Vega VEGAMET 624 Supplementary Instructions Manual

Rs232/ethernet connection

Advertisement

Supplementary instructions
RS232/Ethernet connection
VEGAMET 391/624/625, VEGASCAN 693,
PLICSRADIO C62
Internet
PLC/SPS
Document ID: 30325

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VEGAMET 624 and is the answer not in the manual?

Questions and answers

Summary of Contents for Vega VEGAMET 624

  • Page 1 Supplementary instructions RS232/Ethernet connection VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62 Internet PLC/SPS Document ID: 30325...
  • Page 2: Table Of Contents

    Contents Contents Product description 1.1 Configuration ........................4 1.2 Principle of operation......................4 Adjustment ........................4 Connection Connection requirements ....................5 Connection to PC/network/modem................... 5 Network settings Adjustment with PACTware 4.1 Start PACTware ........................ 8 4.2 Create a project ........................ 8 4.3 Project examples ......................9 Application examples 5.1 Possible application areas ....................
  • Page 3 Contents 12.1 General information ......................59 12.2 Setting of the device trend ....................59 12.3 Start of the recording ...................... 60 12.4 Load recording from device .................... 60 RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 4: Product Description

    Principle of operation Application area RS232 interface The RS232 interface is particularly suitable for easy modem connec- tion for remote enquiry of measured values. For this purpose, external analogue, ISDN and GSM modems can be used. Ethernet interface With the Ethernet interface, instruments can be connected directly to an existing PC network. Each instrument is assigned its own IP ad- dress at which it can be accessed from anywhere in the network. Adjustment To adjust the instrument, the configuration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. In addition, the DTMs can be integrated in other frame ap- plications according to FDT standard. As an alternative, adjustment of the basic functions is possible via the integrated indicating and adjustment module. All device DTMs from VEGA are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistent for simple project configuration simplifies the adjustment considerably. Saving/printing the project as well as import/ export functions are also part of the standard version. In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves. The standard version is available as a free-of-charge download under http://www.vega.com. The full version is available on CD from the agency serving you. The licence regulations allow you to copy a VEGA DTM any number of times and use it on any number of computers, as long as you are using the standard version. The full version must be bought sepa- rately per computer and installation.
  • Page 5: Connection

    2 Connection Connection Connection requirements First of all, the connection to power supply and sensor must be car- ried out as described in the respective operating instructions manual. Connection to PC/network/modem Depending on the application, the connection is made via the optional RS232 or Ethernet interface. Alternatively, the built-in I²C or USB interface can be used for parameter adjustment depending on the instrument type. Details about the interface connection can be found in the respective operating instructions of the instrument. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 6: Network Settings

    If no DHCP server is available (e.g. when connecting directly to a PC via cross-over cable) or automatic addressing is not wanted, the IP addresses are entered manually in each instrument. Your responsible network administrator provides a corresponding IP address with suit- able subnet mask. Subnet mask The subnet mask together with the IP-address determines which network segment the instrument should belong to. Standard gateway A gateway connects individual network segments – data traffic can thus be forwarded to destinations outside the local network segment. Entry of a gateway address is only required when manually assigning an address and for large networks. Host name With automatic addressing via DHCP, it is necessary to assign a host name because the IP address is generally not known or can change. By default, the host name consists of the serial number plus "VEGA-" in front. Anything can be entered as a host name, for example the name of the measurement loop or the plant. The host name must only include letters and numbers without spaces. Only the hyphen "-" is al- lowed as a special character. The first character must be a letter. The host name can have a maximum of 16 characters. DNS server The DNS server is responsible for assigning the IP address and the host name. This is particularly important with automatic addressing because network devices are accessed via the host name. Entry of a DNS server address is only required with manual addressing RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 7: 3 Network Settings

    3 Network settings and large networks. Manual entry of a DNS server address is only possible via DTM – it cannot be done via the integrated display and adjustment unit. Information: A detailled description of these network parameters is available in the online help of the corresponding DTM. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 8: Adjustment With Pactware

    With complex device structures With complex device structures, the connection is carried out via the VEGA project assistant. The VEGA project assistant is a manufactur- er-specific extension module for PACTware. It is part of each VEGA DTM installation package and is installed automatically. By means of the VEGA project assistant, the connected instruments are identified automatically and integrated into the PACTware project. For this, only an online connection to the respective instruments is necessary. The VEGA project assistant is opened from the PACTware menu bar under "Project - VEGA project assistant". The window "VEGA project assistant" opens and you only have to select the requested inter- face for automatic generation of the project. If only one instrument is connected to the selected interface, the DTM parameter adjustment window will be opened automatically and the device data loaded. You can find further information on how to use the VEGA project assistant in the corresponding online help. The online help can be opened directly from the window "VEGA project assistant". RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 9: Project Examples

    View ". Project examples Project creation, VEGAMET 391 with sensor Connection VEGAMET The following example shows a typical project containing a sensor 391 via USB connected to a VEGAMET 391. Communication with VEGAMET 391 is carried out in this example via USB. We recommend using the "VEGA project assistant" as this considerably facilitates project crea- tion and helps avoid errors. All participating components are found automatically and integrated into the project. In case the planning is nevertheless carried out manually, e.g. when creating an offline project, the following DTMs must be added to the project tree: RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 10 2. Select the VEGAMET 391 DTM from the category "Gateways" and transfer it into the project window. 3. Select the requested sensor DTM from the category "Device" and transfer it to the project window. 4. Now open the DTM by double clicking on the sensor in the project window and carry out the desired settings, see chapter "Param- eter adjustment". Fig. 2: Project with VEGAPULS Project creation, VEGAMET 624 with sensor Connection, VEGAMET The following example shows a typical project containing a sensor 624 via Ethernet connected to a VEGAMET 624. Communication with VEGAMET 624 is carried out in this example via network and Ethernet. We recommend using the "VEGA project assistant" as this considerably facilitates project creation and helps avoid errors. All participating components are found automatically and integrated into the project. In case the planning is nevertheless carried out manually, e.g. when creating an offline project, the following DTMs must be added to the project tree: 1. In the device catalogue, first select the "VEGA Ethernet" DTM...
  • Page 11 4 Adjustment with PACTware 5. Now open the VEGAMET DTM and sensor DTM by double click- ing and carry out the desired settings, see chapter "Parameter adjustment". Fig. 3: Project VEGAMET with VEGAPULS RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 12: Application Examples

