Agilent Technologies 3070 Series Site Preparation

Agilent Technologies 3070 Series Site Preparation

Board test systems site preparation
Hide thumbs Also See for 3070 Series:
Table of Contents

Advertisement

Agilent 3070 / 79000 Board Test
Systems
Site Preparation
June 2002

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3070 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Agilent Technologies 3070 Series

  • Page 1 Agilent 3070 / 79000 Board Test Systems Site Preparation June 2002...
  • Page 2 Acknowledgements, 2 direct, indirect, general, special, incidental or consequential damages in connection with the 3070 Series I DECLARATION OF CONFORMITY furnishing, performance, or use of this material. according to ISO/IEC Guide 22 and EN 45014, 3 U.S.
  • Page 3 300 Vrms. Acoustic Level for Agilent 3070 / 79000 Systems LpA = 69dB am Arbeitsplatz (operator’s position) normaler Betrieb nach EN27779: 1991 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 4 3070 Series I DECLARATION OF CONFORMITY according to ISO/IEC Guide 22 and EN 45014 Manufacturer’s Name: Agilent Technologies, Inc. Manufacturing Test Division Manufacturer’s Address: 815 14th Street S.W. Loveland, Colorado 80537 Declares that the Product Name: 3070 (Series I) Board Test Systems...
  • Page 5 3070 Series II DECLARATION OF CONFORMITY according to ISO/IEC Guide 22 and EN 45014 Manufacturer’s Name: Agilent Technologies, Inc. Manufacturing Test Division Manufacturer’s Address: 815 14th Street S.W. Loveland, Colorado 80537 Declares that the Product Name: 3070 Series II Board Test Systems...
  • Page 6 3070 Series 3 DECLARATION OF CONFORMITY according to ISO/IEC Guide 22 and EN 45014 Manufacturer’s Name: Agilent Technologies, Inc. Manufacturing Test Division Manufacturer’s Address: 815 14th Street S.W. Loveland, Colorado 80537 Declares that the Product Name: 3070 Series 3 Board Test Systems...
  • Page 7 Westfields House, West Avenue Kidsgrove, Stoke-on-Trent Staffordshire, ST7 1TL United Kingdom Certificate Number: CC/KID/009/99 Dated: 22 February, 1999 February 26, 1999 For further information contact your local Agilent Sales and Service office. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 8 (unboot) procedures i.e., executing the testhead power off command. To restore power after pressing the Emergency Shutdown Switch, switch the PDU Off for 15 seconds and then On again. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 9 (b) RIGHTS APPLICABLE TO THE SALE OF THIS software, interfacing, parts or supplies not supplied by PRODUCT TO YOU. Agilent, (c) unauthorized modification or misuse, (d) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 10 50 - causing personal injury. 65 - CAUTION - ESD — Devices sensitive to electrostatic discharge may be present. Follow approved ESD-safe handling practices while working with this equipment. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 11 WARNING NOTE ✸ Calls attention to a procedure, practice, or condition that could cause bodily injury or death. Contains important information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 12 DO NOT perform procedures involving cover or shield removal unless you are qualified to do so. DO NOT operate damaged equipment: Whenever it is possible that the safety protection features built into © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 13: Table Of Contents

    Contents Agilent 3070 / 79000 Site Preparation 03066-90114 Rev. Z 05/2002 Site Preparation – Legal Notices .....................1 Site Preparation – Process The Importance of Site Preparation ..................... 1-2 Site Prep Manual Description ....................1-2 Responsibilities ........................... 1-3 Agilent ’s Responsibilities ..................... 1-3 Customer’s Responsibilities ....................
  • Page 14 Table of Contents System-Dependent Criteria....................3-2 The System Plan Drawing ......................3-3 What To Draw ........................3-3 Planning Aids......................... 3-3 Assigning Specialists ........................3-8 What Are Specialists?......................3-8 Site Coordinator ........................3-8 System Administrator ......................3-8 Structural Specialist ....................... 3-8 Environmental Specialist .......................
  • Page 15 Table of Contents Fixture Storage........................4-13 Site Preparation – Environmental Requirements Introduction..........................5-1 Air Quality Requirements ......................5-2 Introduction..........................5-2 Corrosive Contaminants ......................5-2 Particulate Contaminants ....................... 5-2 Viscid Contaminants......................5-2 Air Temperature Requirements ....................5-3 Introduction..........................5-3 Air Temperature Specifications ..................... 5-3 Cooling Requirements ......................
  • Page 16 Agilent Local Area Network Plan ..................8-2 Sample 3070 Series I Network Scheme................. 8-2 Agilent 3070 Series II Networking ..................8-5 Agilent 3070 Series 3 / 79000 Networking ................8-7 Gathering Configuration Information for UNIX Startup and Networking........ 8-10 © Agilent Techonolgies 1989–2002...
  • Page 17 Inventory The Shipment ......................9-2 Uncrating The System ......................... 9-3 Uncrating an Agilent 3070 Series I System................9-3 Uncrating an Agilent 3070 Series II/3 and Agilent 79000 System........9-8 Uncrating Procedure ......................9-8 Uncrating the Support Bay ....................9-12 Tools Needed to Uncrate the Support Bay: ................. 9-12 Uncrating Procedure: ......................
  • Page 18 E1099A PDU Electrical Requirements................10-114 E1170-80003 PDK Electrical Requirements ..............10-124 E1131A PDU Electrical Requirements................10-125 Agilent 3070 Series II / 3 - E1135A/B Power Distribution Units ........10-128 Connecting Mains Power To The System ................10-143 Introduction........................10-143 44964A/B AC Mains Connection..................10-144 E1099A AC Mains Connection ..................
  • Page 19 Table of Contents Who Should do the Current Reduction Conversion?............11-3 What to Do in Case of Difficulty..................11-3 Series 3 Current Reduction Conversion Procedure ..............11-4 Current Reduction Conversion Preview ................11-4 Schedule the Conversion ..................... 11-5 Determine the System Type and Power Option..............11-5 Shut Down the System......................
  • Page 20: Rev. Z

    1 1 1 1 Site Preparation – Process 03066-90114 Rev. Z 06/2002 In this Chapter... The Importance of Site Preparation, 1-2 Responsibilities, 1-3 The Site Preparation Process, 1-5 Site Prep Checklist, 1-6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 21: The Importance Of Site Preparation

    Agilent Requirements describes floor, access, and 3070 Series 3 or 79000 which was shipped prior storage requirements. to June 1999 to have the same power requirements as 3070 / 79000 systems shipped after June 1999.
  • Page 22: Responsibilities

    CEs will bill up to 2 hours to Agilent the "Installation Visit" and the Agilent CE will bill Technologies. up to 4 hours to Agilent Technologies. ELECTRICAL and ENVIRONMENTAL WARRANTY The Agilent 3070 / 79000 family INSPECTION After primary power has been...
  • Page 23 If you believe that your system was damaged in shipment, call your Agilent representative. 1 It will be difficult to remove the tri-wall box in a room with less than 2.5 meters (8 foot) clearance. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 24: The Site Preparation Process

    Agilent representative. Schedule all site preparation activities. Site Plan Implementation Implement the plan. Use Chapters 4 through 11 and the system plan drawing. Work with your specialists. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 25: Site Prep Checklist

    Site Preparation – [ ] Floor suitable? Structural Requirements [ ] Room to move the big box? If not, do your have a place to unpack it? [ ] Storage sufficient? © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 26 [ ] What system drop breaker size is required? Breaker Size ______ [ ] Power cable installed for mains? Connect Mains Power [ ] Power installed to system? Convenience Outlet Power [ ] Testhead strip printer? [ ] Testhead server? © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 27 [ ] Testhead unpacked (Tri-wall box removed, but not removed from the pallet.). The Agilent service representative will unpack the smaller boxes and move the system as part of installation. 1. See Chapter 6, Site Preparation – Power Requirements for more information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 28 System Overview, 2-2 This chapter presents information that will help you understand the parts of the 3070 Series I, II, 3 and 79000 families of board test systems. There are no tasks presented in this chapter. © Agilent Technologies 1989–2002...
  • Page 29: System Overview

    The main difference between the 3070 Series I and the 3070 Series II systems is the latter has twice the test pin capability, and in the Series II the controller is mounted in the testhead pod. The Series II systems also have redesigned testheads which include space for more equipment.
  • Page 30 Chapter 2: Site Preparation – System Overview: System Overview Figure 2-1 307X Series I system Support Bay Testhead Control Bay © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 31 Chapter 2: Site Preparation – System Overview: System Overview Figure 2-2 317X Series I system Testhead © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 32 Chapter 2: Site Preparation – System Overview: System Overview Figure 2-3 3070 Series II system © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 33 Chapter 2: Site Preparation – System Overview: System Overview Figure 2-4 3070 Series 3 system with HP-UX controller and an optional external support bay © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 34 Chapter 2: Site Preparation – System Overview: System Overview Figure 2-5 3070PC Series 3 system using a Windows controller 88.4 mm (3.5 in) Wider Pod pc-sys.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 35 Chapter 2: Site Preparation – System Overview: System Overview Figure 2-6 79000 (E2190A) Functional Test System Performance Ports DUT Power VXI Mainframe Supplies 3070 Module Controller 79000d.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 36: Testhead

    NFS mounted. If a Series-II system is added to a site with Series I systems, the two The controller for the 3070 Series I system is an HP systems can be networked to share resources.
  • Page 37: Support Bay

    In the 3070 Series II family system, the support bay is present in 307X systems but not in 317X systems. The support bay, when present, houses the DUT power supplies and optional test equipment.
  • Page 38 3 3 3 3 Site Preparation – Planning 03066-90114 Rev. Z 06/2002 In this Chapter... Choosing The Site, 3-2 The System Plan Drawing, 3-3 Assigning Specialists, 3-8 Scheduling Preparations, 3-10 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 39: System-Dependent Criteria

    Difficult criteria are air quality, floor strength, ambient vibrations, and physical access. Use the information about difficult criteria described in this manual to qualify prospective locations. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 40: The System Plan Drawing

