PULSE OXIMETER
INSTRUCTION
MANUAL
V1.0M122

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MD300 and is the answer not in the manual?

Questions and answers

John
April 9, 2025

What is error 7

Summary of Contents for ChoiceMMed MD300

  • Page 1 PULSE OXIMETER INSTRUCTION MANUAL V1.0M122...
  • Page 2 V1.0M122 PULSE OXIMETER INSTRUCTION MANUAL Copyright Our company owns all rights of this unpublished work and intends to maintain it as a confidential work. We may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for the purpose of reference or operation of our device.
  • Page 3: Table Of Contents

    V1.0M122 PULSE OXIMETER INSTRUCTION MANUAL Content 1 Introduction...................3 1.1 Brief Introduction................3 1.2 Safety Information .................3 1.3 Electromagnetism Interference ............5 2 General Descriptions ................6 2.1 Appearance ..................6 2.2 Explanation of Symbols ..............8 2.3 Power Supply..................9 2.4 Product Features .................11 3 Install the Probe..................11 4 Setting ID, Date and Time..............12 4.1 Date &...
  • Page 4: Introduction

    1 Introduction 1.1 Brief Introduction Thank you for purchasing the handheld pulse oximeter. The main functions of the device include SpO and PR measurements, visual and audible alarm, batteries charging, data storage and review and USB cable&Bluetooth transferring, etc. Please read this manual carefully before using the device.
  • Page 5 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL This equipment is for use in the medical field, and measurement results only serve as a reference for any relevant treatment. Connect the probe correctly; please see the directions for use of any accessories. Please follow the doctor’s suggestions when starting to monitor the vital sign parameters.
  • Page 6: Electromagnetism Interference

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Application of this device may influence the measuring accuracy in the background of electromagnetic areas such as electro-surgery environment. measurements may be adversely affected in the presence of high ambient light. Shield the probe area (with a surgical towel, for example) if necessary. Dyes introduced into the bloodstream, such as methylene blue, indocyanine green, indigo carmine, and fluorescein, may adversely affect the accuracy of the SpO reading.
  • Page 7: General Descriptions

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL This apparatus complies with the IEC 60601-1-2 international standard. The requirements of this international standard are: CISPR11, GROP1, and CLASS B. 2 General Descriptions The handheld pulse oximeter adopts 2.8 inch TFT screen, which can display the SpO2% and pulse rate value, other indication parameters, such as time, ID number, pulse amplitude bar and battery power status, alarm limits and the connections of Bluetooth and probe, etc.
  • Page 8 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL it down to continuously increase the value. Down button: press this button to decrease the value by one decrement. Or press it and hold it down to continually decrease the value. Menu/Select button: for entering main menu, or confirming the selection/setting. Alarm inhibit/off button: Press this button to inhibit alarm sound for 120 seconds.
  • Page 9: Explanation Of Symbols

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL ◆It shows two dashes throughout probe off and finger out conditions. 4:99: SpO % upper alarm limit indicator 5:90: SpO % lower alarm limit indicator 6: Pulse amplitude bar. It indicates the dynamic pulse amplitude and rate. As the detected pulse becomes stronger, more bars is illumined with each pulse.
  • Page 10: Power Supply

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Keyboard is unlocked Keyboard is locked USB cable is connected USB cable disconnected probe is inserted probe off Bluetooth activated Bluetooth deactivated 2.3 Power Supply 2.3.1 Powered by batteries Fig.2-4 Batteries Installation : 1) Open the battery cover: Rotate the fixing screw slightly in the rear panel to the position which is marked with “...
  • Page 11 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL . And the batteries are locked. Battery life and replacement There are five shapes of the indicator: the centre with 4 bars(full), 3 bars, 2 bars, 1 bar, empty and the frame in red. That the frame of indicator become red means few of battery capacity remains.
  • Page 12: Product Features

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL 3. It is not recommended to measure SpO during the oximeter is powered by wall outlet or being charged, for the damages and injuries may be caused to the device or users. 4. DO NOT charge when the non-rechargeable batteries are installed, otherwise, the damages or injuries may be caused to the devices or users.
  • Page 13: Setting Id, Date And Time

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL The socket is also used for USB cable connected with the PC for data transmission. For more information on data transmission and data management,refer to corresponding software instruction manual. 4 Setting ID, Date and Time Always set the date and time before using the unit for the first time.
  • Page 14: Id Number Setting

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig.4-2 Fig.4-3 Pick different sub-items to set and press the Select button to highlight it and then using the Up or Down button to adjust the value. At last, press the Select button to confirm. Set other sub-items of data and time according to the “Year setting”...
  • Page 15: Take A Measurement

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig. 4-4 5 Take a Measurement Before taking a measurement: Select the suitable probe in terms of type and dimension. Plug probe into SpO socket on top panel of pulse oximeter. Clip the patient finger to the rational position of the probe as the illustration as Fig.5-1. Fig.5-1 Placement of the probe Note:If the finger is not in the probe, “Finger Out!...
  • Page 16: Other Settings

