Optimizing the Image

M/D Cursor

Description
Adjusting
Values
Benefits
Affect on other
controls

Angle Steer and Fine Angle Steer

Description
Adjusting
Values
Benefits
Bioeffects
5
102
-
Displays the Doppler Mode cursor on the B-Mode image.
To activate/deactivate the Doppler Mode cursor, press M/D
Cursor. Trackball to position sample volume graphic. Adjust SV
Length on the Touch Panel to adjust sample volume gate size.
The M/D Cursor key backlights.
On/Off. If blood flow velocities cannot be presented, the
message PRF unavailable. Change Vel Scale" appears.
Lets you position the cursor before you go into Doppler Mode so
that you can make optimum use of the larger B-Mode image.
If the M-Mode cursor is on, clicking SV Gate or Angle Correct
changes the cursor to a Doppler Mode cursor.
You can slant the ROI of the Color Flow linear image left or right
to get more information without moving the probe. The angle
steer function only applies to linear probes.
To slant the linear image to the left/right, adjust Angle Steer.
Press Angle Steer to access Fine Angle Steer.
Linear probes can be steered left (20 degrees), center, or right
(20 degrees).
Angle Steer values are returned to factory or user preset value
when you change: Probe, Exam Category, Exam Calcs, or New
Patient.
Fine Angle Steer steers the Doppler cursor left/right 10 or 20
degrees.
Provides a Doppler cursor angle suitable for linear probe
orientation. Beneficial in Vascular applications.
Activating angle steer may change the TI and/or MI. Observe
the output display for possible effects.
LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1