3.1.3 KaVo Primus 1058 Life C ..................19 3.1.4 KaVo Primus 1058 Life C with right-side set-up kit ..........19 3.2 Patient chair Standard and COMPACTchair................ 20 3.3 Device body with patient unit ..................21 3.4 Dentist unit versions ...................... 22 3.4.1...
Page 4
Instructions for use Primus 1058 Life Table of contents 3.10.1 Rating plate 1440 ....................36 3.11Technical data ....................... 36 3.12KaVo Service table 1568 (optional accessory) ..............43 4 Operation..........................44 4.1 Switching the device on and off..................44 4.2 Adjusting the dental chair ....................45 4.2.1 Adjusting the arm rest (optional)................
Page 5
Instructions for use Primus 1058 Life Table of contents 6.4 Dentist unit ........................89 7 Safety check - Test instructions ..................90 7.1 Introduction ........................90 7.1.1 General instructions .................... 90 7.1.2 Notes for medical electrical systems..............91 7.1.3 Components of the safety check................92 7.1.4 Testing intervals ....................
Instructions for use Primus 1058 Life 1 User instructions | 1.1 User guide 1 User instructions 1.1 User guide Requirement Read these instructions prior to first use to avoid misuse and prevent damage. 1.1.1 Abbreviations Explanation brevi- ation Instructions for use Care instructions Assembly instructions Technician's instructions...
For further information, please visit: www.kavo.com 1.3 Terms and conditions of warranty KaVo provides the final customer with a warranty that the product cited in the handover certificate will function properly and guarantees zero defects in the material or processing for a period of 12 months from data of purchase, subject...
Terms and Conditions, Art. 28). Outside Germany Note KaVo shall not be held liable for damage arising from transportation. The shipment must be checked on arrival. If the packaging is visibly damaged on delivery, please proceed as follows: 1. The recipient of the package must record the loss or damage on the delivery receipt.
Instructions for use Primus 1058 Life 1 User instructions | 1.4 Transportation and storage 1.4.3 Information on the packaging: Storage and transportation Note Please keep the packaging in case you need to return the product for servi- cing or repair. The symbols printed on the outside are for transportation and storage, and have the following meaning: Transport upright with the arrows pointing upwards!
ISO 7494 equipped with a patient chair in accordance with ISO 6875. This KaVo product is designed for use in dentistry only and may only be used by trained medical personnel. Any other type of use is not permitted.
Page 11
"Proper use" includes compliance with all information in the Instructions for Use and ensuring that all inspections and service tasks are performed. The KaVo service table 1568 is an accessory of KaVo treatment centres that do not feature a dentist element mounted to the unit. The KaVo Service table 1568 serves as a support surface for the dentist during the treatment.
Note The MULTIflex couplings, the current K/KL motors, and the ultrasonic scaler hoses of KaVo are equipped as standard with a protective device to prevent treatment water from being drawn back into the treatment centre via the handpieces. If products from other manufacturers are used at the standard-...
The KaVo X-ray viewers 1440 are designed for radiograph viewing in dentistry and comply with the requirements of DIN 6856-3. These KaVo products are designed for use in dentistry only and must be used by trained medical personnel only.
Page 14
Instructions for use Primus 1058 Life 2 Safety | 2.3 Safety instructions WARNING Inappropriate operating conditions. Impairment of the electrical safety of the device. ▶ It is essential to comply with the operating conditions specified in the "Technical Specifications" chapter. WARNING Application of un-authorised accessories or un-authorised modifica- tions of the product.
Instructions for use Primus 1058 Life 2 Safety | 2.3 Safety instructions CAUTION Malfunctions due to electromagnetic fields. The product meets the applicable requirements regarding electromagnetic fields. Given the complex interactions between equipment and cell phones, the product may be influenced by a cell phone that is in use. ▶...
Page 16
Instructions for use Primus 1058 Life 2 Safety | 2.3 Safety instructions CAUTION Risk of injury by suspended instruments (S table). Patients may get injured by sharp instrument tips. ▶ When you move the dentist's unit, make sure that nobody is injured. ▶...
Page 17
Instructions for use Primus 1058 Life 2 Safety | 2.3 Safety instructions CAUTION Third-party device connection assembly kit (optional): Risk of reverse contamination by stagnant water. Infections. If a water consumer is connected to the third-party device connection as- sembly kit, always perform the following actions on the device: ▶...
