Error messages
Fault
"Er00" displayed
"Er01" displayed
"Er02" displayed
"Er03" displayed
"Er04" displayed
"Er05" displayed
"Er06" displayed
"Er09" displayed
Page 44
Operating Manual Tabeo Sintering Oven Metal
Possible cause
Door open during process
Excessive temperature
Broken sensor, measuring
circuit
Broken sensor
Thermocouple reverse polar-
ity
Thermocouple short-circuit-
ed or heater defective
Defective electronic system
Power failure
Faults and error messages
Troubleshooting
Close door ("Press")
Poss. readjust door locking
Replace the thermocouple.
Retighten the thermocouple con-
nections.
Replace the thermocouple.
Connect thermocouple correctly
+ orange line
- white line
Check the heater/thermocouple
and have it/them replaced if nec-
essary.
Check the electronic system and
have it replaced if necessary.
If there is a brief power failure
during
a
heating-up
or
ing-down phase and the tempera-
ture is below 1000 °C, the process
will be continued once power has
been restored. If the power failure
lasts longer than 10 seconds and
the temperature is above 1000 °C,
the process will be stopped and
the display will flash.
Version 4.0en
Responsi-
bility
User
Service
de-
partment
Service
de-
partment
cool-
Start/
The
Stop
but-
ton must be
pressed
to
acknowledge
error Er09.
10/2018
Need help?
Do you have a question about the Tabeo-2/M/Metal-120 and is the answer not in the manual?
Questions and answers