Advertisement

Quick Links

 
 
 
 
 
 
® 
 
ThunderBolt
User's Manual
 
 
1.  
 
 
 
 
 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ThunderBolt and is the answer not in the manual?

Questions and answers

Summary of Contents for Gold Standard ThunderBolt

  • Page 1             ®    ThunderBolt User’s Manual     1.            ...
  • Page 2: Table Of Contents

    2.4.  Biological Safety .......................... 4  2.5.  Safety Labels .......................... 4  3.  System Description  ........................ 5  3.1.  System  ............................ 5  3.2.  System Details  .......................... 5  3.3.  Component Details ........................ 6  4.  ThunderBolt Software ........................ 7  4.1.  Instrument Graphical User Interface (GUI)  ................... 8  4.2.  Other Software Applications  ...................... 9  5.  Running a Worklist ........................ 10  5.1.  Worklist Preparation  ........................ 10  5.2.  Samples Tab .......................... 11  5.3.  Microtiter Plates Tab  ........................ 13 ...
  • Page 3: Introduction

    The ThunderBolt is designed and manufactured in accordance with the safety requirements for electronic  and medical systems listed in the certifications section above to ensure that the instrument  functions  safely, both electrically and mechanically under normal use conditions.  The ThunderBolt is supplied in a  condition that allows for safe and reliable operation.    2.1. General Safety  The  ThunderBolt  must  only  be  operated  in  accordance  with  the  stated  intended  use.    It  is  recommended that the ThunderBolt is used only with the spare parts and accessories suggested or  provided by Gold Standard Diagnostics.  The use and maintenance activities defined in this manual  are intended to ensure the safety of the operator and the proper functionality of the instrument.  All  system  surfaces  must  be  dry  while  operating  the  ThunderBolt.    Gold  Standard  Diagnostics  TB‐0001.I   ...
  • Page 4: Electrical Safety

      This  should  be  done  cautiously  and  only  when  absolutely  necessary.  2.4. Biological Safety  Any  parts  of  the  ThunderBolt  that  have  come  into  contact  with  samples/test  reagents  are  to  be  treated as being potentially infectious.  Some of the general purpose reagents have the potential to  cause  irritation  of  the  skin  and  mucous  membranes.    It  is  recommended  that  the  operator  use  appropriate personal protective equipment (PPE) such as gloves, lab coat, and eye protection while ...
  • Page 5: System Description

    Instrument Power Specifications  Voltage: 100 – 265 V  Frequency: 50 – 60 Hz  Power Consumption: 12W without incubator, 120W max  It  is  recommended  that  the  instrument  be  operated  off  of  a  dedicated  power  source.    Ideally  the  ThunderBolt  should  be  placed  near  an  outlet  and  connected  to  a  surge  protector  or  an  uninterruptable power supply (UPS).   Instrument Dimensions and Weight  Width: 64 cm (25.2 in)  Depth: 57 cm (22.4 in) ...
  • Page 6: Component Details

    The  fully  integrated  barcode  scanner  supports  the  following  barcode  types:    codabar,  code39,  interleaved 2 of 5, code 93, code 2 of 5, IATA code 2 of 5, matrix 2 of 5, code 11, code 128, telepen,  UPC  A,  UPC  E,  EAN  13,  EAN  8,  MSI,  Plessey,  RSS‐14,  RSS‐14  Limited,  RSS‐14  Expanded,  China  Post  Code, and PDF417.  Incubator  The forced convection incubator heats to a maximum of 45°C.   Reader  The  ThunderBolt’s  on‐board  reader  is  an  ultra‐compact,  fully  integrated  LED  spectrophotometer  which contains the standard wavelengths of 405, 450, 490, 550 and 630 nm.  Custom wavelengths are  also available.  TB‐0001.I    Page 6 of 31 ...
  • Page 7: Thunderbolt Software

