Vane/piston exl/h with explosion proof enclosure (34 pages)
Summary of Contents for Universal Flow Monitors FlowStream FP-EF-2N-360SCFH-CO2-X1B
Page 1
™ Laminar Mass Flowmeters User Manual Series: FP UNIVERSAL FLOW MONITORS, INC. 1755 East Nine Mile Road PO Box 249 Hazel Park, MI 48030-0249 TEL (248) 542-9635 FAX (248) 398-4274 http://www.flowmeters.com...
NAMEPLATE EXAMPLE FP-EF-2N-360SCFH-CO2-X1B FP-EF-2N-360SCFH-CO2-X1B ELECTRICAL SAFETY LABEL Serial numbers are formatted as YY MM ID 000 YY = year, MM = month, ID = product identifier, 000 through 999 = three-digit sequential number. FPMAN-1-190214 Page 4...
GENERAL SPECIFICATIONS Flow Ranges: 2 SLPM/5 SCFH F.S. to 1300 SLPM/2600 SCFH F.S. Turndown Ratio: 400:1 Accuracy: ± 1% of reading Repeatability: ± 0.2% of full-scale Maximum Measurable Flow: Up to 125% of full-scale, gas dependent Pressure Effect on Accuracy: Less than 0.04% of reading/ PSI ±...
OPERATION FlowStream flowmeters accurately measure the mass flow rate of most gases. The flow rate is determined by measuring the pressure drop across a unique internal restriction, known as Laminar Flow Element (LFE). The restriction is designed such that the gas molecules are forced into moving in parallel paths along the entire length of the passage for the entire range of operation of the device.
Since the temperature sensor is embedded inside the sensor block, if ambient temperature is different from gas temperature, there would be a discrepancy between what the sensor reads and the true gas temperature. The flowmeter body would track ambient temperature while gas temperature would heat/cool the body at a different rate.
WIRING DIAGRAMS Power Connector 5 Pins Load Resistor vs. Power Supply 4-20 mA Output Pin Wire color Function Brown +24VDC White 0VDC Blue Black Gray 4-20 mA with Alarm Option Pin Wire color Function Brown +24VDC White 0VDC Blue Alarm 1 Black Common Gray...
KEYPAD OPTIONS UP: Consecutive (short) button press cycles between the following information on top of the display. a. Gas: Current selected Gas b. Pressure: Gas pressure in PSI c. Pressure: Gas pressure in PSIA d. Totalizer: Total volume of gas according to the selected units. e.
CHG: Pressing CHG once shows the user menu. Once the desired menu item is selected, you can do one of the following: To change the set points Press SET to choose the particular menu item. Consecutive pressing of SET will cycle through the different digits of the set point. ...
When finished, keep holding the SET button down until “SAVED” is displayed, then release the SET button. FLOW LO: Low set point for flow alarm. Range is 0 to FLOW HI – 2% FS. If set to 0, alarm is disabled.
AVG: Moving average array size. Use SET, UP, DOWN buttons as explained above to change the set point. This is the number of samples that are averaged when calculating the flow. Range is 1-100. Samples are added to the internal moving average array and a new output is displayed once every 100 msec.
RESET: Reset the parameters back to the original factory settings. Hold the SET button until the letters RESET will be displayed on the LCD one by one. This will provide a 5 second delay that is intended to prevent and accidental reset.
AVAILABLE FLOW SIZES Each size is offered in several optimized maximum flow rates as shown in the table below. These configurations provide the highest accuracy and turndown ratios stated in the specifications. Please refer to the model code in the Ordering Information section to select the desired flow rate.
NOTICE RETURN MATERIAL AUTHORIZATION Please read the following UFM policy information carefully. By following the guidelines outlined below you will assist in providing a timely evaluation and response regarding the status of your flowmeter. UFM evaluates all AUTHORIZED RETURNED MATERIALS in a timely manner and will promptly provide notification regarding the status of the related materials and/or a written quotation indicating the total charges and description of the necessary repairs.
Page 19
If no disposition is received by UFM within 30 days, the goods will be scrapped and the $90.00 evaluation fee will be billed. WARRANTY RETURNS Warranty returns must be shipped prepaid to UFM. UFM will review the goods and advise the customer of the evaluation and validity of the warranty claim. Valid warranty claims will be repaired or replaced at no charge.
1) ACCEPTANCE AND INTEGRATION CLAUSE: This Sales Order Acknowledgment and the sales order information that Universal Flow Monitors, Inc. ("Universal") attaches to or associates with it (herein "Acknowledgment"), constitutes an acceptance by Universal of an offer by the buyer upon the conditions and terms and at the prices stated in this Acknowledgment.
Page 21
(b) Aircraft Applications. Any application involving direct or indirect installation in or on, or use in connection with, any aircraft or aircraft product. (c) Definitions. As used in this section, the following definitions apply, whether the terms use initial capitals or not.
Need help?
Do you have a question about the FlowStream FP-EF-2N-360SCFH-CO2-X1B and is the answer not in the manual?
Questions and answers