Page 3
Copyright 2018 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Release time: July 2018 Revision 3.0 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 4
This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties.
Page 5
Mindray’s option, at the factory or at an authorized distributor, for any product which shall under normal use and service appear to Mindray to have been defective in material or workmanship. Recommended preventative maintenance, as prescribed in the service manual, is the responsibility of the user and is not covered by this warranty.
Page 6
Exemptions Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel.
Page 7
Upon request, Mindray provides circuit diagrams, component part lists, descriptions, calibration instructions, or other information which assist the user’s appropriately qualified technical personnel to repair those parts of the equipment which are designated by Mindray DS USA, Inc. as repairable. Contact Information Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Page 8
This manual is for biomedical engineers, authorized technicians or service representatives responsible for troubleshooting, repairing and maintaining the patient monitors. Contact your local Mindray Service Organization for information on product courses which address service and support for this product. Passwords A password may be required to access different modes within the monitor.
3.3.7 C.O. Test ......................................3-15 3.3.8 Mainstream CO Tests ................................... 3-15 3.3.9 Sidestream and Microstream CO Tests ..........................3-16 3.3.10 AG Tests ......................................3-16 3.3.11 EEG Test ......................................3-16 3.3.12 BIS Test ......................................3-18 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 10
5.3.5 Upgrading Independent Display Function ..........................5-4 6 Repair and Disassembly ..............................6-1 6.1 Tools............................................. 6-1 6.2 Preparations for Disassembly ..................................6-1 6.3 Main Unit Disassembly ....................................6-2 6.3.1 Basic Disassembly..................................... 6-2 6.3.2 Disconnecting the Base ................................. 6-3 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Provides application tips or other useful information. 1.1.1 DANGER There are no dangers that refer to the product in general. Specific “Danger” statements may be given in the respective sections of this manual. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Refer to Operation Manual for detailed operation and other information. 1.2 Equipment Symbols See the N series Operator’s Manual (P/N: 046-011259-00) for information about the symbols used on this product and its packaging. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
The MPAN and WiFi modules are optional. The built-in recorder is optional. The N15/N17 can connect to the external module rack and TDS; the N12 can connect to the TDS. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
USB 4*port Independent display Audio RMII1 Speaker board/iView codec/AMP FPGA LCDC (N15/N17) Power key board Power M0 UART transfer interface RJ-45 Back light driver Figure 2-2 Diagram of the main control board BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Windows OS running on other modules due to an unexpected power failure of the AC power supply. The battery supports the main unit to stop the iView module in the normal power-off mode. In the case of battery power supply, the iView module cannot start. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Meanwhile, they can also be sent to the central station or other monitors through the wired or wireless network. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
(and once a year for CO and AG modules). See the following sections for detailed test procedures and contents. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
4. AG leakage test should be performed before AG measurement. Leakage test AG tests Performance test Calibration EEG test BIS test Interconnecting function CCO/SvO tests Output calibration Performance test NMT tests Sensor check BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Inspect all external connections for loose connectors, bent pins or frayed cables. Inspect all connectors on the equipment for loose connectors or bent pins. Make sure that safety labels and data plates on the equipment are clearly legible. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Record the current pressure value and meanwhile count time with a timer. Then, record the pressure value after counting to 60 seconds. Compare the two values and make sure the difference is not greater than 6 mmHg. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 23
You can use an NIBP simulator to replace the squeeze bulb and the reference manometer to perform the test. You can use an appropriate cylinder and a cuff instead of the rigid vessel. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Select Main Menu → Maintenance → enter the required password → Module → CO2 → Calibration. Connect the test system as follows: BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 25
Check the airway for leaks and perform a leakage test as well to make sure that the airway has no leakage. Select Main Menu → Maintenance → enter the required password → Module → CO2. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
If the alarm message continues, it indicates that the module does not leak. If the alarm message disappears, or the flow rate is greater than or equal to 10 ml/min, it indicates that the module leaks. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 27
N2O, Bal O2, P/N: 0075-00-0048-01 and flow regulator P/N: 0119-00-0235). Gas concentration should meet the following requirements: AA ≥ 1.5%, CO2 ≥ 1.5%, N2O ≥ 40%, O2 ≥ 40%, of which AA represents an anesthetic agent. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 28
If the calibration is finished successfully, the message Calibration Completed! is displayed. If the calibration failed, the message Calibration Failed! is displayed. In this case, perform another calibration. If the calibration fails several times, return the module to Mindray for repair. 3-10...
. A square wave appears on the screen and the message " " is displayed. Compare the amplitude of the square wave with that of the scale. The difference should be with 5%. After completing the calibration, select Stop Calibration. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 3-11...
sensor. Instead, it can only verify that whether the monitor is functional. The accuracy of a pulse oximeter monitor or a SpO sensor needs to be verified by clinical data. 3-12 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Pressure Calibration Method 1: Tools required: Medsim300B patient simulator, MPS450, or other equivalent equipment Dedicated IBP adapter cable (300B, P/N 00-002199-00) (use P/N 00-002198-00, if the simulator is MPS450) BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 3-13...
