After replacing the reagent, reboot the Trident or unplug the 1LINK cable to the Trident.
Error Condition

Test C Fail

Explanation of Error
Test C Fail
The "Test C Fail" error indicates there is an error in the titration associated with reagent C.
Recommended Course of Action
Test C Fail
This most often occurs because there is no reagent in the reagent bottle, but there is still reagent level
according to the Trident tile and Apex.
To resolve this, run the reagent C replacement task in Apex Fusion or see the guide to manually replace
reagents.
After replacing the reagent, reboot the Trident or unplug the 1LINK cable to the Trident.
Error Condition

Volts

Explanation of Error
Volts
A "Volts" error usually indicates an issue with the 24VDC power supplied to the Trident from the 1LINK or
power supply.
Recommended Course of Action
Volts- 1LINK
If you are using a 1LINK cable with the Trident, this error can be resolved by simply removing the 1LINK
cable from the Trident and then reconnecting it.
If the VOLT error continues, disconnect the far end of the 1LINK cable from the EB832 or 1LINK module
then reconnect it.
Volts- Aquabus and Power Supply