Each meter has the set of special functions: automatic zeroing, pieces counting, comparing with threshold values, etc., which may be available for the user or not – according to the order. In the further manual part the SE-01 meter operation is described, as the part of complete balance. 2 Certificates Test certificate of SE-01 indicator (No PL CB 1) was issued by Main Office for Measurements in Warsaw (Notified Body No 1440).
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 5 Security rules To avoid electrical shock or damage of a scale or connected peripheral devices, it is necessary to follow the security rules below. 1. All repairs and necessary regulations can be made by authorized personnel only. 2.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 6 Technical data Parameter SE-01/A SE-01/N/18 SE-01/N/25 Max verification units number 6000e (3000e/range - in multi range scale) Part of error(pi) Accuracy class 1, 2, 5, 10, 20 (g. dkg, kg) Readability (d) free Verification unit (e)
Page 7
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ The metrological parameters of the balance are indicated on the rating plate. Installation SE-01/A SE-01/N/18 SE-01/N/25 Installation system Fixing 2 holes Ф5, 2 holes Ф5, 2 holes Ф5, spacing 59mm spacing 6.30in spacing 7.99in...
1. To build a scale basing on SE-01 indicator contact authorized manufacturer service point or use Installation Guide delivered with the indicator. 2. The manufacturer gives a full guaranty for SE-01 indictor only when the indicator was mounted by AXIS Sp. z o.o. In other cases the guaranty obligation is taken over by the final contractor of the weighing device.
Page 9
- 2 coil turns signal wires - 4 coil turns 1. Connect the external devices cables to the meter sockets, or in the SE-01/N/.. meters to the board strip inside the meter, using the hermetic penetrations in the housing (the SE-01/N.. strip drawing is shown below).
Page 10
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Diagram of common wires and shields in the platform balances: ~230V 50Hz Caution: The galvanic connection of sensors and adder housing is necessary.
DE-9 casing casing In SE-01/A/18 indicator an interface is placed on indicator’s housing. In SE- 01/N/... indicators interfaces are ends of wires which go out from an indicator. Attention: To connect scale’s wire with RS232C interface (described above) to a...
The input of external keys allows to place (make double) selected scale keys into control box or operator’s workstation. As a standard the input is taken out with a wire for direct connect to a control panel. SE-01/A/18 indicators can be equipped with external key interfaces (as an option on demand).
Page 13
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Marker numbers and outer wires colors of SE-01/N/18 (SE-01/N/25) indicators: Marker External keys inputs Wire color* yellow MENU (F) brown green →0← white →T← (T/ON) +24V (voltage of external feeder) pink I/ ( ) blue violet * colors can be changed External keys connecting –...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 7.4 Strip for connecting external devices Inside SE-01/N/18 and SE-01/N/25 indicators housing there is a strip with communication ports, transmitters and external keys connections on it. The draft of SE-01/N/18 (SE-01/N/25) indicators strip: Strip description: IN-A(+), IN-B(-),...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 8 General Rules 1. After switching-on the power, the scale proceeds with self-tests and zeroing the scale. During start-up the scale should not be loaded (does not apply to tank scales, where scales are not zeroed after start-up). 2.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 10 Cooperation with computer or printer In indicators equipped with one serial RS232C (RS485) interface there are two possible modes of interface work: Standard mode The scale sends weighing result when initial signal form computer appears or key of the scale is pressed.
Page 17
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Scale →Computer: without response, → ← scale zeroing (calling key press): Computer→ Scale: S Z CR LF (53h 5Ah 0Dh 0Ah), Scale →Computer: without response, entering to special function menu (calling MENU key press): Computer→ Scale: S F CR LF (53h 46h 0Dh 0Ah), Scale →Computer: without response, setting low threshold value: Computer→...
