This manual must be given to the Owner/User of this bed and should be read carefully before putting this product into use. Keep manual in an accessible location for future reference. Drive Medical 99 Seaview Blvd Port Washington, NY 11050 Phone: 1.877.224.0946 Fax: 516.988.4601...
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 TABLE OF CONTENTS STANDARD SYMBOLS USED IN THIS MANUAL ................ 2 WARNING LABELS ........................3 INTENDED USE ........................... 4 CONTRA-INDICATIONS ......................4 GENERAL SAFETY ........................4 LEFT/RIGHT REFERENCE GRAPHIC ..................9 ELECTROMAGNETIC EMMISSION AND IMMUNITY .............. 10 FEATURES AND ACCESSORIES ....................
P1752 Owner's Manual| REV 002 April 21, 2015 STANDARD SYMBOLS USED IN THIS MANUAL This manual includes important information about the safety of personnel and equipment. As you read through this manual be aware of the symbols and their meanings. DANGER Information that appears under the DANGER symbol concerns the protection of personnel from direct and pending hazards that, if not avoided, will result in...
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 PRODUCT SYMBOLS Symbols Description Symbols Description According to IEC 60417-5840. Type B Change head deck position applied part equipment According to IEC 60529. Rating for dust- protection and identified as equipment Change thigh/foot deck position that is protected against splashing water.
P1752 Owner's Manual| REV 002 April 21, 2015 INTENDED USE This Primus Medical by Drive Bariatric bed is intended for use within an institutional healthcare environment, Application Environment 3 (i.e.: Skilled Nursing, Transitional Care, Rehabilitation Care, Assisted Living). Compliance with the regulations and guidelines as specified by your facility is recommended.
Page 6
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 DANGER SHOCK HAZARD This bed is equipped with a three-prong grounding plug for protection against possible shock hazard. DO NOT under any circumstances cut or remove the grounding prong. DO NOT open any actuators, control boxes or pendants. Service is only to be performed by authorized service personnel.
Page 7
P1752 Owner's Manual| REV 002 April 21, 2015 WARNING DO NOT operate this product without first reading and understanding this user manual. Damage or injury may result from improper use of this product. Possible injury or Death may occur if accessories are not provided by Primus Medical by Drive.
Page 8
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 CAUTION This medical bed is not intended to work with patient lifts other than those specified by Primus Medical by Drive as being compatible. DO NOT use assist rails as handles for moving the bed. ...
Page 9
The FDA Guidance document identifies specific dimensional criteria on potentially injury-threatening gaps and spaces that can occur between bed system components, such as side rails when improperly installed. Drive Medical’s Long Term Care beds and approved accessories are manufactured to be in conformance with these guidelines.
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 MATTRESS SPECIFICATIONS WARNING Possible ENTRAPMENT Hazard may occur if you DO NOT use the recommended specification mattress. Resident entrapment may occur leading to injury or death. A mattress may not be included with this bed. Mattresses sized to the dimensions of the mattress support platform must be used.
P1752 Owner's Manual| REV 002 April 21, 2015 ELECTROMAGNETIC EMMISSION AND IMMUNITY This MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the table below. Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT WARNING...
Page 12
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 Recommended separation distances between portable and mobile RF communications equipment and the PrimePlus™P1752 The PrimePlus™P1752 is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the PrimePlus™P1752 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the PrimePlus™P1752 as recommended below, according to the maximum output power of the communications equipment.
Page 13
P1752 Owner's Manual| REV 002 April 21, 2015 Guidance and manufacturer's declaration: Electromagnetic Immunity The PrimePlus™P1752 is intended for use in the electromagnetic environment specified below. The customer or the user of the PrimePlus™P1752 should ensure that it is used in such an environment. Immunity test IEC 60601 test Compliance...
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 FEATURES AND ACCESSORIES Standard Features Accessories and Options Travel Range: 16.3”‐29” Pressure Reduction or Low‐Air‐Loss Mattress Width Extension: 37”, 39”,42”,48” Extra Padded Assist Rails ǂǂ Locking System: 5” Locking Casters Fall Mat Left and Right side Pendant Connections Battery Back‐Up Slat Deck...
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 COMPLIANCE INFORMATION Matching the correct bed components to meet regulatory specifications can be complicated. Primus Medical by Drive offers a wide variety of compliance options. Primus Medical by Drive can assist your facility in selecting components or accessories that are recommended for the specific bed model.
P1752 Owner's Manual| REV 002 April 21, 2015 Remove the mattress retainers from the bed frame. They are wrapped in foam and zip tied to the frame. Remove any remaining zip ties or foam left on bed frame. ...
Page 18
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 Install the mattress retainers into 4 locations on the sides of the bed in the appropriate holes for the width mattress being used. Width options are detailed in the technical specification section. HEADBOARD AND FOOTBOARD INSTALLATION AND ADJUSTMENT TOOLS REQUIRED ...
Page 19
P1752 Owner's Manual| REV 002 April 21, 2015 Place the headboard/footboard mounting brackets into the slots of the bed frame support assembly. NOTE: The mounting bracket on the foot board should have the push/pull bar facing away from the bed. ...
Page 20
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 Place assist rail on side of bed sliding one bracket into the locating bale making sure the other bracket goes over the locating sticker. The holes on the bracket of the assist rail should be below mattress deck. ...
