High Priority Alerts
The following high priority alerts are accompanied by a five beep pattern, repeated every
10 seconds and a flashing yellow light.
Description
Battery Empty
Attach Plug
Battery HOT
System HOT
System COLD
System Error

6. Troubleshooting

Problem
Any problem accompanied
by information on
concentrator display,
indicator lights and/or
audible signals
17
Display Icons
Possible Cause
Refer to Chapter 5
Condition/Action/Explanation
Concentrator has insufficient battery power to
produce oxygen. Attach external power supply
or exchange battery, then restart unit if
necessary by pressing On/Off button.
Battery has exceeded temperature limit while
concentrator is running on battery power.
Concentrator has stopped producing oxygen.
If possible, move concentrator to a cooler
location, then turn power off and back on.
Ensure air intake and outlet vents have clear
access and particle filters are clean. If condition
persists, switch to a backup source of oxygen and
contact your equipment provider.
Concentrator temperature is too high and
oxygen production is shutting down. Ensure air
intake and outlet vents have clear access and
particle filters are clean. If condition
persists, switch to a backup source of oxygen
and contact your equipment provider.
This may result from the concentrator being
stored in a cold environment (below 0ºC (32ºF)).
Move to a warmer environment to allow the
unit to warm up before starting it. If condition
persists, switch to a backup source of oxygen
and contact your equipment provider.
Concentrator has stopped producing oxygen
and is shutting down. You should:
1. Switch to backup oxygen source
2. Contact your equipment provider
Recommended Solution
Refer to Chapter 5

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the One G5 and is the answer not in the manual?

Questions and answers

Maria Norma Bordine
January 19, 2025

what is replace columns

joe rempe
January 23, 2025

can plug in inogen g5 oxygen concentrator into jackery power station while inogen battery in the concentrator so it runs portable concentrator & charges inogen battery at the same time

1 comments:
Mr. Anderson
February 10, 2025

Yes, the Inogen One G5 oxygen concentrator can be plugged into a Jackery power station using the Inogen cigarette lighter adapter, which is compatible with 12V DC outlets, including those on Jackery power stations. The AC power supply can charge the battery when connected to an AC power source, but when using DC power, it is not explicitly stated whether the battery will charge while running the concentrator.

This answer is automatically generated

Maggie
February 13, 2025

how do I read the chart on page 26

Judy Chapman
March 18, 2025

02 with arrow pointing down

1 comments:
Mr. Anderson
March 18, 2025

The downward arrow on the Inogen One G5 indicates that the concentrator is producing oxygen at a slightly low level. Cleaning the intake filters (also called particle filters) may help resolve the issue.

This answer is automatically generated

Rob Poling
March 30, 2025

G5 will noy cut off when the [ower icon is pressed

1 comments:
Mr. Anderson
March 31, 2025

The Inogen One G5 may not turn off if the power button is not pushed and held. To turn it off, you must press and hold the power button until the machine begins powering down. Simply pressing the button briefly will not shut it off.

This answer is automatically generated