Page 32
Inflate using the squeeze bulb until the reading of sphygmomanometer approximates the preset calibration value. Adjust the calibration value in the Maintain IBP menu until it is equal to the reading of sphygmomanometer 3-14 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
≤ 0.01 (where a = absolute gas concentration accuracy, c = gas concentration) balance gas N2 T-shape connector Tubing Flowmeter Plug the module into the module rack and connect the sensor. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 3-15...
3.3.9 Sidestream and Microstream CO Tests See section 3.1.6Sidestream and Microstream CO2 Tests. 3.3.10 AG Tests See section 3.1.7AG Tests. 3.3.11 EEG Test You can choose either of the following methods to perform the test: 3-16 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 35
Adjust the resistance box to 1 kΩ, verify the resistance value displayed on the monitor is 1kΩ. Test the lead type of the monitor to B+, C+ and D+ respectively instead of lead A+. Set Montage Type: Monopolar Mode , then repeat the step 3~4. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 3-17...
After the cyclic impedance check is finished, check that the result for each electrode is passed. Method 3: Tools: Signal generator, (Maker: NF, Model:WF1946B) Covidien Signal simulator (Covidien PN:189-0137) 3-18 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Set the resistance value to 1kOhm. Connect the stimulation electrodes to the two wiring terminals. Set the multimeter to operate in DC mode. Connect the multimeter sensors to the NMT stimulation electrodes, making sure that the sensor and electrode connected have the same polarity. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 3-19...
Page 38
If the check fails, check whether the sensor is placed correctly as instructed, and perform the sensor check again. NOTE Stop NMT measurement or calibration before starting NMT sensor check. Avoid forcefully striking the sensor. 3-20 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Connect the patient simulator to the monitor using an ECG or IBP cable and connect the oscilloscope to the Auxiliary Output Connector of the MPM module of the patient monitor. Verify that the waves displayed on the oscilloscope are identical with those displayed on the monitor. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 3-21...
Electrical safety tests must be carried out under normal environmental conditions (that is, normal temperature, humidity and barometric pressure). See Appendix A Electrical Safety Inspection for electrical safety tests. 3-22 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Set the print resolution to 300dpi or 600dpi as required. Click Print Test Page to check whether the output of the printer's test page is normal. If not, recheck the connection and configuration of the printer. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 3-23...
Check that the trim strip is properly installed after the display is disassembled and repaired. Check that the monitor handle is not loose. Check that the VHM bracket can place the monitor at any height as required. 3-24 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
4.4 Software Version Check Some troubleshooting tasks may involve software version compatibility. For information about the configuration and software version of your patient monitor, contact Mindray service. To check the software version, do as follows: ...
Replace the Power switch power switch and indicator board indicator board damaged Power module Replace the power module. defective Replace the main control board. Main control board failure BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Touch the screen stably for 20 seconds. If the touch firmware is running screen still does not work, please restart the monitor. wrong. If the touch screen could not work yet, replace the front case assembly. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
COM board to the main control main control board to board is connected reliably. the internal module rack Verify the connecting cables and connectors COM board are not damaged. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Speaker failure Replace the speaker. Cable defective or Verify the cable connecting the speaker to the improperly connected main control board is properly connected. Main control board Replace the main control board. failure BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Touch screen failure of Replace the display. the external display the external display Touch pad firmware of the external display Contact the display manufacturer to fix the inconsistent with the problem. system software BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
View Others mode patient monitor according to the user manual. Check for IP conflict in the network. If conflict Incorrect network settings is found, reset the network. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
The BeneLink module The patient monitor application of the BeneLink module with application software is corrupted has no response when the network upgrading tool. loading the ID adapter BeneLink module damaged Replace the module. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Main control board Replace the main control board. failure Printing paper thermal Replace the printing paper. coating failure Poor printing effect Thermal head dirty Clean the thermal head. Recorder failure Replace the recorder. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 52
FOR YOUR NOTES 4-10 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
During removal, note to avoid breaking the connection line by pulling or damaging the connector. Place the removed screws and other parts separately by category so that they can be used in the re-installation. Do not drop, contaminate or lose them. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Recorder module You can insert and remove all the external modules during patient monitoring. Refer to the BeneVision N Series Patient Monitor Operator's Manual (P/N: 046-011259-00) for the use of parameter modules. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
2D Barcode scanner (support RFID) kit This monitor is configured with wireless network functions and can be connected to network through wireless AP. Contact Mindray Technical Support for assistance in connecting to a network. 5.3.1 Installing an SMR The SMR can be connected to the patient monitor through the SMR connector via the SMR cable.
