Warning code Cause
The dispersion tool is not secured
no
properly.
tool
The medium temperature exceeds
hot
the maximum temperature setting.
(only for S 25 EC-T...).
tool
Any malfunction during operation may be identified by an error message on the display.
Proceed as follows in such cases:
 Switch off device using the power switch
 Carry out corrective measures
 Restart the device.
Error code
Cause
E 3
Inside temperature is too high
E 4
Motor can't run normal (can't start
or not reach the target speed)
E 25
Internal hardware failure
E 86
E 47
Overload
E 48
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.

Warning message

Effect
Solution
motor off - Secure the dispersion tool properly.
- Restart the device by pressing the "Start/Stop" knob (F).
motor off - Switch off the device using on/off button (L).
- Switch on the device again using on/off button (L).
- Let the medium and dispersion tool cool down or set
higher maximum temperature.
- Restart the device by pressing the "Start/Stop" knob (K).

Error codes

Effect
Solution
motor off
- Switch off the device and let it cool down.
- Restart the device.
motor off
- Switch off the device.
- Check the structure for possible blockage
motor off
- Send the device for repair.
motor off
- Switch off the device
- Decrease the speed setting or the load
14