Instructions for use Primus 1058 Life 3 Product description | 3.1 Treatment unit versions 3 Product description 3.1 Treatment unit versions 3.1.1 KaVo Primus 1058 Life TM 3.1.2 KaVo Primus 1058 Life S 18 / 126...
Instructions for use Primus 1058 Life 3 Product description | 3.1 Treatment unit versions 3.1.3 KaVo Primus 1058 Life C 3.1.4 KaVo Primus 1058 Life C with right-side set-up kit 19 / 126...
Instructions for use Primus 1058 Life 3 Product description | 3.2 Patient chair Standard and COMPACTchair 3.2 Patient chair Standard and COMPACTchair ① Headrest ② Backrest ③ Chair base ④ Seat ⑤ Arm rest (optional) 20 / 126...
Instructions for use Primus 1058 Life 3 Product description | 3.3 Device body with patient unit 3.3 Device body with patient unit ① Patient element ② Unit body The central control is housed in the unit body. ③ Pressurised water bottle ④...
Instructions for use Primus 1058 Life 3 Product description | 3.4 Dentist unit versions 3.4 Dentist unit versions 3.4.1 TM/C table ① Handle ② Turbine (multiflex coupling) ③ INTRA LUX motor KL 703 or ④ Ultrasonic scaler INTRA LUX motor KL 701 ⑤...
Instructions for use Primus 1058 Life 3 Product description | 3.4 Dentist unit versions 3.4.2 S table Note The holder assignment and arrangement of the instruments can be changed as needed and does not have to follow the picture. ① Control element ②...
Instructions for use Primus 1058 Life 3 Product description | 3.5 Assistant element – Versions 3.5 Assistant element – Versions 3.5.1 Standard assistant unit ① Three-function or multifunctional ② Spray mist suction handpiece ③ Control element ④ Saliva ejector ⑤ Satelec Mini LED ⑥...
Instructions for use Primus 1058 Life 3 Product description | 3.5 Assistant element – Versions 3.5.2 Assistant element right, left (optional, only in conjunction with patient chair Standard) ① Triple function handpiece ② Spray mist suction ③ Control element ④ Saliva ejector ⑤...
Instructions for use Primus 1058 Life 3 Product description | 3.6 Three function handpiece (3F handpiece) 3.6 Three function handpiece (3F handpiece) ① MF handpiece hose ② Gripping sleeve ③ Media buttons (air/water) ④ Labelled blue: Three-function hand- piece (3F handpiece) ⑤...
3 Product description | 3.8 X-ray viewer 1440 3.8 X-ray viewer 1440 X-ray viewer 1440 Note The KaVo X-ray viewer 1440 is a type 1 radiological viewing device in accord- ance as defined in DIN 6856-3. 3.9 Controls 3.9.1 Dentist elements...
Instructions for use Primus 1058 Life 3 Product description | 3.9 Controls Dentist element S-table Group of keys for menu selection Group of keys for the timer (MEMOspeed optional) Group of keys for the handpieces Group of keys for the dental chair Group of keys for illumination Group of keys for hygiene 3.9.2 Assistant unit...
Page 29
Instructions for use Primus 1058 Life 3 Product description | 3.9 Controls Assistant element key Dentist element key Name "Chair up" key "AP 0" key (automatic position 0) "Chair down" key "SP" key (rinsing position) "LP" key (last position) "AP" key (activate automatic posi- tion) "Backrest down"...
Instructions for use Primus 1058 Life 3 Product description | 3.9 Controls Name Control element "Bowl rinsing" key Dentist element and assistant element "Intensive germ reduc- Assistant element (op- tion" key tional) "HYDROclean" key Assistant element Group of keys for the handpieces/timer Name Control element "Preselected spray"...
Instructions for use Primus 1058 Life 3 Product description | 3.10 Rating plate and identification plate Item Name Function with hand- Function with hand- piece mounted piece removed "Spray pre-selection/ Moves the dental chair Sets the spray pre-se- ① AP" footswitch into automatic posi- lection.
Page 32
Instructions for use Primus 1058 Life 3 Product description | 3.10 Rating plate and identification plate Internal attachment site for rating plate Outside attachment site for rating plate Serial number Read and take note of the content of accompanying documents Please note the instructions for use Follow the instructions for use! Type B applied part...
