English By purchasing the seca 869 electronic In addition to the conventional determina- tion of weight, the seca 869 also has a personal scale, you have acquired a high- ly-accurate and sturdy piece of equip- function for determining body mass index.
• 6 mignon-type AA 1.5 V batteries Desk-top stand If you prefer, you can stand the display part rather than hang it on the wall: Alter the angle of the display part to enable the display to be read most easily. Model 869...
Fitting the wall bracket The display part is intended for mounting on the wall. Use the mounting material supplied to do so. – Screw the wall bracket to the wall as shown. Be aware that you have a length of 2 m of cable between the scale and the dis- play part.
– Then turn the scale back up the right way. Note: the scale is prepared for use with a seca mains unit which is availa- Connection for mains unit ble as an accessory. Connecting the connecting cable to the weighing platform...
Setting up and aligning the scale The screwed feet can be used to level off any slight unevenness in the floor. – Place the scale on a level surface. spirit level – Align the scale by screwing the foot screws. The air bubble in the spirit level must be right in the centre of the circle.
1.5 kg. The display flashes until a stable weight has been measured. The display is then frozen until the next weighing op- eration. HOLD appears in the display. – Read the weighing result off the digital display. Model 869...
– Ask the patient to step off the scale. 0. 0 0 The weighing result has been erased. The scale is ready for the next patient. – If a load greater than 250 kg, 550 lbs or St0P 39 sts is placed on the scale, pears in the display.
8. 8 . 8 . 8 . 8 . 0. 0 appear con- secutively in the display. The scale is then automatically set to zero and ready for operation. – Press the bmi key. On the display, you will see the last height entered. Model 869...
– You can use the arrow keys to adjust the value in increments according to the graduations displayed. – Once the correct value is set, press the bmi key again. The BMI function is now activated. – Step on the scale and keep still. –...
Page 11
Service department. 7. Servicing Your seca scale leaves the factory with an accuracy greater than ± 0.15 %. To ensure that this accuracy is maintained, the product must be carefully set up and regularly maintained.
Changing the batteries bAtt When appear in the display, you should change the batteries. You need 6 type AA 1.5 V batteries. – Carefully turn over the scale so that the base of the scale is accessible. – Open the cover. –...
Switchmode power adapter, 100-240V~ / 50-60Hz / 12V= / 0.5A seca no. 68-32-10-265 Attention! Use only seca original plug-in mains units with 9 V or controlled 12 V output voltage. Commercially-available mains units may supply a higher voltage than stated and thus lead to damage to the scale.
12. Declaration of conformity seca gmbh & co. kg hereby declares that the product meets the terms of the applicable European directives. The unabridged declaration of conformity can be found at: www.seca.com.