KERN YKT-01 Operating Instructions Manual

Statistics printer
Table of Contents

Advertisement

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the YKT-01 and is the answer not in the manual?

Questions and answers

Reinold
March 12, 2025

How to change the currency from Euro to Norwegian Krone

Summary of Contents for KERN YKT-01

  • Page 2: Table Of Contents

    Table of contents page Introduction ....................42 Technical Data, Standard accessories, Denominations ......43 Technical Data ....................43 Scope of delivery ................... 43 Denominations ....................44 Putting into operation ................. 45 Power supply ....................45 3.1.1 Mains power supply ..................45 3.1.2 Battery power supply ..................
  • Page 3 Edit parameter ..................... 64 Editor function ....................64 Edit time interval .................... 65 Edit weight unit ....................66 Edit Date, Time ....................67 Additional functions ..................68 Initialisation of the internal memory ............... 68 Printer Self-test ....................68 Appendix Error messages and references ..............69 Available optional accessories ..............
  • Page 4: Introduction

    Introduction Prior to first use of the Statistic Printer YKT-01, we recommend that you read these operating instructions very carefully. The statistics printer YKT-01 is fitted with a thermal printing unit. It has an interface for connecting electronic scales. Range of application...
  • Page 5: Technical Data, Standard Accessories, Denominations

    Technical Data, Standard accessories, Denominations Technical Data Measured values max. Characters per line Paper width 58 mm LxWxH 215 x 116 x 85 mm Battery powered by 1100 mAh > 7000 printed lines Datalogger by accu. operation ca. 24 hours Protection class IP 40 Total weight incl.
  • Page 6: Denominations

    Denominations 8 not proved by documents 1 Housing 2 Control panel 9 Data-input from measuring instrument 3 Printer cover RS 232 (INPUT) 4 ON/OFF-switch 10 not proved by documents 5 Printout 11 not proved by documents 6 Alphanumerical display 12 Connector to mains power supply 7 Tolerance-LED 13 Battery compartment cover...
  • Page 7: Putting Into Operation

    Putting into operation Power supply Connect the statistics printer with the supplied mains power adapter to a mains power source or insert 6 standard batteries AA 1,2 V. 3.1.1 Mains power supply - Connect the mains power adapter to the mains (observe the voltage) and with the 12 pin connector socket.
  • Page 8: Loading The Paper Roll

    Loading the paper roll - Switch off instrument. - Move printer cover 3 backwards (a) and take it off. - Remove the plastic shaft and guide the new paper roll core onto it (when changing paper: Take out plastic shaft and remove the old roll core and if necessary any paper parts that have remained in the printer).
  • Page 9 - Slide the end of the paper roll through the slot in the printer cover (e) and replace the printer cover onto the housing (f). - The Statistics Printer YKT-01 is now ready to print. Automatic Paper feed - Clamp (d) of the printer is closed.
  • Page 10: First Steps

    First steps => Establish a power supply, see Chapter 3.1 => Loading a paper roll, see Chapter 3.2 => Basic set-up Setting up of Language, Measuring unit and Date/Time Press following appears in the display: e.g Protocol ENTER Statistics L C D - c o n t r a s t Press and in the display appears L a n g u a g e...
  • Page 11 With select yes or no and confirm with Of course the YKT-01 Statistics Printer offers a wide range of possibilities and ranges to optimize and support your measuring and quality requirements. A detailed explanation and reference guide can be found on the following pages.
  • Page 12: Operation Mode

    Switching ON and OFF, supply voltage display 5.1.1 Switching on (long pressing). The following appears in the display after switching on: YKT-01 V 1 . 4 W e l c o m e If there is no previous measuring series, then after 2 seconds appears in display: F r 1 2 .
  • Page 13: Tolerance Display/Changing Tolerance Value (Tol)

    Tolerance display/changing tolerance value (TOL) 5.3.1 Displaying tolerance short pressing: Current tolerances will appear in the display. The cursor will appear on the sign of the Upper Tolerance. O . T o l ± 1 0 0 . 0 0 0 0 U .
  • Page 14: Transferring Measuring Values

    Transferring measuring values DATA Measuring values are accepted either using the function(print) key (data transfer) at the scales or using the DATA key on the YKT -01. They will then be shown on the display and also printed if „print measuring value“ has been activated. N o .
  • Page 15: Deleting A Measuring Value (All Measured Values)

    5.5.3 Deleting a measuring value (all measured values) Press, in the display appears D e l e t e a l l m e a s. v a l u e s ? ENTER By pressing it is possible to abandon this menu without deleting any of the measuring values D e l e t e a l l m e a s.
  • Page 16: Documenting The Adjustment (Glp)

    Documenting the adjustment (GLP) The YKT-01 is able to produce a GLP adjustment report. STAT Keep the key pressed for approx. 3 seconds. The following report will be printed and can be completed by hand. Documentation of Adjustment (GLP) Date:...
  • Page 17: Setting Mode

    Setting mode Menu guide With ENTER you get from the working mode into the setup mode. The currently selected interface will always be displayed. e.g.: Main menu I n t e r f a c e 440/572/C/D/KB Parameter With the main menu is selected L L L L L M M M M M With...
  • Page 18: Overview: Setting Mode

    Overview: Setting mode Main menu Parameter menu Chapter Edit parameter Chapter L L L L L M M M M M STAT Report Statistics Stat. / Histogr. Sample chart Print value Send value not documented Report header Sample size 1 . . . 25 Auto.
  • Page 19: Protocol

    Protocol The desired protocol can be selected. After having made the selection the following appears on the display: Protocol S t a t i s t i c s With L L L L L M M M M M it is possible to switch to Protocol S t a t .
  • Page 20: Statistics With Histogram (Stat./Histogr.)