Page 21
42” Width Setting 48” Width Setting NOTE: This bed can accommodate 39”, 42” and 48” mattresses. The appropriate mattress MUST be used, such as a Drive Medical Pressure Reduction Mattress for each width setting. Page 20 The information in this document is subject to change without notice.
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 LENGTH EXTENSION (80” and 84”) TOOLS REQUIRED 3/16” Allen wrench Remove the two 5/16” bolt holding the headboard support bracket in position. Adjust the headboard support bracket to the position required for the mattress length being used.
Page 23
P1752 Owner's Manual| REV 002 April 21, 2015 WALL GUARD (No tools required) NOTE: To install wall guard locate the mount on the cross bar under the head of the bed. Place the Wall Guard into the holes of the mount underneath the bed. Place pins on each side of the bar into the holes to keep the Wall Guard in place.
Page 24
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 Slide the battery into place; make sure it is positioned securely into the mounting bracket. Attach the cord from the battery into the control box (black color). The battery back-up will now work if the bed becomes unplugged from the wall outlet or there is a power failure.
Page 25
P1752 Owner's Manual| REV 002 April 21, 2015 CAUTION If improperly performed, equipment or property DAMAGE or resident INJURY is possible during maintenance. The battery is not a toy for children to play with. Maintenance Notes: Following longer periods of storage, the battery may need to be charged and discharged several times so that performance reaches a peak.
Page 26
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 BED MOUNTED TRAPEZE TOOLS REQUIRED: 7/32” Hex Key 9/16” Hex Key 13/16” Wrench or socket 9/16” Wrench or socket (C) Cross Bar (A) Vertical Trapeze arms (2) Vertical Trapeze arms (2) Mounting...
P1752 Owner's Manual| REV 002 April 21, 2015 FREQUENTLY USED BED FUNCTIONS 4 Function Hand Pendant This pendant can be plugged into the left or right side of the bed for the convenience of the resident. Hand held pendant can be plugged into the left or right side of the bed frame for the convenience of the resident.
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 BED MOBILIZATION AND STABILIZATION WARNING Unintended bed movement may occur if the caster locks are left unlocked. Unintended bed movement may lead to property DAMAGE or resident INJURY. Never leave a bed unattended while the caster locks are not engaged. Caster Lock To lock the bed frame, push down on all 4 caster locks with your foot.
April 21, 2015 Contact Primus Medical by Drive if you need any help with installation, usage or maintenance of the product. Corporate Office: Drive Medical, 99 Seaview Blvd., Port Washington, NY 11050. Ph: 1-877-224-0946 Fax: 516-988-4601 Website: www.drivemedical.com Manufacturer: Land America Health & Fitness Co., Ltd Cleaning Instructions ...
Page 30
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 Control Box Check power cord for chafing, cuts or wear. Ensure all attachment hardware and brackets are tight. Check electrical connections for wear or fractures. Verify that all actuator connections are tight. ...
P1752 Owner's Manual| REV 002 April 21, 2015 ANNUAL INSPECTION Perform all Quarterly and Semi-Annual inspections plus: Mattress Support Surface, Frame and Base Assemblies Inspect welds on the mattress support surface, frame and base assemblies for stress fractures. ...
Page 32
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 Replacing Control Box Unplug mains cable from the wall outlet Unplug connections for all actuators and accessories Separate the control box from the actuator Reinstall the control box on the actuator ...
P1752 Owner's Manual| REV 002 April 21, 2015 Replace tie-wraps to hold power cord to frame Plug mains cable into wall outlet Test function WARNING When routing cables in the medical bed, precautions shall be taken to avoid squeezing those between parts of the medical bed.
Page 34
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 Unlock panel by Hi/Lo lockout may be The light on the attendant pushing lock control panel is on button and light will go off Actuators not working Disconnect power cord from bed that is not Faulty mains cable functioning and use on Contact Primus...
This warranty is extended only to the original owner who purchased this product new and unused from Drive Medical or a Drive Medical Authorized dealer/Distributor. Warranty is not extended and is not transferable or assignable to any subsequent purchaser or future owners.
Sales and Service Contact Drive Medical if you need any help with installation, usage or maintenance of the product. Corporate Office: Drive Medical, 99 Seaview Blvd., Port Washington, NY 11050.
P1752 Owner's Manual| REV 002 April 21, 2015 APPENDIX SPECIAL NOTE For your convenience, we have provided the April 2010edition of the FDA’s guide to bed safety. This information from the FDA’s brochure, published by Hospital Bed Safety Workgroup, is replicated verbatim; the latest version is available at http://www.fda.gov A Guide to Bed Safety Bed Rails in Hospitals, Nursing Homes and Home Health Care: The Facts...
Page 38
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 More serious injuries from falls when patients climb over rails. Skin bruising, cuts, and scrapes. Inducing agitated behavior when bed rails are used as a restraint. Feeling isolated or unnecessarily restricted. ...
Page 39
P1752 Owner's Manual| REV 002 April 21, 2015 For more information about this brochure, contact Beryl Goldman at 610-335-1280 or by e-mail at bgoldman@kendaloutreach.org. She has volunteered to answer questions. For information regarding a specific hospital bed, contact the bed manufacturer directly.
Page 40
P1752 Owner's Manual| REV 002.04.21.2015 April 21, 2015 Drive Medical, 99 Seaview Blvd., Port Washington, NY 11050 Ph: 1-877-224-0946 Fax: 516-988-4601 www.drivemedical.com Page 39 The information in this document is subject to change without notice.