Page 33
Instructions for use Primus 1058 Life 3 Product description | 3.10 Rating plate and identification plate Operating mode: Operating time of the patient chair: 25 seconds Pause time of the patient chair: 400 seconds (The permissible operating times correspond to common dental pro- cedure.) Fuse ratings: The "?????"...
Page 34
Instructions for use Primus 1058 Life 3 Product description | 3.10 Rating plate and identification plate Nameplate dentist element (e.g. table TM) / marking of the applied parts of type BF Type Device type Year of manufacture - serial number Material number Labelling and marking of the three-function handpiece and multifunctional handpiece...
Page 35
Instructions for use Primus 1058 Life 3 Product description | 3.10 Rating plate and identification plate Rating plate Service table 1568 Site for affixing the rating plate ① Device type ② Material number ③ Year and month manufactured - ④ HIBC Code Serial number ⑤...
Instructions for use Primus 1058 Life 3 Product description | 3.11 Technical data 3.10.1 Rating plate 1440 Rating plate X-ray viewer 1440 ① Device type ② SN: Year and month manufactured - Serial number ③ Material number ④ Supply voltage, frequency ⑤...
Page 37
Instructions for use Primus 1058 Life 3 Product description | 3.11 Technical data Electrical system Electrical lead 3 x 2.5 mm Free end above the floor 1 000 mm Input voltages 100/110/120/130/220/230/240 V AC Frequency 50/60 Hz Input voltage set by the manufac- See rating plate turer Power consumption at 100 to 240 V...
Page 38
Note In conjunction with the "DVGW water block with integrated water germ re- duction" a water germ reduction unit is installed in dental units from KaVo. The germ reduction liquid OXYGENAL 6 is continually added to the water at a concentration which is harmless for persons, but hygienically effective to maintain the quality of the treatment water.
Page 39
(Mat. no. 0.489.3451). ▶ If the Water bottle kit is used with the enclosed dosing attachment (Mat. no. 1.002.0287), add the proper amount of KaVo OXYGENAL 6 (Mat. no. 0.489.3451) with each filling. For the correct amount, please refer to the instructions of the dosing attachment for water germ reduction.
Page 40
Instructions for use Primus 1058 Life 3 Product description | 3.11 Technical data Air supply WARNING Non-compliance with national guidelines concerning the quality of the dental air. Infection hazard. ▶ Note and comply with the national guidelines concerning the quality of the dental air - if any.
Page 41
Instructions for use Primus 1058 Life 3 Product description | 3.11 Technical data Cannula connecting pieces according to DIN EN ISO 7494 - 2 Diameter of cannula of saliva ejector 7 mm and surgical suction Diameter of cannula of spray mist 15 mm suction Suction air quantity at spray mist...
Page 42
Treatment centre (TM) with COM- 206 kg gross, 158 kg net PACTchair With steel mounting plate and pa- 316 kg gross, 263 kg net tient communication Package KaVo Service table 1568 Length approx. 1040 mm Width approx. 810 mm Height approx.
Instructions for use Primus 1058 Life 3 Product description | 3.12 KaVo Service table 1568 (optional accessory) X-ray viewer 1440 Input voltage 24 V AC Frequency 50/60 Hertz Power consumption max. 35 VA ON-time 100 % Lights 2x Osram Lumilux de Luxe G5 Day-...
Instructions for use Primus 1058 Life 4 Operation | 4.1 Switching the device on and off 4 Operation 4.1 Switching the device on and off Note Always switch the machine off before leaving the office. ▶ Switch on the device using the main switch. ð...
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair 4.2 Adjusting the dental chair 4.2.1 Adjusting the arm rest (optional) Armrest for the standard dental chair To make it easier for the patient to sit in the chair, the armrest can be swung CAUTION The patient's hands are in a bad position when the chair is rising Danger of crushing fingers between the backrest and arm rest.
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair 4.2.2 Adjust head rest Adjust double-jointed knob headrest with knob CAUTION Adjusting the headrest. Injury to the neck muscles. ▶ Make sure that the patient is aware of the headrest being adjusted. ▶...
Page 47
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair Setting the push button of 2-joint headrest (optional) CAUTION Adjusting the headrest. Injury to the neck muscles. ▶ Make sure that the patient is aware of the headrest being adjusted. ▶...
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair 4.2.3 Positioning the dental chair manually CAUTION Danger of injury from overload or dynamic load. Patient chair may be damaged by overloading it. ▶ Do not subject the patient chair to a load exceeding its limit (Standard pa- tient chair 185 kg/ COMPACTchair patient chair 135 kg).