    5 Application examples Application examples Possible application areas • VMI (Vendor Managed Inventory) • Stock management of a tank farm via network and web browser • Automatic transmission of levels, thresholds and fault messages via e-mail or SMS • Remote enquiry of several tank farms via modem • Measured value enquiry via Modbus-TCP • Measured value enquiry via ACSII protocol • Read out of measured value files via http Inventory enquiry via network and browser Requirement The stock of a tank farm should be measured and monitored continu- ously. The measured values should be available to the scheduler and the sales department. In addition, a message should be triggered automatically if a level drops below a certain value. Solution One or more signal conditioning instruments with Ethernet interface enquire the respective sensors cyclically. The measured values are processed in the signal conditioning instrument and transferred in the requested form and unit to the integrated web server. The measured values can now be displayed on the computer of every user in the company network. In addition, the required minimum quantity is en- tered for each vessel. An e-mail is sent to the appropriate person via the company-internal mail system if a level falls below a certain value.
  • Page 13: Inventory Enquiry/Vmi Via Web-Vv

    5 Application examples • Input of IP address/Host name, date/time on the signal condition- ing instrument • Installation of PACTware and DTMs on individual network PC • Parameter adjustment of the sensors (e.g. false echo memory) via PACTware • Parameter adjustment of the signal conditioning instrument (adjustment, scaling, linearization) via PACTware • Setup of the web and e-mail server • Indication of measured values via web browser by entering the Host name/IP address of the signal conditioning instrument Inventory enquiry/VMI via WEB-VV A supplier would like to keep track of the tank stock levels of his Requirement customers and replenish them on his own whenever necessary. Via a display that is updated several times daily, he has access to the level data of the previous days or weeks. The supplier can thus assess the requirements/consumption of his customers and plan his deliveries accordingly. This allows him to do anticipatory purchasing and also better utilize the capacity of his truck fleet. In addition, a message should be sent if the level in any tank falls below a certain predefined minimum value. Using this method, the supplier can always guaran- tee his customers a sufficient supply of raw materials for production without their having to bother with purchasing and ordering. And he has the added benefit of long-term customer retention and continu- ous orders. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 14 5 Application examples Fig. 5: Remote enquiry via WEB-VV Solution A signal conditioning instrument with serial interface and modem (conventional telephone network or GSM) is installed at each cus- tomer. The measured values are transmitted automatically from each signal conditioning instrument to the central WEB-VV server at VEGA. As an alternative, the measured value transmission can be also car- ried out via the Ethernet interface and the existing company network. The measured values can now be retrieved worldwide quickly and easily by any number of (authorized) persons via Internet and web browser. The current measured values as well as the history data are available as a line graph. A certain alarm threshold can be defined for each measurement loop. If a level falls below a predetermined value, an e-mail or SMS can be sent to certain persons. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 15: Pc/Dcs Values

    6 PC/DCS values PC/DCS values General information PC/DCS outputs are used as digital outputs for transmission of meas- ured value information. The data can be transmitted to a superordi- nate PC, DCS or processing system via the RS232/Ethernet interface. The values can be sent e.g. as e-mail or read out via Modbus TCP in case of an existing Ethernet interface. The HTML pages of the signal conditioning instrument also show the measured values of the PC/ DCS outputs. The adjustment of the data format and the reference value is carried out with PACTware and the corresponding DTM. In the parameter adjustment page "PC/DCS" the reference value and the data format are selected; in addition, the option "In case of error: transmit error code instead of measured value" can be activated. Reference value With the reference value you determine which measured value is used as input signal for the PC/DCS output. The following reference values are available depending on the instrument: • Sensor value • Percent • Lin. percent • Scaled • Totalizer With "Data format", the format of the PC/DCS value is determined.
  • Page 16: Measured Value Enquiry Via Web Browser/Http 7.1 General Information

    The measured value enquiry can be also carried out via an individual http-capable software, e.g. Excel. Access protection To prevent unauthorised access to the measured values, the signal conditioning instrument can be provided with access protection. Use the web browser and enter the host name or the IP address of the instrument. Under "Settings - General" you can define the user name and the password and activate the access protection. When the page is called up, the default user name and password are asked for first of all. Enter "VEGA" for user name and password. Now you can activate the access protection for measured value enquiry and enter your own password. The access protection applies also to measured value and instrument trend files which can be called up via http. Note: This access protection only prevents measured value enquiry via web browser. If the configuration of the signal conditioning instrument also needs to be protected against unauthorized access, a second access protection is available. This protection is configured with PACTware or the respective DTM. Measured value enquiry via web browser/ Ethernet • Prerequisites Signal conditioning instrument with Ethernet interface •...
  • Page 17: Measured Value Enquiry Via Web Browser/Rs232

    7 Measured value enquiry via Web browser/http sensors as described in the respective operating instructions manual. Further information is available in chapter "Parameter adjustment with PACTware" as well as in the online help of PACTware or the DTM. Measured value indica- Open the web browser (e.g. Internet Explorer) on any PC in your tion network. Enter the host name or the IP address in the field called "Address" or "URL". Now the HTML chart generated in the signal conditioning instrument is displayed with all measured values in your browser window. This measured value enquiry can be carried out on any number of PCs in your network. Fig. 6: Measured value indication by means of Internet Explorer Measured value enquiry via web browser/ RS232 Measured value enquiry via RS232 interface and PPP (Point to Point Protocol) is used if no internal company network or other direct con- nection is available. PPP is a transmission standard for serial connec- tion between two computers (points). A dial-up line (telephone line) with a modem is usually used for this. Measured value enquiry by means of web browser is a dial-in connec- tion. The signal conditioning instrument accepts an incoming call and reacts to the calling host like an Internet Service Provider (ISP). This allows data exchange with the signal conditioning instrument via http.
  • Page 18 7 Measured value enquiry via Web browser/http • Prerequisites Signal conditioning instrument with RS232 interface • PACTware with suitable device DTM • Modem with RS232 interface • Telephone connection at the location of the signal conditioning instrument (not necessary with GSM modem) • Windows PC with modem, telephone connection and web browser (e.g. Internet Explorer) Connection Connect the signal conditioning instrument to the PC with PACTware via the RS232 interface (see chapter "Connection"). After the configu- ration, you can connect the modem instead of the PC. If changes are necessary later, you can access the instrument remotely via modem and carry out your settings. Configuration of the RS232 interface of the signal conditioning instrument Start PACTware with the suitable DTM and carry out the following configuration. Communication protocol With this setting, you define which mode the RS232 interface should operate in. The following options are available: • VVO protocol: Direct serial connection between signal condition- ing instrument and PC for parameter adjustment and enquiry (e.g. with PACTware and DTM) • PPP: Dial-up connection between signal conditioning instrument and modem for independent transmission of e-mails (dial-out con- nection) or enquiry via web browser (dial-in connection) • ASCII protocol: Direct serial connection between signal condi- tioning instrument and PC for enquiry with terminal programs, e.g.
  • Page 19: Measured Value Enquiry Via Http Software