    Site Preparation – system on the system plan drawing. Figure 3-1 Structural Requirements for the exact dimensions of page 3-4 shows three layout suggestions for 3070 Series individual testheads and support bays. I system. Figure 3-2 on page 3-5 shows layout...
  • Page 41 12 ft. 12 ft. (3.7 m) (3.7 m) 11 ft. (3.4 m) Control 9 ft. Testhead Conveyor (2.7 m) Strip Monitor & Printer Keyboard SiteP6.wpg Left-to-Right Layout with Board Handler © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 42 POTS Control Control Testhead 9 ft. Testhead 9 ft. 2.7 m 2.7 m Strip Strip Printer Printer Monitor & Keyboard Keyboard 12 ft. 14 ft. 4.3 m 3.7 m sitepx.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 43 Chapter 3: Site Preparation – Planning: The System Plan Drawing Figure 3-3 Recommended 3070 Series II system layouts Right-to-Left Layout Left-to-Right Layout Support 3 m (9 ft.) Testhead Support 3 m (9 ft.) Testhead Strip Printer Strip Printer Monitor &...
  • Page 44 Chapter 3: Site Preparation – Planning: The System Plan Drawing Figure 3-4 Recommended 3070 Series 3 system layouts Right-to-Left Layout Left-to-Right Layout Controller Controller Support 3 m (9 ft.) Support 3 m (9 ft.) Testhead Testhead Monitor & Strip Printer Monitor &...
  • Page 45: Assigning Specialists

    Chapter 5, Site Preparation – Environmental Requirements contains The HP-UX System Administrator Manual is a good information for this specialist. tool for your system administrator; it will arrive with the © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 46: Electrical Specialist

    The communications specialist will plan your local area network and install the LAN cables to your system. LAN cables and telephone lines should be marked on the system plan drawing. Chapter 8, Site Preparation – © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 47: Scheduling Preparations

    Make sure that you schedule extra time; if any delays occur, you can use the extra time to finish preparations. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 3-10...
  • Page 48: Site Preparation - Structural Requirements

    4 4 4 4 Site Preparation – Structural Requirements 03066-90114 Rev. Z 06/2002 In this Chapter... Floor Requirements, 4-2 Moving Access Requirements, 4-6 Storage Space Requirements, 4-12 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 49: Floor Requirements

    Table 4-1 System weights System Type Product Weight 3070 Series I Testhead 370 kg (750 lb) Support Bay 205 kg (450 lb) Testhead and Support Bay on Pallet 725 kg (1600 lb) 307X Series I...
  • Page 50 327X Series II Testhead 340 kg (750 lb) Testhead on pallet 477 kg (1050 lb) 327X, 3070PC Series 3 Testhead 340 kg (750 lb) Testhead on pallet 477 kg (1050 lb) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 51: Anti-Static Surface

    Quick Verify on pallet 150 kg (330 lb) Controllers 3070 - 300 Series Computer (in cabinet) 82 kg (180 lb) 3070 Series II - C110 Series Computer 15 kg (33 lb) Controllers – B180L 14 kg (36 lb) 3070 Series 3 and 79000...
  • Page 52: Immobilization For The Efs Board Handler

    If you mount the system to a different type of floor, you can still use the brackets provided with the kit, but you will need to furnish bolts © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 53: Moving Access Requirements

    1080 mm (42.5 in) Height 1600 mm (63 in) 317X Series I Testhead Length 1740 mm (68.25 in) (without Support Bay) Width 1080 mm (42.5 in) Height 1600 mm (63 in) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 54 Width 1080 mm (42.5 in) Height 902 mm (33.5 in) 3X79 Series II POTS Bay Length 1120 mm (44 in) Width 860 mm (34 in) Height 1860 mm (73 in) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 55: Dimensions Of The Uncrated System

    870 mm (34 in) Height 900 mm (35 in) 307X and 317X Series II Length 1860 mm (73.5 in) Testhead (without Support Bay) Width 820 mm (31 in) Height 900 mm (35 in) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 56 307XPC and 317X Length (88.4 mm (3.5 in) longer than 307X, 317X) 1765 mm (69.5 in) Series 3 Testhead Width 794 mm (31.3 in) (without Support Bay) Height 907 mm (33.5 in) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 57 1410 mm (55.5 in) Width 795 mm (31.3 in) Height 902 mm (33.5 in) 3X79 POTS Bay Length 1080 mm (42.5 in) Width 840 mm (33 in) Height 1600 mm (63 in) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 4-10...
  • Page 58: Ramp Requirements

    If you cannot move the testhead crate all the way, you must remove it from its crate in a receiving area and push it on its casters to the destination. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 4-11...
  • Page 59: Storage Space Requirements

    3070 system, and on CD-ROM. building. They are not available printed, therefore no storage space is needed. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 4-12...
  • Page 60: Fixture Storage

    Don’t store fixtures on wooden shelves. Wood splinters can be carried on the bottom of the fixture to the testhead, thereby contaminating testhead interface pins. Don’t stack fixtures on top of one another. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 4-13...
  • Page 61: Site Preparation - Environmental Requirements

    This chapter provides site environment requirements for reliable use of 3070 / 79000 family systems, including air quality, air temperature, and humidity. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 62: Air Quality Requirements

    Corrosives generally cannot be filtered out of the air by normal filtration methods, and the techniques that must © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 63: Air Temperature Requirements

    The system has a built-in temperature sensor inside the testhead, located on the ASRU card. This sensor constantly monitors the temperature of the air that has been drawn through the testhead after it cools the © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 64: Cooling Requirements

    3700 BTU (10.9 kW) 327X / 79000 1 Module 10500 BTU (2.5 kW) 3X79CT POTS Bay 8 Channels 8500 BTU (2.5 kW) 44900A EFS Board Handler 1000 BTU (0.3 kW) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 65: Humidity Requirements

    If the system is subjected to condensation, as if moved from a cold loading dock into a warm (and damper) environment, allow at least 24 hours for the system to recover before powering up. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 66: Site Preparation - Power Requirements

    3PD — 200–240 V 3-phase Delta 3PY — 208–220 V 3-phase Wye 3PN — 380–415 V 3-phase Wye with neutral © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 67: Identifying Your System

    PDU (see Is PDU Series I and Series II systems. Re-wiring Necessary? on page 6-11. Figure 6-2 on page 6-4 shows rear views of the 3070 Series 3 and 79000 systems. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 68 Chapter 6: Site Preparation – Power Requirements: Identifying Your System Figure 6-1 Rear views of 3070 Series I and Series II systems 317X Series I Testhead E1170-80003 PDK 307X Series I Testhead with Support Bay PDU #2 (Branch): E1135A or...
  • Page 69 Chapter 6: Site Preparation – Power Requirements: Identifying Your System Figure 6-2 Rear views of 3070 Series 3 and 79000 systems 317X Series 3 Testhead Port1 Port3 Port0 Port2 Mains Disconnect E1135B/C PDU (Main) 307X Series 3 Testhead with Support Bay...
  • Page 70: Power Requirements

    Be installed as close to the system as possible, where it can be easily reached by the system operator. If a circuit breaker is used, it must meet all of the above requirements plus: © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 71 6-7 shows the full-load amps (FLA) for each system type. FLA was calculated as follows: For the 79000: a fully-loaded system with 6 STC cards in the module. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 72 79000 317X 307X xDSL/POTS Bay 200–240V 3-Phase Delta 208–220V 3-Phase Wyel 120 / 208 127 / 220 380–415V 3-Phase Wye w/Neutral 220 / 380 230 / 400 240 / 415 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 73: Connecting Power To The Pdu

    North America — E1135-61611: uses #10 AWG 5-conductor cord with NEMA plug that mates with NEMA L21-30 (see Figure 6-5 page 6-10). The female receptacle is available locally. 1 Service with Oil resistant jacket © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 74 Chapter 6: Site Preparation – Power Requirements: Power Requirements Figure 6-4 3070 Series 3 - International 3-phase systems Site ac voltage source SO-Plug-International.wpg 2.5 m (8.2 ft) of 5x6mm SO cord, cord and male receptacle supplied by Agilent (E1135-61610) 30A / 3-Ph...
  • Page 75 Chapter 6: Site Preparation – Power Requirements: Power Requirements Figure 6-5 3070 Series 3 - North American 3-phase systems Site ac voltage source SO-Plug-NorthAmerica.wpg 2.5 m (8.2 ft) of #10 AWG 5 conductor SO Cord, cord and male receptacle supplied by Agilent...
  • Page 76: Is Pdu Re-Wiring Necessary

    Re-wiring is necessary because the internal circuits If you determine that re-wiring the PDU is necessary, go are wired phase-to-neutral and the internal PDU Wiring Diagrams on page 6-12. components would be under-powered. Whenever you © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-11...
  • Page 77: Pdu Wiring Diagrams

    Table 6-4 on page 6-13 directs you where to find wiring Diagrams diagrams in this chapter for 3070 Series 3 and 79000 testheads and E2197A xDSL / POTS bays. If you have a Series I or Series II system, refer to Chapter 10, Site Preparation –...
  • Page 78 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Table 6-4 3070 Series 3 / 79000 PDU wiring diagram Voltage Power line-to-neut Testheads xDSL/POTS Bay Option / line-to-line 327X Outlet 79000 Outlet 317X Outlet 307X Outlet Outlet locations Locations...
  • Page 79 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Table 6-4 3070 Series 3 / 79000 PDU wiring diagram Voltage Power line-to-neut Testheads xDSL/POTS Bay Option / line-to-line 208–220V 3-Phase Wye 120 / 208 Figure 6-7 Figure 6-12 Figure 6-17...
  • Page 80: 327X 1-Module System Ac Outlets

    Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams 327X 1-Module System AC Outlets Figure 6-6 327X 1-module system Outlet 2 Outlet 3 Outlet 1 Testhead (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-15...
  • Page 81 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-7 327X Options 3PY, 3PD 32-111.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-16...
  • Page 82 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-8 327X Options 3PY, 3PD © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-17...
  • Page 83 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-9 327X Option 3PN 32-111n.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-18...
  • Page 84 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-10 327X Option 3PN © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-19...
  • Page 85: 79000 1-Module System Ac Outlets

    Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams 79000 1-Module System AC Outlets Figure 6-11 79000 1-module system Outlet 2 Outlet 1 Outlet 3 Testhead (Rear View) E2190_ol.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-20...
  • Page 86 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-12 79000 Options 3PY, 3PD 79-111.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-21...
  • Page 87 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-13 79000 Options 3PY, 3PD © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-22...
  • Page 88 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-14 79000 Option 3PN 79-111n.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-23...
  • Page 89: 317X 2-Module System Ac Outlets

    Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-15 79000 Option 3PN 317X 2-Module System AC Outlets Figure 6-16 317X 2-module system Outlet 0 Outlet 2 Outlet 1 Outlet 3 Testhead (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-24...
  • Page 90 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-17 317X Options 3PY, 3PD 31-111.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-25...
  • Page 91 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-18 317X Options 3PY, 3PD © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-26...
  • Page 92 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-19 317X Option 3PN 31-111n.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-27...
  • Page 93: 307X 4-Module System Ac Outlets

    Figure 6-20 317X Option 3PN 307X 4-Module System AC Outlets Figure 6-21 307X 4-module system Outlet 0 Outlet 2 Outlet 1 Outlet 3 Support Bay Testhead (Rear View) (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-28...
  • Page 94 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-22 3070 Options 3PY, 3PD 30-111-1.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-29...
  • Page 95 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-23 3070 Options 3PY, 3PD © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-30...
  • Page 96 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-24 3070 Options 3PY, 3PD (support bay) 30-111-2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-31...
  • Page 97 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-25 3070 Option 3PN 30-111n-1.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-32...
  • Page 98 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-26 3070 Option 3PN © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-33...
  • Page 99 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-27 3070 Option 3PN (support bay) 30-111n-2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-34...
  • Page 100: E2197A Xdsl/Pots Bay Ac Outlets

    With the E2197A xDSL/POTS bay the 6811B ringer supply is plugged into the E1135C PDU and the isolation transformer is removed. See the Telecom Theory and Repair Manual, Chapter 4, for additional information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-35...
  • Page 101 Figure 6-28 E2197A xDSL / POTS Bay (shown with POTS options) Outlet 1 - Customer Instruments 200-240 Volts See Table To Gnd See Table See Table To Gnd See Table Line Mains Filter Input POTS-3ph.WPG © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-36...
  • Page 102 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-29 E2197A xDSL / POTS Bay Options 3PD, 3PY 21-111.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-37...
  • Page 103 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-30 E2197A xDSL / POTS Bay Options 3PD, 3PY © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-38...
  • Page 104 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-31 E2197A xDSL / POTS Bay Option 3PN 21-111n.wpg AC mains input cord/plug provided © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-39...
  • Page 105 Chapter 6: Site Preparation – Power Requirements: PDU Wiring Diagrams Figure 6-32 E2197A xDSL / POTS Bay Option 3PN © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-40...
  • Page 106: Facts About Pdus, Mpus And Power Options

    Chapter 6: Site Preparation – Power Requirements: Facts about PDUs, MPUs and Power Options Facts about PDUs, 79000 systems shipped prior to the 3070 Series 3 Beginning with the release of Series 3, in 2-PDU MPUs and Power used one E1135B PDU in the testhead and those...
  • Page 107 AWY (100/200 V Single-Phase Center-Tap Neutral) AWX (200 V Single-Phase Non-earthed) Series 3 current requirements were re-calculated and empirically checked, verifying lower current requirements when all the above enhancements were implemented. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-42...
  • Page 108 230 / 400 V 3-phase Wye with Neutral 240 / 415 V 3-phase Wye with Neutral 1. Regardless of the power option, the system always operates at 200–240 volts. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 6-43...
  • Page 109: Site Preparation - Compressed Air And Vacuum Requirements

    This chapter describes those requirements. For more about compressed air and vacuum principles and terminology, see the short Compressed Air and Vacuum Primer page 7-19. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 110: Compressed Air Requirements

    When the system is installed, connect the air to the Install compressed air lines using rigid line to a point testhead with a flexible line that has a Hansen Series near the testhead. An air cutoff valve is recommended © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 111: Air Quality

    The compressed air must be oil-free and must meet the humidity specifications in Table 7-1. We recommend that the air be filtered to remove contaminants such as oil and aerosols. The filter should filter 0.3-micron particles. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 112: Vacuum Requirements

    1. l/s is liters per second. STP is standard temperature and pressure. SCFM is standard cubic feet per minute. psi is pounds per square inch. See the Compressed Air and Vacuum Primer on page 7-19. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 113: Connecting Vacuum To The Testhead

    Chapter 7: Site Preparation – Compressed Air and Vacuum Requirements: Vacuum Requirements Connecting Vacuum to the Testhead Agilent 3070 Series I and Series II testheads do not have vacuum-switching solenoids (valves) inside the testhead, so external vacuum solenoids must be installed near the testhead to control the vacuum.
  • Page 114 307X Series I, II vacuum ports Vacuum Outlet 2 Vacuum Outlet 0 Vacuum Outlet 3 Vacuum Outlet 1 Vacuum Ports, 1-inch, male Top of Testhead connectors Rear of Testhead site_prep5.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 115 Chapter 7: Site Preparation – Compressed Air and Vacuum Requirements: Vacuum Requirements Figure 7-3 317X Series I, II system vacuum ports Vacuum Outlet 2 Vacuum Outlet 3 Top of Testhead Vacuum Ports, 1-inch male connectors Rear of Testhead site_prep18.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 116 3279 Series 3 Testhead 79000 Testhead Optional Optional Port3 Port2 Port2 Port3 E1135C PDU (Main) E1135B or E1135C PDU (Main) *Note: The E9998-63200 manifold will not work on the 3279 or 79000 testhead © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 117: Vacuum Guidelines

    If the vacuum manifold is used, a 2-inch vacuum hose should be used. If vacuum ports are connected individually, a 1-inch vacuum hose must be used. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 118 Chapter 7: Site Preparation – Compressed Air and Vacuum Requirements: Vacuum Requirements Figure 7-5 24-Vdc air-piloted solenoid vacuum example Support Bay Testhead 1-inch flexible tubing Air Line Filter Vacuum Source 24 Vdc Pressure site_prep3.wpg Air-Piloted Meter Solenoid © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-10...
  • Page 119 Chapter 7: Site Preparation – Compressed Air and Vacuum Requirements: Vacuum Requirements Figure 7-6 Individually activated vacuum port control example Support Bay Testhead Air Line Vacuum Source Pressure Meter site_prep2.wpg Filters 24 Vdc Solenoid © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-11...
  • Page 120 Figure 7-7 Vacuum with manual cutoff example Support Bay Testhead Flow 24 Vdc Manual Meter Solenoid Filter Valve Vacuum Source Manual Pressure Valve Meter Filter 24 Vdc site_prep4.wpg Solenoid © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-12...
  • Page 121 1 Vacuum valve inside diameter (ID) is 0.75 inch and outside diameter (OD) is 1.5 inches. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-13...
  • Page 122 E9945A Vacuum components shown above the dotted line Vacuum components below the dotted line are customer supplied Vacuum Source Pressure Pressure Meter Meter S3SiteP4 Manual Cut-off Valve 2-inch Vacuum Hose © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-14...
  • Page 123 4-Module Manifold E9946A Vacuum components shown above the dotted line Vacuum components below the dotted line are customer supplied Vacuum Source Pressure Meter S3Sitep3.wpg Manual Cut-off Valve 2-inch Vacuum Hose © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-15...
  • Page 124 Standard/Optional Vacuum components shown above the dotted line Vacuum components below the dotted line are customer supplied 2-inch Vacuum Hose Vacuum Source Pressure Pressure Meter Meter 79KSitep.wpg Manual Cut-off Valve © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-16...
  • Page 125: Vacuum Control Assembly

    The E4047A Vacuum Control Assembly is a pre-assembled metal housing. Features include two 24 Vdc, air-assisted, 3-way vacuum solenoid-actuated valves; in-line vacuum filters; air pressure regulator; manual valve for 1-channel or 2-channel operation. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-17...
  • Page 126 Chapter 7: Site Preparation – Compressed Air and Vacuum Requirements: Vacuum Requirements Figure 7-11 E4047A vacuum control assembly P neum at i cs P neum at i cs © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-18...
  • Page 127: Compressed Air And Vacuum Primer

    “standard temperature the customer’s fixture. and pressure” (STP). The standard temperature is zero © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-19...
  • Page 128 Agilent has found that a flow rate of 19 l/s (40 SCFM) will pull down most fixtures. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 7-20...
  • Page 129: Site Preparation - Communications Cabling Requirements

    Site Preparation – Communications Cabling Requirements 03066-90114 Rev. Z 06/2002 In this Chapter... Local Area Network Requirements, 8-2 Gathering Configuration Information for UNIX Startup Networking, 8-10 Gathering Configuration Information for MS Windows Networking, 8-13 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 130: Local Area Network Requirements

    IEEE 802.3 Local Area Networking. All the necessary software and The 3070 Series I family systems are shipped with a licensing comes with the system, but you will need to recommended list of host names and IP addresses. The purchase and install the LAN cabling and associated ’/etc/hosts’...
  • Page 131 LAN interface cabling. Figure 8-1 on page 8-4, there are three 3070 Series I systems. hp3070a, hp3070b, and hp3070c are systems with disks. These could be either testhead controllers or Test Development Centers; from the LAN standpoint it does not matter.
  • Page 132 Chapter 8: Site Preparation – Communications Cabling Requirements: Local Area Network Requirements Figure 8-1 An example of 3070 Series I family networkp3070a5 hp3070a hp3070b1 hp3070b1 hp3070b hp3070c1 hp3070c2 hp3070c3 hp3070c4 hp3070c5 hp3070c sitep13.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 133: Agilent 3070 Series Ii Networking

    IP address of the testhead controller and every other new controller in your Series II network. An example of the 3070 Series II internal testhead and external LAN connections is shown in Figure 8-2 page 8-6.
  • Page 134 Chapter 8: Site Preparation – Communications Cabling Requirements: Local Area Network Requirements Figure 8-2 An example of 3070 Series II network Router / Public LAN Public LAN Gateway /Backbone /Backbone as needed Test Development System 2..n on Hub Center Test Development...
  • Page 135: Agilent 3070 Series 3 / 79000 Networking