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair Positioning the chair and backrest manually using the foot control The cross switch of the foot control assumes the function of the button wheel on the dentist element during manual positioning of the patient chair. Requirement All instruments are in their holder.
Page 50
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair The four chair positions can be saved on the control panels. Two of these four positions can be saved with the foot control. It is for example recommendable to save the sitting down and getting up posi- tion using the "AP 0"...
Page 51
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair Note The memory is erased when you turn off the device. After turning on the device again (for example in the morning or after lunch), the chair does not execute a specific movement when you press the "LP"...
Page 52
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair ① Spray preselection/AP footswitch ② Blown air/AP footswitch ③ Foot pedal The chair positions can be recalled with two foot switches; the standard setting is as follows: ▪...
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair Saving an automatic position with the foot control ① Spray preselection/AP footswitch ② Blown air/AP footswitch ③ Foot pedal The chair positions can be saved on two footswitches; the standard setting is as follows: ▪...
Page 54
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair CAUTION Pinching from the treatment chair. The safety shutoff of the treatment chair is activated by lifting the respective component. Depending on the patient's body weight and the leverage, more force can be exerted on the object to be triggered than is necessary to trigger the switching function.
Page 55
Instructions for use Primus 1058 Life 4 Operation | 4.2 Adjusting the dental chair Safety shutoff for the COMPACTchair patient chair ① Patient element pivoted over dental ② Assistant element chair ③ Backrest ④ Bracket on the foot control ⑤ Bench support / seat cushion ⑥...
Instructions for use Primus 1058 Life 4 Operation | 4.3 Moving the patient chair Note The chair's position cannot be changed with the key wheels when a safety shutoff is activated. Exception: The patient unit safety switch only stops the upward and down- ward movement of the patient chair.
Instructions for use Primus 1058 Life 4 Operation | 4.4 Move the dentist's unit 4.4 Move the dentist's unit NOTICE Damage from overloading the dentist element. Exceeding the maximum weight of more than 2 kg by adding handpieces, ac- cessories, etc., can cause damage. ▶...
Instructions for use Primus 1058 Life 4 Operation | 4.4 Move the dentist's unit 4.4.1 Move the cart CAUTION Moving and overloading the cart. Danger of tipping and damaging the cart. ▶ Only use the card on a continuously smooth floor. ▶...
Instructions for use Primus 1058 Life 4 Operation | 4.5 Move the patient unit 4.5 Move the patient unit 4.5.1 Swing the patient unit by hand The swinging range is about 250 CAUTION The left armrest can collide with the manually adjusted patient's unit when the chair moves.
Instructions for use Primus 1058 Life 4 Operation | 4.6 Moving the assistant element ▶ To set a lower level, pull the assistant unit all the way up until the lock re- leases, and then lower the assistant unit. Mounting the tray holder ▶...
Page 61
Instructions for use Primus 1058 Life 4 Operation | 4.6 Moving the assistant element Pivoting range of assistant element r, l (optional) ▶ Move the backrest up before swinging the assistant element. ▶ Move the assistant element to the desired position in its swinging range. Adjusting the height of the assistant element right, left (optional) Note...
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu 4.7 Using functions through the menu 4.7.1 Using the user menu The following options can be opened in the user menu: Feature Description tion Firmware Display current firmware version.
Page 63
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu Option 1: Displaying the firmware version The firmware version is displayed Option 2: Setting the time of day ▶ Press the "SET" (S2) key to change the values of minutes and hours. ð...
Page 64
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu Option 3: Set the date ▶ Press the "SET" (S2) key to change the values of day, month, and year. ð The value to be changed flashes. ▶...
Page 65
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu Option 5: Setting the language ▶ Press the "Decrease value" or "Increase value" key to select the menu lan- guage. The following languages can be selected: Deutsch, English, Italiano, Francais.
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu 4.7.2 Standby menu Standby menu as default setting The unit starts in the standby menu. Select function The display shows display fields with symbols for the operating functions. Below each display field, there is a key for selecting the displayed operating function.
Page 67
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu Note The device acts like on level E when level switching is deactivated. ▶ To switch between levels, keep the "Direction of motor rotation" and the "Bowl flushing"...
Page 68
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu ▶ Press the selection keys "+" ② and "-" ① to switch between multiple status messages. ▶ Press the "OK" selection key ③ to exit from the display of status messages. Error messages in the status display See also: 2 9 Troubleshooting, Page 114...