    7 Measured value enquiry via Web browser/http Password/Password 5. Enter here an individual password for your connection. This pass- word must be used later on for the dial-up connection. IP address 6. Enter the preferred IP address. In general you can use the stand- ard setting "192.168.200.200". Host name 7. Enter an individual name in the provided field. Create a dial-up connection on the PC First of all create a dial-up connection. Under Windows 2000/Xp there is the "Assistent for new connection" which you can reach via "Start - Settings - Network connection". Proceed in the way you are creating a connection to the Internet via modem. Enter under Call number the telephone number of the local modem. Enter under User name/ Password the specifications already used for the signal conditioning instrument. Adjust a fix baud rate of 9600. Measured value indica- Now start your dial-up connection and establish the communication to tion the signal conditioning instrument. Open the web browser (e.g. Internet Explorer) and enter the previ- ously defined IP address in the field called "Address" or "URL". Now the HTML chart generated in the signal conditioning instrument is now displayed with all measured value in your browser window. Fig. 7: Measured value indication by means of Internet Explorer Measured value enquiry via http software To enquire measured value files, an individual http capable software can be used. With Excel, the measured values can be readout auto-...
  • Page 20 7 Measured value enquiry via Web browser/http • PACTware with suitable device DTM • Ethernet connection at the site of the signal conditioning instru- ment • IP address/Subnet mask for each signal conditioning instrument suitable for the network • Windows-PC with Ethernet connection and http-capable software Measured value files The PC/DCS values can be retrieved in four different file formats via http. Use the following command: http://ip-address or Host name/file name, e.g. http://192.168.200.200/val.htm. • val.txt (text file) • val.csv (CSV file) • val.htm (HTML file) • val.xml (XML file) Status files In addition to the measured value file there is also a status file with status, relay and current output information. For retrieval via http, use the following command: http://ip-address or Host name/file name, e.g. http://192.168.200.200/state.htm. • state.txt (text file) • state.csv (CSV file) • state.htm (HTML file) •...
  • Page 21: Measured Value Transmission Via E-Mail

    8 Measured value transmission via e-mail Measured value transmission via e-mail General information Via the integrated e-mail server, the current measured values can be sent to anyone at individually defined times or intervals. A message can also be sent via e-mail in case a certain level is exceeded or underrun. Status-controlled E-mail transmission is also possible, e.g. in case of a fault signal. There are two different ways to send e-mails: • Transmission via internal company network and mail server (option Ethernet interface) • Transmission via dial-up connection with modem and external Internet/e-mail provider (option RS232 interface) Both options and their different application areas are explained more detailled in the following subchapters and the online help. Clock With the time-controlled transmission of e-mails, the current time must be available. For this purpose, there is a buffered clock in the signal conditioning instrument. Via the DTM menu item "Date/Time", you can set the date and time. Via the button "Accept PC system time", the current PC system time and the date are taken over in the fields "Time" and "Date". Via the button "Write data in device", the time set in the DTM is written into the device. Both buttons are only available in online mode. Note: In case of power loss, the time is backed up by battery for a certain period (at least a few days, depending on instrument type); in case of a long interruption of the power supply, the time must be reset! For details on the exact bridging time, please see the respective instru-...
  • Page 22: E-Mail Transmission Via Network And Mail Server

    8 Measured value transmission via e-mail Content of the message Under "Contents of messages" you determine the content of the message. • Measured value: The actual measured values are sent in text, CSV, HTML or XML format. A detailled description of the file types is available in chapter "Measured value/Status files". • Visual VEGA file (only with VEGAMET/VEGASCAN series 600): The actual measured values are sent in a VEGA-specific format. These mails can be read in automatically by the software Visual VEGA. • Device trend file (only with VEGAMET/VEGASCAN): If the trend recording was activated in the signal conditioning instru- ment, a file with this recording is sent. You find further information in chapter "Device trend" and in the online help. In the event list, e-mail addresses of up to 15 recipients can be en- List of the addressees tered. The column "active" marks the list entries that are currently acti- vated. The signal conditioning instrument sends an e-mail in case of a pertinent event. In the context menu you have the option of sending a test mail. The format and configuration of the test mail corresponds to that of a real mail which is sent if the event occurs. Thus, you can test in advance if the e-mail contains the requested information and if it will be sent to the correct recipient. With "Info" you get an overview of...
  • Page 23: E-Mail Transmission Via Modem/Dial-Up

