15.
SELECTION OF THE PANEL SWITCH LAYOUT
The panel switch layout can be selected from among the following five types according to user
preference.
(1)
TMSJ abdomen A
(2)
TMSJ abdomen
B
(3)
Cardiac
(4)
TMSE/TAMS abdomen
(5)
TAMS left-handed
15.1
Standard Panel Switch Layout
The switches on the main panel are arranged as shown below at the factory.
15.2
Changing the Panel Configuration
File
(1)
Start Explorer from the service menu.
(2)
Open \\config\TUSPanel\TUSPanelPattern.
The following batch files are displayed.
Batch file name
I
Purpose
Standard.bat
1 Sets the panel switch layout to "Standard".
TMSJ-Abd-A.
bat
TMSJ-Abd-B.bat
Sets the panel switch layout to "TMSJ abdomen A".
1 Sets the panel switch layout to "TMSJ abdomen
B".
Cardiac.bat
TMSE-TAMS-Abd.bat
TAMS-Left-Hand.bat
Sets the panel switch layout to "Cardiac".
Sets the panel switch layout to "TMSE/TAMS
abdomen".
1 Sets the panel switch layout to "TAMS left-handed".
(3)
Execute the desired batch file.
(4)
Reboot the system.