    The concept of a diskless work workstation has been eliminated with 3070 Series 3 and the 79000. The test servers are connected to the LAN the same way as any other servers.
  • Page 136 Chapter 8: Site Preparation – Communications Cabling Requirements: Local Area Network Requirements Figure 8-3 An example 3070 Series 3 / 79000 network Router / Public LAN Public LAN Gateway /Backbone /Backbone as needed Test System 2..n on Hub Server Test...
  • Page 137 Chapter 8: Site Preparation – Communications Cabling Requirements: Local Area Network Requirements Figure 8-4 An example 3070 Series 3 3070 MS Windows network Router / Public LAN Public LAN Gateway /Backbone /Backbone as needed Test System 2..n on Hub Server...
  • Page 138: Gathering Configuration Information For Unix Startup And Networking

    Do not configure to a DHCP network server. Factory tests have shown this configuration to be unstable. NOTE DNS and NIS are the only supported network protocols for UNIX systems. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 8-10...
  • Page 139 Task 6 Gateway IP Address The address designated to connect outside of the subnet. If there is not a designated address for this, use the IP Address from Task © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 8-11...
  • Page 140 Do not complete if configuring for DNS. 12 Will this TDS be a Font Server? [ ] Yes [ ] No Select either Yes or No. Ask yourself, will X-terminals be served by this system? © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 8-12...
  • Page 141: Gathering Configuration Information For Ms Windows Networking

    Workgroup or Domain? [ ] Domain The Domain name is the identifier for the server that controls and manages a Workgroup or Domain name group of computers on a client/server network. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 8-13...
  • Page 142 Gateway(s) for any backup routers of network traffic. Gateways 7 Will DNS be Used? [ ] Yes [ ] No If no, go to Task 12 (Will WINS Be Used?). (Domain Name System) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 8-14...
  • Page 143 [ ] Enable DNS for Windows Check or uncheck either of these boxes as applicable. Resolution Checkboxes CAUTION [ ] Enable LMHOSTS Lookup ✸ If configuring for DNS, check Enable DNS for Windows Resolution. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 8-15...
  • Page 144 Configured? [ ] DLC Protocol [ ] NetBEUI Protocol [ ] NWLink IPX/SPX Compatible Transport [ ] Point To Point Tunneling Protocol [ ] Streams Environment © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 8-16...
  • Page 145 Site Preparation – Receiving and Moving Instructions 03066-90114 Rev. Z 06/2002 In this Chapter... Inventory The Shipment, 9-2 Uncrating The System, 9-3 Placing The System, 9-15 Re-shipping a 3070 and 79000 System, 9-16 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 146: Inventory The Shipment

    Use the packing list that came with your system to Shipment determine if the system was fully received. Carefully examine the boxes for shipping damage. If you suspect the system was damaged in shipment, contact your Agilent representative. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 147: Uncrating The System

    ✸ Follow the unpacking instructions carefully and Uncrating an Agilent 3070 Series I System do not work alone. A fully loaded 307X Series I system with the shipping pallet might weigh 725...
  • Page 148 Remove all four pallet support bolts and all Remove 3.5 inch bolts with the sixteen bracket bolts from the pallet with the 9/16-inch open-end wrench 9/16-inch open-wrench or a socket inst17.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 149 Safety glasses 1 If you have a long floor jack (two meters or six feet long), use it on the RAMP END to move the pallet. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 150 If tester. 9 Retrieve the hex wrenches shipped with the system to help you unbolt the shipping brackets from the tester. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 151 For a 307X system, replace the foam that was placed between the testhead and the support bay to prevent scratching and denting of the cabinets. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 152: Uncrating An Agilent 3070 Series Ii/3 And Agilent 79000 System

    Chapter 9: Site Preparation – Receiving and Moving Instructions: Uncrating The System Uncrating an Agilent 3070 Series II/3 and Agilent 2 Remove two 150-millimeter (6-inch) bolts from a 79000 System plastic bag attached to the ramp. Figure 9-3 : Remove the “Removable End” board...
  • Page 153 Alternately lower each side a little at a time. Then Screw the bolts down (clockwise) until the platform remove the 6-inch bolts. rises slightly. Figure 9-4 : Remove two foam blocks from the space between the platform and pallet. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 154 Place two bolts that you removed from the sides of the pallet in step 3 through the holes in the ramp to keep the ramp from sliding off the pallet. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 9-10...
  • Page 155 The only safe way to ship the testhead is using the pallet and shipping materials provided. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 9-11...
  • Page 156: Uncrating The Support Bay

    Support Bay. Figure 9-6 : Remove the two shippin clamps (one on each side) from the bottom of the Support Bay (1, 2). © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 9-12...
  • Page 157 Chapter 9: Site Preparation – Receiving and Moving Instructions: Uncrating The System Figure 9-6 Removing the shipping hardware © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 9-13...
  • Page 158 (1) and pull the block out from under the bay (2). bay down the ramp (3). Position the ramp so the block of wood under the Figure 9-7 Installing the ramp © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 9-14...
  • Page 159: Placing The System

    If you are installing an 44990A EFS Board Handler into an automated test cell, the integration specialist will need to bolt the testhead to the floor. Refer to the 44990A EFS Board Handler Manual, 44990-90004, for more information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 9-15...
  • Page 160: Re-Shipping A 3070 And 79000 System

    44813A Opt. 572 3070 (Win), 79000 Series 3 327X, 327X (Win) 44813A Opt. 571 44813A Opt. 572 1. 3070 MS Windows system must use a pallet received after March 2000. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 9-16...
  • Page 161: Special Option To Part Number Listing

    44813A Opt. 565 = E4000-69901 44813A Opt. 566 = E4000-69902 44813A Opt. 567 = E4000-69903 44813A Opt. 570 = E9900-69901 44813A Opt. 571 = E9997-69901 44813A Opt. 572 = E9900-69902 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 9-17...
  • Page 162: Site Preparation - Power Requirements For Older Systems

    Series 3 Power Requirements and Connections (Prior to May 2001), 10-5 Series 3 Power Requirements and Connections (Prior to May 2000), 10-78 Tester Power Requirements (Series I/II), 10-109 Connecting Mains Power To The System, 10-143 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-1...
  • Page 163: Introduction

    3070 and Figure 10-1 same way. The power connections are very similar for 3070 Series II family testheads and support bays. the E1135C, except the input power is connected to a Following are important points to note about Figure mains disconnect switch.
  • Page 164 Quick Verify cabinet. Each PDU requires its own ac mains supply. The 327X Series II system uses one E1135B PDU in the testhead. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-3...
  • Page 165 E1135A or E1135B PDU E1135B or E1135C PDU (Main) 307X Series II Testhead with Support Bay View with rear door open SYS_ID.WPG E1135A or E1135B PDU (Main) E1135A or E1135B PDU (Branch) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-4...
  • Page 166: Series 3 Power Requirements And Connections (Prior To May 2001)

    3-phase system, depending on the country where the system will be installed: E1135-61611 - uses #10 AWG 5-conductor cord with NEMA plug that mates with NEMA L21-30. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-5...
  • Page 167 POTS and xDSL bays always require a separate power drop. Use the POTS/xDSL power option that is on the same line as the E1135C power option. Use E1135-61611 power cord with NEMA L21-30 receptacle/connector body. Use E1135-61610 power cord with IEC 60309 32A 3-phase receptacle/connector body (included with the system) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-6...
  • Page 168 Single-Phase Wye w/ 120 / 208 AWY, 16 Neutral 127 / 220 AWZ, 16 220 / 380 05M, 14 230 / 400 05F, 14 240 / 415 05J, 14 Single-Phase Earthed © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-7...
  • Page 169 2. DUT Power supplies are configured for 200-240 volts ac. 3. POTS and xDSL bays always require a separate power drop. Use the POTS/xDSL power option that is on the same line as the system power option. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 170: Series 3 / 79000 Power Distribution

    Your new system power option System type (the number of modules) Location of outlets Number of PDUs, as well as the figure page number which describes your power option wiring © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-9...
  • Page 171 Figure 10-24 Figure 10-35 Bay Outlet page 10-17 on page 10-28 on page 10-39 on page 10-50 locations Opts/Current Figure 10-53 Figure 10-52 on page 10-69 on page 10-68 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-10...
  • Page 172 Figure 10-27 Figure 10-39 Figure 10-55 05J, Figure page 10-20 on page 10-31 on page 10-42 on page 10-54, on page 10-71 10-60 Figure 10-41 page 10-76 on page 10-56 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-11...
  • Page 173 Rewiring of the PDU may be required. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-12...
  • Page 174 Figure 10-9 Figure 10-20 Figure 10-31 Figure 10-42 05J, Figure page 10-24 on page 10-35 on page 10-46 on page 10-57, 10-60 Figure 10-44 page 10-76 on page 10-59 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-13...
  • Page 175 Figure 10-11 Figure 10-22 Figure 10-33 Figure 10-49 Figure 10-51 Figure 10-60 on page 10-26 on page 10-37 on page 10-48 on page 10-64 on page 10-66 on page 10-76 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-14...
  • Page 176 0EH, Figure Figure 10-7 Figure 10-18 Figure 10-29 Figure 10-46 Figure 10-48 10-58 page 10-22 on page 10-33 on page 10-44 on page 10-61 on page 10-63 page 10-74 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-15...
  • Page 177 10-74 1. POTS and xDSL bays always require a separate power drop. Use the POTS/xDSL power options that is on the same line as the system power option. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 178: 327X 1-Module System Ac Outlets

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) 327X 1-Module System AC Outlets Figure 10-2 327X 1-module system Outlet 2 Outlet 3 Outlet 1 Testhead (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-17...
  • Page 179 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-3 327X 0ED, AWW, 05E, 05C, 05H, 0E6 32LLL.wpg Unswitched (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-18...
  • Page 180 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-4 327X 0ED, AWW, 05E, 05C, 05H, 0E6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-19...
  • Page 181 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-5 327X 0E5, 05L, 0E7 Unswitched 32LLLN.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-20...
  • Page 182 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-6 327X 0E5, 05L, 0E7 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-21...
  • Page 183 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-7 327X AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH 32LL.wpg Unswitched (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-22...
  • Page 184 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-8 327X AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 185 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-9 327X 05B, 05G, 05K 32LLN.wpg Unswitched (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-24...
  • Page 186 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-10 327X 05B, 05G, 05K © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-25...
  • Page 187 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-11 327X 05M, 05F, 05J Unswitched 32LN.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-26...
  • Page 188 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-12 327X 05M, 05F, 05J © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-27...
  • Page 189: 79000 - 1-Module System Ac System