    8 Measured value transmission via e-mail then be reached from everywhere in the network via its Host name or the IP address. The system administrator must create an e-mail address on the mail server from which the measured values, messages and failures are sent. The names or IP addresses of the mail server for inbox and outbox must also be provided by the system administrator. If user name and password are required by the mail server, these must also be entered. Note: Some mail servers are configured in such a way that the password expires after a certain and must be reedited. If possible, have this function deactivated otherwise you have to reedit the passwords in the set cycle and transfer them to the signal conditioning instrument. Configuration of the e-mail account Here you can enter a name which should be displayed instead of the Name (alias) e-mail address. The name is for a better readability at the addressee because the real account name is using from time to time a not very meaningful expression. E-mail address The e-mail address constitutes the actual account name that is set up for the signal conditioning instrument on the mail server. If no alias is given, this address is shown to the addressee as the sender of the e-mail. Answering address Since the signal conditioning instrument cannot receive e-mails, there is the possibility to state an alternative e-mail address. An answering mail is sent to this address if the addressee is answering to the e-mail of the signal conditioning instrument. Configuration of the e-mail server Address of the outbox Enter the name of the outbox server provided by the system adminis- server (SMTP) trator. Alternatively, its IP address can also be entered.
  • Page 24 8 Measured value transmission via e-mail a telephone connection with direct outward dialing as well as an Internet/e-mail provider are requisites, which will create additional costs. If a telephone connection is not available or possible, a GSM wire- less modem with RS 232 interface can be also used. In this case, a GSM mobile phone contract with the option of data transmission is required. Make sure that your location is sufficiently covered by the network of the mobile network operator. Also the PIN of the SIM card used must be deactivated. Please note that the dial-in numbers to the Internet provider are usually not the same as those for the conven- tional telephone network. Prerequisites: • Signal conditioning instrument with RS232 interface • Adjustment software frame application, e.g. PACTware and device- • VEGACONNECT for parameter adjustment of the signal condition- ing instrument (only with series 600) • Modem with RS232 interface • Available telephone connection with direct outward dialing at the location of the signal conditioning instrument • External Internet provider (e.g. T-Online, MSN) with user name/ password for authentication • External e-mail account/address for transmission of measured values with name of the inbox and outbox server (POP/SMTP) as well as user name/password for authentification • E-mail address(es) of the requested measured value addressees Connection Connect the signal conditioning instrument via the RS232 interface to the PC with PACTware (see chapter "Connection"). After the configu- ration, you can connect the modem instead of the PC. For probable later changes, it is now possible to address the instrument remotely via modem and carry out your settings. To adjust the parameters of the signal conditioning instrument and at the same time check the e-mail transmission, an additional con- nection is recommended. For this purpose, use for series 600 a...
  • Page 25 8 Measured value transmission via e-mail • ASCII protocol: Direct serial connection between signal condi- tioning instrument and PC for enquiry with terminal programs, e.g. Hyperterminal 1. When connecting a modem for e-mail transmission, choose the option "PPP". Modem initialisation When connecting a modem, you should switch on this option so that it will be provided with parameters necessary for data transmission. Point to Point Protocol The Point to Point Protocol (PPP) enables the transmission of LAN protocols (e.g. http) via point-to-point connection. PPP connections are e.g.: • Switched connections via the analogue telephone network with analogue, ISDN and GSM modem • Standard connections 2. For sending e-mails select under "Dialing directions" the option "Dial-out (outgoing connections)". 3. Enter under "Access data for connection to the Internet service provider" the following data which you get from your Internet provider: Telephone connection...
  • Page 26 8 Measured value transmission via e-mail User name/Password Enter the user name and the corresponding password for the inbox server provided by the e-mail provider. Many providers also request an authentification for the outbox server. Depending on the provider, the same access data may also be intended for the inbox server. In principle, completely different access data may also be required. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 27: Measured Value Transmission To A Web-Vv Server

    9 Measured value transmission to a WEB-VV server Measured value transmission to a WEB-VV server Prerequisites Signal conditioning The measured values dermined by the sensors are further processed instruments by the signal conditioning instrument and can be transferred to a WEB-VV server. For this purpose, one of the following signal condi- tioning instruments, including one of the optional interfaces (RS232 or Ethernet), is required. • VEGAMET 391/624/625 • VEGASCAN 693 • PLICSRADIO C62 Connection types The transmission of readings is generally carried out via an Internet connection, independent of the interface type or access mode. The connection alternatives can be subdivided into the following three groups: • Signal conditioning instrument with RS232 interface and ana- logue/GSM/GPRS modem • Signal conditioning instrument with Ethernet interface and MoRoS (ModemRouterSwitch) • Signal conditioning instrument with Ethernet interface directly con- nected to the company network The selection of the most suitable connection alternative is depend- ent on the local conditions and the frequency of the measured value...
  • Page 28 9 Measured value transmission to a WEB-VV server modem) is the simplest choice. Alternatively, when there is no tel- ephone connection and a wireless modem is used, the data transmis- sion is carried out via GSM or GPRS. For cost reasons transmission via GPRS is recommended, as charges are based on the transmitted volume. The following modems can be used for measured value transmission: • Phoenix Contact PSI-Data/Fax modem (analogue telephone network), art. no.: MODEM.JX • Siemens TC35i (GSM radio modem), art. no.: MODEM.FX • Insys GPRS 5.0 serial (GPRS radio modem), art. no.: MODEM.GX To access the Internet, access data from an Internet service provider (ISP) must be obtained. These data are entered into the signal condi- tioning instrument via PACTware. When GSM or GPRS is implement- ed, a SIM card is additionally required. Signal conditioning instrument with analogue modem Prerequisites for setup: • PC with PACTware and suitable DTM as well as access to the signal conditioning instrument via VEGACONNECT or USB • Access data from an Internet provider (ISP) must exist (call by call or contractual relationship) • Analogue telephone connection with direct outward dialing (note: dialout number “0” as prefix may be necessary) • No telephone number blocking for the number used to call the ISP Start PACTware and make a connection to the signal conditioning in- strument. Select the menu item "Instrument settings - RS232 interface ". Carry out the following steps as shown in the following illustrations.
  • Page 29 9 Measured value transmission to a WEB-VV server Fig. 9: Dial-in connection Dial-out connection Enter here the access data you received from your Internet provider. Do not enter anything in the entry field "APN". Fig. 10: Dial-out connection Measured value transmis- Define a WEB-VV event as described in chapter "Set up measured sion value transmission". Signal conditioning instrument with GSM modem Prerequisites for setup: • PC with PACTware and suitable DTM as well as access to the signal conditioning instrument via VEGACONNECT or USB • Modem location with sufficient wireless coverage • SIM card with activated data transmission (data card CSD) • PIN of SIM card deactivated •...
  • Page 30 9 Measured value transmission to a WEB-VV server Start PACTware and make a connection to the signal conditioning in- strument. Select the menu item "Instrument settings - RS232 interface ". Carry out the following steps as shown in the following illustrations. RS232 configuration Enter the parameters for the RS232 interface as shown. Fig. 11: RS232 configuration Dial-in connection Enter the parameters for the dial-in connection as shown. These data make access from the outside possible (remote parameterisation). The password is "webvv". Fig. 12: Dial-in connection Dial-out connection Enter here the access data you received from your Internet provider. Nothing may be entered in the entry field "APN". RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 31 9 Measured value transmission to a WEB-VV server Fig. 13: Dial-out connection Define a WEB-VV event as described in chapter "Set up measured Measured value transmis- sion value transmission". Signal conditioning instrument with GPRS modem Prerequisites for setup: • PC with PACTware and suitable DTM as well as access to the signal conditioning instrument via VEGACONNECT or USB • Modem location with sufficient wireless coverage • SIM card with activated data transmission (data card CSD+GPRS) • PIN of SIM card deactivated • Internet access data with APN (Access Point Name) from the mobile communications provider must be known. • To keep transmission costs low, we recommend using a pure data rate (M2M) with low data block rounding (1 kB) Start PACTware and make a connection to the signal conditioning in- strument. Select the menu item "Instrument settings - RS232 interface ". Carry out the following steps as shown in the following illustrations.
  • Page 32: Signal Conditioning Instrument (Ethernet) With Moros