Page 18
SW-1 SW-2 SW-3 SW-4 SW-5 SW-6 SW-7 SW-8 AXIS offers computer programs for cooperation with scales which description and demo versions are available on www.axis.pl website. Furthermore a free program for testing of scale serial interface is also placed there.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 11 Basic functions To make clear how to manage with each function, in further part of instruction descriptions are replaced with pictures. - put a load on the pan - remove the load from the pan - press the key when indication is 2.30 displayed - forced change...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 11.2 Weighing with tare The scale is equipped with tare equal 2.30 to its range. Joint value tare and mass net can not 0.00 cross a maximum of scale. To display gross weight press B/G key. 40.01 Note: Press B/G key to return to net weight indication.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12 Special functions List of available functions: menu customization function (ACtIV), removal of all function from menu (dEFAUL). autozeroing function (AutoZE), pieces counting function (PCS), function for setting serial port working (PrInt) function for setting serial port (Port) entering tare function (tArE) recipe weighing function (rECIPE), weighing large animals function (LOC)
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.1 Menu customization function (ACtIV and dEFAUL) Among available user functions it is 0.00 possible to select these, which should be displayed after pressing MENU key. It allows avoiding displaying whole list of available functions, which makes ACtIV operation time longer.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.2 Autozeroing function (AutoZE) When F..-Aut function is activated, 0.01 the scale automatically ensures stable zero indication if the pan is empty or if zero indication was acquired by → ← pressing key. AutoZE To leave the function press MENU →...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.3 Pieces counting (PCS) This function enables to count identical pieces, e.g. turnbuckles or 0.00 buttons. A measurement is performed in two phases: - first phase - single piece weight calculation on the basis of defined 23.40 pieces amount (5, 10, 20, 50, 100, 200 or 500 pieces), - second phase –...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.4 Printer cooperation settings (PrInt) Activate the function for automatic 0.00 serial port working mode (cooperation with a printer. After activation the scale prints a header. Weighing result with PrInt a successive measurement number is printed automatically after result stabilisation (without using key).
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.5 Serial port parameters setting (Port) 0.00 The function enables to set the following transmission parameters (standard parameters underlined: Port - transmission protocol (Prot): LonG - printer, ELtron – label printer, - transmission speed (bAud: 1200, Port-0 4800, 9600, ...), - the number of bits in a byte (bit: 7, Port-1...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.6 Constant tare (tArE) This function enables to measure gross weight of a sample placed in a container of a known weigh value (stored in the memory) and to display calculated net weight → ← of the sample. Tare value is recalled from the memory with key when the pan is empty.
Page 28
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Weighing with constant tare: -1.70 When tAr function is activated, press → ← key to zero the indication and to recall tare value from the memory. Tare value is displayed with ”-” sign. 10.00 enables instant switching between net and gross weight.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.7 Average calculation function (AVErAG) The function allows for calculating average value of performed measurement series. During series of measurements successive results are registered automatically when scale indication is stabilised. Press MENU key and select AUErAG 0.00 →...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.8 Recipe weighing function (RECIPE) 0.00 This function allows for separate MENU weighing of several ingredients in one container with a possibility of control total weight of all weighed rECIPE components. The function has the following options: rECI- 0 - rECI-0 –...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.9 Percentage weighing function (PErC) This function allows displaying weighing result in percents. A measurement is performed in two phases: 0.00 - first phase – weighing a reference sample (100%), 100% - second phase – measuring specific sample percentage 1.70...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.10 Anti-disturbance filter function (FILtEr) This function allows using digital filter 0.00 with selected intensivity during weighing. Filter reduces the influence of mechanical vibrations (air blasts, FILtEr base vibrations) on measurement result. Press MENU key and select FILtEr →...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.11 Function for maximum value indication (UP) This function allows holding on display maximum value shown by the scale in a while. Before measurement scale should be 0.00 tared. After using MENU key and selecting UP function the highest mass result will be hold on display.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.12 Force measuring function (nEWto) Function activation causes displaying result in force units (N). Press MENU key. nEWto → ← Using key choose NEWto function, and then NEW-1. Note: 1N≈0,1019kg nEW -0 nEW -1 0.000...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.13 Function for weighing large animals (LOC) The function allows weighing animal moving on the scale. Press MENU key. 0.00 MENU When LOC function is displayed press → ← key. The following options appear on display successively: - LOC-0 –...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.14 Checkweighing function (thr) This function allows comparing weighing result with two programmed reference values: lower and upper threshold. Comparison result is signalled with indicators (MIN, OK, MAX) and sound signal generated when threshold values are exceeded. If comparison result is: - smaller than lower threshold –...
Page 37
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Relays output is the open collector transoptor output with load capacity 100mA / 24V. Transmitter inputs must be protected with diodes, e.g. 1N4148. It is advised to use MS3K/P electronic board (sold separately), consisting of RM96P transmitters, with DC24V input voltage and AC250V, 3A output. Important notes: 1.
Page 38
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Operation sequence: Press MENU key and choose thr 0.00 → ← pressing key. The following options are displayed successively: - thr-0 – deactivate the function, - thr-1 – activate the function, - thr-P – check last threshold values (press key several times), thr -0...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.15 Total weight function (totAL) The function allows calculating total weight for series of measurements, which can be greater than scale capacity. It allows calculating total weight as well as average value. Press MENU key. 0.00 MENU →...