NOTE During installation, use a blade to cut off the DVI seal at the rear case so that the interface can stretch out from the rear case. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 57
BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 58
FOR YOUR NOTES BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Be sure to follow the correct sequence when disassembling the monitor. Before removing assemblies, make sure that all the cables have been unplugged. During removal, note to avoid breaking the cable by pulling or damaging the connector. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
2. Use a pair of tweezers to pry up the four screw covers at the four corners of the rear case and loosen and remove the four M3×8 screws under them. (Screw covers) (Screws) BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
6 for the N15N17 series) from the bottom case; Pull out the base. For the N12 series machines, the base assembly can be removed when the base is pulled out. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
For the N12 series machines, remove the cable for the LCD screen and the cable for the small board of front case. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 63
As shown in the following figure, there are two types of front case assemblies (NLT or SHARP). Cable connections differ between NLT and SHARP. (N12 rear case assembly) (N12-NLT front case assembly) (N12-SHARP front case assembly) BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 64
(N15 rear case assembly) (N15-NLT front case assembly) (N15-SHARP front case assembly) (N17 rear case assembly) (N17-NLT front case assembly) (N17-SHARP front case assembly) BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
During installation, press the board to the alarm lamp shade direction by aligning with the front case board positioning rib. Use care to avoid overtightening the screws and stripping the holes when remounting the board. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Prevent pressure on the front face of display screen during disassembly. Remove the LCD screen assembly in an environment as dust-free as possible; the display screen and touch screen are integrated materials and cannot be disassembled. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
NOTE Use care when disconnecting the cable to avoid damage. 6.5.2 Further Disassembly of the Recorder First remove one PT2X6 screw, and remove the grounding piece at the same time. M2X6 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 68
Remove the other PT2×6 screw and take out the thermal printhead. Loosen and remove the two PT2×6 screws and remove the button board of recorder. Recorder button Thermal print head NOTE Use care when disconnecting the cable to avoid damage. 6-10 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
MPAN antenna connector inserted on the Bluetooth board; unscrew one M3×6 screw to remove the MPAN board. Tear the MPAN antenna fixing piece. Then, the MPAN antenna can be removed. NOTE Use care when disconnecting the cable to avoid damage. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-11...
Remove the adhesive tape fixing the antenna, and tear off the Wi-Fi sticker antenna. Pull out the Wi-Fi antenna plug from the board; unscrew the three M2X4 screws, and separate the Wi-Fi module from the Wi-Fi support board. Wi-Fi module Wi-Fi support board 6-12 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
COMM board cable. Remove four M3x6 screws and lift up to remove the battery holder assembly. Cable of battery interface board Cable of internal module rack COMM board Loosen and remove the two M3X6 screws, and remove the battery interface board. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-13...
Page 72
SMR interface, the two PT3X8 screws, and the six M3X6 screws in turn, and then remove the internal module rack COMM board. M3X6 M2.5X SMR interface sheet metal PT3X8 Turn over the removed internal module rack COMM board, and remove the four POGO PIN silicon cases. 6-14 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 73
Place the face of the removed module rack assembly board up. First unscrew the two M2.5X6 screws on the SMR interface, the two PT3X8 screws, and the seven M3X6 screws in turn, and then remove the internal module rack COMM board. M2.5X PT3X8 M3X6 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-15...
Disconnect the Speaker cable, unscrew the two M3X6 screws, and remove the speaker assembly. Speaker cable Loosen and remove the five M3X6 screws, and vertically take out the main support assembly. 6-16 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 75
AC input to the ACDC power supply board. . Unscrew four M3X6 screws, and remove the power board. Cable from the main control board to the ACDC Power Supply Cable between the AC input to the ACDC power supply board BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-17...
Page 76
Disconnect the speaker cable. Loosen and remove the two M3X6 screws, and remove the speaker assembly. Speaker cable As shown below, for the N15 series machine, loosen and remove the five M3X6 screws, and lift out the main support assembly. (N15) 6-18 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 77
AC input to the ACDC power supply board. Unscrew and remove the four M3X6 screws, and remove the power board. Cable from the main control board to the ACDC Power board AC input to ACDC power cord BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-19...
Connector to the ACDC Power Supply. NOTE Cable tie Use care not to damage the cable when cutting the tie. 6.5.7 Removing the iView Board Assembly (N17) First unscrew the two DVI stud screws. 6-20 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 79
As shown below, place up the face of the iView board side with the computer module. Loosen and remove the four M2X6 screws, and separate the computer module from the iView support board. Computer module iView support board BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-21...
Remove the independent display board cable and the internal module rack cable. Loosen and remove the four M3X6 screws, and then remove the independent display board assembly. Loosen and remove the four M3X6 screws, and remove the independent display board. 6-22 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Pry the two handle pins towards the center of the case, then remove the handle. BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-23...
