Monarch Instrument’s Limited Warranty applies. Track-It is a trademark of Monarch Instrument. Android and Google Play are trademarks of Google, Inc. The Android robot is reproduced or modified from work created and shared by Google and used according to terms described in the Crea ve Commons 3.0 A ribu on License.
TABLE OF CONTENTS: INTRODUCTION ............1 GETTING STARTED ............2 2.1 Install USB Driver and Track-It So ware ....... 2 2.2 Connect Logger and Start Track-It So ware ....3 2.3 Configure Logger ............3 INSTALLATION............4 POWER / TRANSMITTER CONNECTIONS ...... 5 BUTTONS AND MENU ..........
INTRODUCTION The Track-It Pressure Logger with Display and Current Transmi er is a 24V DC powered water ght (IP65) compact data logger with an isolated 4 to 20mA current transmi er output for pressure. It can be configured to record both Pressure and Temperature up to 130,000 samples or Pressure alone, to maximize data storage space.
Install USB Driver and Track-It So ware IMPORTANT: Do NOT plug the Track-It Data Logger into the USB port un l AFTER the USB Driver and Track-It So ware have been installed. Unzip the downloaded file to a local directory on your PC.
Connect Logger and Start Track-It So ware Plug the Track-It logger into an available USB port on the PC using the supplied cable. If the Found New Hardware Wizard appears, follow the prompts to complete the installa on. Start the Track-It So ware by double-clicking the Track-It icon (shown right) on your PC desktop or from the Start menu.
Review default factory configura on se ngs and make any necessary changes—sample rates, triggers, alarms, etc. Be sure that the me and date is set properly. Click the “Send Config” bu on under the Device Configura on sec on of the window. Your data logger is now opera onal! INSTALLATION The logger can be ordered with various pressure sensor ranges to cover most applica ons.
POWER / TRANSMITTER CONNECTIONS The Logger has a 4 pole DIN43650 solenoid valve style connector on the rear panel for power current connec ons. A ma ng plug (Hirschmann GDS 307) and flat gasket (Hirschmann GDS 307-4) are supplied. The plug is a Form C connector with 3 Poles + Ground, 9.4 mm pin spacing, with a black fiberglass reinforced polyamide housing, and a PG 7 cable entry which accommodates a...
Page 9
To connect, remove the screw and plug base from the housing. Thread the cable through the PG7 gland (A), metal washer (B) and rubber washer (C). Insert the wires into the plug housing strip wire ends ~0.24” (6mm). plug four compression screw connec ons.
10 seconds. The content of each menu is set up in the Track-It PC So ware. In addi on, the admin menu can be locked in which case the user at 1.800.561.8187...
Setup. Menu Op ons (In alphabe cal order – the actual order in which these items appear is dependent on se ngs in the Track-It PC So ware) Back Light BKLT – AUtO/OFF – In Auto Mode the backlight will come on whenever the display is on. Press bu on to toggle auto/off...
Page 12
Record On/Off rEC – ON/OFF – Will toggle the record mode of the unit (Only visible if record mode set to “Bu on Press” in the Track-It PC So ware). Press bu on to toggle on/off status. Zero Mode ZERO – OFF/EnAb – Will enable the zero offset feature.
AM or PM – 08:30 AM), and memory usage as a % of available memory. Values shown are set in the display setup using the Track-It PC So ware. See sec on 8.4 Display Setup.
Page 14
There are various icon indica ons on the display that indicate the following: Indicates the unit is currently recording. This icon will blink at a 2 second rate. Indicates the unit is currently triggered and wai ng to record. Record is ini ated by me delay, alarm or bu on depending on configura on.
You must first download and install the USB Drivers and PC So ware. Please refer to sec on 2.0 Ge ng Started. The free Track-It So ware runs on a PC and gives the user complete control in programming the logger and allows for the upload, examina on and archiving of data recorded on the logger and se ng the current output.
® Export user selected data in CSV formats for import into Excel . Filter data to be exported. Scale output current to user specified pressure range There are a number of unique op ons specifically for this Display Pressure Logger with Current Transmi er which are covered in the following sec ons.
If the “Units from Logger” box is checked - the selec on of engineering units here is ignored and the engineering units used will be as set by default in the Logger. If the “Units from Logger” box is unchecked, then the unit selected here will override that set in the Logger.
the pressure sensor, as some values will be too large to display on the Logger. Note also that these units may be changed locally at the Logger if the menu is set up to allow this. See sec on 8.5 Menu Setup.
output. The logger will scale the output linearly to pressure between the two set points. If “Default” is checked the output will be 4mA at 0 pressure and 20mA at the full scale pressure of the device. The scaling cannot be changed in default mode.
The “Display List” column shows the available func ons that are available to be displayed on the Logger. The “LCD Display” column are the func ons that will be displayed on the Logger. The user can move the op ons between the columns by selec ng the op on (clicking on it) and using the arrow keys.
(display offset) will not be available in the Normal Menu. TRACK-IT TRANSPORT APP Track-It Transport is a free Applica on for Android that allows you to use your device for Android to start and stop recording and transfer data using a USB On-the-Go cable.
10.0 SPECIFICATIONS Note: Product specifica ons are subject to change without no ce. 10.1 Measurement Pressure Ranges: Various pressure ranges absolute and gauge. Check ordering informa on Accuracy: ±0.25% Full Scale typical (incl. linearity, repeatability and hysteresis) Resolu on: 0.01%. Display 4½ Digits. Max 2 decimal places.
Page 23
Alarms: 2 user programmable alarms. High or Low Communica on: Mini USB connec on So ware: Track-It So ware—Program device, view data (historic or real me), export to spreadsheet Ba ery: Internal rechargeable Li-Ion. Not user serviceable. Allow 24 hours to fully recharge when connected to power.
Page 24
Isolated Current Out: Range: 4 to 20mA (Min. 3.6mA, Max. 21.5mA) User scalable for pressure. Isola on: 500V to 24Vdc power Compliance Voltage: 12Vdc @24Vdc supply (Min @21.6Vdc supply) Accuracy: ±0.06% Resolu on: 6μA Update rate: 125 msec typical 2.94in 1.82in [74.6mm] [46.2mm]...
10.3 Opera ng Environment Temperature: -20 °C to +85 °C (-4 °F to 185 °F) Humidity: 100% Environmental: IP65. Note: USB connector cover must be in place. 10.4 Compliance 10.4.1 Ba ery Compliance The Lithium-Ion ba ery used in this product meets the requirements of UN DOT 38.3.
USB Cable for use with Android devices USB Extension Cable 3-foot USB Extension Cable Track-It So ware Track-It So ware on CD DIN46350 Ma ng Plug Hirshmann GDS 307 ma ng plug and GDS with Gasket 307-4 gasket for DC Power and Current Out connec on CAL-N.I.S.T...
Page 27
Check out our other product lines… Handheld Panel Portable Machine Vision Tachometers Tachometers Stroboscopes Stroboscopes Speed Sensors Temperature/ Vibra on Meters Humidity Sensors Paperless Recorders Track-It™ Data Loggers Printed in the U.S.A. Copyright 2016 Monarch Instrument, all rights reserved 1071-4396-210 -1216 1.800.561.8187 information@itm.com www. .com...