G100 CO 0-20% G110 CO 0-100% G150 CO 0-10,000ppm VIASENSOR 850 South Via Lata Suite 112 Colton, CA 92324 Toll Free: 855 VIASENSOR Tel: +1 (909) 783-9472 Fax: +1 (909) 825-0591 Email: sales@viasensor.com Website: www.viasensor.com LANDTEC North America, Inc., DBA VIASENSOR...
Page 2
G100 Range OMG100N1.17 This page is intentionally left blank Page 2 LANDTEC North America, Inc., DBA VIASENSOR...
TAKING READINGS ..................27 Preliminary Checks - Best Practice ..............27 Gas Measurement Process – Best Practice ............28 Alternative Reading Methods ................29 6.3.1 Logged Reading ..................29 Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 3...
Page 4
Technical Specification - G150 ............... 49 12.0 EVENT LOG ....................51 13.0 SAMPLE CERTIFICATE OF CALIBRATION ............52 14.0 EC DECLARATION OF CONFORMITY ............. 53 15.0 GLOSSARY OF TERMS – G100 RANGE ............54 Page 4 LANDTEC North America, Inc., DBA VIASENSOR...
Important/useful information and instructions are shown clearly throughout the manual in a note format. For example: Note: For further information, please contact Sales at VIASENSOR Toll Free (855) VIASENSOR, + 1 (909) 783-9472 or email Sales@viasensor.com. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 5...
Large data storage and user friendly software and download Easy-to-read, large well lit display Built-in gas moisture removal Applications: IVF/Medical Research Laboratories Brewing Atmosphere Control Incubators Gas mixing Page 6 LANDTEC North America, Inc., DBA VIASENSOR...
The G100 analyzer range has a number of optional products for purchase which enhance the usability and enable further analysis of data and reading information. Note: For further information, please contact Sales at VIASENSOR Toll Free (855) VIASENSOR, + 1 (909) 783-9472 or email Sales@viasensor.com. 3.1.1...
A unit calibrated with dry air could cause the readout to read low by up to 0.5% by volume. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 9...
G100 Range OMG100N1.17 Instrument Accessory Products Optional accessory and replacement parts may be purchased for the G100 from VIASENSOR direct. Please refer to the website www.viasensor.com for further information. Description Part Number USB Communication Cable G-USB Filters (Pack of 5)
Key 8 – Scroll Down Menu Key Enter Key Key 2 – Scroll Up Key 6 – Scroll Right Back View: Reference: Moisture Removal Tube Serial Number Instrument Stand Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 11...
Note: Do NOT attempt to remove the cover off the back of the analyzer which houses the moisture removal tube. Do NOT cover the moisture removal tube with your hand when holding the analyzer to take readings. Page 12 LANDTEC North America, Inc., DBA VIASENSOR...
Note: Temperature connectors are fitted with snap rivets to prevent dust ingress, remove before use. Grip back cap and pull upwards to release. Refit when not in use. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 13...
Switching the Instrument On To switch on the instrument, press the ‘On/Off’ key briefly. There will be a short beep and a slight pause followed by the VIASENSOR logo. The power on self-test will then commence (approximately 15 seconds) including warm-up time.
Page 15
G100 Range OMG100N1.17 Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 15...
For example, to enter a new time 09:25:00 the user would type in 092500 using the numeric keypad in the following sequence:- : : 0 : :09 : 0:92 :09:25 Page 16 LANDTEC North America, Inc., DBA VIASENSOR...
Press the ‘Menu’ key on the front of the instrument panel and the following screen is displayed: Main Menu Press the soft-key ‘Exit’ to exit the ‘Main’ menu. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 17...
From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel. Press ‘Key 2’ to display the ‘Utilities’ menu. Press ‘Key 1 – Time & Date’ and the ‘Set Time & Date’ menu is displayed: Page 18 LANDTEC North America, Inc., DBA VIASENSOR...
Press ‘Key 4 - Scroll Left’ and ‘Key 6 - Scroll Right’ to adjust the value displayed. Press the soft-key ‘Accept’ or ‘Reject’ accordingly to accept or reject the changes. Note: The manually set contrast setting is retained when the instrument is switched off. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 19...
