INSTRUCTIONS FOR USE
Parker
04.AL.01_13EN • 09/2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Parker and is the answer not in the manual?

Questions and answers

Summary of Contents for Arjo Parker

  • Page 1 INSTRUCTIONS FOR USE Parker 04.AL.01_13EN • 09/2019...
  • Page 2 Mandatory to read the Instructions for Use Design Policy and Copyright ® and ™ are trademarks belonging to the Arjo group of companies. © Arjo 2019. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.The content of this publication may not be copied either whole or in...
  • Page 3: Table Of Contents

    Contents Foreword ................... 4 Intended Use ................5 Safety Instructions ..............6 Preparations ................7 Parts Designation ..............8 Control Panel ................10 Hand Controls ................. 11 Product Description/Functions ..........12 Drain Control Knob .............. 12 Auto Fill Function ..............12 Scalding Protection .............
  • Page 4: Foreword

    Foreword Definitions in this IFU Thank you for purchasing Arjo Equipment WARNING Please read this Instructions for Use CAUTION thoroughly! Instructions for Use (IFU) Parker NOTE Serious incident Service and Support Parker Care and Preventive Maintenance on page 42...
  • Page 5: Intended Use

    Installation Instructions Assembly Instructions for Use (IFU). and Installation Instructions If there is anything you do not completely understand, ask your Arjo representative. Care and Preventive Maintenance on page 42 Parker Instructions for Use (IFU)
  • Page 6: Safety Instructions

    Safety Instructions WARNING To avoid injury to both the patient and the caregiver, never modify the equipment or use incompatible parts. WARNING To avoid explosion or fire, never use the equipment in oxygen rich environment, in presence of heat source or flammable anaesthetic gases.
  • Page 7: Preparations

    Preparations Site Preparations WARNING Prior to the installation of the Parker bath the To prevent scalding, make sure that the floor and walls within the required area must water mixer has been calibrated. be finished. Confirm that responsible organization has established calibration routines.
  • Page 8: Parts Designation

    Parts Designation Sound & Vision Sound & Vision Sound & Vision Sound & Vision Sound & Vision Sound & Vision Type B Applied part. Degree of Protection - Electrical shock Sound & Vision Sound & Vision Sound & Vision...
  • Page 9 21 22...
  • Page 10: Control Panel

    Control Panel 14 15...
  • Page 11: Hand Controls

    Hand Controls Hand Control Sound & Vision® System Hand Control (option) Sound & Vision...
  • Page 12: Product Description/Functions

    Product Description/Functions Drain Control Knob Fig. 1 Auto Fill Function Fig. 2 Scalding Protection °C Temperature Display Parker Fig. 3...
  • Page 13: Emergency Stop

    Emergency Stop Fig. 4 Shower Handle Fig. 5 Lock Button Trigger Functions Lock Fig. 6 Door Lock Parker...
  • Page 14: Air Spa System (Option)

    Air Spa System (Option) Parker NOTE Operating the Air Spa system with soapy water can cause excessive foaming. Sound & Vision® System (option) WARNING To avoid electric shock, only use the USB memory stick in the connector on the bath.
  • Page 15: Disinfection Unit (Option)

    The Sound & Vision system will stop automatically 30 minutes after activation. Press the Sound & Vision system light ON/OFF button on the Sound & Vision system hand control for continued lighting. Disinfection Unit (Option) Fig. 8 Parker Replace Disinfectant Bottle on page 30...
  • Page 16: Battery Back-Up

    Ensure that the bath is connected to a power outlet or the battery might run out of power. Contact your local Arjo technician if problem occurs. Battery charge level when the main power supply is not connected.
  • Page 17: Bathe Patient

    Bathe Patient To complete a full patient bath follow these 36 steps. WARNING Fig. 1 To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged.
  • Page 18 Fig. 4 WARNING To avoid falling, make sure the patient uses the grab rails for support when entering or exiting the bath. Parker WARNING To avoid falling, make sure the patient always remains in a seated position. Fig. 5 WARNING...
  • Page 19 WARNING Fig. 6 To avoid entrapment, make sure to keep the patients hair, arms and feet close to the body and use designated grab supports during any movement. Fig. 7 Shower the Patient (Step 15-24 of 36) Continue with the steps on the next page.
  • Page 20 WARNING To prevent scalding, always check the water temperature with your naked hand before Fig. 8 directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the flow of the water away from the patient.
  • Page 21 Air Spa System (option) (Step 25-25 of 36) Fig. 10 Sound & Vision® System (option) (Step 26 of 36) Sound & Vision Sound & Vision End the bath cycle Fig. 11 (Step 27-35 of 36) Fig. 12 Continue with the steps on the next page.
  • Page 22 Fig. 13 Fig. 14 WARNING To avoid falling, make sure the patient uses the grab rails for support when entering or exiting the bath. Disinfect after each Patient (Step 36-36 of 36) Fig. 15 Cleaning and Disinfection Instructions on page 28. NOTE To preserve the door seal, leave the door slightly open when not in use.
  • Page 23: Bathe Patient Using A Standing & Raising Aid