    9 Measured value transmission to a WEB-VV server Dial-in connection Enter the parameters for the dial-in connection as shown. These data make access from the outside possible (remote parameterisation). Fig. 15: Dial-in connection Dial-out connection Enter here the access data from your Internet provider. Fig. 16: Dial-out connection Measured value transmis- Define a WEB-VV event as described in chapter "Set up measured sion value transmission". Signal conditioning instrument (Ethernet) with MoRoS This alternative is used if several signal conditioning instruments have to be connected to WEB-VV at the same site and no company network with Internet access is available. The signal conditioning instruments must be equipped with an Ethernet interface option. The option of remote parameterisation is also available, i.e., the signal conditioning instrument and the sensors can be accessed from out- side the system. This allows settings for remote transmission as well RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 33 9 Measured value transmission to a WEB-VV server as parameterisation of the signal conditioning instrument/sensor to be carried out or changed from a distant location. A MoRoS with integrated modem is used for connection to the Internet. The built-in router takes over dialing into the Internet. The device also contains a 4-port switch for direct connection to up to four signal conditioning instruments with Ethernet connection. More signal conditioning instruments can be connected via an additional external switch. The choice of the modem integrated in the MoRoS depends on the lo- cal conditions. If a telephone connection (analogue or ISDN) is avail- able on site, the analogue (PSTN) or ISDN modem is the simplest choice. Alternatively, a wireless modem is used when no telephone connection exists; the data transmission is carried out in this case via GSM or GPRS. For cost reasons transmission via GPRS is recom- mended, as charges are based on the transmitted volume. The following MoRoS versions are available: • MoRoS with PSTN modem, art. no.: ROUTER.AXX • MoRoS with ISDN modem, art. no.: ROUTER.IXX • MoRoS with GSM/GPRS modem, art. no.: ROUTER.GXX To access the Internet, access data from an Internet service provider (ISP) must be obtained. These data are entered via Web browser into the MoRoS. A SIM card is additionally required when GSM or GPRS is implemented. The MoRoS has the standard IP address 192.168.1.1. Since it is an independent network that is involved here, this address should not be changed. Every connected signal conditioning instrument must be assigned a unique address from the same address range, e.g. 192.168.1.2 for the first signal conditioning instrument, 192.168.1.3 for the second, etc. Signal conditioning instrument with MoRoS/PSTN/ISDN modem Prerequisites for setup: •...
  • Page 34 9 Measured value transmission to a WEB-VV server Fig. 17: IP addressing DNS addressing Enter the IP address of the MoRoS under "Instrument settings - DNS Server". Fig. 18: DNS addressing MoRoS setup The access data of the ISP must be written into the MoRoS. You can find the procedure in chapter "Commissioning the MoRoS" as well as in the operating instructions manual. Measured value transmis- Define a WEB-VV event as described in chapter "Set up measured sion value transmission". Signal conditioning instrument with MoRoS/GPRS modem Prerequisites for setup: •...
  • Page 35 9 Measured value transmission to a WEB-VV server • To keep transmission costs low, we recommend using a pure data rate (M2M) with low data block rounding (1 kB) Start PACTware and make a connection to the signal conditioning instrument. Select the menu item "Instrument settings - LAN/Internet". Carry out the following steps as shown in the following illustrations. IP addressing Give every signal conditioning instrument a unique IP address from the address range of the MoRoS (192.168.1. xxx); the address of the MoRoS itself (192.168.1.1) may be not used. The subnet mask remains unchanged (255.255.255.0). Type in the IP address of the MoRoS under Standard gateway. Fig. 19: IP addressing DNS addressing Enter the IP address of the MoRoS under "Instrument settings - DNS Server". Fig. 20: DNS addressing MoRoS setup The access data of the mobile telecommunications provider must be written into the MoRoS. You can find the procedure in chapter "Commissioning the MoRoS" as well as in the operating instructions manual. Measured value transmis- Define a WEB-VV event as described in chapter "Set up measured sion...
  • Page 36: Signal Conditioning Instrument (Ethernet) In Lan

    9 Measured value transmission to a WEB-VV server Signal conditioning instrument (Ethernet) in This alternative is used if a company network with internet access is available. Any number of signal conditioning instruments can be connected to WEB-VV. The signal conditioning instruments must be equipped with the Ethernet interface option. When addressing is done via DHCP, a suitable Host name should be assigned. As an alterna- tive, a unique IP address suitable to the network can be assigned. You can find further information in chapter "Network settings". Prerequisites for setup: • PC with PACTware and suitable DTM as well as access to the signal conditioning instrument via VEGACONNECT or LAN • Subnet mask, standard gateway and DNS server dof the company network • http port 80 must be activated for an Internet connection Start PACTware and make a connection to the signal conditioning instrument. Select the menu item "Instrument settings - LAN/Internet". Carry out the following steps as shown in the following illustrations. 1. Give every signal conditioning instrument a unique IP address that you have received from your responsible network administra- tor. The subnet mask matching your network and the standard gateway also come from your network administrator. Alternatively, you can use the DHCP function if a DHCP server is available in your network. Fig. 21: Allocate IP address 2. Enter the IP address of your DNS server under "Instrument set- tings - DNS Server".
  • Page 37 9 Measured value transmission to a WEB-VV server Fig. 22: Allocate DNS address 3. A proxy server is usually a computer connected between the sig- nal conditioning instrument and the Internet. With a proxy server it is possible to check and limit the access of the signal conditioning instrument to the Internet. Such access of the signal condition- ing instrument to the Internet is, for example, carried out when there is a WEB-VV data transmission. To ensure that these data are transmitted via the existing proxy server, the settings shown below must be made in the signal conditioning instrument. Your network administrator knows if the Internet access should be carried out via a proxy. He will also provide the necessary infor- mation such as IP address, port, user name and password. Fig. 23: Configuration Proxy server 4. In conclusion, define a WEB-VV event as described under chap- ter "Set up measured value transmission". RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 38: Setup Moros

