The fault is shown by an error code on the display (C) as following if the error occurs, e.g. Err. 5.
Proceed as follows in such cases:
F Switch the device switch (A) off.
F Remove the stirrer tool and remove the instrument from the assembly.
F Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.
Error code
Err. 5
Err. 4
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
In accordance with IKA
period is 24 months. For claims under the warranty please
contact your local dealer. You may also send the machine
direct to our factory, enclosing the delivery invoice and giv-
ing reasons for the claim. You will be liable for freight costs.
R 1825
Stand
R 1826
Stand
R 1827
Stand
R 182
Cross sleeve
R 1311 Turbine stirrer
R 1312 Turbine stirrer
R 1342 Propeller stirrer
R 1381 Propeller stirrer
R 1382 Propeller stirrer
R 1389 Propeller stirrer (PTFE)
R 1300 Dissolver stirrer
R 1303 Dissolver stirrer
R 1330 Anchor stirrer
R 1352 Centrifugal stirrer
Find Quality Products Online at:
Cause
Motor overload
Motor blocked or overloaded
warranty conditions, the warranty
®
Permitted IKA
recommended speed
(rpm)
≤ 2000
≤ 2000
≤ 2000
≤ 2000
≤ 2000
≤ 800
≤ 2000
≤ 2000
≤ 1000
≤ 2000
GlobalTestSupply
www.
Error codes
Effect
Solution
motor off
- Switch off the instrument
- Decrease the motor's load and restart again
motor off
- Switch off the instrument
Warranty
The warranty does not cover worn out parts, nor does it
apply to faults resulting from improper use, insufficient care
or maintenance not carried out in accordance with the in-
structions in this operating manual.
Accessories
RH 3
Strap clamp
FK 1
Flexible coupling
R 301
Stirrer shaft protection
R 301.1
Support holder
stirrer tools
®
R 3000.1 Moebius stirrer
R 3001.1 Moebius stirrer
R 3003
Spiral stirrer
R 3003.1 Spiral stirrer
R 3003.2 Spiral stirrer
R 3004
Blade stirrer
R 3004.1 Blade stirrer
R 3004.2 Blade stirrer
.com
recommended speed
(rpm)
≤ 800
≤ 800
≤ 800
≤ 800
≤ 800
≤ 1000
≤ 1000
≤ 1000
10
sales@GlobalTestSupply.com
Need help?
Do you have a question about the RW 20 digital and is the answer not in the manual?
Questions and answers