    Bathe Patient using a Standing & Raising Aid Parker Allowed Combinations Technical Specifications on page 49 Patient Transfer from Sara 3000 to Parker Bath (30 steps) Fig. 1 WARNING To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient.
  • Page 24 To avoid entrapment, make sure to keep the patients hair, arms and feet close to the body and use designated grab supports during any movement. Bathe the Patient on page 18 Parker Fig. 6 Continue with the steps on the next page.
  • Page 26: Bathe Patient Using A Ceiling Lift

    Bathe Patient using a Ceiling Lift Parker Allowed Combinations Technical Specifications on page 49 Patient Transfer from a Ceiling Lift to Parker Bath (16 steps) Fig. 1 WARNING To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient.
  • Page 27 Fig. 4 Fig. 5 Bathe the Patient on page 18 Fig. 6 Fig. 7 Cleaning and Disinfection Instructions on page 28.
  • Page 28: Cleaning And Disinfection Instructions

    To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use. WARNING To avoid eye or skin irritation, never disinfect in the presence of a patient. CAUTION To avoid damage on the equipment only use Arjo branded disinfectants.
  • Page 29 Microbiological Formation Prevention Installation precautions In order to limit the growth of bacterial flora, it is necessary to take action at four levels: • Avoid stagnating water and make sure proper water circulation; • Take precautions against the formation of scale and corrosion based on the water quality;...
  • Page 30 To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use. Parker Fig. 2 Fig. 3 Continue with the steps on the next page. Arjo Disinfectant Cleanser IV (60) Arjo General Purpose Disinfectant (35) Arjo Clean (30) Cen-Kleen IV (30) Arjo Guard (15)
  • Page 31 Fig. 4 Clean and Disinfect with a Spray Bottle (31 steps) WARNING To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use. Remove visual residue (Step 1-5 of 31) Continue with the steps on the next page.
  • Page 32 Clean (Step 6-17 of 31) Fig. 1 CAUTION Never use green or brown Scotch-Brite nor steel wool on the bath as it may scratch the surface. Fig. 2 Parker Continue with the steps on the next page.
  • Page 33 Disinfect (Step 18-31 of 31) Fig. 3 Parker Fig. 4...
  • Page 34 Instructions for Use. Arjo Disinfectant Remove visual residue Cleanser IV (60) (Step 1-10 of 66) Arjo General Purpose Disinfectant (35) Arjo Clean (30) Cen-Kleen IV (30) Arjo Guard (15) Fig. 2 Continue with the steps on the next page.
  • Page 35 Clean Fig. 3 (Step 11-38 of 66) Arjo Disinfectant Cleanser IV (60) Cleaning and Disinfection Instructions on page 28. Arjo General Purpose Disinfectant (35) Arjo Clean (30) Cen-Kleen IV (30) Arjo Guard (15) Fig. 4 Parker Parker Fig. 5 Continue with the steps on the next page.
  • Page 36 NOTE Disinfection should be distributed through all Air Spa system nozzles or proper disinfection can not be guaranteed. Contact an Arjo technician if problem occurs. Parker Fig. 6 Continue with the steps on the next page.
  • Page 37 (Step 39-66 of 66) Fig. 7 Parker Parker Fig. 8 NOTE Disinfection should be distributed through all Air Spa system nozzles or proper disinfection can not be guaranteed. Contact an Arjo technician if problem occurs. Continue with the steps on the next page.
  • Page 38 Parker Fig. 9...
  • Page 39 Clean and Disinfect Parker Bath using Fig. 1 MultiClean (33 steps) WARNING To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Arjo Disinfectant Use. (See Fig. 1) Cleanser IV (60) Follow steps 1- 30 after bathing each patient.
  • Page 40 NOTE Disinfection should be distributed through all Air Spa system nozzles or proper disinfection can not be guaranteed. Contact an Arjo technician if problem occurs. Multi Clean Fig. 4 Multi Clean Parker Parker Fig. 5 Multi Clean Continue with the steps on the next page.
  • Page 41 Fig. 6...
  • Page 42: Care And Preventive Maintenance

    Local regulations and standards may be more stringent than the recommended maintenance schedule. NOTE Product cannot be maintained and serviced while in use with the patient. PREVENTIVE MAINTENANCE SCHEDULE: Parker bath Between Every Every Every...
  • Page 43 Caregiver Obligations Parker Between Patients Parker Cleaning and Disinfection Instructions on page 28 Fig. 1 Every week Fig. 2 Fig. 3 Parker Continue with the steps on the next page.
  • Page 44 Fig. 4 Fig. 5 Fig. 6 Continue with the steps on the next page.
  • Page 45 Fig. 7 Every month Fig. 8 WARNING Pressurised system and risk of scalding. Make sure that the valve (A) is closed before removing the cap, this is especially important for the hot water valve. Fig. 9...
  • Page 46 Yearly Checks by Qualified Personnel Only Parker Maintenance and Repair Manual WARNING To avoid injury and/or unsafe product, the maintenance activities must be carried out at the correct frequency by qualified personnel using correct tools, parts and knowledge of procedures.
  • Page 47: Troubleshooting