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu 4.7.3 Operating the MEMOspeed menu (optional) The MEMOspeed menu is used to display and set handpiece-specific values. The display depends on which instrument was withdrawn. To save the handpiece-specific values, there are 3 memory levels (1, 2, 3) each available for two dentists (dentist 1 and dentist 2).
Page 70
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu ▶ Press the "Increase value" key to increase the speed. ð The intensity is shown on the display. ▶ Press the "Save" key to save the values. You can save after setting each value, or after setting all values.
Page 71
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu Setting the speed Note The speed can be set at level E only with the foot pedal. The speed cannot be saved in level E. Motor KL 701/KL 703 COMFORTdrive 200XD Minimum...
Page 72
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu Feature One LED is on: Spray air cooling status Both LEDs are on: Spray cooling status ▶ Press the "Save" key to save the values. You can save after setting each value, or after setting all values.
Page 73
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu Settings menu for minimal/maximal intensity ▶ Press the "min/max" (S4) keyed to toggle between the settings menus for minimal and maximal intensity in the set operating mode. ▶...
Page 74
Instructions for use Primus 1058 Life 4 Operation | 4.7 Using functions through the menu Feature No LED is on: No cooling Both LEDs are on: Spray cooling status ▶ Press the "Save" key to save the values. You can save after setting each value, or after setting all values.
The function of the CONEXIOcom menu is to control the display of previously recorded and saved images and videos. In order to use the function, the unit must have access to the data of the KaVo Software "CONEXIO" software. For details on the configuration, please refer to the "CONEXIO" installation instruc- tions.
Instructions for use Primus 1058 Life 4 Operation | 4.8 Using function through the dentist or assistant unit Icon Setting Next image/video To be able to communicate efficiently with the pa- tient, individual images can be selected and dis- played directly. This uses a rolling system that ad- vances from left to right and from top to bottom.
Page 77
Instructions for use Primus 1058 Life 4 Operation | 4.8 Using function through the dentist or assistant unit Name Feature Control ele- ment "Intensive germ Intensive germ reduction" key reduction/rinsing Assistant element/ function See also: 2 Servicing in- structions "HYDROclean" HYDROclean function Assistant element/...
Instructions for use Primus 1058 Life 4 Operation | 4.8 Using function through the dentist or assistant unit ð Once the key is released, the current value is saved. 4.8.2 Using lamp and X-ray image viewer The following buttons are available for the light functions: Feature Control element Turning the operating...
Instructions for use Primus 1058 Life 4 Operation | 4.8 Using function through the dentist or assistant unit Setting the timer Requirement Standby menu is selected. ▶ To set a timer time (e.g. Timer 1), press the "Timer" key until you hear a beep.
Page 80
Instructions for use Primus 1058 Life 4 Operation | 4.8 Using function through the dentist or assistant unit Note The speed is not shown in the display cannot be saved. The minimum and maximum speed depends upon the type of turbine that is used.
Page 81
Instructions for use Primus 1058 Life 4 Operation | 4.8 Using function through the dentist or assistant unit ▶ Press the "Direction of motor rotation" foot pedal. ð The direction of motor rotation is reversed each time the cross-switch and/ or the "Direction of motor rotation"...
Instructions for use Primus 1058 Life 4 Operation | 4.9 Operating the foot switch 4.9 Operating the foot switch 4.9.1 General functions The footswitches of the foot control have two functions. The function of the con- trol depends on whether an instrument is in its holder or whether it has been removed.
Instructions for use Primus 1058 Life 4 Operation | 4.9 Operating the foot switch 4.9.5 Setting the cooling condition ▶ Remove the handpiece (e.g turbine, motor) from the holder. ð The handpiece is active. ▶ Press the "Preselected spray" footswitch. ð...
Instructions for use Primus 1058 Life 4 Operation | 4.10 Service table 1568 (optional accessory) 4.9.9 Using CONEXIOcom (fee-based additional option) Setting ① U-shaped switch Discard image/video Press briefly - deletes the selected image/video Press long - all images/videos in the clipboard are deleted ②...
Instructions for use Primus 1058 Life 4 Operation | 4.10 Service table 1568 (optional accessory) CAUTION Over-travel beyond swivel range. Property damage. ▶ Comply with rotary knob swivel range of 180°. The service table 1568 can be locked in 4 snap-in positions using the rotary button ①.
