8.8 Restore Factory Settings
A reset to factory settings must only be carried out by technicians or qualified medical staff who
have been trained by the operator to do so.
As a result of this procedure, all individually programmed special reverse-Trendelenburg position,
normal position, washing position and mattress base height limits will be reset to the factory default
settings.
A reset to factory settings can be carried out on the LCD handset or on the control panel (outside)
or on the control box / locking box, depending on the equipment available on the bed.
8.8.1 On the LCD Handset
1 Press the UP and DOWN buttons simul-
taneously, and keep them pressed while
you hold the top end of the handset briefly
against the orange magnetic unlocking
key.
3 The OK ! shown on the display acknowl-
edges the change to the technician con-
trol level.
62
Instruction manual for Evario hospital bed
2
Keep both the UP and DOWN buttons
pressed for a further 5 seconds while the
warning triangle symbol is displayed.
4
Select the factory reset function by pressing
the toggle button. The factory reset symbol
will appear on the display.