    NOTE Before starting the troubleshooting check: Power supply, Water supply and the LEDs on control panel. If the problem can not be solved with the problem solving actions below, please contact an Arjo authorized technician. PROBLEM ACTION Sound & Vision...
  • Page 48 PROBLEM ACTION Care and Preventive Maintenance on page 42. Care and Preventive Maintenance on page 42. Cleaning and Disinfection Instructions on page 28 Cleaning and Disinfection Instructions on page 28 Sound & Vision Sound & Vision Sound & Vision Sound & Vision...
  • Page 49: Technical Specifications

    Technical Specifications Parker TECHNICAL DATA General Tub length in the upright position 1855 mm / 73 inches Tub length in reclined position 1960 mm / 77 1/8 inches Tub width 750 mm / 29 1/2 inches Height of door with bathtub in highest position, feet in lowest...
  • Page 50 IP Class IP X4 IP Class hand control and Sound & Vision System Hand control IP X7 Weight Parker bath depending on selected options (excluding packaging) 124 - 131 kg 273 - 289 lbs End of Life Disposal Bathtub •...
  • Page 51 Allowed Combinations Parker bath integrates with the following Arjo products: Stedy, Sara Lite, Sara 3000, Sara Plus, Maxi Sky 600 (with 2-point spreader bar), Mesh sling with head support (MLA4060 XS-L, XL) and, Mesh sling without head support (MLA4531 S-L, XL)
  • Page 52 Electrical Data: EU and North America Voltage 230 V (EU) 120 V (North America) Frequency 50 Hz (EU) 60 Hz (North America) Power consumption (maximum) 900 VA Number of phases Single phase AC Fuses 230 V 10 A 120 V 13 A Protective earth Potential equalisation Power supply cord replaced by qualified personnel...
  • Page 53 WATER SUPPLY Drain requirement Flow capacity 100 l/minute 26.4 US gallons/minute Position (floor, Wall) either Water (Hot and Cold) Connection 3/4” B.S.P female Upper limit static pressure 6 bar / 87 PSI Recommended dynamic pressure, at 8 °C (46 °F) and at 3 bar / 43 PSI 80 °C (176 °F) flow of 58 l/minute (15.3 US gallons/minute) Lower limit dynamic pressure at 8 °C and at 80 °C and...
  • Page 54: Labels

    Labels Label explanation Parker Assembly and Installations Instructions Symbol explanation...
  • Page 55 (located on the control (located on the control panel before first use) panel before first use) Approval Label 8360890 Sound & Vision ARJO HOSPITAL EQUIPMENT AB, ESLÖV SWEDEN Identification Label Product Feature Label Applied Part Approval Label Equipotential Bonding Lug Label...
  • Page 56: List Of Standards And Certificates

    List of Standards and Certificates STANDARDS/ DESCRIPTION CERTIFICATES...
  • Page 57: Electromagnetic Compatibility

    Electromagnetic Compatibility Parker WARNING Stacking or placing other electrical equipment next to this device is not recommended, it can interfere with the equipment’s operation and safety. Portable and mobile radio-frequency (RF) communications equipment can interfere with this equipment operation and safety.
  • Page 58 Guidance and manufacturer’s declaration – electromagnetic immunity Parker Parker Electromagnetic Immunity test IEC 60601 test level Compliance level environment – guidance...
  • Page 59 Guidance and manufacturer’s declaration – electromagnetic immunity Parker Parker IEC 60601 Compliance Immunity test Electromagnetic environment – guidance test level level Parker Parker Parker Parker...
  • Page 60 Recommended separation distances between portable and mobile RF communications equipment and the Parker bath Parker Parker Parker Separation distance according to frequency of transmitter Rated maximum output power of transmitter 150 kHz to 800 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz...
  • Page 61: Parts And Accessories

    Parts and Accessories 400.00.00.02- Grey 400.00.00.09 - Grey 8257376-033 400.00.00.15 - Blue 400.00.00.16 - Blue Leg Rest Back Support Cushion Head Cushion For disinfectant contact local Arjo representative...
  • Page 62 Intentionally left blank...
  • Page 63 AUSTRALIA FRANCE POLSKA Arjo Australia Pty Ltd Arjo SAS Arjo Polska Sp. z o.o. 78, Forsyth Street 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 O’Connor CS 70133 PL-62-052 KOMORNIKI (Pozna ) AU-6163 Western Australia FR-59436 RONCQ CEDEX...
  • Page 64 At Arjo, we are committed to improving the everyday lives of people affected by reduced mobility and age-related health challenges. With products and solutions that ensure ergonomic patient handling, personal hygiene, disinfection, diagnostics, and the effective prevention of pressure ulcers and venous thromboembolism, we help professionals across care environments to continually raise...