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) 79000 - 1-Module System AC System Figure 10-13 79000 1-module system Outlet 2 Outlet 1 Outlet 3 Testhead (Rear View) E2190_ol.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-28...
  • Page 190 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-14 79000 0ED, AWW, 05E, 05C, 05H, 0E6 Unswitched 79LLL.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-29...
  • Page 191 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-15 79000 0ED, AWW, 05E, 05C, 05H, 0E6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-30...
  • Page 192 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-16 79000 0E5, 05L, 0E7 Unswitched 79LLLN.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-31...
  • Page 193 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-17 79000 0E5, 05L, 0E7 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-32...
  • Page 194 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-18 79000 AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH Unswitched 79LL.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-33...
  • Page 195 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-19 79000 AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 196 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-20 79000 05B, 05G, 05K 79K LLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-35...
  • Page 197 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-21 79000 05B, 05G, 05K © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-36...
  • Page 198 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-22 79000 05M, 05F, 05J 79KLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-37...
  • Page 199 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-23 79000 05M, 05F, 05J © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-38...
  • Page 200: 317X 2-Module System Ac Outlets

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) 317X 2-Module System AC Outlets Figure 10-24 317X 2-module system Outlet 0 Outlet 2 Outlet 1 Outlet 3 Testhead (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-39...
  • Page 201 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-25 317X 0ED, AWW, 05E, 05C, 05H, 0E6 CHECK COUNTRY CODE FOR FREQUENCY Unswitched 31LLL.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-40...
  • Page 202 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-26 317X 0ED, AWW, 05E, 05C, 05H, 0E6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-41...
  • Page 203 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-27 317X 0E5, 05L, 0E7 CHECK COUNTRY CODE FOR FREQUENCY Unswitched 31LLLN.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-42...
  • Page 204 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-28 317X 0E5, 05L, 0E7 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-43...
  • Page 205 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-29 317X AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH Unswitched 31LL.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-44...
  • Page 206 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-30 317X AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 207 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-31 317X 05B, 05G, 05K Unswitched 31LLN.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-46...
  • Page 208 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-32 317X 05B, 05G, 05K © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-47...
  • Page 209 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-33 317X 05M, 05F, 05J Unswitched 31LN.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-48...
  • Page 210 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-34 317X 05M, 05F, 05J © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-49...
  • Page 211: 307X 4-Module 1-Pdu System Ac Outlets

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) 307X 4-Module 1-PDU System AC Outlets Figure 10-35 307X 4-module 1-PDU system Outlet 0 Outlet 2 Outlet 1 Outlet 3 Support Bay Testhead (Rear View) (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-50...
  • Page 212 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-36 3070 1-PDU 0ED, AWW, 05E, 05C, 05H, 0E6 CHECK COUNTRY CODE FOR FREQUENCY Unswitched 30LLL_1.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-51...
  • Page 213 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-37 3070 1-PDU 0ED, AWW, 05E, 05C, 05H, 0E6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 214 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-38 3070 1-PDU 0ED, AWW, 05E, 05C, 05H, 0E6 (support bay) 30LLL_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-53...
  • Page 215 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-39 3070 1-PDU 0E5, 05L, 0E7 Unswitched 30LLLN_1.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-54...
  • Page 216 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-40 3070 1-PDU 0E5, 05L, 0E7 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-55...
  • Page 217 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-41 3070 1-PDU 0E5, 05L, 0E7 (support bay) 30LLLN_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-56...
  • Page 218 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-42 3070 1-PDU 05B, 05G, 05K Unswitched 30LLN_1.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-57...
  • Page 219 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-43 3070 1-PDU 05B, 05G, 05K © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-58...
  • Page 220 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-44 3070 1-PDU 05B, 05G, 05K (support bay) 30LLN_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-59...
  • Page 221: 307X 4-Module 2-Pdu System Ac Outlets

    Outlet 0 Outlet 2 See Table To Gnd Main See Table See Table To Gnd See Table Outlet 1 Outlet 3 Branch Support Bay Testhead (Rear View) (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-60...
  • Page 222 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-46 3070 2-PDU AWX, AWY, AWV, AWZ, 0EB, 0E3, 0EJ, 0EG, 0EC, 0EH CHECK COUNTRY CODE FOR FREQUENCY Unswitched 30LL_1.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-61...
  • Page 223 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-47 3070 2-PDU AWX, AWY, AWV, AWZ, 0EB, 0E3, 0EJ, 0EG, 0EC, 0EH © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 224 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-48 3070 2-PDU AWX, AWY, AWV, AWZ, 0EB, 0E3, 0EJ, 0EG, 0EC, 0EH (support bay) CHECK COUNTRY CODE FOR FREQUENCY 30LL_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-63...
  • Page 225 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-49 3070 2-PDU 05M, 05F, 05J 30LN_1.wpg Unswitched (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-64...
  • Page 226 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-50 3070 2-PDU 05M, 05F, 05J © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-65...
  • Page 227 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-51 3070 2-PDU 05M, 05F, 05J (support bay) 30LN_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-66...
  • Page 228: E2195/E2197A Xdsl/Pots Bay Ac Outlets

    Kikusui power supply, are plugged into the E1135C PDU. With the E2197A xDSL/POTS bay the 6811B ringer supply is plugged into the E1135C PDU and the isolation transformer is removed. See Communications Test for additional information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-67...
  • Page 229 Figure 10-52 E2195A xDLS / POTS Bay (shown with POTS options) Outlet 1 - Customer Instruments See Table To Gnd See Table See Table To Gnd See Table Line Mains Filter Input OTS1 WPG © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-68...
  • Page 230 Figure 10-53 E2197A xDLS / POTS Bay (shown with POTS options) Outlet 1 - Customer Instruments 200-240 Volts See Table To G nd See Table See Table To Gnd See Table Line Mains Filter Input © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-69...
  • Page 231 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-54 E2197A xDLS / POTS Bay AWW, 0ED, 05C, 05E, 05H, 0E6 Switched Ringer Supply E2197LLL.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-70...
  • Page 232 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-55 E2197A xDLS / POTS Bay 0E5, 05L, 0E7 Switched Ringer Supply E2197LLLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-71...
  • Page 233 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-56 E2195A xDLS / POTS Bay AWW, 05E © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-72...
  • Page 234 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-57 E2195A xDLS / POTS Bay AWW, 05E Switched E2195LLLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-73...
  • Page 235 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-58 E2195A xDSL / POTS Bay AWY, AWV, AWZ, 0E3, 0EG, 0EH E2195LLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-74...
  • Page 236 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-59 E2195A xDSL / POTS Bay AWY, AWV, AWZ, 0E3, 0EG, 0EH Switched © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-75...
  • Page 237 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-60 E2195A xDSL / POTS 05M, 05F, 05J E2195LN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-76...
  • Page 238 Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2001) Figure 10-61 E2195A xDSL / POTS 05M, 05F, 05J Switched © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-77...
  • Page 239: Series 3 Power Requirements And Connections (Prior To May 2000)

    Requirements and Introduction, 10-78 This remaining sections of this chapter describe the Connections (Prior electrical power requirements for 3070 Series 3 or Series 3 / 79000 Power Requirements, 10-80 79000 family board test systems which have serial to May 2000)
  • Page 240 Table 10-7 Serial numbers of Series 3 / 79000 systems shipped with E1135B PDUs. System Type 79000 327X 317X 307X Testhead Product E2190A E9997A E9998A E9900A Serial Numbers US38050108 and US38240114 and US38240138 and US38240113 and between US38050100 US38240100 US38240100 US33240100 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-79...
  • Page 241: Series 3 / 79000 Power Requirements

    See the 3070 / 79000 Repair Manuals , Chapter 1B for the locations where instruments are plugged in for load balancing for Series II and Series 3. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-80...
  • Page 242 Part number for cord and plug - E1135-61611 for North America and E1135-61610 internationally. Cord is Hard Service (SO) Multi-Strand AWG 12/5 Mating receptacle is a NEMA# L21-30 208Plug-20A m p.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-81...
  • Page 243: Power Recommentations

    20 amps at 100–120 volts or be quickly removed in case of an emergency. The 10 amps at 200–240 volts, whichever is E1135C PDU, used with 3070 Series 3 and 79000 applicable. systems, contains a mains disconnect. This input...
  • Page 244 Sizing The Input Wires And Circuit on page 10-89 and Breakers Connecting Power To The Testhead on page 10-90 to determine the SO Cord size and breaker size. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-83...
  • Page 245 [23] 18 [42] 28 AWY, 16 [15] 14 [23] 18 [42] 28 0E3, 16 [15] 14 [23] 18 [42] 28 03G, 16 [15] 14 [23] 18 [42] 28 0EH, 16 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-84...
  • Page 246 If 3-phase Delta power (0ED) is not available, use the Single-Phase power option (AWY). 307X single-phase option AWY systems will require two power drops. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-85...
  • Page 247 [37] 22 05J, 14 Single-Phase Earthed [22] 20 [36] 24 [42] 28 [24] 15 [22] 20 [36] 24 [42] 28 [24] 15 [22] 20 [36] 24 [42] 28 [24] 15 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-86...
  • Page 248 [43] 24 [42] 28 [24] 15 1. Calculated with a fully-loaded 3X7X system with 3 STC cards in every module. 2. DUT Power supplies are configured for 200-240 volts ac. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-87...
  • Page 249 4. POTS and xDSL bays always require a separate power drop. Use the POTS/xDSL power options that is the same line as the system power option. 5. Values in brackets [] apply to Series 3 systems with serial numbers less than those listed in Table 10-6 on page 10-79. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-88...
  • Page 250: Sizing The Input Wires And Circuit Breakers

    CAUTION ✸ All power configurations using a neutral are unbalanced, so there will be neutral current. Therefore, the neutral conductor must be the same size as the phase conductors. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-89...
  • Page 251: Connecting Power To The Testhead