    Photometric range: 0 to 3.0 OD  Spectral range: 400 nm to 700 nm  Read time: approx. 2min/plate  Precision: 0 to 2.000 OD: +‐0.003 OD or CV < 1%, 2.001 OD to 3.000 OD: CV < 1.5%  Resolution: 0.001 OD  Linearity: < 1% 0 to 2.000 OD (typical), < 2% 2.001 OD to 3.000 OD  ThunderBolt  instruments  equipped  with  the  chemiluminescence  option  also  contain  a  chemiluminescence reader and can handle both EIA and CLIA assays.  Detection Type: glow  Spectral Range: 300nm‐500nm  Dark Count: 50 counts per second (standard)  Measuring time: 100ms (adjustable 100‐1000ms)  Dynamic range: 6 decades (orders of magnitude)  Orbital Shaker  The integrated orbital shaker features a no‐spill design which closely replicates manual shaking.  The  ThunderBolt can dispense and/or incubate while the shaker is in motion.  Amplitude: 1 mm  Adjustable speed: up to 900 RPMs  Capacity: 3 MTP (2 reaction, 1 predilution)  Voltage: 24 V  Netbook PC  The included netbook PC is attached to the instrument via a USB connection.  The PC also contains  ethernet, RS232, and additional USB ports.  PC Minimum requirements are listed below:  Processor: 1 GHz 32‐bit (x86) or 64‐bit (x64)  Memory (RAM): 1 GB (32‐bit) or 2 GB (64‐bit)  Memory (hard disk): 20 GB available (32‐bit) or 24 GB available (64‐bit)  Ports: USB 2.0 (RS232, Ethernet ports optional)  Integrated monitor: DirectX 9 graphics device with WDDM 1.0 driver, 800x600 resolution  Software  The pre‐installed EIA/CLIA Graphical User Interface software runs on the Windows based Netbook  device included with the instrument. The software provides the ability to create and run worklists, ...
  • Page 8: Instrument Graphical User Interface (Gui)

    5. Report Templates: Any custom report templates (if applicable) are stored in the Report  Templates subfolder.  GUI Initialization  1. Verify  that  computer  with  installed  GUI  is  attached  to  the  ThunderBolt  instrument,  power  is  connected to instrument and to computer, and instrument power button is in the on position.   Turn  on  computer  and  click  the  GUI  icon  (if  GUI  does  not  initialize  automatically  at  computer  startup) to initialize the system.  2. Wait  for  the  ThunderBolt  to  initialize.    The  instrument  requires  approximately  30  seconds  to ...
  • Page 9: Other Software Applications

    4. Racks: Allows the user to enable the Reagent Loading Wizard, (a window which prompts the user  to check the volume of each reagent before starting a worklist), enable/disable intelligent rack  and tube detection, and set reagent rack’s (and intelligent racks’, if applicable) size/definition.  5. Reports:  Allows  user  to  select  custom  laboratory  information  and  logo  for  worklist  reports.   Options for custom report templates are also available.  Default report template and GSD logo  will be used on worklist reports if no changes are made.  Contact local ThunderBolt distributor if  additional information about custom report options is required.  6. LIS  Communication:  LIS  connection  options  are  set  here  (described  in  further  detail  in  the  LIS  Configuration section of this manual).  7. LIS Test Mappings: Assigning of LIS test IDs to instrument test names is performed here (described ...
  • Page 10: Running A Worklist

    If  the  instrument  has  been  moved,  probe‐related  maintenance  has  been  performed,  or  if  different‐sized  racks  have  been  placed  on  the  instrument  since  the  previous  alignment,  also  perform instrument alignment prior to use.  2. Check  Settings:  Be  sure  to  adjust/view  the  GUI  Settings  (described  in  ThunderBolt  Software  section above) as necessary prior to running the instrument.  3. Prepare Kit: It is essential that all package insert instructions for reagent handling and storage are ...
  • Page 11: Samples Tab

