Screen indicated the probe is too far. How do I change this?
1 comments:
Mr. Anderson
February 11, 2025
To change the probe distance setting on the Icare HOME device, follow these steps:
1. Without touching the patient’s eye, set the distance between the tip of the probe and the center of the cornea to 4-8mm (5/32-5/16”). 2. Adjust the forehead and cheek support positioning by turning the knobs as needed. 3. Write down the settings on a support position tag for the patient. 4. Repeat the process for the patient’s other eye if necessary.
Need help?
Do you have a question about the Home and is the answer not in the manual?
Questions and answers
Screen indicated the probe is too far. How do I change this?
To change the probe distance setting on the Icare HOME device, follow these steps:
1. Without touching the patient’s eye, set the distance between the tip of the probe and the center of the cornea to 4-8mm (5/32-5/16”).
2. Adjust the forehead and cheek support positioning by turning the knobs as needed.
3. Write down the settings on a support position tag for the patient.
4. Repeat the process for the patient’s other eye if necessary.