    9 Measured value transmission to a WEB-VV server Setup MoRoS Adjustment of the MoRoS is carried out with the help of any Web browser, e.g. the Internet Explorer. Connect your PC via Patchkabel with one of the four Ethernet ports on the front side of the MoRoS. In order to establish a connection, the PC must be able to obtain an IP address from the MoRoS automatically (DHCP). Alternatively, a fixed IP address can be used if it lies in the same address range of the MoRoS (192.168.1. xxx). You can get further information also from the operating instructions manual of the MoRoS. MoRoS with PSTN/ISDN modem Login Start your Web browser and enter the IP address of the MoRoS (192.168.1.1) in the field marked "Address" or "URL". Authenticate yourself with the user name "insys" and the password "moros". Fig. 24: MoRoS user authentication Modem settings Carry out the modem settings for remote parameterisation shown in the following illustrations. Select from the list box the country code of the country in which the modem will be used.
  • Page 39 9 Measured value transmission to a WEB-VV server Fig. 25: Modem configuration Dial-in settings Make the dial-in settings for remote parameterisation shown in the following illustrations. Fig. 26: Dial-in settings Dial-out settings Enter here the access data from your Internet provider. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 40 9 Measured value transmission to a WEB-VV server Fig. 27: Dial-out settings Internal clock The MoRoS contains an internal clock for controlling time-dependent events and dating system messages. Set the time, select the proper time zone and activate the time synchronization. Fig. 28: Time synchronisation MoRoS with GPRS modem Login Start your Web browser and enter the IP address of the MoRoS (192.168.1.1) in the field marked "Address" or "URL". Authenticate yourself with the user name "insys" and the password "moros". RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 41 9 Measured value transmission to a WEB-VV server Fig. 29: MoRoS user authentication GSM/GPRS settings Carry out the GSM-GPRS settings shown in the following. Enter the ID of your mobile communications provider under "Log in exclusively at this provider". You get this by clicking on the option "Read out pro- vider list from modem". You can get further information also from the operating instructions manual of the MoRoS. Fig. 30: GSM/GPRS settings Dial-in settings Make the dial-in settings for remote parameterisation shown in the following illustrations. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 42 9 Measured value transmission to a WEB-VV server Fig. 31: Dial-in settings Dial-out settings Enter here the access data from your mobile communications pro- vider. Fig. 32: Dial-out settings RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 43: Set Up Measured Value Transmission

    The MoRoS contains an internal clock for controlling time-dependent events and dating system messages. Set the time, select the proper time zone and activate the time synchronization. Fig. 33: Time synchronisation Set up measured value transmission Select the menu item "Instrument settings - Event list" and press the button "New". An assistant will now lead you through the configura- tion. 1. Select in the first window under "Message type" the list entry "WEB-VV" or "WEB-VV (encoded)". 2. In the second step enter the name or the IP address of the WEB- VV server. If you’re using the VEGA servers (VEGA Hosting/ WEB-VV VH), enter the URL "web-vv.vega.com" here. If you have a server of your own (Local Hosting/WEB-VV LH), you can get the server name or the IP address from your responsible network administrator. 3. In the last step, define when or how often the measured values are transmitted. Please remember that costs arise with every measured value transmission, depending on the type of connec- tion. 4. Transfer your settings to the signal conditioning instrument. After completing the configuration, you should start a test transmis- sion to the WEB-VV server. Then contact the WEB-VV administrator.
  • Page 44 9 Measured value transmission to a WEB-VV server Fig. 34: Event list for setting up the measured value transmission. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 45: Measured Value/Status Files

    10 Measured value/Status files 10 Measured value/Status files 10.1 Description of the measurement data files The PC/DCS values can be sent in four different file formats from the instrument via e-mail or retrieved via http. Enter the following line into your browser: http://ip-address or Host name/file name, e.g. http://192.168.200.200/val.htm. The files have the following name: • val.txt (text file) • val.csv (CSV file) • val.htm (HTML file) • val.xml (XML file) Configuration The contents of the files consists of a header with the following infor- mation: • #System: VEGAMET 625 (instrument type) • #Version: 1.91 (Firmware version) • #TAG: Device Name (device-TAG) •...
  • Page 46 10 Measured value/Status files File content Description 001,"TAG-No. 1",98,75,% DCS value 1: no., measurement loop-TAG, value, unit 002,"TAG-No. 2"98,65,% DCS value 2: no., measurement loop-TAG, value, unit 003,"TAG-No. 3",0.10,% DCS value 3: no., measurement loop-TAG, value, unit CSV file In the CSVfile (Character Separated Values) the values are sepa- rated by a certain character. Often a comma is used as separator. As separators, the signs "Comma" (default setting), "Semicolon" and "Tabulator" can be set. The individual data sets are separated by a line break character. CSV files are particularly suitable for import into spread sheet pro- grams, such as Microsoft Excel. Example of the content of a "val.csv" file of a VEGAMET 625. File content Description #System: VEGAMET 625 Instrument type #Version: 1.91 Firmware version #TAG: Device Name Device-TAG #SNR: 11162431 Instrument serial number #Date: 27.10.04 Transmission date #Time: 14:42:34 Transmission time #Ontime: 02:58:51 Switch on period #PLS;TAG;VALUE;UNIT...
  • Page 47 10 Measured value/Status files Fig. 35: Measured value file XML file With the XML file, information is transmitted in XML format. These files are particularly suitable for automated reading in of device data. The XML document is divided into different elements. These ele- ments structure the XML document and start each with a so called "Start Tag" and finish with the "End Tag". Example of the content of a "val.xml" file of a VEGAMET 624. <?xml version="1.0" XML version encoding="iso-8859-1"? <data> Start-TAG for data <device> Start-TAG device information <system>VEGAMET624</sys- Instrument type tem> <version>1.91</version> Firmware version <devicetag>Device Name</de- Device-TAG vicetag> <snr>1230985</snr> Instrument serial number <date>13.09.2004</date> Transmission date <time>12:00:00</time> Transmission time <ontime>02:58:51</ontime> Switch on period </device>...
  • Page 48: Description Of The Status Files