    Multiple Tests: Multiple tests can be added if settings are compatible; only compatible  tests will be available in the Add Tests drop down menu once the first test file has been  chosen.  c. Test Files: New assays can be added to the Add Test drop down menu by copying and  pasting the test files into the Tests folder within the ThunderBolt EIA documents folder  before launching the GUI software.  d. Remove  Test:  Any  added  test  can  be  removed  from  the  worklist  by  selecting  it  in  the  Remove Test drop down list.  2. Add  samples:  Click  the  Add  Samples  button.    The  Add  Samples  window  will  appear  in  the ...
  • Page 12 Follow the detailed instructions for sample placement as included in the Add Samples window for  each Sample Entry Method while adding samples.    Important Notes:  Intelligent  Racks:  If  intelligent  rack  tube  detection  is  enabled,  samples  can  be  placed  in  any  desired locations in the sample racks.  If intelligent rack tube detection is not enabled, samples  must be placed in the designated blinking positions.  The Preloaded Sample Rack option can only  be used when intelligent rack detection is disabled.  Tube Placement: While placing samples, verify that samples are recognized (spaces with loaded  samples display yellow or blue; unoccupied spaces will remain white or grey).  If sample tubes are  removed prior to starting the worklist (and intelligent rack tube detection is enabled), they will be  removed from the worklist, and their identification will be lost.  To be included in the worklist,  they must be added again using one of the Sample Entry Methods described below.   Sample IDs: Sample IDs cannot be edited after placement.  To edit a sample ID, remove the sample  from the worklist and add it again with a new ID.  Duplicate sample IDs are not permitted; a suffix  will be added to the sample ID if a sample with a duplicate ID is added.  Sample Entry Methods:  a. Manual: Use for non‐barcoded samples with custom IDs.  Type the sample ID into  designated field and click Add (or press Enter on the netbook keyboard). Place the  sample in the rack as prompted and repeat for all manually‐labeled samples.  b. Auto Increment: Use for Auto‐numbered samples.  Change the Sample ID Template and  Start Number if desired and click Add.  Place samples in the sample rack, one by one as  prompted.  Samples will automatically be numbered consecutively, beginning with  Sample #1 and continuing with Sample #2, #3, #4 etc. (according to ID Template and  Start Number).  c.
  • Page 13: Microtiter Plates Tab

    When all samples have been added, click Done to exit the Add Samples window.  Display will return  to the Samples tab.  3. Additional Sample Tab Options (optional): Click on the sample name to select sample(s)  (selection is highlighted).  Use the CTRL and SHIFT keys to highlight multiple samples.      a. Remove Samples: Click this button to remove highlighted sample(s).  b. Multiply Samples: Click this button to change the replicate of highlighted sample(s).   Use the + and – buttons to adjust the indicated sample multiplication.  c. Drag and Drop: Samples will be pipetted in the order listed on the Samples tab.  Drag  and drop samples to adjust processing order as desired.  4. Query LIS (optional): If LIS‐prompted sample selection is desired and LIS connectivity has been  correctly configured, click the Query LIS button at this time.  LIS will select appropriate samples  for the chosen test(s) in the worklist.  5. Select Samples: If LIS‐prompted sample selection is not used, click on the box(es) next to the  samples to select them for test(s).  Check marks appear to indicate selection.    a. Multiple samples can be selected for a test simultaneously by highlighting multiple  samples, then clicking a box next to a sample to select all highlighted samples for a test.  b. To select all samples for a test, right click on the desired test name at the top of the  column.  All samples in the test column will be selected.  c. If the same testfile is run twice in the same worklist, Use the Smart Fill Tests button to  fill all available sample positions on the first testfile and select remaining samples to run  in the second testfile automatically.  Following is an example of a Samples tab that has been properly prepared for a worklist run:    Note: Rack Position is indicated for each sample (rack #: tube position #).  5.3. Microtiter Plates Tab  This tab is used to determine the number of wells/strips/plates required for the run by displaying all  well locations on the two microtiter plates (MTP) and the predilution plate.  Additional information  about each well is displayed on the screen when mouse cursor is placed over the well.  This tab also  includes predilution plate options.  The following steps are taken to complete this tab:  1. Verification: Visually verify that all desired test wells and samples appear on the MTP layout in  the correct plate locations. ...
  • Page 14: Racks Tab

    5.4. Racks Tab  This  tab  displays  the  reagent  rack  and  sample  racks  and  is  used  to  load  reagents,  calibrators,  and  controls, and to verify sample placement.  Additional information about each position can be seen by  hovering the mouse over the desired location.  The following steps are taken to complete this tab:  1. Reagents:  Load  reagents  into  the  ThunderBolt  reagent  rack  using  appropriately  sized  reagent  adapters and following the colors/descriptions indicated.    a. Be sure the notched edge of each reagent adapter is facing forward.  b. When a worklist is started, the Reagent Loading Wizard will also guide the user through  the process of adding reagents (if wizard is enabled in the GUI Settings). ...
  • Page 15 a. Use the Special Start button to start the worklist from a stopped/aborted position, from  a  specific  step  of  the  selected  testfile(s),  or  to  reread  the  plate(s)  (plates  can  only  be  reread if worklist has already been run).  6. Check: Check required reagent and calibrator/control volumes as prompted by the Reagent  Loading Wizard (if wizard is enabled).  If multiple bottles of a single reagent are required to  accommodate the indicated volume, use the Split Bottle option within the wizard window to  load additional bottle(s).  A split bottle can also be removed if desired, using the Remove Split  Bottle button.  7. Run:  Close  the  lid  as  prompted  and  wait  for  tests  to  run.    Worklist  actions  can  be  monitored  throughout the run using the log file and the live camera feed.  The bar at the bottom of the screen ...
  • Page 16: Tools Tab