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL integrity and correct positioning and adhesion of the probe. More frequently inspection should be taken depend on different patients if necessary. Inaccurate measurements may be caused by: Incorrect application or use of probe Significant levels of dysfunctional hemoglobins (such as carboxyhemoglobin or methemoglobin) Intravascular dyes such as indocyanine green or methylene blue Exposure to excessive illumination, such as surgical lamps (especially ones with a...
  • Page 17 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig.6-1 Set the low limit for SpO2 as the above steps of SpO2 high limit settings. The range of SpO2 lower limit is 70%-99%. 2) PR alarm setting Also, the limits settings of PR are performed as similarly as the SpO2 limits. The range of High limit is 31bpm-235bpm.
  • Page 18: Data Management

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL are divided by built-in module and can not be changed by user. Assignment of priority: High Medium Paramter Value Yellow Alarm lamp Flashing with yellow Flashing with yellow Lamp 1.5Hz 0.5Hz Frequency Audiblesound Di- Di – Di ----- Di - Di Di - Di - Di Alarm cycle 20 s...
  • Page 19 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig.6-3 Press the Up button or Down button select the sub-item to set, and then press the Select button to confirm or Return button to return to the previous screen. 6.2.1 Data Review Pick ”Data Review” sub-item and enter by pressing the Select button. The screen as in fig.6-4 will appear.
  • Page 20 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Select button to confirm, you can delete or review all the SpO trends saved under the ID. Fig. 6-6 Fig. 6-7 6.2.3 PR Trend The operation is referred to 6.2.2 SpO Trend. Fig.6-8 Fig.6-9 6.2.4 Clear Data Pick ”Clear Data”...
  • Page 21: System Setting

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig.6-10 6.2.5 In the data management screen, the upper half screen will display the maximum, minimum and average values of SpO2 and PR for the current ID. Refer to fig.6-11. Fig.6-11 6.3 System setting From the main menu, select and enter the System Setting screen, refer to Fig.6-12&6-13.
  • Page 22 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Brightness level: 1-7 Bright time: ON, 15, 30, 45, 60, 75, 90,105,120 seconds. 6.3.2 Bluetooth Enter BlueTooth screen from the System Setting screen, and set the Bluetooth on or off before transferring data to software for reviewing or analyzing data. Notes: For details, refer to attached Software Instruction Manual.
  • Page 23: Maintain And Repair

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Enter Default Setup screen from the System Setting screen, and you can restore the default configurations. Default configuration includes: low limit: 90% high limit: 100% SpO alarm: on PR low limit: 60 bpm PR high limit: 100bpm PR alarm: on Brightness Level:3 Bright time: On Bluetooth: Off...
  • Page 24: Safety Checks

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL delivered with the accessories. 7.2 Safety Checks Before every use, or after your pulse oximeter has been used for 6 to 12 months, or whenever your pulse oximeter is repaired or upgraded, a thorough inspection should be performed by qualified service personnel to ensure the reliability.
  • Page 25: Calibration And Verification

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL The applied parts touching the patients’ body are required to be disinfected once after each use. The recommended disinfectants include: ethanol 70%, isopropanol 70%, glutaraldehyde-type 2% liquid disinfectants. Disinfection may cause damage to the equipment and is therefore not recommended for this pulse oximeter unless otherwise indicated in your hospital’s servicing schedule.
  • Page 26: Trouble Shooting

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL 7.3.3 SpO & Pulse Rate Measurement Value Verification a).Connect SpO Probe to the SpO connector on the oximeter. b).Insert the operator’s finger into the finger probe, the SpO measured value of healthy person should be from 95% to 99%, and the pulse rate is same as heart rate, c).If SpO Simulator is available, verify the accuracy of Oxygen Saturation Value with probes as follows:...
  • Page 27 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL a) Our company isn’t responsible for such damage caused by force majeure. For example: fire, thunder flash, flood, cyclone, hail, earthquake, house collapse, commotion, plane failing and traffic accident, deliberate damage, lack of fuel or water, labor and capital bother, strike and stop-working etc.
  • Page 28: Storage And Transportation

    V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL professionals or authorized personnel disassemble oximeter and lead to oximeter failure. ●Not carefully read manual and so wrong operation lead to oximeter damage and failure. 7.5.5 User’s Special Request for Guarantee Time Our guarantee constitution for oximeter complies with electronic product after-sale service standard regulated by national laws.
  • Page 29 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Alarm mode: audible alarm, visual alarm and information Alarm limits range: SpO 70%-100%, PR 30-235 bpm Default limits: SpO High 100%, low 90%; PR High 100 bpm; low 60 bpm Display range: 0%~100% Measurement range: 70%~100% Resolution: Accuracy: 70%-100%...
  • Page 30 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL IPX1 Outline of product Dimension: 125mmX60mmX30mm Weight: 195g (excluding the batteries) Accessories: Standard accessories: 1. Three AA alkaline batteries. 2. One instruction manual 3. One adult finger probe: Model:M-50E. 4. The key for screwing the screw. 5.
  • Page 31 V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL...
  • Page 32 FCC ID: WWIMD300M122 This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.