Page 86
Instructions for use Primus 1058 Life 4 Operation | 4.10 Service table 1568 (optional accessory) Dimensions and swing ranges (in mm) 86 / 126...
Instructions for use Primus 1058 Life 5 Preparation methods DIN EN ISO 17664 5 Preparation methods DIN EN ISO 17664 Note The preparation methods can be found in the care instructions. 87 / 126...
Monitor support arm The monitor support arm is either affixed to the lamp mount pole or a Centro 1540. Monitor KaVo Screen One and KaVo Screen HD Service table 1568 It can be mounted on a device stand (cart ver- sion).
Assembly kit INTRA LUX motor KL 701 Brushless motor with light. KaVo COMFORTdrive 200 XD Dental handpiece for the high speed range up to 200,000 rpm. It can be attached only to the KaVo COMFORTbase coupling. Triple function handpiece Multifunctional handpiece featuring air, water, no heating, and no cold light.
Note If several medical electrical devices (ME device) or electrical devices from several manufacturers combined into a system are connected to the KaVo dental unit, the manufacturer data contained in the instructions for use for all products subject to the safety checks must also be noted.
7 Safety check - Test instructions | 7.1 Introduction Note KaVo offers a medical device book for keeping an inventory and recording es- sential master data on the medical device. The medical device book is only available in German (Mat. no. 0.789.0480).
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.1 Introduction Note If it should be impossible to check single ME devices that are functionally connected to an ME system individually for technical reasons, the ME system must be checked as a whole.
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check 7.1.6 Notes on repeat testing Note The value determined in these tests must be documented and evaluated to- gether with the measuring processes. The measured values may not over- shoot the specified values.
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check 7.2.2 Visual inspection (inspection by examination) Check the following items in advance: ▪ Has the equipment of the ME device or the ME system been changed since the last inspection? ▪...
Page 95
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check ▪ Condition of air and water connections ▪ Any damage on the sight window and the casing of the camera ERGOcam ▪...
Cables and wires, e.g. power supply cords, measuring circuits and data lines, must be arranged appropriately such that their influence on measurements is minimised. Note The following measuring aids can be ordered: KaVo measuring cable (Mat. no. 0.411.8811) 96 / 126...
Page 97
▶ Remove plug X2 from the power input board and plug it into the matching connector X2 of the KaVo- measuring cable (Mat. no. 0.411.8811). ▶ Plug the second plug X2 of the KaVo measuring cable into the network card (X2).
Page 98
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Connecting the safety tester without the KaVo measuring cable to the treatment centre ▶ Switch L + N of the on-site power supply cord to be voltage-free.
Page 99
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Note Additional measuring points AP X must be taken into consideration in the presence of accessories: e.g. accessories such as PIEZO ultrasonic scaler, etc.
Page 100
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Note In this measurement the resistance of the protective earth connection of the supply network can be taken into consideration. Note If applicable: all removable supply connection lines, which are kept handy for possible use, should be taken into consideration and the respective PE meas- ured.
Page 101
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scan the treatment centre with the test tip Measuring points on the device base ① Main switch holding plate ② Stand cover base plate ③...
Page 102
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scanning the patient chair with the test tip ① Top part of the chair ② Base plate of chair base ③ Chair switched mode power supply COMPACTchair measuring points ①...
Page 103
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scanning the operating elements with the test tip ① Dentist element S: Fixing screw on ② Dentist element TM: Fixing screw the underside of the dentist ele- on the underside of the dentist ele- ment...
Page 104
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Scan the treatment lamp with the test tip Operating light KaVoLUX 540 LED U ① Fastening screw of the handle sup- port when the gripping sleeve has been removed Operating light EDI/MAIA...
Page 105
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Equipment leakage current - Alternative measuring method Limit < 10 mA (maximum value!) Protection class 1 WARNING Electrical power. Death or injury from electric shock. ▶...
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Note In the testing of ME devices with several application parts, the parts must be connected in succession. The measured results must be evaluated using the threshold values.
Page 107
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Patient chair Standard Safety shutoff for the standard patient chair Item Safety switch-off ac- LED on assistant LED on dentist ele- tuated element ment...
Page 108
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Patient chair COMPACTchair Safety shutoff for the COMPACTchair patient chair If a person or object actuates a safety shutoff, the chair immediately stops moving.