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000) Connecting Power To The Testhead WARNING ✺ The power options available for Agilent 3070 Series 3 / Hazardous voltages, capable of causing injury or 79000 with E1135C PDUs are shown in Table 10-10 death, are present inside the PDU as long as page 10-91.
  • Page 252 10-102 230 / 400 Table 10-13 on page 10-96 Figure 10-63 on page 10-102 240 / 415 Table 10-13 on page 10-96 Figure 10-63 on page 10-102 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-91...
  • Page 253 10-102 230 / 400 Table 10-15 on page 10-98 Figure 10-63 on page 10-102 240 / 415 Table 10-15 on page 10-98 Figure 10-63 on page 10-102 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-92...
  • Page 254 Table 10-17 on page 10-100 Figure 10-68 on page 10-108 Table 10-17 on page 10-100 Figure 10-68 on page 10-108 Table 10-17 on page 10-100 Figure 10-68 on page 10-108 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-93...
  • Page 255 10-106 115 / 230 Table 10-17 on page 10-100 Figure 10-66 on page 10-106 120 / 240 Table 10-17 on page 10-100 Figure 10-66 on page 10-106 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-94...
  • Page 256 1. Breaker size based on 125 percent of maximum rms current 2. Wire size based on multiconductor cables * Three current-carrying conductors ** Four current-carrying conductors *** Pigtail required (Appendix 10, Connecting a Pigtail to the E1135C PDU © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-95...
  • Page 257 [4 (12)] 4(12)** [4 (12)] 4(12)** [10 (8)] 6(10)** 1. Breaker size based on 125 percent of maximum rms current 2. Wire size based on multiconductor cables ** Four current-carrying conductors © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-96...
  • Page 258 1. Breaker size based on 125 percent of maximum rms current 2. Wire size based on multiconductor cables * Three current-carrying conductors ** Four current-carrying conductors *** Pigtail required (Appendix 10, Connecting a Pigtail to the E1135C PDU © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-97...
  • Page 259 1. Breaker size based on 125 percent of maximum rms current 2. Wire size based on multiconductor cables * Three current-carrying conductors *** Pigtail required (Chapter 10, Connecting a Pigtail to the E1135C PDU © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-98...
  • Page 260 1. Breaker size based on 125 percent of maximum rms current 2. Wire size based on multiconductor cables * Three current-carrying conductors *** Pigtail required (Appendix 10, Connecting a Pigtail to the E1135C PDU © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-99...
  • Page 261 1. Breaker size based on 125 percent of maximum rms current 2. Wire size based on multiconductor cables * Three current-carrying conductors *** Pigtail required (Chapter 10, Connecting a Pigtail to the E1135C PDU © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-100...
  • Page 262: Mains" Wiring Diagrams

    See the E1135C Power Distribution Unit Operating and Service Manual E1135-90001 for more information. NOTE Set the Voltage Selector switch (S1) on the Control Board to ’230V’ for all voltage options. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-101...
  • Page 263: 220/380-240/415 V 3-Phase Wye With Neutral (0E5, 05L, 0E7)

    Figure 10-63 220/380-240/415 3-phase Wye with neutral (0E5, 05L, 0E7), 220/380-240/415 Single-phase Wye with neutral (05B, 05G, 05K) 10A 10A Connect Input Line here Note: L3 is available only on 3-phase power © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-102...
  • Page 264: 120/208-127/220 V 3-Phase Wye With Neutral (Aww, 05E)

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000) 120/208–127/220 V 3-Phase Wye with Neutral (AWW, 05E) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-103...
  • Page 265: 120/208-127/220 V Single-Phase Wye With Neutral (Awv, Awz)

    (AWV, AWZ) 10A 10A Connect Input Line here Note: Install jumper between T2 and T3 for Single-Phase option for E1135C. * Neutral is required only for POTS and xDSL Bay © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-104...
  • Page 266: 3-Phase Delta (0Ed, 05C, 05H, 0E6)

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000) 200-240 3-phase Delta (0ED, 05C, 05H, 0E6) Figure 10-65 200-240 3-phase delta (0ED, 05C, 05H, 0E6) 10A 10A Connect Input Line here © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-105...
  • Page 267: 100/200-120/240 V Single-Phase Center-Tap Neutral (Awy, 0E3, 0Eg, 0Eh)

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000) 100/200-120/240 V Single-Phase Center-Tap Neutral (AWY, 0E3, 0EG, 0EH) Figure 10-66 100/200-120/240 V Single-phase center-tap neutral (AWY, 0E3, 0EG, 0EH) 10A 10A Connect Input Line here Jumper L2-L3 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-106...
  • Page 268: Single-Phase Earthed (05M, 05F, 05J)

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000) 220–240 V Single-Phase Earthed (05M, 05F, 05J) Figure 10-67 220–240 V Single-phase earthed (05M, 05F, 05J) 10A 10A Connect Input Line here © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-107...
  • Page 269: Single-Phase Non-Earthed (Awx, 0Eb, 0Ej, 0Ec)

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000) 200–240 V Single-Phase Non-Earthed (AWX, 0EB, 0EJ, 0EC) Figure 10-68 200–240 V Single-phase non-earthed (AWX, 0EB, 0EJ, 0EC) 10A 10A Connect Input Line here Jumper L2-L3 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-108...
  • Page 270: Tester Power Requirements (Series I/Ii)

    Requirements, 10-124 NOTE E1131A PDU Electrical Requirements, 10-125 You cannot accurately measure the input current Agilent 3070 Series II / 3 - E1135A/B Power of the system’s power supplies using conventional Distribution Units, 10-128 50/60-hertz current probes because the input current waveforms are rich in harmonics.
  • Page 271 10-163 Shipped after 9/4/98: E1135C PDU See Chapter 6, Site Preparation – Power Requirements POTS Bay: Shipped before 7/7/95: has an additional E1131A PDU Figure 10-75 on page 10-155 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-110...
  • Page 272 10-163 E1085A POTS Bay or E2195A xDSL Bay See Chapter 6, Site Preparation – Power Requirements or the E2195A xDSL and E1085A POTS Installation and Upgrade Manual, E1085-90000 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-111...
  • Page 273: Connecting A Pigtail To The E1135C Pdu

    1 It is difficult to fit conductor size 16 square millimeters (AWG 6) into the mains disconnect switch, and 25 square millimeters (AWG 4) will not fit. With an 450 mm (18-inches) to 690 mm (24-inches length pigtail, 10 square millimeters (#8 AWG) will safely carry the 3070 load) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-112...
  • Page 274: 44964A/B Pdu Electrical Requirements

    24 A 2 modules 36 A 3 modules 42 A 4 modules 50 A 1 See the 44964B Power Distribution Unit Installation and Service Manual , 03066-90032) for more information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-113...
  • Page 275: E1099A Pdu Electrical Requirements

    120/240 V (218–266 V rms) 1 phase neutral center tap Maximum Total Current: 1 module 41 A 2 modules 60 A 3 modules 77 A 4 modules 96 A © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-114...
  • Page 276 Note how the devices are allocated to different outlets to try to balance the currents of the center-tapped phase; as in this example, these currents sometimes can’t be balanced exactly. The total current required is © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-115...
  • Page 277 Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II) 36.2 amps (27.2 amps for Phase A to B, plus 9 amps for the larger of phase B to N or phase A to N currents. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-116...
  • Page 278 DUT Power #3 DUT Power #4 Aux. Equipment Controller/Peripherals 40.0 * Total current: 27.2 + 9 = 36.2 Phase A - B Phase B - N Phase A - N © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-117...
  • Page 279 +5 V STC P.S. #3 +5 V STC P.S. #4 DUT Power #1 DUT Power #2 DUT Power #3 DUT Power #4 Aux. Equipment Controller/Peripherals 20.0 * Total current: © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-118...
  • Page 280 +5 V STC P.S. #3 +5 V STC P.S. #4 DUT Power #1 DUT Power #2 DUT Power #3 DUT Power #4 Aux. Equipment Controller/Peripherals 20.0 * Total current: © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-119...
  • Page 281 DUT Power #1 DUT Power #2 DUT Power #3 DUT Power #4 Aux. Equipment Controller/Peripherals 40.0 * Total current: Phase A - B Phase B - N Phase A - N © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-120...
  • Page 282 DUT Power #1 DUT Power #2 DUT Power #3 DUT Power #4 Aux. Equipment Controller/Peripherals 40.0 * Total current: Phase A - B Phase B - N Phase A - N © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-121...
  • Page 283 DUT Power #1 DUT Power #2 DUT Power #3 DUT Power #4 Aux. Equipment Controller/Peripherals 40.0 * Total current: Phase A - B Phase B - N Phase A - N © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-122...
  • Page 284 DUT Power #1 DUT Power #2 DUT Power #3 DUT Power #4 Aux. Equipment Controller/Peripherals 40.0 * Total current: Ph B-C Ph C-A Ph B - N Ph A - N © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-123...
  • Page 285: E1170-80003 Pdk Electrical Requirements

    1 See the E1099A Power Distribution Unit Installation and Service Manual , E1099-90000, for more information. 2 The E1170-80003 PDK was only used with 3170X systems, not 307X systems. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-124...
  • Page 286: E1131A Pdu Electrical Requirements

    The voltage drop to the load must not exceed 5-percent of nominal for Single-phase and 3-phase power. With 3-phase options only two phases are loaded and the third is a without a load. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-125...
  • Page 287 3070 system does not exceed 5-percent of the nominal voltage. Table 10-29 3070 Series I, E1131A power requirements for all power options 317X 307X...
  • Page 288 Support Bay (PDU #2) Full Load Amps (FLA) Breaker Size in Amps SO Cord Wire Size mm2 (American Wire Gauge - AWG) 6 (10) 10 (8) 4 (12) 4 (12) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-127...
  • Page 289: Agilent 3070 Series Ii / 3 - E1135A/B Power Distribution Units