    Tools  1. Home Instrument: Use this button to “home” the instrument, just as is done automatically at GUI  startup.  2. Park  Instrument:  Use  this  button  to  reset  the  ThunderBolt  back  to  its  starting,  or  “parked”  position (with the probe assembly above the wash cup).  3. Prime Instrument: In the pop up window, select desired wash bottle(s) and number of priming  cycles  and  click  Start.    If  both  washes  are  selected,  the  set  number  of  priming  cycles  will  be ...
  • Page 17: Alignment And Calibration

    read using a cursor to highlight.  CTRL and SHIFT buttons can also be used to highlight specific wells  on either/both the left and right plate(s).    Set the applicable options for the read type:  1. Manual Read Plate ODs: Set measurement wavelength and reference wavelength (optional).  2. Manual Read Plate RLUs: Set integration time.  Label the read data using the comment field (optional).  Click start.  Measurements will appear on the  screen as the plate is read.  When read is complete, export readings using the Export to CSV button  or print using the Print button.  Note: Manual readings are NOT saved automatically by the software.  6.3. Alignment and Calibration  These  instructions  are  intended  to  provide  an  overview  of  these  functions;  detailed  step  by  step  instructions  are  included  in  pop‐up  windows  within  the  instrument  GUI  when  each  function  is  performed.  ...
  • Page 18 Instructions: Align Instrument: The user aligns the X and Y‐axis positions of the probe with the  sample racks and microtiter plates.  The Z‐axis position, which moves the probe up and down, is not  manually aligned by the user: it is aligned automatically by the instrument during this procedure.   1. Click on the Align Instrument button.  A pop up window will appear and provide instructions for  each alignment step.   2. To align the probe at each position, use the buttons or keys to position the probe needles over  the center of each location as prompted.  Some example alignment positions are shown below:  a. Wash cup alignment step: Align the probe to the middle of the wash cup, then move it  down to verify that the probe can reach the bottom of the wash well without resistance.   Proper alignment is shown here:    Note: After the aligning the wash cup and clicking Next, the instrument will automatically  calibrate the z‐sensor to the bottom of the wash cup.  b. Sample tube alignment steps: Align the center of the probe needles to the center of the  sample tube (X and Y axes).  Proper alignment is shown here:    3. After  completing  all  probe  alignment  positions,  remove  the  Alignment  Plates  or  MTPs  as  prompted, prior to continuing.    Note: If these are not removed, the Z axis position cannot be properly set by the GUI.  4. At the final alignment step, ensure that the probe is centered above the left MTP carrier position,  as prompted.  The correct placement of the probe is shown below with the tip centered between  four holes and roughly above the word “Hot” printed on the incubator surface:    TB‐0001.I ...
  • Page 19   5. Click Finish when prompted to complete and save the alignment and exit the window (instrument  will return to its home position).  Instructions: Auto Align Instrument: After the user sets the starting position of the probe on the  MTP carrier, the instrument aligns all axis positions (X, Y, and Z) automatically during this procedure.  1. Click on the Auto Align Instrument button.  A pop up window will appear and provide instructions  for each alignment step.   2. Align the probe needles to the edge of the MTP carrier as prompted.  Y axis (front to back) position  of the probe must be exactly above the rear carrier lip.   X axis (left to right) and Z axis (up and  down) positioning are not important at this step.  Proper alignment is shown here:    3. Verify proper plate and plug positioning when prompted and allow the instrument to complete  the rest of the alignment automatically (This step takes approximately 10‐15 minutes).  4. Click Finish when prompted to complete and save the alignment and exit the window (instrument  will return to its home position).  Align Reader  The reader alignment process is performed to ensure that the reader is in the correct position for  reading MTPs.  It is recommended that the reader alignment be performed and recorded as a part of  the routine instrument maintenance.  General Information  1. Remove  all  microtiter  plates  and  Alignment  Plates  from  the  MTP  carrier  prior  to  starting  the  reader alignment procedure. ...
  • Page 20: Maintenance