    10 Measured value/Status files <device> Information within the "<device>"-TAG is information about the signal conditioning instrument. It is stated once at the beginning of the docu- ment. <dataitem> "<dataitem>"-TAG there is measured value information from a PC/ DCS output. For each assigned PC/DCS output there exists such a TAG, i.e. up to six of these TAGs can exist with a VEGAMET/PLICS- RADIO, up to 30 with a VEGASCAN. 10.2 Description of the status files In addition to the measured value file there is also a status file with status, relay and current output information. For retrieval via http, enter the following line into your browser: http://ip-address or Host name/file name, e.g. http://192.168.200.200/state.htm. The files have the following name: • state.txt (text file) • state.csv (CSV file) • state.htm (HTML file) • state.xml (XML file) Configuration The contents of the files consists of a header with the following infor- mation: • #System: VEGAMET 625 (instrument type) • #Version: 1.91 (Firmware version) • #TAG: Device Name (device-TAG) •...
  • Page 49 10 Measured value/Status files File content Description #Fault relay: 0 Status fail safe relay #Loop,Tag,State,Description Header measurement loop status information 1,TAG-No. 1,OK,- Measurement loop no., measurement loop TAG, status, description #Input,State,Description Heading input status information 1,OK,- Input no., status, description #Current,Value,Unit,State,De Heading current output status information scription 1,19.798,mA,OK,- Current output no., current value, unit, status, description #Relay,State,Unit,Description Heading relay status information 1,1,-,OK,- Relay no., status: 1=On 0=Off, status, de- scription #PC-PLS,Value,Unit,State,De Heading PC/DCS output status information scription 1,98.73,%,OK,- PC/DCS no., value, unit, status, description 2,TAG-No. 2,OK,- Measurement loop no., measurement loop TAG, status, description #Input,State,Description Heading input status information 2,OK,- Input no., status, description #Current,Value,Unit,State,De Heading current output status information scription...
  • Page 50 10 Measured value/Status files File content Description #PC-PLS,Value,Unit,State,De Heading PC/DCS output status information scription 3,0.08,%,OK,- PC/DCS no., value, unit, status, description CSV file In the CSVfile (Character Separated Values) the values are sepa- rated by a certain character. Often a comma is used as separator. As separators, the signs "Comma" (default setting), "Semicolon" and "Tabulator" can be set. The individual data sets are separated by a line break character. CSV files are particularly suitable for import into spread sheet pro- grams, such as Microsoft Excel. Example of the content of a "state.csv" file of a VEGAMET 625. File content Description #System: VEGAMET 625 Instrument type #Version: 1.91 Firmware version #TAG: Device Name Device-TAG #SNR: 11162431 Instrument serial number #Date: 10.10.04 Transmission date #Time: 10:26:58 Transmission time #Ontime: 02:58:51 Switch on period #Fault relay: 0 Status fail safe relay #Loop,Tag,State,Description...
  • Page 51 10 Measured value/Status files File content Description #Current,Value,Unit,State,De Heading current output status information scription 2,19.784,mA,OK,- Current output no., current value, unit, status, description #Relay,Value,Unit,State,Des Heading relay status information cription 2,1,-,OK,- Relay no., status: 1=On 0=Off, status, de- scription #PC-PLS,Value,Unit,State,De Heading PC/DCS output status information scription 2,98.65,%,OK,- PC/DCS no., value, unit, status, description 3,TAG-No. 1,OK,- Measurement loop no., measurement loop TAG, status, description #Input,State,Description Heading input status information 3,OK,- Input no., status, description #Current,Value,Unit,State,De Heading current output status information scription 3,19.987,mA,OK,- Current output no., current value, unit, status, description #Relay,Value,Unit,State,Des Heading relay status information cription 3,1,-,OK,- Relay no., status: 1=On 0=Off, status, de- scription #PC-PLS,Value,Unit,State,De Heading PC/DCS output status information...
  • Page 52 10 Measured value/Status files Fig. 36: Status file XML file With the XML file, information is transmitted in XML format. These files are particularly suitable for automated reading in of device data. The XML document is divided into different elements. These ele- ments structure the XML document and start each with a so called "Start Tag" and finish with the "End Tag". The XML-file provided by VEGAMET/VEGASCAN has the following configuration: Example of the content of a "state.xml" file of a VEGAMET 624. File content Description <?xml version="1.0" XML version encoding="iso-8859-1"? <data> Start-TAG for data <device>…</device> TAGs device information <loopitem>…Start-TAG measurement loop status in- formation <input>…</input> TAGs input status information <measurement>…</measure- TAGs measurement loop status information ment> RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 53 10 Measured value/Status files File content Description <output> Start-TAG for output status information <current>…</current> TAGs current output status information <relay>…</relay> TAGs relay status information <pcpls>…</pcpls> TAGs PC/DCS output status information </output>End-TAG measure- End-TAG for output status information ment loop status information </loopitem> </data> End-TAG for data <device> The information within the "<device>"-TAG is information on VE- GAMET/VEGASCAN/PLICSRADIO. It is stated once at the beginning of the document. <loopitem> Within the "<loopitem>"-TAG there is status information about a measurement loop. Such a TAG exists for each measurement loop, e.g. up to 3 such TAGs can be available with a VEGAMET, 15 with a VEGASCAN 15 and six with a PLICSRADIO C62. <input> In the " <input> "-TAG you find the status information on the input of a measurement loop. <measurement> In the " <measurement> "-TAG you find the status information on the measurement loop processing of a measurement loop. <output> In the " <output> "-TAG you find the status information on the outputs of the measurement loop.
  • Page 54 10 Measured value/Status files File content Description <device> Start-TAG device information <system>VEGAMET624</sys- Instrument type tem> <version>1.91</version> Software version <devicetag>Device Name</de- Device-TAG vicetag> <snr>1230985</snr> Instrument serial number <date>10.09.2004</date> Transmission date <time>12:00:00</time> Transmission time <ontime>02:58:51</ontime> Switch on period <faultrelay>0</faultrelay> Status fail safe relay </device> End-TAG device information <loopitem> Start-TAG measurement loop status infor- mation <nr>001</nr> Measurement loop no. <tag>TAG No. 1</tag> Meas. loop TAG <input> Start-TAG input status information <nr>001</nr> Input no. <state>OK</state>...
  • Page 55 10 Measured value/Status files File content Description <unit>mA</unit> Unit <state>OK</state> Current output status <desc>-</desc> Description status </current> End-TAG current output status <current> Start-TAG current output status <nr>003</nr> Current output no. <value>12.00</value> Value of the current output <unit>mA</unit> Unit <state>OK</state> Current output status <desc>-</desc> Description status </current> End-TAG current output status <relay> Start-TAG relay output status <nr>001</nr> Relay output no. <value>1</value> Switching status 0=Off, 1=On <unit>-</unit> Unit <state>OK</state> Relay output status <desc>-</desc> Description status </relay>...
  • Page 56 10 Measured value/Status files File content Description <desc>-</desc> Description status <pcpls> End-TAG PC/DCS output status </output> End-TAG for output status information </loopitem> End-TAG measurement loop status informa- tion </data> End-TAG for data RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 57: Measured Value Transmission Via Sms

