Chapter 6 | Troubleshooting

Troubleshooting

Troubleshooting Table

PROBLEM
ViroVac
©
Surgical Smoke
Evacuator is on
but suction is
minimal or none.
ViroVac
©
Surgical Smoke
Evacuator does
not function
even though On/
Standby Button
is depressed. No
lights or locked
lights.
www.buffalofilter.com
POTENTIAL CAUSE
Filter is not seated
completely.
Filter is clogged.
Vacuum tubing is
clogged.
Motor/blower is
obstructed.
Not plugged into an
electrical outlet.
Fuse(s) are blown.
Electronic ViroVac
©
Surgical Smoke
Evacuator failure.
Filter life has expired
or invalid filter installed.
PCB is not functioning.
+1 716.835.7000 (International)
CORRECTIVE ACTION
Reinstall the filter, press firmly into place.
Replace the filter with a new filter from the
manufacturer.
Replace the vacuum tubing with the manufacturer's
products.
Call the facility's biomed department or the
manufacturer's Technical Services.
Check the grounded electrical outlet connection and
the power receptacle located at the rear of the
ViroVac
Surgical Smoke Evacuator.
©
Disconnect the power. Remove the fascia and the
body. Check if one of the two fuses on the circuit
board (PCB) are blown. Replace fuse(s).
Call the facility's biomed department or the
manufacturer's Technical Services.
Replace filter.
Apply the power directly to the motor. If the motor spins,
the PCB is not functioning. Replace the PCB. Perform
a hipot test and current leakage test after replacing the
PCB or the motor.
800.343.2324 (United States)
6
bf.techsupport@filtrationgroup.com
6-1

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ViroVac VV120 and is the answer not in the manual?

Questions and answers

Sally Hout
February 3, 2025

Virovac is staying on even when cautery not in use. **** just keeps running

1 comments:
Mr. Anderson
February 10, 2025

The Buffalo Filter ViroVac VV120 continues to run when the cautery is not in use because it does not automatically turn off. It will keep operating until it is manually turned off. Additionally, an Automatic Activation Device can be used to control the suction on/standby modes in conjunction with ESU activation, which helps in saving filter life.

This answer is automatically generated

This manual is also suitable for:

Virovac seriesVirovac vv220Virovac dkvv220

Table of Contents

Save PDF