Page 40
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ To leave the function with clearing total register, select total function from menu and choose tot-0 option. When It will cause the scale prints the communicate informing about clearing registers. The form of receipt after each measurement: Date: Time.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.16 Function for setting date and time (dAtE) The function allows setting current 0.00 date and time of scale internal clock and mode of its use. The function has the following dAtE options: - dAt-0 – deactivate date and time during printout of current weighing dAt- 0 result,...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.17 Function for selecting label number (LAbEL) The function appears in scales with ELTRON data transmission protocol. This protocol allows printing scale indication and optionally date and time on label printer, as variable texts. Other data, e.g. company address, product name, its bar code can appear on label as constant fields.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.18 External calibration (CALIb) Note: This function is enabled in non legalized scales only. Calibration of sensitivity should be make when aaccuracy of scale is not satisfactory. Could use standard mass equal of maximum range of scale (Max). Operation sequence: Press MENU key to call functions 0.00...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.19 Setting time of stabilization function (Stb) The function allows changing stabilisation time of scale indication and connected with it the time of waiting for starting result printout on a printer connected to the scale. Press MENU key. 0.00 →...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.20 Entering reference zero function (Zero) Note: This function is enabled in non legalized scales only. ZER function allows entering new value of reference zero (value referred to empty pan) without need of contacting with authorised service centre. Press MENU key.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.21 Automatic switching off the scale function (AutoOF) The function is helpful in scales supplied from accumulator. The function causes scale to switch off automatically after c.a. 10 minutes of not using it. Switching function on causes last entered zero and tare values are remembered in scale memory.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.22 Density determining function (dEnSIt) The function allows determining 0.00 solid density basing on its weight in the air and in liquid of known density using the formula below: dEnSIt g= * g liquid dEn -1 where: m -weight in the air - weight in liquid...
Page 48
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ If a printer is connected to the scale, printout of solid density measurement results will be performed in the following form: Date: ... Time..MEASUREMENT No. = WEIGHT in air WEIGHT in liquid HANGER WEIGHT LOAD DENSITY g/cm LIQUID DENSITY g/cm...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.23 Calculator for good packaging control (tP) Note: This function is enable for special orders substitutable with other functions. The function allows performing not destroying control of 60, 100 or 160 samples of packed goods. Calculation formula complies with the description of reference method described in act of law from 26.07.2001 (with later changes).
Page 50
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ - "F..-1"- measurements, - "F..-2"- enter product data: product code and batch size, - "F..-3"- enter control parameters: nominal mass Qn and measurement number, - “F..-4”- select mode for data transmission (automatic or manual). [Entering product data] →...
Page 51
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ → ← 11.Press key when "F..-4" is displayed. The following options appear successively on display: - „Auto” – automatic data transmission after scale indication is stable, - „Recz” – manual data transmission by operator using key, - „out” – exit. →...
Page 52
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Date: Time: MEASUREMENT COUNT = QUALIFYING AVERAGE = STANDARD DEVIAT. S *HISTOGRAM* <Qn-2T1 -n Qn-2T1 ---------- 19.In order to finish working with the function and reset results register, press MENU key, and when "F..-tP" and "F..-0" is displayed →...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.24 Statistical calculations function (StAt) Attention: Function is available on demand and it replaces other special functions. This function evaluates from series of measurements (max 500) statistical parameters of weighting process. Adding successively measurements to register is automatic and it occur after the scale is loaded and its indications stabilize.
Page 54
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Data: Hour. LSL - allowable lower value, SAMPLES USL - allowable upper value, TOTAL MASS = A, B, C, .. – measurement intervals, AVER MASS = ... – amount of measurements in A MIN MASS interval; MAX MASS measurement is in A interval if it is MAX –...
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 12.25 Paperweight calculation (PAP) Note: This function is enable for special orders substitutable with other functions. This function enables to calculate paperweight of 1m of paper basing on samples of known area. For quick access, the function is accessible directly by pressing MENU key.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ 13 Maintenance and troubleshooting 1. The scale should be kept clean. 2. Take care that no dirt gets between the casing and the pan. If found any, remove the pan (lift it up), remove dirt and then replace the pan. 3.
ENGINEERING DOCUMENTATION _____________________________________________________________________________________________ Declaration of Conformity AXIS Spółka z o.o. 80-125 Gdańsk, ul.Kartuska 375B declare with full responsibility, that the balance meters: SE-01 marked with CE mark are consistent with: 1. Standard PN-EN 61010-1:2004 Safety requirements electrical equipment measurement, control and laboratory use. General requirements harmonized with the directive 73/23/EEC (Low Voltage Directive), 2.
Need help?
Do you have a question about the SE-01 and is the answer not in the manual?
Questions and answers