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II) Agilent 3070 Series II / 3 - E1135A/B Power requirements of the system (see Table 10-31 Distribution Units page 10-130and on page 10-131). Table 10-32...
  • Page 290 50 percent less than actual. Also, current demand The power requirements for 3070 Series II family depends on the tests that are running on the systems with E1135A/B PDUs are shown in Table system.
  • Page 291 Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II) Table 10-31 3070 Series II family power requirements with E1135A/B PDU System Power Option Voltage Full-Load Amps (FLA) for: 327X 317X 307X xDSL/POTS Testhead Support Bay...
  • Page 292 4. POTS and xDSL bays always require a separate power drop. Use the POTS/xDSL power options that is in the column below on the same line as the system power option. Table 10-32 3070 Series II family power requirements with E1135A/B PDU System...
  • Page 293 Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II) Table 10-32 3070 Series II family power requirements with E1135A/B PDU (continued) System Power Option Voltage Full-Load Amps (FLA) for: Single-Phase Earthed © Agilent Technologies 1989–2002...
  • Page 294 Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II) Table 10-32 3070 Series II family power requirements with E1135A/B PDU (continued) System Power Option Voltage Full-Load Amps (FLA) for: 327X 317X 307X xDSL/POTS Testhead...
  • Page 295 4. POTS and xDSL bays always require a separate power drop. Use the POTS/xDSL power options that is in the column below on the same line as the system power option. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 296 CAUTION ✸ All power configurations using a neutral are unbalanced, so there will be neutral current. Therefore, make the neutral conductor the same size as the line conductors. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-135...
  • Page 297 10-139 230 / 400 Table 10-36 on page 10-139 240 / 415 Table 10-36 on page 10-139 2-Phase Open Delta 100/200 Table 10-37 on page 10-140 w/CT Neutral © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-136...
  • Page 298 Table 10-38 on page 10-140 240 / 415 Table 10-38 on page 10-140 Single-Phase Earthed Table 10-40 on page 10-141 Table 10-40 on page 10-141 Table 10-40 on page 10-141 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-137...
  • Page 299 Table 10-39 on page 10-141 Neutral 110 / 220 Table 10-39 on page 10-141 115 / 230 Table 10-39 on page 10-141 120 / 240 Table 10-39 on page 10-141 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-138...
  • Page 300 Table 10-36 3-phase wye with neutral (0E5, 05L, 0E7) System Type 327X 317X 307X Breaker Size in Amps SO Cord Conductor Size mm2 (American Wire Gauge - AWG) 4 (12) 4 (12) 6 (10) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-139...
  • Page 301 Table 10-38 Single-phase wye w/neutral (05B, 05G, 05K) System Type 327X 317X 307X Breaker Size in Amps SO Cord Conductor Size mm2 (American Wire Gauge - AWG) 4 (12) 6 (10) 6 (10) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-140...
  • Page 302 317X 307X Testhead Support Bay Breaker Size in Amps SO Cord Conductor Size mm2 (American Wire Gauge - AWG) 4 (12) 6 (10) 10 (8) 10 (8) 6 (10) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-141...
  • Page 303 317X 307X Testhead Support Bay Breaker Size in Amps SO Cord Conductor Size mm2 (American Wire Gauge - AWG) 4 (12) 6 (10) 10 (8) 10 (8) 6 (10) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-142...
  • Page 304: Connecting Mains Power To The System

    The electrician will not be able to measure true current to the system because power consumption is dependent on the testplan, and the crest factor of switching power supplies causes a non-sinusoidal waveform. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-143...
  • Page 305: 44964A/B Ac Mains Connection

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System 44964A/B AC Mains Connection Figure 10-69 44964A/B AC mains connection 1 See the 44964B Power Distribution Installation and Service Manual , 03066-90032, for more information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-144...
  • Page 306: E1099A Ac Mains Connection

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System E1099A AC Mains Connection Figure 10-70 E1099A ac mains connection 1 See the E1099A Power Distribution Installation and Service Manual , E1099-90000, for more information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-145...
  • Page 307: Connecting Power To The E1170-80003 Pdk

    Rotate the testhead, if necessary, to access the top of the PDK. Using a T20 Torx driver, remove 6 screws to access the mains terminals as shown in Figure 10-71 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-146...
  • Page 308 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System Figure 10-71 Accessing the mains terminals in the E1170-80003 PDK © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-147...
  • Page 309 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System Using a 1/4-inch flat blade screwdriver, connect the mains conductors to the mains terminals as shown in Figure 10-72 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-148...
  • Page 310 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System Figure 10-72 Wiring the mains to the E1170-80003 PDK © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-149...
  • Page 311: Connecting Power To The E1131A Pdu

    Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System Connecting Power to the E1131A PDU Figure 10-73 shows the location of PDU #1 and PDU #2 in a 307X support bay. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-150...
  • Page 312 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System Figure 10-73 Locating E1131A PDU #1 and PDU #2 (Branch) (Main) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-151...
  • Page 313 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System Using a T20 Torx driver, remove 5 screws to access the mains terminals as shown in Figure 10-74 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-152...
  • Page 314 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System Figure 10-74 Accessing the mains terminals in the E1131A PDU © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-153...
  • Page 315 Using a 1/4-inch flat blade screwdriver, connect the mains conductors to the mains terminals as shown in Figure 10-75 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-154...
  • Page 316 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System Figure 10-75 Wiring the mains to the E1131A PDU © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-155...
  • Page 317 The voltage selector switch (S1) inside the PDU DOES NOT set the output voltage of the PDU. It sets the PDU’s remote shutdown circuitry for the appropriate input voltage range: 100–127 volts or 200–240 volts. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-156...
  • Page 318: Connecting Power To The E1135A/B Pdu

    E1131A or E1135A/B/C PDUs; older PDUs do not have this capability. For wiring instructions, see the E1135C Power Distribution Unit Operating and Service Manual , E1135-90001, section The Branch Control Function . © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-157...
  • Page 319 See the E1135C Power Distribution Unit Manual , E1135-90001, for the switch location. NOTE For all of the power options set the Voltage Selector switch (S1) on the Control Board to ‘230V’. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-158...
  • Page 320 Figure 10-76 220/380–240/415 V 3-phase wye with neutral (0E5, 05L, 0E7), 220/380–240/415 V single-phase wye with neutral (05B, 05G, 05K) Connect Input Line here Note: L3 is available only on 3-phase power © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-159...
  • Page 321 200V 2-Phase Open Delta with Phase Center-Tap Neutral (05A) Figure 10-77 208 or 220 V 3-phase wye with neutral (0EF, AWW, 05E), 200V 2-phase open delta with phase center-tap neutral (05A) Connect Input Line here © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-160...
  • Page 322 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System 200–240V Single-Phase Center-Tap Neutral (AWY, 0E3, 0EG, 0EH) Figure 10-78 200–240V single-phase center-tap neutral (AWY, 0E3, 0EG, 0EH) Connect Input Line here © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-161...
  • Page 323 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System 220–240V Single-Phase Earthed (05M, 05F, 05J) Figure 10-79 220–240V single-phase earthed (05M, 05F, 05J) Connect Input Line here © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-162...
  • Page 324 Chapter 10: Site Preparation – Power Requirements for Older Systems: Connecting Mains Power To The System 200–240V Single-Phase Non-Earthed (AWX, 0EB, 0EJ, 0EC) Figure 10-80 200–240V single-phase non-earthed (AWX, 0EB, 0EJ, 0EC) Connect Input Line here © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-163...
  • Page 325: Site Preparation - Series 3 / 79000 Current Reduction Conversion

    Site Preparation – Series 3 / 79000 Current Reduction Conversion 03066-90114 Rev. Z 06/2002 In this Chapter... Introduction, 11-2 Series 3 Current Reduction Conversion Procedure, 11-4 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-1...
  • Page 326: Introduction

    PDUs, MPUs and Power Options on page 6-41 for a list of important changes concerning PDUs, MPUs and power options which have occurred with 3070 Series 3 / 79000 products. Table 11-1 Series 3 / 79000 current requirements reduction serial numbers...
  • Page 327: Who Should Do The Current Reduction Conversion

    For most problems, contact Agilent’s Measurement System Knowledge Center (MSKC). In the U.S. call 1-800-593-6635 with your MSKC support handle. Outside the U.S. call your local Agilent Service Representative to access the MSKC. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-3...
  • Page 328: Series 3 Current Reduction Conversion Procedure

    Reconnect the Loads Reconnect the Loads, 11-84 Verify the System Verify the System, 11-84 Complete the Conversion Complete the Conversion, 11-84 Connecting a Pigtail to the E1135C PDU, 11-84 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-4...
  • Page 329: Determine The System Type And Power Option

    11-6, to identify the testhead product number which may be used to locate the correct tables in Chapter 6, Site Preparation – Power Requirements and throughout this procedure. 1 SO - Service with Oil-resistant jacket © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-5...
  • Page 330 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-1 Identifying the merchandising number Series3+79K.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-6...
  • Page 331 Mark out the old value and add the new value on the A line when the upgrade is complete. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-7...
  • Page 332: Shut Down The System

    Volts AC of Phases Full Load Amperes MainsLabel1.wpg Shut Down the System Execute Shut down the testhead power off. controller. 1 Log in as superuser. Type: shutdown -h 0 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-8...
  • Page 333: Set The Dut Power Supply's Input Voltage

    Only the electrician who installed the lock or the electrician’s supervisor (under certain circumstances) may remove the lock and turn the “mains” power switch on. 2 In some cases this may mean unplugging the system. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-9...
  • Page 334 3-Phase Wye w/ Neutral 120 / 208 220V 127 / 220 220V 220 / 380 220V 230 / 400 240V 240 / 415 240V Three-Phase Delta 220V 220V 240V 240V © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-10...
  • Page 335 Single-Phase Wye w/ 120 / 208 220V Neutral 127 / 220 220V 220 / 380 220V 230 / 400 240V 240 / 415 240V Single-Phase Earthed 220V 240V 240V © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-11...
  • Page 336 DUT Supply Option Line Voltage Single-Phase 220V Non-Earthed 220V 240V 240V Single-Phase Center-Tap 100 / 200 220V Neutral 110 / 220 220V 115 / 230 240V 120 / 240 240V © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-12...
  • Page 337 The desired line voltage must be visible through the PULL to the left and remove the fuse (see Figure module window when the board is installed. 11-4 on page 11-14). © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-13...
  • Page 338 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-4 6621A and 6624A voltage select PC board Line input power module Voltage Select PC Board Operating voltage is shown in module window 6628pow.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-14...
  • Page 339: Agilent 6634A Line Voltage Settings

    (S2 and S3) and the rear panel fuse (F1). To convert the supply from one line voltage to another, proceed as follows: © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-15...
  • Page 340 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-5 6634A line voltage selection GPIB Connector Terminal Block Line Receptacle Fuse (F1) 6632ps.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-16...
  • Page 341: 6634B Line Voltage Settings

    1 Turn off power to the supply and disconnect the power cord from the power source. 2 Remove the four screws that secure the two carrying straps and outer cover. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-17...
  • Page 342: Agilent 6642A Line Voltage Settings