      It  is  recommended that the startup and shutdown steps be performed any time the instrument is idle for  four or more hours to prevent build‐up in the fluidics lines.  Startup  1. Ensure wash bottle(s) are properly connected to the ThunderBolt.   2. Prime  the  instrument  using  the  wash  buffer(s)  that  will  be  used  for  testing  a  minimum  of  10  priming cycles per line.    3. Ensure that all wash buffers are within their expiration date prior to each use. ...
  • Page 21: Weekly Maintenance

    Shutdown  1. Prime the instrument using DI water a minimum of 50 priming cycles per line.    2. Ensure all instrument surfaces are clean and dry.  3. Remove any remaining reagents and samples from the ThunderBolt and close the lid.    4. Shut down the instrument by shutting down the attached computer.  5. Empty waste bottle.  7.2. Weekly Maintenance  Weekly maintenance includes cleaning and alignment steps that are to be performed approximately  every seven days.  It is recommended that the same day of the week be used for consistency.  Care  should be taken to ensure that no excess liquid comes into contact with the instrument surfaces while  cleaning.  Cleaning should only be performed with the recommended reagents.  Weekly Maintenance  steps should be performed as follows:  1. Remove sample racks and inspect each one for cleanliness and damage.    a. Use isopropyl alcohol to gently clean all sample racks.  2. Clean the sliding reagent tray deck with isopropyl alcohol.  3. Clean the outside of both probe needles by wiping them down gently with an alcohol wipe.  4. Perform instrument and reader alignments as described in further detail in the Alignment and  Calibration section of this manual.  5. Decontaminate the instrument tubing by priming the instrument 20 priming cycles per line with  Liquinox solution.   Note: The Liquinox solution is prepared by making a 1:100 dilution of the Liquinox Concentrate  supplied with the instrument, using DI water.  6. Rinse the tubing by priming the instrument 50 priming cycles per line with DI Water.   7. It is recommended that wash bottles be cleaned weekly to avoid contamination.  7.3. Monthly Maintenance ...
  • Page 22: Periodic Maintenance

    8. If  the  liquid  level  fluctuates  significantly  or  the  cup  fills  with  liquid  in  the  two  previous  steps,  readjust tubing and monitor priming again.  If problem persists, contact technical service.  7.4. Periodic Maintenance  Approximately every year (time will vary depending on instrument usage) a professional maintenance  procedure  should  be  performed  by  a  trained  service  engineer.    Service  agreements  are  typically  laboratory‐dependent and established to meet the specific needs of each client.  Contact the local  ThunderBolt distributor for more information.  7.5. Storage, Transportation, and Disposal  Storage, transportation, and disposal of the ThunderBolt must be completed by a trained technician.   A complete reinstallation must be performed following any move or storage of the instrument.  The recommended storage and transportation conditions for the instrument should be similar to the  use conditions.  If the instrument is moved from its installed location or removed from service for an ...
  • Page 23: Laboratory Information System Integration

    8. Laboratory Information System Integration  The LIS implementation of the ThunderBolt completely conforms to the NCCLS LIS2‐A2 and NCCLS LIS01‐ A2 (formerly known as ASTM E1394‐91 and ASTM E1381‐91) specifications.  The ThunderBolt can  communicate successfully with LIS systems that also conform to these specifications.  Please be aware  that some LIS companies will need to create special drivers before they will be able to be integrated  with the ThunderBolt.  In addition to the TCP‐IP implementation of NCCLS LIS01‐A2, the software also supports a “clean” TCP‐IP  transmission.  In this mode the software will send the unmodified NCCLS LIS2‐A2 frames including the  <CR> at the end of the frame.  No checksum or control characters are added.  All LIS Settings are found within the LIS Communication and LIS Testname Mappings sections of the GUI  Settings.  LIS is queried from the Samples tab during worklist setup and results are sent to the LIS from  the Worklist tab after run is complete.  8.1. LIS Records  Data is sent to the LIS according to the following:  Header Record: All communications begin with the header record.  In the “Sender Name or ID” field,  the name of the software and the version are displayed.  Example:  H|\^&|||Thunderbolt EIA^0.42.0.73|||||||P|LIS2‐A2|20101022162157  Request Information Record Specifics: The software will query the LIS system using “Request  Information Records”.  The field used to transmit the Sample ID to the LIS is the “Starting Range ID  Number” field. In this field the first component is used: “Patient ID”  Example:  Q|1|S001^^||ALL  In this example, the Sample ID is “S001”  Patient Information Record: The LIS system will respond to the software using “Patient Information  Records”.  In the “Laboratory Assigned Patient ID” field the Sample ID is stored.  Example:  P|1||S001||  In this example, the Sample ID is “S001”  Test Order Record: The LIS system will give specific Patient/Test information using “Test Order  Records”.  In the “Specimen ID” field the Sample ID is stored.  From the “Universal Test ID” field the  “Manufacturer defined test code” component is used to store the test information. This can be  defined by the lab; a lookup table is used to match those codes to the real test names.  Example:  O|1|S004||^^^CMVIgG|R  In this example, the Sample ID is “S004”, the test name is “CMVIgG”.  Result Record: The instrument will send the results of patients back to the LIS using “Result  Records”.  In the “Data or Measurement Value” field, the numerical result of a patient is stored. ...
  • Page 24: Lis Query And Transmission Examples

