Instructions for use
QUATTROcare PLUS 2124 A
- 1.008.3805

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the QUATTROcare PLUS 2124 A and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo QUATTROcare PLUS 2124 A

  • Page 1 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805...
  • Page 2 Distributed by: Manufacturer: KaVo Dental Technologies, LLC Kaltenbach & Voigt GmbH 11727 Fruehauf Drive Bismarckring 39 Charlotte, NC 28273 USA D-88400 Biberach Phone: 847 550 6800 www.kavo.com Fax: 847 550 6825...
  • Page 3: Table Of Contents

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 Table of contents Table of contents 1 User instructions ........................ 5 1.1 User guide ........................5 1.1.1 Abbreviations ..................... 5 1.1.2 General marks and symbols................. 5 1.1.3 Target group ...................... 6 1.2 Service .........................
  • Page 4 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 Table of contents 6.2 Safety checks ("STK") ....................31 7 Troubleshooting ......................... 33 8 Tools and consumables ...................... 35 9 Terms and conditions of warranty ..................38 4 / 40...
  • Page 5: User Instructions

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 1 User instructions | 1.1 User guide 1 User instructions 1.1 User guide Requirement Read these instructions prior to first startup to avoid misuse and prevent dam- age. 1.1.1 Abbreviations Explanation...
  • Page 6: Target Group

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 1 User instructions | 1.2 Service WARNING In cases which – if not prevented – could lead to death or severe in- jury. CAUTION In cases which – if not prevented – could lead to minor or moderate injury.
  • Page 7: Information On The Packaging: Storage And Transportation

    Freight Forwarders' Terms and Conditions, Art. 28). Outside Germany Note KaVo shall not be held liable for damage arising from transportation. The shipment must be checked on arrival. If the packaging is visibly damaged on delivery, please proceed as follows: 1.
  • Page 8: Disposal

    If you have any questions regarding proper disposal of the KaVo product, please contact the KaVo branch. 1.5 Disposal of electronic and electrical devices...
  • Page 9 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 1 User instructions | 1.5 Disposal of electronic and electrical devices Email: eom@enretec.de and Postal address: enretec GmbH, Geschäftsbereich eomRECYCLING® Kanalstraße 17 D-16727 Velten 3. A unit that is not permanently installed will be picked up at the office.
  • Page 10: Safety

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 2 Safety | 2.1 Explosion hazard 2 Safety The instructions for use are a component of the product and must be read care- fully prior to use and be accessible at all times.
  • Page 11: Accessories And Combination With Other Equipment

    ▶ Have a safety check done on the product every 5 years. For this purpose, send the product to the KaVo Customer Service Centre Warthausen or a test centre approved by KaVo.
  • Page 12: Product Description

    3.1 Purpose – Proper use Indications for use: The intended use of the KaVo QUATTROcare is for internal cleaning, i.e, purging of old lubricant, for the maintenance of rotating dental and surgical instru- ments. KaVo QUATTROcare spray can is a component of the KaVo QUATTRO- care unit and is specifically designed to fit this unit only.
  • Page 13: Quattrocare Plus 2124 A

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 3 Product description | 3.2 QUATTROcare PLUS 2124 A 3.2 QUATTROcare PLUS 2124 A ① 2x INTRAmatic service couplings, short ② 2x MULTIflex service couplings ③ Control panel and display ④ Door of spray can chamber ⑤...
  • Page 14: Rating Plate

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 3 Product description | 3.3 Rating plate 3.3 Rating plate Type QUATTROcare PLUS 2124 A Year of manufacture - serial number Material number Fuse-protected with T1000 mA Comply with the Instructions for use...
  • Page 15: Scope Of Delivery

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 3 Product description | 3.5 Scope of delivery Electrical supply Line voltage 100 to 240 VAC Mains frequency 50 or 60 Hz Power consumption 41 VA Overvoltage category Mains voltage fluctuation ±10 %...
  • Page 16 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 3 Product description | 3.5 Scope of delivery ① Chuck service adaptor ② QUATTROcare Plus Spray canister ③ Compressed air coupling ④ Instructions for use ⑤ O-ring MULTIflex (4 units) ⑥ O-ring INTRAmatic (3 units) ⑦...
  • Page 17: First Use

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 4 First use | 4.1 Select location 4 First use 4.1 Select location The QUATTROcare Plus Spray canister contains explosive propellant gases. The following instructions and setup conditions therefore need to be observed when setting up the unit.
  • Page 18: Connect The Quattrocare Plus 2124 A

    (on the compressor) or blow-out the lines before connecting them. NOTICE Use of power cables other than original KaVo parts. Damage to or malfunction of the product. ▶ Use original KaVo power cables only (for material numbers, see chapter on "Consumables and tools"). 18 / 40...
  • Page 19: Replacing The Quattrocare Plus Spray Can

