    V2.0 Technician’s Manual TABLE OF CONTENTS bibag Details ........................2 General Warnings ....................... 3 Hydraulic Flow Diagram ....................4 Hydraulic Component Descriptions ..................5 Hydraulic Operation......................7 Electronic Description ......................9 Electronic Block Diagram....................11 Calibrations ........................12 Annual Maintenance ......................14 Troubleshooting ........................
  • Page 4: Bibag Details

    A door handle locks the bibag door in place over the bibag disposable. Figure 1 – bibag connector: door closed and bibag disposable inserted with door open Page 2 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 5: General Warnings

    Spills can cause slips and falls; clean up spills immediately. Caution: Only the bags manufactured by Fresenius Medical Care may be used in the bibag connector. Note: When the bibag connector is installed, the online pressure holding test becomes mandatory.
  • Page 6: Hydraulic Flow Diagram

    Hydraulic Flow Diagram Figure 2 Page 4 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 7: Hydraulic Component Descriptions

    113 – bibag Conductivity Cell The bibag conductivity cell is used to measure the conductivity of the sodium bicarbonate concentrate leaving the bibag disposable and the conductivity of the pre-mixed concentrates. Page 5 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 8 (not possible if a bag is hanging from the connector). The door must be in the closed position (3) for rinse, cleaning, and jug dialysis mode. Position (2) should not be used. bibag disposable Door Figure 3 Page 6 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 9: Hydraulic Operation

    Valves 104 and 108 alternate opening every 3 seconds. Valves 100 and 103 alternate opening every 3 seconds. Valve 101 is also opened periodically when conductivity is low. Valve 105 is open. Page 7 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 10 When the empty is complete, the operator is notified, normal balance chamber switching resumes, but the hydraulics remain in bypass until a new bag is installed and correct conductivity of the dialysate returns. Page 8 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 11: Electronic Description

    20 pin ribbon cable from the bibag hydraulics the bibag distribution box 2 ribbon cable plugs in here. plugs in here (see Figure 5) (see Figure 6). Figure 4 – bibag Interface Board Page 9 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 12 Cell (117) Acid Port Valve (105) Acid Pressure Transducer (106) 20 Pin Ribbon Cable Connection to the bibag Interface Board Figure 6 - bibag Distribution Box 2 – Distribution Board Page 10 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 13: Electronic Block Diagram

    Conductivity bibag Vent Valve Cell (101) (113) bibag Temp Hydrochamber Sensor Outlet Valve (114) (103) Air Sensing bibag Pressure Probes Transducer (112) (110) bibag Hydraulic Assembly P/N 190591 Figure 7 Page 11 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 14: Calibrations

    During the calibration process, the 90XL will communicate with the machine through the RS232 port. When the calibration process is complete, the screen will display Bicarb Cond Cell calibration is complete. Page 12 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 15 Any port is ok, if only the conductivity probe is connected to the 90XL. MESA SERIAL CABLE P/N 368402-10 Figure 8 – Null Modem Connection to the 90XL Page 13 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 16: Annual Maintenance

    Figure 9 – bibag Inlet Filter (M30225) bibag Connector Maintenance Annually replace the two (2) o-rings (P/N 640919) on the bibag connector door. Figure 10– bibag Connector Door O-Rings (640919) Page 14 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...
  • Page 17: Troubleshooting

    2008T Debug Screens P/N 490139 ™ 2008K@home Debug Screens P/N 490084 Spare Parts A list of bibag spare parts can be found in the 2008 Series Hemodialysis Spare Parts Manual P/N 490124. Page 15 bibag V2.0 Technician’s Manual P/N 490188 Rev. I...