    Examples:  R|1|^^^CMVIgG|1.33|ng/ml||||F  R|1|^^^CMVIgG|1.33^high|ng/ml||||F  R|1|^^^CMVIgG^Subtest1|1.33|ng/ml||||F  8.2. LIS Query and Transmission Examples  Example 1: Query of sample IDs   Instrument  LIS  H|\^&|||Thunderbolt EIA^0.42.0.73|||||||P|LIS2‐A2|20101022162157  Q|1|S001^^||ALL  Q|2|S002^^||ALL  Q|3|S004^^||ALL  L|1|N  LIS  Instrument  H|\^&|||LIS|||||||P|LIS2‐A2|20101022162157  P|1||S001  O|1|S001||^^^CMVIgG|R  P|2||S002  O|1|S002||^^^CMVIgG|R  O|2|S002||^^^HPLIgG|R  P|3||S004  O|1|S004||^^^CMVIgG|R  O|2|S004||^^^HPLIgG|R  L|1|N  Example 2: Transmission of results  Instrument  LIS  H|\^&|||Thunderbolt EIA^0.42.0.73|||||||P|LIS2‐A2|20101022162157  P|1||S001  O|1|S001||^^^CMVIgG|R  R|1|^^^CMVIgG|1.33|ng/ml||||F  L|1|N  Example 3: Transmission of multiple test results per patient  Instrument  LIS  H|\^&|||Thunderbolt EIA^0.42.0.73|||||||P|LIS2‐A2|20101022162157  P|1||S001  O|1|S001||^^^CMVIgG|R  R|1|^^^CMVIgG|1.33|ng/ml||||F ...
  • Page 25: Lis Configuration

    P|1||S001  O|1|S001||^^^CMVIgG|R  R|1|^^^CMVIgG^ST1|1.33|ng/ml||||F  R|2|^^^CMVIgG^ST2|1.24|ng/ml||||F  L|1|N  8.3. LIS Configuration  1. Verify that instrument is correctly installed and all power correctly connected to instrument and  netbook.  a. Verify that tests have been added to the Tests folder of instrument netbook.  2. Ensure that all relevant cords and cables for LIS system are connected to the instrument.  a. TCPIP: Ethernet cord; RS232: serial cord  3. Select correct settings within the GUI software LIS Communication section.  These must be  entered before any communication with LIS is attempted.  Select appropriate settings according  to LIS being used:  a. In the RS‐232 Connection section, set the baud rate, data bits, parity, and stop bits of the  port.  b. In the TCP/IP Connection section, set the Server address and the Server port of the LIS  TCP Server.  i. The “Use Frame Pooling” option allows the system to put more than one ASTM  message (for example a Header Record) into a  TCP Packet.  This  can speed the  transfer.  ii. If the “Use LIS01‐A2 Encoding” option is used, the software will use a “telnet”‐  like system (same data as through the RS‐232 connection).  4. Assign LIS test IDs to instrument test file names within the GUI software settings in the LIS Test  Mappings section.  List all LIS names and corresponding test file names, being careful to copy  codes exactly as they are used in the LIS system.  a.
  • Page 26: Rs-232 Pin Layout

