n Installing from the USB Memory
1.
Remove the LAN cover and open the blind plate.
CAUTION
When touching the blind plate of the USB connector, be sure to use a wristband
to ground yourself. Otherwise the equipment may not operate properly.
2.
Connect the USB memory to the USB connector of the CPU board.
3.
Select the installation type.
• [Upgrade]:
Version upgrade
• [Full Install]: Full installation
4.
[Are you sure?] →
→ T he data in the USB memory is loaded in the equipment. When installation
completes normally, "Completed." is displayed.
5.
Check that "Completed." is displayed and remove the USB memory.
6.
Reinstall the removed plate and the LAN cover.
7.
Reboot the equipment so that the changes made in the main unit
software become effective.
016-201-04E
02.15.2007 FM5075
8.
If full installation has been performed, perform the following procedure
as well.
(i) Restore the individual data.
(ii) Reboot the equipment.
(iii) Set the system date and time.
(iv) Set the number of remaining films.
(v) Set the total film counter.
<IMPLEMENTATION OF CHANGES>
The changes made will be effective after the equipment is rebooted.
DRYPIX 2000 Service Manual
MU-25.2
MU-25.2

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DRYPIX 2000 and is the answer not in the manual?

Questions and answers

BANGET
February 13, 2025

sur DRY TIX 2000 nous avons un message d'erreur IN-0 A quoi cela correspond-il ?

This manual is also suitable for:

Drypix lite

Table of Contents

Save PDF