    6.3.2 Statistics with XX.XXX to histogram (Stat./Histogr.) XX.XXX to XX.XXX XX.XXX to XX.XXX XX.XXX to X.XXX 20.500 XX.XXX to 19.900 XX.XXXXX XX.XXX to XX.XXXXX Subgr. s i z e XX.XXX X.XXXXX — — — — — — — — — — — — X.XXXXX [ g ] —...
  • Page 21: Print Values

    Print values After having made the selection the following appears on the display: P r i n t values y e s L L L L L P r i n t values With be selected. M M M M M When measurement list are accepted an acoustic signal will be issued.
  • Page 22: Protocol Header

    Protocol header Once selected the following will appear on the display: Protocol h e a d e r ( S t a t ) y e s L L L L L Protocol h e a d e r With be selected.
  • Page 23: Ascii - Printer

    ENTER With the time-controlled automatic data transfer is terminated A u t o . D a t a and goes back to ( S t a t ) y e s With L L L L L M M M M M A u t o .
  • Page 24: Acoustic Signal

    6.10 Acoustic signal Here the acoustic signal can for all functions be switched on or off. There are 3 acoustic signals available: 1 x short for measuring value transfer 1 x long for deleting measuring value/measuring list 3 x short for Error messages After having made the selection the following appears on the display: A c o u s t ic S i g n a l With...
  • Page 25: Measuring Unit

    6.13 Measuring unit The desired measuring unit can be selected. After having made the selection the following appears on the display: M e a s . u n i t With L L L L L M M M M M M e a s .
  • Page 26: Edit Parameter

    Edit parameter Editor function The display shows in the right hand corner the current editor mode. P a r t n o . : [ > ] DATA With it can be switched as follows: [>] refers to upper case ( A to Z ) [<] refers to lower case ( a to z and ä, ö, ü, ß)
  • Page 27: Edit Time Interval

    Edit time interval After having made the selection the following appears on the display: A u t o . D a t a ( S t a t ) y e s The key STAT allows the selected parameter to be edited. A u t o .
  • Page 28: Edit Weight Unit

    Edit weight unit After having made the selection the following appears on the display: M e a s . u n i t ( S t a t ) < x x x x > With the Meas. unit can be freely edited. STAT M e a s .
  • Page 29: Edit Date, Time

    Edit Date, Time After having made the selection the following appears on the display: D a t e , T i m e ( S t a t ) STAT With can Date and Time be edited. F r 1 2 . 1 1 . 0 1 1 0 : 1 3 : 4 0 With the cursor can be moved to the next position...
  • Page 30: Additional Functions

    7 . 5 V Following this press for a short period. FEED YKT-01 V 1 . 4 In the display appears : S e l f t e s t ! The available characters, date, time and voltage are printed, the LED is active and an acoustic...
  • Page 31: Appendix A Error Messages And References

    Error messages and references The task is always shown on the LCD-display. An error message will be visible for approx. 1 or 2 seconds. An acoustic signal (3 x) is issued. Then the previous display will reappear. Some error messages which require a decision or signal system failure have to be confirmed. Note: When the acoustic signal is switched off, no error tone will be sounded.
  • Page 33 Message/Error Cause Remedy The measuring value may include a Check on the measuring instrument that the W r o n g maximum of 6 places after the decimal format conforms to the default. point. The maximum number of digits f o r m a t ! including the prefix and the decimal point is 10 digits.
  • Page 34: Available Optional Accessories

    Appendix B Overview scales and data cables Weighing model series Interface cables 440,572,CB,DE,DS,KB 572-926 470,880,770,GS,GJ,CGB 70-926 474,EG,EW * 474-926* AR,PR PR-A23 ABS,ABJ ABS-A05 ITB, ITT, ITS, FTB, FTC* ITB-A15 * Print signal can only be triggered by the scale, a signal request via YKT is not possible Appendix C Available optional accessories Printpaper...
  • Page 35: Scale Configuration

    Appendix D Scale configuration This additional description contains information on the required settings to be made on the scales in order to enable a communication between scale and printer. Where a scale type is selected under the printer automatically accepts the with all INTERFACE INTERFACE PARAMETERS...
  • Page 36: Collection Of Formulas

    Appendix E Collection of formulas : Number of measuring values : Maximum value of populations : Minimum value of populations : Range of populations (max. value – min. value) : Mean value of all measuring ranges : Sum of all measuring ranges : Standard deviation of population : Standard deviation of a sample >OT...
  • Page 37 Machine Potential OGW–UGW Cm = σ Machine Capability Index OGW–Xm Xm–UGW Cmk = Minimum_of_ _resp._ σ σ Process Potential − σ ˆ Process Capability OGW–Xm Xm–UGW Cpk = Minimum_of_ _resp._ 3σ 3σ Estimated value for the standard deviation , whereby “d2” is representing a constant dependent on the sample size (table) σ= Mean value for the standard deviation +...+R...
  • Page 38: Operation Flow Chart

    Appendix F Statistics Printer YKT-01 Flow chart (schematic) Operation flow chart L L L L L Select with in menu M M M M M ENTER ENTER Save with or when changing to the next menu item Statistics L L L L L Protocol Stat.
  • Page 39 from page 1 not documented Send meas. list not documented ASCII-Printer L L L L L Acoust. Signal M M M M M 440,572,CB,DE,DS,KB 770,GS,GJ L L L L L AR/PR M M M M M Interface <47x,EG,EW> <ABS,ABJ> ITx/FTx <...

Table of Contents