Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
4.7.5 NIBP Buttons/Leds
The selections Auto ON/OFF, Manual ON/OFF, STAT ON/OFF, and Measur. ON/OFF have effect
only on the NIBP module, M-NIBP.
Buttons/Leds Data
For information on general items Pressure, Zero,
Protect handle, Calibr. prot., +15 V power, AD0 to
AD7 as well as Timeouts etc., see section
Module".
Buttons Checking
The front panel keys function is confirmed by
pressing and releasing the key and observing that
OFF turns to ON at Auto On/Off, and Start Cancel.
Set Cycle Time and STAT On/Off are only in use with
the NIBP module, M-NIBP.
36
.
Document no
M1137266-02
"4.7 NIBP