Correlation window*
Use this window to create and display correlation charts and the data that displays in the
charts.
When needed, use the correlation data to create a real time correction formula in the
Analytical Parameters (Chemistry) window (or ISE Parameters of Setting window).
Perform any of the following tasks at this window:
•
display charts
•
display chart data
•
create charts (add data automatically)
Use this procedure if the measurement data comes from the system
•
create charts (add data manually)
Use this procedure if the measurement data comes from another blood chemistry system
•
print charts
•
delete charts
Calibration windows
Select
Calibration
View Calibration Curve window
Use the View Calibration Curve window to review the detailed calibration curve data for
a method, to restore a calibration, to monitor calibration trends, to obtain a summary of
all calibration information and to view RBL and calibration check information.
NOTE:
window is closed and cannot be viewed until the analyzer returns to the Ready or Wait
mode.
Calibration/RBL History
The system stores data for up to 100 calibration/RBL events for each of 2
different reagent lot pairs (R1 and R2), for a total of 200 data sets per method.
Use the Calibration/RBL History window to:
•
review the calibration history for all methods run on the system
•
print the displayed calibration data
•
generate a CSV file for displayed calibration data
System Overview
to display the following Calibration menu.
While running calibrations or processing samples, the View Calibration Curve
45