From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel. Press ‘Key 2’ to display the ‘Utilities’ menu. Press ‘Key 4’ to select settings and the following screen is displayed: Page 20 LANDTEC North America, Inc., DBA VIASENSOR...
From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel. Press ‘Key 2’ to display the ‘Utilities’ menu. Press ‘Key’ 5 to select flow fail and the following screen is displayed: Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 21...
Page 22
Flow fail alarm? Press pump cancel alarm Blocked or excessive restriction Decrease displayed value by 1 Refer to Section 8.0 Problem Solving Accept Page 22 LANDTEC North America, Inc., DBA Flow Fail Set-up Process – Best Practice G100 Range VIASENSOR...
10 seconds if it takes 30 seconds to draw in a new sample. 3-ID 00000000 - Press ‘Key 3’ to create an 8-digit numeric ID Code. 4-Start/Stop Logging - Press ‘Key 4’ to start and stop data logging. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 23...
Press ‘Key 2 - Scroll Up’ and ‘Key 8 - Scroll Down’ to switch between the first (CO & Baro) and second (T & humidity) group of reading parameters. Press the soft-key ‘More’ to refine or filter the readings to view. Page 24 LANDTEC North America, Inc., DBA VIASENSOR...
To clear the readings press the soft-key ‘More’ followed by ‘Key 1 - Delete all’. 5.9.5 Diagnostics The ‘Diagnostics’ option enables VIASENSOR Technical Support to identify and resolve issues with the instrument and readings. If required, the operator may be asked to confirm the diagnostics displayed.
8-10 hours. When the instrument is already powered on, the operation is slightly different as the battery icon changes to a flashing plug symbol. This will stop flashing when the charge is complete. Page 26 LANDTEC North America, Inc., DBA VIASENSOR...
Don’t get the instrument wet, for example exposure to rain. Warning Always ensure that the exhaust gases exit in a safe manner into a well ventilated area. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 27...
The ID code prompt can be switched on or off. This option is accessed via the ‘Settings’ menu, then press ‘Key 1 - Prompt for ID: Yes or No’. If the ID prompt is set to ‘No’ the reading is stored using the default logging ID. Page 28 LANDTEC North America, Inc., DBA VIASENSOR...
The peak value is reset after a reading is stored or by exiting the peak mode using the appropriate soft-key. The current mode of operation can be identified by the status of the soft-key, either ‘Normal’ or ‘Inverse’, where inverse indicates peak mode is active. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 29...
The current mode of operation can be identified by the status of the soft-key which is inversed while in the hold phase. G150 Screen Shown: Hold Reading Page 30 LANDTEC North America, Inc., DBA VIASENSOR...
Only use gases with a known certified gas concentration. Note: Certified calibration gases can be supplied by VIASENSOR. Warning For each gas used the appropriate material safety data sheet must be read and understood before proceeding. Calibration gases and the use of pressure regulators can be dangerous.
Calibration Equipment The diagram below displays the regulator and tubing equipment for user calibration: Certified calibration gas cylinders can be supplied by VIASENSOR . Please refer to the VIASENSOR website www.viasensor.com for further information. The regulator supplied with the calibration kit is recommended as flow and pressure rates are factory set.
Press the ‘Start’ key and wait for the reading to stabilise. This can take a couple of minutes. Press the ‘Pump’ key to draw sample gas. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 33...
User Calibration - Reset To prevent the user calibration data being accidentally erased the user must confirm the action by pressing the soft-key ‘Accept’, or soft-key ‘Reject’ to exit without change. Page 34 LANDTEC North America, Inc., DBA VIASENSOR...
Note: If the calibration failed, then try again using a longer purge time or different target gas. This event log can only be downloaded and viewed via the optional Analyzer Data Manager software. It cannot be viewed on the analyzer screen. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 35...
This section outlines various warning and error messages which the operator may receive during general operation of the instrument. For further assistance please contact Technical Support at VIASENSOR at +1 (909) 783-9472 Ext. 7130. Warnings and Errors When switched on the instrument will perform a predetermined self-test sequence taking approximately 15 seconds.
