voltage power supply network that supplies buildings used IEC 61000-3-2 for domestic purposes. Voltage fluctuations Compliance /flicker emissions IEC 61000-3-3 © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 14 IEC levels characteristic of 61000-4-8 a typical location in a typical commercial or hospital environment. NOTE UT is the a.c. mains voltage prior to application of the test level. © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 15 If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the bed. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 16 NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 17 Model reference: FloorBed 2 Serial No: Refer to product label conforms to the Requirements of Directive 93/42/EEC relating to Medical Devices Class 1 Product. Signed by: 01.01.2017 Name: Laurie Drake Position: Operations Director © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com...
  • Page 18 © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com IFU-FB2-001 EN Rev 1 JAN 2016...
  • Page 19 • •...