Download Print this page

Operating Cautions - Sony MT10 Instruction Manual

Advertisement

Available languages

Available languages

Operating Cautions

Outline
The MT10 and MT11 are 5 µm resolution detectors for use exclusively with the DT12N/
DT12P scale unit. The MT10 is used to connect a LT20 Series or LY Series display unit.
(The DZ51 is required.) The MT11 has frayed-tip cables and outputs a differential A/B
phase signal.
Parts name and Dimensions
MT10
Scale unit connector
25/0.98"
MT11
Scale unit connector
25/0.98"
Do not route the connecting cable through the same duct as the machine power line.
Take preventive steps when the noises from other equipment may disturb the power
supply line to the units.
When providing DC power, be sure to use within the specified voltage range.
Do not pull the cable forcibly or pull at the cable when you connect or disconnect it.
Such handling may cause a break in the line.
Be sure to ground the probe frame to prevent misoperation due to noise or static
electricity.
For installation of the detector, avoid a location exposed to chips, cutting oil, or
machine oil.
If unavoidable, take adequate countermeasures.
The ambient temperature should be in the range of 0°C to 50°C (32°F to 122°F).
Avoid exposure to direct sunlight, hot air currents, or heated air.
MT10
98/3.86"
MT11
98/3.86"
Cable
Cable length : 300/11.81"
Cable
Cable length : 300/11.81"
Unit : mm/inch
Display unit
connector
1 (E)

Advertisement

loading

This manual is also suitable for:

Mt11