The oxygen cell has not been changed for at least 3 years; its Cell performance and accuracy may be impaired. The instrument will need to be returned to the manufacturer for service. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 37...
Page 38
G100 Range OMG100N1.17 Page 38 LANDTEC North America, Inc., DBA VIASENSOR...
Check the span target is set to the correct value, if not, correct and retry spanning the channel. Repeat the entire procedure, including zeroing the channel and then calibrate the span. Ensure the reading is stable before spanning the channel. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 39...
Page 40
This optimizes the instrument at the span calibration concentration [D] for the current operational conditions and variations in user calibration gasses. It corrects the span point leaving the zero unadjusted and should be done at the concentration of normal operation. Page 40 LANDTEC North America, Inc., DBA VIASENSOR...
Press and hold the ‘On/Off’ key for 10 seconds; wait at least 15 seconds and the instrument should restart automatically. Note: Performing a hardware reset may cause loss or corruption of currently stored data including the time/date. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 41...
Page 42
G100 Range OMG100N1.17 Page 42 LANDTEC North America, Inc., DBA VIASENSOR...
Do NOT apply any moisture to the moisture removal tube on the rear of the instrument as this may damage the membranes. Do NOT use solvents or any other chemical cleaners. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 43...
12 months from the date of the shipment to the user. During this period VIASENSOR will repair or replace defective parts on an exchange basis. The decision to repair or replace will be determined by VIASENSOR.
±1.5%RH across the range Visual and audible alarm User selectable CO and O alarm levels Communications USB type B mini-connector, HID device class Data Storage 1000 reading sets + 270 events Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 45...
Page 46
Gas sample filters Built-in gas dryer tube to remove moisture User replaceable PTFE water trap filter Note: Due to VIASENSOR’S continuous program of improvement, this specification is subject to change without prior notice. Page 46 LANDTEC North America, Inc., DBA...
±1.5%RH across the range Visual and audible alarm User selectable CO and O alarm levels Communications USB type B mini-connector, HID device class Data Storage 1000 reading sets + 270 events Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 47...
Page 48
Gas sample filters Built-in gas dryer tube to remove moisture User replaceable PTFE water trap filter Note: Due to VIASENSOR’S continuous program of improvement, this specification is subject to change without prior notice. Page 48 LANDTEC North America, Inc., DBA...
±1.5%RH across the range Visual and audible alarm User selectable CO and O alarm levels Communications USB type B mini-connector, HID device class Data Storage 1000 reading sets + 270 events Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 49...
Page 50
Gas sample filters Built-in gas dryer tube to remove moisture User replaceable PTFE water trap filter Note: Due to VIASENSOR’S continuous program of improvement, this specification is subject to change without prior notice. Page 50 LANDTEC North America, Inc., DBA...
Attempt to store when readings memory full Readings memory nearly full Limit, Actual Change logging mode Status, Interval, Pump time Change logging mode ID Flow fail current limit exceeded Limit, Actual Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 51...
G100 Range OMG100N1.17 13.0 Sample Certificate of Calibration This is a sample certificate of calibration supplied at the time of purchase and updated when the instrument is serviced. Page 52 LANDTEC North America, Inc., DBA VIASENSOR...
OMG100N1.17 14.0 EC Declaration of Conformity Products G100 range of Gas analyzers VIASENSOR declares that the item(s) described above is/are in compliance with the following standards: EMC Directive 2004/108/EC EN 50270 :2006 Electromagnetic compatibility. Electrical apparatus for the detection and measurement of combustible gases, toxic gases or oxygen.
It can also be used as a diagnostic tool if there is a problem with the analyzer. The event log can be viewed via the – Analyzer Data Manager software. It cannot be viewed on the analyzer screen. Page 54 LANDTEC North America, Inc., DBA VIASENSOR...
Page 55
This can be a simple absorption filter as per a moisture trap, a catch-pot or simple return tube depending on the application. Oxygen gas typically 20.9% in air, measured via electro-chemical cell. Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 55...
Page 56
Zero The point at which the gas analyzer is calibrated when there is none of the target gas present. Page 56 LANDTEC North America, Inc., DBA VIASENSOR...
Page 57
G100 Range OMG100N1.17 Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 57...