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 4 First use | 4.3 Replacing the QUATTROcare Plus Spray can Back of QUATTROcare PLUS 2124 A A T1000 mA fuse is installed ④. ▶ Plug the power cable into the electric socket ③.
  • Page 20 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 4 First use | 4.3 Replacing the QUATTROcare Plus Spray can Note The safety data sheet of the QUATTROcare Plus spray can be viewed at www.kavo.com. ▶ Open the side door of the unit.
  • Page 21: Remove The Spray Canister

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 4 First use | 4.3 Replacing the QUATTROcare Plus Spray can Note If any oil or gas leaks, you need to use a different spray can. ▶ Swing the spray can and the adaptor back into the device.
  • Page 22: Dispose Of The Spray Canister

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 4 First use | 4.3 Replacing the QUATTROcare Plus Spray can 4.3.3 Dispose of the spray canister ▶ Dispose of the spray can in empty condition. Note Waste code according to the European waste catalogue: 15 01 04 - metals, packaging, empty.
  • Page 23: Operation

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 5 Operation | 5.1 Control panel 5 Operation 5.1 Control panel ① LED spray can control ② LED control light main switch (on/ off) ③ LED display ④ Start button for chuck service ⑤...
  • Page 24: Mounting Instruments

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 5 Operation | 5.3 Mounting instruments 5.3 Mounting instruments Note Remove the handpieces from the unit as soon as the servicing is completed. Note The LED display shows the following: Adaptor on the far left corresponds to LED display 1.
  • Page 25 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 5 Operation | 5.3 Mounting instruments ▶ Mount straight and contra-angle handpieces on the INTRAmatic service coupling, short, (service coupling on the left in the middle ② or on the far left ①) and lock them into place.
  • Page 26: Start The Treatment

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 5 Operation | 5.4 Start the treatment 5.4 Start the treatment Note The LED of the spray canister control① lighting up indicates that the spray can is empty. When the main switch is on, the LED ② lights up.
  • Page 27: Reposition The Treatment Couplings

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 5 Operation | 5.6 Reposition the treatment couplings ▶ Pull the turbine in the axial direction of the service coupling, and take it off. ▶ Press on the latch with a thumb and take the straight or contra-angle hand- piece off by simultaneously pulling in the direction of the axis of the service coupling.
  • Page 28: Chuck Servicing

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 5 Operation | 5.7 Chuck Servicing ▶ Unscrew the union nut clockwise in the direction of the arrow, and take the service coupling off. ▶ Attach the service coupling as selected and fasten it turning the union nut counterclockwise against the direction of the arrow.
  • Page 29: Set The Blow-Out Time

    The spray is then briefly injected through the chuck, completing the service. Quit the chuck service mode: ▪ Place the handpieces in the QUATTROcare PLUS 2124 A, close the front door and start theservice procedure. ▪ Alternatively, after three minutes with no service procedure running, the device automatically switches back to normal service mode.
  • Page 30 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 5 Operation | 5.8 Set the blow-out time ▶ Hold down the start key ②. ▶ Turn the main switch③ on. ð LEDs 1 to 4 of the LED display① light up one after the other.
  • Page 31: Maintenance

    6.2 Safety checks ("STK") The QUATTROcare PLUS 2124 A must be subjected to a service check including safety check ("STK") every 5 years. The safety check may only be done by a professional trained by KaVo or in a shop trained by KaVo.
  • Page 32 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 6 Maintenance | 6.2 Safety checks ("STK") Perform the safety check ("STK") as described in the KaVo technician's instruc- tions. 32 / 40...
  • Page 33: Troubleshooting

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 7 Troubleshooting 7 Troubleshooting Note Heads of the "22" series cannot be serviced because their necks are too thin. These heads must be serviced together with the corresponding shanks. Malfunction Cause Remedy The red LED ①...
  • Page 34 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 7 Troubleshooting Malfunction Cause Remedy LED ② 1 and 3 flash in Fan is blocked or Replace the fan. alternating manner with 2 defective. Notify the KaVo Service and 4. A signal is emit- Team.
  • Page 35: Tools And Consumables

    Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 8 Tools and consumables 8 Tools and consumables Presentation Material summary Mat.No. Service coupling 1.011.7380 INTRAmatic, short Service coupling 1.009.6142 MULTIflex Service coupling for IN- 0.411.7941 TRA heads includes actuating sleeve Service coupling 1.005.1707...
  • Page 36 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 8 Tools and consumables Presentation Material summary Mat.No. Service coupling 1.000.7585 YOSHIDA Turbine handpieces without light Service coupling 1.008.2083 MIDWEST High Speed Service coupling 1.000.8459 MIDWEST Low Speed Service coupling 1.000.8786...
  • Page 37 Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805 8 Tools and consumables Presentation Material summary Mat. no. Power cable China 1.004.3850 37 / 40...
  • Page 38: Terms And Conditions Of Warranty

    12 months from the purchasing date, subject to the following condi- tions: Upon justified complaints of flaws or a short delivery, KaVo will make good its warranty by replacing the product free of cost or repairing it according to the customer's wishes.