Preparations Before Use; Moisture Measurement - Oasis LDS-IG User Manual

Computerized moisture meter
Table of Contents

Advertisement

1. Take out protective foam blocks filled around the measurement sensor of the moisture meter ( "the
meter" for short), and meanwhile confirm that the sensor is free of any substance, otherwise the meter
will not work properly;
2. Open the cover of battery compartment at the bottom of the meter, load in four 5# alkaline (1.5V)
dry batteries (rechargeable batteries must not be used) according to polarity indication, or plug into AC
power source (200V±10%, 50Hz) with AC adapter which is provided together with the meter;
3. Put the meter on a wind-free and shake-proof horizontal platform and make the funnel closely
attached to sample container.
4. Get ready the sample to de tested: conduct preliminary sample screening, get rid of impurities, and
put the sample in the meter for as long as possible to achieve temperature balance;
5. Check the list to identify category code: for ease of operation, the calibration parameters for
representative categories have been preset during shipment of the meter. During application, all you
need to do is check the "category-code list" attached to this User's Manual and chooses the
corresponding category code for moisture measurement.
Note: when the meter is used for trade settlement and safe storage which needs higher precision, we
strongly recommend that you adopt standard sample of the category to be measured and conduct
calibration and error correction of the meter so as to ensure measuring precision and safeguard your
interest.
1. Press down the power switch and the meter starts self-check. If no abnormality is found, the category
code will be displayed;
测量:measure
Note
2. Press the " " or " " key to choose the code of category to be measured (refer to the attached list
for details);
3. Put the sample into the blanking cylinder to the lower edge of funnel for later use (see Figure A);
4. Make the blanking cylinder closely attached to the meter's sensor, support the blanking cylinder with
one hand and press gently the switch for discharge gate (shown in Figure B) to make the sample fully
and evenly fall into the measuring sensor. Without pressing any key, the meter will automatically start
the measurement and the moisture value will be displayed after the decimal point flickers several times;
样品放至下沿口:Put the sample to the lower edge
Note
A:Figure A
B:Figure B