    8.4.  RS‐232 Pin Layout        TB‐0001.I    Page 26 of 31 ...
  • Page 27: Troubleshooting Guide

    9. Troubleshooting Guide  Issue  Details  User Action  Instrument GUI  Is "No Instrument Found"  Check that power switch on the back of the  software will not  error displayed?  instrument is turned on.  Check that power cord is  start up  plugged into the instrument and the outlet.  Check  that USB cable from instrument to computer is  secure at both ends.  Try rebooting.  If problem  persists, contact technical service.  Run will not start  Does instrument display  Check that MTP plate(s) and predilution plate is/are    an MTP plate error?  correctly loaded and seated securely.    Check that sufficient wells are placed.    Check positions of probe during instrument  alignment.  If misaligned, realign instrument and  restart.  Does instrument display a  Prime 10x, lift probe briefly (1‐2sec) to check for  clog error?  steady dispense flow.  If dispensing appears  correct, restart run.  If not, carefully use stylet to  clean probe.        Check if blue liquid detection light is functioning    properly by placing a small tube of liquid beneath    the needles.  Blue light should go on when in liquid ...
  • Page 28 Issue  Details  User Action  Run stops in the  Is there an IO or  Power off computer and instrument, unplug and  middle of a  read/write error message  replug power and  USB connections, and restart.   worklist  displayed?  Load worklist and rerun from aborted position.  If  problem persists, contact technical service.    Is the machine jammed  Turn off machine, carefully remove plate and move  during the read step?  reader back, load worklist, place plate back in, start  worklist from the read step.    Is the probe assembly  Make sure there are no obstructions to probe  jammed?  movement.  Make sure probe tubing is zip‐tied to  flat cable and not being caught anywhere.    Is there a drop between  Pause the worklist, wipe off any visible drops or  needles error?  clots, clean with an alcohol wipe.  Resume worklist.   When worklist finishes, carefully clean the probe  with the stylet and perform Liquinox  decontamination.  Waste Bottle  Are there bubbles or liquid  Wipe connector and needles completely dry.  If  alarms during run  on the waste sensor  alarm persists, disable waste bottle alarm in    needles or the white ...
  • Page 29 Issue  Details  User Action  Failed run (cont.)  Do points on curve appear  Check if physical positioning of calibrators and    out of order on report?   controls in sample rack matches Racks tab.  If not,  Do high and low controls  contact technical service to produce a manual  appear switched?  report.  If test is in duplicate: Do  Contact technical service to produce a manual  calibrator/control  report, omitting the outlier control/calibrator  replicates show large  replicate(s) or use the Drop Bottles Option in the  differences?  Test Properties window.    Does it appear that only a  Check for sufficient volume and bubbles in    single calibrator or control  reagents/calibrators/controls.  Rerun with sufficient    is missing?  volume, bubbles removed.  Note: Even if no    bubbles are observed at the end of the run, a    bubble could have caused the problem and later    popped.     Do all OD values on report  Check position of red light during reader alignment.     appear elevated? ...
  • Page 30 Issue  Details  User Action  Flooded Run  Is the flooding outside the  Check positions of probe during instrument  (cont.)  wells on the MTP carrier?  alignment.  If misaligned, realign instrument and    rerun worklist.    Perform a manual shake to check if shaker homes  properly at the end of the shake.  Repeat manual  shake several times to confirm.      Check if there is a visible clog on the probe.  Even if    no visible clog is seen, carefully use stylet to clean  both probe needles.      Check that MTPs are correctly loaded and seated  securely.      Check that the correct number of wells are loaded  in the MTP using the Microtiter Plates Tab.  Need to reprint or  Does user know which  Retrieve data from the Worklist or Worklist Report  find past data  date the desired data was  folders of the instrument netbook.  run?  Dripping Probe  Is the dripping coming  While priming the instrument, follow tubing at the  Needles  from the dispense needle?  top of the probe and check for any droplets.    ...
  • Page 31: Contact Information

      10. Contact Information  10.1. Device Manufacturer  Company Name  Gold Standard Diagnostics  Company Address  2851 Spafford Street  Davis, CA 95618  Country  United States  Phone  530‐759‐8000  Fax  530‐759‐8012  Website  www.gsdx.us  ThunderBolt®  Catalog #  00300;  00300‐CL  10.2. Authorized Representative  Company Name  Emergo Europe  Company Address  Molenstraat 15   2513 BH, The Hague  Country  The Netherlands  Phone  +31 70.345.8570  Fax  +31 70.346.7299  Website  www.emergogroup.com     TB‐0001.I   ...

Table of Contents