    11 Measured value transmission via SMS 11 Measured value transmission via SMS 11.1 General information Sending an SMS is only possible with signal conditioning instruments with serial interface in conjunction with modems supported by VEGA. A GSM modem and a conventional telephone network modem are available for the analogue telephone connection. To send an SMS, the signal conditioning instrument dials into an SMS Service Center (SMSC): Via this SMS Service Center the SMS is the transferred to the subscriber (e.g. mobile phone). Note: To set up the SMS function, PACTware with the corresponding DTMs is required. 11.2 Settings for SMS transmission Open the suitable DTM in PACTware and select the menu item "Device settings - SMS" for configuration of the SMS function. As basic setting, the connected modem and the requested SMS provider (SMSC) are selected. Modem If the SMS should be transmitted via the conventional telecommuni- cation network, you will need an analogue telephone connection with direct outward dialing and the Phoenix Contact PSI data/fax modem (order no.: MODEM.JX) If transmission is made via GSM, the Siemens TC35i GSM cellular phone modem must be used (order no.: MODEM.FX). For this you will need a GSM contract. Make sure that there is sufficient network...
  • Page 58 11 Measured value transmission via SMS For e-mail transmission there is an assistant available for simple con- figuration. To start the assistant, select the button "New". For further information see the online help. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 59: Device Trend/Data Logger (Only Vegamet/Vegascan)

    12 Device trend/Data logger (only VEGAMET/VEGASCAN) 12 Device trend/Data logger (only VEGAMET/ VEGASCAN) 12.1 General information With the device recording the measured values of a VEGA instrument can be monitored. There are two options for trend recording: Device trend The data are saved over a certain period in the VEGA instrument and can be read out again later on. Via PACTware and DTM different recording conditions can be set and the process can be started or stopped. The max. recording period is limited by the memory of the signal conditioning instrument. The max. recording period differs de- pending on the recording type (interval or measured value difference). When recording with interval, a maximum of 200 000 measured val- ues can be recorded, when recording with measured value difference, the maximum is reduced to 140 000. If a measured value is saved every minute, a period of e.g. 138 days results (with measured value difference). If more than one curve is recorded, this period is reduced accordingly. Note: The recorded data are written in a period of two to five minutes from the internal RAM to a flash disk. In case of voltage failure, recordings of the last minutes can get lost. DTM trend The data are only saved on the PC during an online connection with PACTware and DTM. The max. recording period is limited only by the storage space on the hard disc. 12.2 Setting of the device trend The settings are carried out in the DTM under the menu item "Diag-...
  • Page 60: Start Of The Recording

    12 Device trend/Data logger (only VEGAMET/VEGASCAN) Start/Stop conditions To enable a specific monitoring, the start and stop conditions can be set for the recording in dependence on the current measured value. If a start condition is met, the recording starts in the VEGA instrument. The current recording stops as soon as the stop condition is reached. If the start condition is fulfilled again, the recording does not switch on again. The recording can be stopped automatically if the memory in the VEGA instrument is full. Select the option "Stop recording when memory is full". If this option is selected, the older measured values will be overwritten automatically (ring buffer). 12.3 Start of the recording After having adjusted the recording mode and the start/stop condi- tions, these settings have to be transferred to the VEGA instrument ("Device data - Save in device"). Activate the trend recording by push- ing the button "Start recording". The current utilisation of the memory and the number of recorded measured values is now displayed under "Status". Modification of the recording conditions is not possible dur- ing a recording. 12.4 Load recording from device Retrieval of the recorded measured values can be carried out via PACTware/DTM, http or e-mail. The Ethernet as well as the RS232...
  • Page 61 12 Device trend/Data logger (only VEGAMET/VEGASCAN) Information: The presentation of large recordings can take a long time depend- ing on the number of curves, the interface and the retrieval method. Reading out via Ethernet is much quicker than via RS232. Display via http is also much quicker than via the DTM. When transmitting device trends via e-mail, the same files as for http Transmission via e-mail data retrieval are used. In the e-mail configuration assistant you can select and configure the respective curves and formats. You can find further information in chapter "Measured value transmis- sion via e-mail" and in the online help of the respective DTM. RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 62 INDEX INDEX MoRoS 27, 30, 32 Access protection 16 APN 31 Network 6 ASCII 12, 18, 24 Offline mode 9 Communication protocol 18 CSD 29, 31 CSV file 46, 50 PACTware 4, 8 – Project 8 Password 8, 16 Data logger 59 Password/Password 16 Device trend 59 PC/DCS 15, 45 DHCP 6 PLC 15 Dial-In 30, 32 POP 22, 23...
  • Page 63 INDEX XML file 47, 52 RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62...
  • Page 64 Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2013 VEGA Grieshaber KG Phone +49 7836 50-0 Am Hohenstein 113...

This manual is also suitable for:

Vegamet 391Vegascan 693Plicsradio c62Vegamet 625

Table of Contents