KM 1320 Earth Resistance Soil Resistivity Tester is specially
designed and manufactured for measuring earth resistance, soil
resistivity, earth voltage, AC voltage. Adopting the latest digital
and micro-processing technology, precise 4-pole, 3-pole and
simple 2-pole method for earth resistance measurement,
importing FFT and AFC technology, with a unique function of
anti-interference capability and the ability to adapt to the
environment, consistency of repeat testing, to ensure high
precision, high stability and reliability for prolonged measure,
which is widely used in electric power, telecommunications,
meteorology, oil field, construction, lightning protection,
industrial electrical equipment and other earth resistance, soil
resistivity, earth voltage, AC voltage measurement.
KM 1320 Earth Resistance Soil Resistivity Tester is composed
of host machine, monitoring software, testing wires, auxiliary
ground pillars, communication wires and others. The large LCD
display of host machine is with blue backlight & bar graph
indicating that can be seen clearly. At the same time it can
store 300 sets of data, fulfilling historical inquiry and online real-
time monitoring through monitoring software, dynamic display,
alarm indicator, and with the functions like historical data
access, reading, preservation, report forms, printing and so on.
GENERAL SPECIFICATIONS :
Function
Measurement Range
Measuring Mode
Measuring Method
Test Frequency
Short-circuit Test Current : AC 20mA max
Open-circuit Test Current : AC 40V max
Test Voltage Wave
Electrode Distance Range : Can be set 1m-100m
Shift
Backlight
Display Mode
Measuring Indicator
LCD Frame Dimension
LCD Window Dimension : 124mm×67mm
Standard Test Wire
Simple Test Wire
Auxiliary Earthing Rod
Measuring Rate
Line Voltage
D:\Chhaya\My Documents\Chhaya\backup\catlog\New Cat\KM 1320-3000B.cdr
®
Earth Resistance Soil Resistivity Tester
2 / 3 / 4 Wire
: Measurement of 2/3/4-pole earth resistance, soil resistivity, earth voltage, AC voltage
: Earth Resistance: 0.00Ω-30.00
Soil Resistivity: 0.00Ω -9000kΩ
: Precise 4-pole measurement, 3-pole measurement, simple 2-pole measurement
: Earth Resistance: rated current change-pole method, measurement current 20mA Max
Soil Resistivity: 4-pole measurement
Earth Voltage: average rectification(between P(S)-ES)
: 128Hz/111Hz/105Hz/94Hz(AFC)
: Sine wave
: Earth Resistance: 0.00Ω-30.00
Soil Resistivity: 0.00Ω -9000kΩ , automatic shift
: Blue screen backlight, suitable for dim places
: 4-digital super-large LCD display, blue screen backlight
: During measurement, LED flash indicator, LCD count down display, progress bar indicator
: 128mm×75mm
: 4 wires: each for red 20m, black 20m, yellow 10m, and green 10m
: 2 wires: each for red 1.6m and black 1.6m
: 4 rods: Φ10
m m×150mm
: Earth voltage: about 3 times/second
Earth resistance, soil resistivity: about 5 seconds/time
: below AC 600V
kΩ
kΩ, automatic shift
All Specifications are subject to change without prior notice