    The 6642A supply can be set to accept 100, 120, 220 or 2 Remove the four screws that secure the two carrying 240 volts ac input by setting the line voltage select straps and outer cover. switches. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-18...
  • Page 343 Table 11-6 6642A Fuse Requirements Line Voltage Fuse Needed Part Number 100 V 6 AF/250V 2110-0056 120 V 5 AF/250V 2110-0010 220 / 230 / 340 V 3 AF/250V 2110-0003 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-19...
  • Page 344: Rewire The Inputs To The Outlets

    This will be done after the voltages connected to the outlets are tested. Use the tables to verify the voltage from outlets or that devices are plugged into the correct outlets. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-20...
  • Page 345 11-29 on page 11-40 on page 11-51 on page 11-62 POTS or 1-PDU: E2195A Figure 11-41 xDSL page 11-62 Figure 11-58 2-PDU: page 11-80 Figure 11-51 page 11-72 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-21...
  • Page 346 Figure 11-22 Figure 11-33 Figure 11-45 with testhead Figure on page 11-32 on page 11-43 on page 11-54 on page 11-66, 11-61 Figure 11-47 page 11-83 on page 11-68 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-22...
  • Page 347 Figure 11-9 Figure 11-20 Figure 11-31 Figure 11-42 with testhead Figure page 11-30 on page 11-41 on page 11-52 on page 11-63, 11-61 Figure 11-44 page 11-83 on page 11-65 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-23...
  • Page 348 11-29 on page 11-40 on page 11-51 on page 11-62 E2195A xDSL 1-PDU Figure 11-41 Figure 11-58 page 11-62 on page 11-80 2-PDU Figure 11-51 page 11-72 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-24...
  • Page 349 Figure 11-15 Figure 11-26 Figure 11-37 Figure 11-48 with testhead Figure on page 11-36 on page 11-47 on page 11-58 11-61 page 11-69, page 11-83 Figure 11-50 on page 11-71 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-25...
  • Page 350 Figure 11-17 Figure 11-28 Figure 11-39 Figure 11-55 Figure 11-57 Figure 11-61 on page 11-38 on page 11-49 on page 11-60 on page 11-76 on page 11-78 on page 11-83 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-26...
  • Page 351 Figure 11-13 Figure 11-24 Figure 11-35 Figure 11-52 Figure 11-52 Figure on page 11-34 on page 11-45 on page 11-56 on page 11-73 on page 11-73 11-60 page 11-82 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-27...
  • Page 352 Figure 11-13 Figure 11-24 Figure 11-35 Figure 11-52 Figure 11-52 Figure 11-60 on page 11-34 on page 11-45 on page 11-56 on page 11-73 on page 11-73 on page 11-82 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-28...
  • Page 353: 327X 1-Module System Ac Outlets

    Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure 327X 1-Module System AC Outlets Figure 11-8 327X 1-module system AC outlets Outlet 2 Outlet 3 Outlet 1 Testhead (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-29...
  • Page 354 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-9 327X 0ED, AWW, 05E, 05C, 05H 0E6 CHECK COUNTRY CODE FOR FREQUENCY 327xLLL.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-30...
  • Page 355 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-10 327X 0ED, AWW, 05E, 05C, 05H 0E6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-31...
  • Page 356 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-11 327X 0E5, 05L, 0E7 327xLLLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-32...
  • Page 357 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-12 327X 0E5, 05L, 0E7 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-33...
  • Page 358 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-13 327X AWX, AWY, AWV,AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH 327xLL.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-34...
  • Page 359 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-14 327X AWX, AWY, AWV,AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 360 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-15 327X 05B, 05G, 05K 327xLLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-36...
  • Page 361 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-16 327X 05B, 05G, 05K © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-37...
  • Page 362 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-17 327X 05M, 05F, 05J 327xLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-38...
  • Page 363 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-18 327X 05M, 05F, 05J © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-39...
  • Page 364: 79000 1-Module System Ac Outlets

    Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure 79000 1-Module System AC Outlets Figure 11-19 79000 1-Module System Outlet 2 Outlet 1 Outlet 3 Testhead (Rear View) E2190_ol.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-40...
  • Page 365 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-20 79000 0ED, AWW, 05E, 05C, 05H, 0E6 Unswitched 79LLL.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-41...
  • Page 366 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-21 79000 0ED, AWW, 05E, 05C, 05H, 0E6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-42...
  • Page 367 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-22 79000 0E5, 05L, 0E7 Unswitched 79LLLN.wpg (Fused) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-43...
  • Page 368 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-23 79000 0E5, 05L, 0E7 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-44...
  • Page 369 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-24 79000 AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH 79KLL.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-45...
  • Page 370 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-25 79000 AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 371 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-26 79000 05B, 05G, 05K 79KLLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-47...
  • Page 372 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-27 79000 05B, 05G, 05K © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-48...
  • Page 373 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-28 79000 05M, 05F, 05J 79KLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-49...
  • Page 374 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-29 79000 05M, 05F, 05J © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-50...
  • Page 375: 317X 2-Module System Ac Outlets

    Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure 317X 2-Module System AC Outlets Figure 11-30 317X 2-module system Outlet 0 Outlet 2 Outlet 1 Outlet 3 Testhead (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-51...
  • Page 376 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-31 317X 0ED, AWW, 05E, 05C, 05H, 0E6 317xLLL.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-52...
  • Page 377 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-32 317X 0ED, AWW, 05E, 05C, 05H, 0E6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-53...
  • Page 378 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-33 317X 0E5, 05L, 0E7 317xLLLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-54...
  • Page 379 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-34 317X 0E5, 05L, 0E7 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-55...
  • Page 380 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-35 317X AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH 317xLL.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-56...
  • Page 381 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-36 317X AWX, AWY, AWV, AWZ, 0EB, 0EJ, 0EC, 0E3, 0EG, 0EH © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 382 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-37 317X 05B, 05G, 05K 317xLLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-58...
  • Page 383 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-38 317X 05B, 05G, 05K © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-59...
  • Page 384 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-39 317X 05M, 05F, 05J 317xLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-60...
  • Page 385 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-40 317X 05M, 05F, 05J © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-61...
  • Page 386: 307X 4-Module 1-Pdu System Ac Outlets

    Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure 307X 4-Module 1-PDU System AC Outlets Figure 11-41 307X 4-module 1-PDU system Outlet 0 Outlet 2 Outlet 1 Outlet 3 Support Bay Testhead (Rear View) (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-62...
  • Page 387 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-42 3070 1-PDU 0ED, AWW, 05E, 05C, 05H,0E6 3070LLL_1.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-63...
  • Page 388 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-43 3070 1-PDU 0ED, AWW, 05E, 05C, 05H,0E6 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-64...
  • Page 389 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-44 3070 1-PDU 0ED, AWW, 05E, 05C, 05H, 0E6 3070LLL_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-65...
  • Page 390 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-45 3070 1-PDU 0E5, 05L,0E7 3070LLLN_1.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-66...
  • Page 391 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-46 3070 1-PDU 0E5, 05L,0E7 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-67...
  • Page 392 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-47 3070 1-PDU 0E5, 05L, 0E7 3070LLLN_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-68...
  • Page 393 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-48 3070 1-PDU 05B, 05G, 05K 3070LLN_1.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-69...
  • Page 394 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-49 3070 1-PDU 05B, 05G, 05K © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-70...
  • Page 395 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-50 3070 1-PDU 05B, 05G, 05K 3070LLN_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-71...
  • Page 396: 307X 4-Module 2-Pdu System Ac Outlets

    Outlet 0 Outlet 2 See Table To Gnd Main See Table See Table To Gnd See Table Outlet 1 Outlet 3 Branch Support Bay Testhead (Rear View) (Rear View) © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-72...
  • Page 397 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-52 3070 2-PDU AWX, AWY, AWV, AWZ, 0EB, 0E3, 0EJ,0EG, 0EC, 0EH T1 T2 3070LL_1.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-73...
  • Page 398 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-53 3070 2-PDU AWX, AWY, AWV, AWZ, 0EB, 0E3, 0EJ,0EG, 0EC, 0EH © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation...
  • Page 399 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-54 3070 2-PDU AWX, AWY, AWV, AWZ, 0EB, 0E3, 0EJ, 0EG, 0EC, 0EH CHECK COUNTRY CODE FOR FREQUENCY 3070LL_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-75...
  • Page 400 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-55 3070 2-PDU 05M, 05F, 05J 3070LN_1.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-76...
  • Page 401 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-56 3070 2-PDU 05M, 05F, 05J © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-77...
  • Page 402 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-57 3070 2-PDU 05M, 05F, 05J 3070LN_2.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-78...
  • Page 403: E1085 Pots Bay And E2195 Xdsl Bay Ac Outlets

    E1085A POTS bay was modified by plugging the instruments in the POTS bay, except the Kikusui power supply, directly into the PDU. See the Telecom Theory and Repair Manual , Chapter 4, for additional information. © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-79...
  • Page 404 Figure 11-58 E1085A POTS / E2195A xDSL Bay with E1135C PDU Outlet 1 - Customer Instruments See Table To Gnd See Table See Table To Gnd See Table Line Mains Filter Input © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-80...
  • Page 405 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-59 E1085A POTS / E2195A xDSL AWW, 05E 120/127 Volt External Instrum ents E1085-E2195LLLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-81...
  • Page 406 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-60 E1085A POTS / E2195A xDSL AWY, AWV, AWZ, 0E3, 0EG, 0EH 100/127 Volt External Instruments E1085-E2195LLN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-82...
  • Page 407 Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure Figure 11-61 E1085A POTS / E2195A xDSL 05M, 05F, 05J 220/240 Volt External Instruments E1085-E2195LN.wpg © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-83...
  • Page 408: Reconnect The Loads

    (24-inches), called a pigtail be created using 10 square millimeters (#8 AWG) wire, and the recommended larger wire size cable be connected to the receptacle which connects to the pigtail, as shown in Figure 11-62 © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-84...
  • Page 409 Added P ow er Receptac le on a short pigtail if the E1135C PD U is used and the existing conductor size of the pow er drop exceeds 10 mm (#8 A WG). Pigtail.wp g © Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 11-85...
  • Page 410 Printed in USA 03066-90114 Rev. Z 06/2002...