Quick Guide
How to measure on GE Senographe Pristina with Black Piranha
This Quick Guide refers to use of the Black Piranha on the GE Senographe Pristina.

1. Before starting

a.
The Black Piranha
after May 1, 2016.
b. The Ocean software version should be 2016.12.27.243 or higher.
c.
The GE Quick Check templates, version 1.1D, for Ocean must be installed.
d. Make sure to have the Piranha Pristina holder, GE part # 5758284, available (image
to right).
1
Piranha ordered under GE Part # 5758293 is Black, and must have a calibration for
mammography performed at RTI after May 1, 2016. Red Piranhas ordered under GE
Part # 5401610, or 5458262 must have been upgraded to Black Piranha with calibration after May 1, 2016.
If the Piranha has a calibration date from May 1, 2016 to March 9, 2017, it also must have been updated for correct CF
calibration using the IQST phantom (see internal GE service notes).

2. Measure using GE Quick Check Templates

Setup
e. Place the Black Piranha in the holder as shown in image (right).
The Black Piranha will only fit in one way.
f.
Slide the holder into the GE dosimeter support.
g.
The Black Piranha can be left in this positon for all measurements described below.
h. Follow the job card instructions when it comes to image detector protection, IQST
phantom set-up, compression paddle set-up, etc...
Measure
When starting Ocean the measure guide as shown in image (right) should appear. If
not, the option "Auto-start Quick Check" is deactivated. This function can be activated
under Options.
i.
Select GE Mammography...
j.
In next screen select GE Mammography[Senographe Pristina]...
k.
In next screen (image to right) select the proper application: Conversion
factor, HVL, or kVp.
9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx
v.2017.3D
1
must have a calibration for mammography performed at RTI
1(4)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Black Piranha and is the answer not in the manual?

Questions and answers

Summary of Contents for RTI Black Piranha

  • Page 1 RTI after May 1, 2016. Red Piranhas ordered under GE Part # 5401610, or 5458262 must have been upgraded to Black Piranha with calibration after May 1, 2016. If the Piranha has a calibration date from May 1, 2016 to March 9, 2017, it also must have been updated for correct CF calibration using the IQST phantom (see internal GE service notes).
  • Page 2 The template is built for the sequence to measure small and large focus in the same sequence, but can be used also for one at the time. The template can be used for one shot HVL as well as for traditional measurements by adding layers of aluminum sheets. 9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 2(4) v.2017.3D...
  • Page 3 Ocean. 3. Measure using GE Mammo Quick Check Setup n. Place the Black Piranha in the holder as shown in image (right) with the left pin inside the tripod mounting hole. The Black Piranha will only fit in one way.
  • Page 4 When ready for next test, just press the New Quick Check button. You shall now be back to the selection guide described above. For deeper understanding and guidance, please see User’s Manuals for Piranha and Ocean. *** END *** 9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 4(4) v.2017.3D...

This manual is also suitable for:

Ge senographe pristina

Table of Contents