Table of Contents Introduction…………………………………………………………………2 Technical Information ………………………………………………………3 Operating Instructions ……………………………………………………4 Guidance for treatment with Handy Cure s’ ………………………………5 Method of treatment ………………………………………………………5 Treatment Area ……………………………………………………………6 Laser Safety precautions…………………………………………………….7 General Safety…………………………………………………………….…8 Care and Cleaning …………………………………………………………8 Packing List …………………………………………………………………8 Troubleshooting ……………………………………………………………9 Declarations…………………………………………………………………9 Warnings …………………………………………………………………13 ………………………………………………………………………13 Warranty…………………………………………………………………14...
• Non-invasive treatment • Safe to use • Painless • Easy to use • No side effects • Compact Technical Information Average power: Total radiation: 60-90 mW Laser radiation: 0.4-6.25 mW Infrared radiation: 30-90 mW Red LEDs: 2-10 mW Permanent magnet induction: 25-45 mT Wavelength of: Laser radiation:...
Operating Instructions 1. Press the Start/Stop button to power the device. Handy Cure S’ will run a self-test* (3 lights will blink).When 2 lights will turn green Handy Cure S’ is ready for use. 2. Press the Mode button to select treatment program as indicated in the guidelines.
Guidance for treatment with Handy Cure s’ The user can choose between 3 programs which have been pre-set to treat different pain conditions. A program is selected by pushing the mode button. Each program uses different pulse frequencies, as displayed in the table below.
Treatment Area • Where possible, treatment should be provided over bare skin. • During treatment, the device should be held in a hovering state over the area where the pain is felt. • If the treatment area is wider, the device should be moved (scanned) around the treatment area.
Laser Safety precautions 1. Laser beam parameters: The product contains a class 3B near IR laser having following beam specifications. Wavelength 905 nm Pulse energy Pulse width 100 ns Average power Max 15 mW Beam Diameter 1.3x3.1 mm Beam divergence >...
We strongly recommend you use the device in strict compliance with the safety pre-cautions and operating instructions in this manual. Handy Cure S’ should be maintained and serviced by Medical Quant Limited personnel or other qualified personnel approved in writing by Medical Quant Limited.
Troubleshooting Falut Possible cause Solution Battery light is red Handy Cure S’ Connect Handy Cure S’ to rechargeable battery the electricity until the battery has run out of power light turns green Handy Cure S’ can’t be operated while recharging All 3 program Wrong mode of device Power the device as specified...
Page 10
Declaration – electromagnetic immunity IMMUNITY test IEC 60601 Compliance Electromagnetic environment test level level guidance Electrostatic 8 kV contact 8 kV contact Floors should be wood, concrete discharge (ESD) 2, 4, 8, 15kV 2, 4, 8, 15kV or ceramic tile. If floors are IEC 61000-4-2 covered with synthetic material, the relative humidity should be...
Page 11
Declaration – electromagnetic immunity IMMUNITY test IEC 60601 Compliance Electromagnetic environment test level level guidance Conducted Portable and mobile RF 3V, 6V 3Vrms, 6V communications equipment IEC 61000-4-6 should be used no closer to any part of the Handy Cure S’, including cables, than the Radiated RF recommended separation...
Page 12
Recommended separation distances between portable and mobile RF communications equipment and the Handy Cure S’ Rated maximum Separation distance according to frequency of transmitter m output power of 150 kHz to 80 MHz 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz transmitter outside ISM in ISM bands...
WARNINGS 1. “WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.” 2.
Please bring this warranty with the Handy Cure S’ in the event of a complaint Medical Quant Limited shall not be liable in any manner in respect of bodily injury and / or property damage arising from the use of Handy Cure S’ where such use is not in strict compliance with the instructions and safety precautions contained in the Handy Cure S’...
Symbols Symbol Symbol meaning Phrase Medical Quant Ltd, 14 Manufacturer Hata'as St. Kfar- Saba, Israel 4442514, Israel Obelis S.A.- Av. De Authorized Representative in the Tervueren 34 bte 44, European Community B-1040 Brussels, Belgium Symbol for "Caution, consult accompanying documents" or "Attention, see instructions for use."...
Page 16
Manufactured by Medical Quant Ltd. 14 Hataash St., Kfar-Saba 4425144, Israel info@medical-quant.com European Representative UM300720-EN.HC version Nov/2018 “Obelis S.A” Av. De Tervueren 34 1023 BTE 44, B-1040 Brussels Belgium MADE IN ISRAEL...
Need help?
Do you have a question about the Handy Cure S’ and is the answer not in the manual?
Questions and answers
I need the power cord adapter