▶ Name, date and signature of test engineer There is a copy of a test report template at the end of the chapter on Safety Checks. KaVo recommends the use of this template. Note Following testing, repair or adjustment, it must be verified whether the ME equipment or ME system has been restored to the state that is required for the intended usage before it is employed once again.
Page 110
Instructions for use Primus 1058 Life 7 Safety check - Test instructions | 7.2 Instructions for the safety check Unit leakage current - replacement measurement Leakage current of applied part - replacement measurement 110 / 126...
Instructions for use Primus 1058 Life 8 Appendix - Additional measuring sites | 8.1 Additional scanning sites SL X in the protective conductor measure- ment 8 Appendix - Additional measuring sites Note Regarding accessories, which are not listed here, comply with the specifica- tions of the relevant instructions for use.
Instructions for use Primus 1058 Life 8 Appendix - Additional measuring sites | 8.2 Additional measuring sites AP X for EUL/EPL measurement Ceiling adapter for operating light assembly kit ① Base plate for the ceiling adapter ② Surroundings of the protective con- ductor connector ③...
Instructions for use Primus 1058 Life 8 Appendix - Additional measuring sites | 8.3 Additional connection sites ACP X (additional earth connections) Exemplary presentation of the measuring point on the PiezoLEDultrasonic scaler ① Test probe on ultrasonic scaler tip in ultrasonic scaler handpiece Note The switch on the handpiece must be activated during the EPA measure- ment.
Instructions for use Primus 1058 Life 9 Troubleshooting 9 Troubleshooting Note In case of malfunctions, consult the separate instructions for the use and care of the individual instruments (such as the turbine, motor, camera, Sate- lec Mini LED, etc.). Malfunction Cause Remedy ▶...
Page 115
Instructions for use Primus 1058 Life 9 Troubleshooting Malfunction Cause Remedy ▶ Replace the high-pressure lamp or Multi LED. No cold light on the The high-pressure lamp handpieces. or Multi LED on the See also: handpiece is defective. 2 Instructions for Use of the handpiece ▶...
Page 116
Instructions for use Primus 1058 Life 9 Troubleshooting Malfunction Cause Remedy ▶ Refill the Oxygenal container. Beeps every The Oxygenal container 10 seconds. is empty. See also: The LED on the "In- 2 Servicing instructions tensive germ reduc- tion" key flashes (green).
Page 117
Instructions for use Primus 1058 Life 9 Troubleshooting Malfunction Cause Remedy ▶ Restart the software. Camera image freezes Electrical or electro- without the release magnetic interference button or foot control by other equipment. having been triggered. Taking the camera out again did not solve the problem.
Page 118
Instructions for use Primus 1058 Life 9 Troubleshooting Malfunction Cause Remedy ▶ Call a technician. Display shows: ID XX This error is not de- scribed in this chapter. ▶ Turn the unit off and on again, consult a ser- Display shows: CAN fail Internal communication vice technician according to need.
Instructions for use Primus 1058 Life 10 Data on electromagnetic compatibility according to EN 60601-1-2 | 10.1 Electromagnetic Transmissions 10 Data on electromagnetic compatibility according to EN 60601-1-2 10.1 Electromagnetic Transmissions The treatment unitPrimus 1058 Life is designed for operation in an environment as specified below.
Instructions for use Primus 1058 Life 10 Data on electromagnetic compatibility according to EN 60601-1-2 | 10.3 Recommended safe distance between portable and mobile HF telecommunications equipment and the treatment unit Interference im- EN 60601 test level Compliance level Electromagnetic en- munity tests vironment - Guidelines...
Instructions for use Primus 1058 Life 10 Data on electromagnetic compatibility according to EN 60601-1-2 | 10.4 Immunity to electromagnetic interfer- ence Rated power P of the Safe distance depending on the transmission frequency in m transmitter in W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz d=1.17...
Page 122
Instructions for use Primus 1058 Life 10 Data on electromagnetic compatibility according to EN 60601-1-2 | 10.4 Immunity to electromagnetic interfer- ence Interference im- EN 60601 test level Compliance level Electromagnetic environment munity tests - Guidelines Wire-based HF inter- Handheld and mobile wireless ference according to 150 kHz to 80 MHz 3 V/m...
Page 123
Instructions for use Primus 1058 Life 10 Data on electromagnetic compatibility according to EN 60601-1-2 | 10.4 Immunity to electromagnetic interfer- In the frequency range of 150 kHz to 80 MHz, the field strength should be less than 3V V/m. 123 / 126...