Loosen and remove the two M3X6 screws, and remove the battery interface board. As shown below, open the battery door, insert a pad between the battery door and the base and press down firmly to remove the battery door. 6-24 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
As shown in the following figure, loosen and remove the eight M3X10 cross recessed screws, and remove the hooks. M3X10 As shown in the following figure, loosen and remove the three M3×8 cross recessed screws, and remove the handle. M3X8 screw BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-25...
Page 84
Loosen and remove the two M3X8 cross recessed pan head screws, disconnect the cable between the interface board and the internal module rack COMM board, and then remove the interface board. M3X8 screw Cable from the interface board of external module rack to the 8-slot board 6-26 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 85
Loosen and remove the seven M3X8 cross recessed pan head screws on the internal module rack COMM board, and remove the eight POGO PIN silicon cases from the module rack. Module rack antenna M3X6 screw POGO PIN silicon cases of module rack BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-27...
As shown in the following figure, loosen and remove the board screws. Remove pull out the pump and valve cable and NIBP air tube. Then, the parameter board can be removed. Parameter board screws and cable Remove along this direction 6-28 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 87
Disconnect the valve power cable and NIBP air tube, use a flat-bladed screwdriver to release the clips on the sides of the valve, and then remove the valve assembly. Release the clips Cut the cable ties Disconnect power lines BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 6-29...
Page 88
FOR YOUR NOTES 6-30 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Alarm LED and Light Sensor Board PCBA 051-002693-00 Cross recessed pan head screws, PT3X8 N12 front shell plate cable 009-006396-00* N12 Mainboard to LCD cable (NLT) 009-006410-00* Power Switch and Indicate LED Board PCBA 051-002711-00 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
MK power button (P+R) 043-007956-01 Earth plate (N12) 042-018667-00* Conductive cloth pad 0.08m CORE O.D=21.5mmI.D=9.5mm *: included in No.1 N12-Sharp front housing assembly (FRU) (115-044503-00) 7.1.4 N12 Rear Housing 7.1.4.1 Exploded View BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Page 93
Cross recessed pan head screw, M3X6 with pad N12 main bracket assembly (with DVI) TR6F recorder Handle Right Damping Block, N12 Self-Clinching Pilot Pins.TPS-3-18 Screw cap 2(T8) 049-000650-00 Cross recessed pan head crews,PT3X8 Conductive foam 2.0*7.0mm BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
ITEM No. Description FRU Part No. N12 battery cavity Cross recessed pan head screw GB/T818-2000 M4X30 plated with green color zinc Battery bar 9211-20-87256 Battery bar sleeve Cushion pad.Bumpon SJ5302, white and transparent BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
PT3X8 plated with green color zinc N15N17 front shell plate cable 009-006736-00* CORE O.D=21.5mmI.D=9.5mm 1 N15 Mainboard to LCD cable (NLT) 009-006408-00* N15N17 touch screen cable (NLT) 009-006397-00* Power Switch and Indicate LED 051-002711-00 7-10 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
7.2.6 N15 Base Assembly (FRU)(115-044543-00) 7.2.6.1 Exploded View 7.2.6.2 Parts List ITEM No. Description FRU Part No. N15 bottom case N15 Battery Door (Tether) 043-007966-00 Rubber foot pad BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 7-15...
7.3.6 N17 Base Assembly (FRU) (115-044550-00) 7.3.6.1 Exploded View 7.3.6.2 Parts List ITEM No. Description FRU Part No. N17 bottom case Rubber foot pad N15 Battery Door (Tether) 043-007966-00 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 7-23...
7.3.7.1 Exploded View 7.3.7.2 Parts List ITEM No. Description FRU Part No. Independent display assembly 115-044554-00 Small cross recessed pan head GB9074.5-88 M3X6 with pad and plated with green color zinc 7-24 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
FRU Part No. N17 heat sink (iView or independent display) 041-026485-01 Independent Display Board PCBA 051-002712-00 Small cross recessed pan head GB9074.5-88 M3X6 with pad and plated with green color zinc BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 7-25...
7.3.9.2 Parts List ITEM No. Description FRU Part No. N17 heat sink (iView or independent display) 041-026485-01 IVIEW CPU heat conducting pad IVIEW memory heat conducting pad Cross recessed pan head screw, M2X6 7-26 BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
No physical damage to meters, switches, connectors, etc. The enclosure and accessories No residue of fluid spillage (e.g., water, coffee, chemicals, etc.). No loose or missing parts (e.g., knobs, dials, terminals, etc.). BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...
Test items: 1, 2, 3, 4, 6, 7, 8 PCBA is repaired or replaced When both power supply PCBA and Test items: 1, 2, 3, 4, 5, 6, 7, 8 patient electrically- connected PCBA are repaired or replaced BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual...