kENDALL SCD EXPRESS Operation And Service Manual

kENDALL SCD EXPRESS Operation And Service Manual

Compression system
Table of Contents

Advertisement

tyco
kENDALL
SCD EXPRESS
Compression System
Operation and Service Manual
SYSTÈME DE COMPRESSION
Manuel D'utilisation Et D' e ntretien
KOMPRESSIONSSYSTEM
Betriebs- Und Serviceanleitung
SISTEMA DI COMPRESSIONE
Manuale Operativo E Di Manutenzione
SISTEMA DE COMPRESIÓN
Manual De Operación Y Servicio
KOMPRESSIONSSYSTEM
Användar- Och Servicemanual
COMPRESSIESYSTEEM
Handleiding Voor Bediening En Onderhoud
SISTEMA DE COMPRESSÃO
Manual De Funcionamento E Assistência
KOMPRESSIOJÄRJESTELMÄ
Käyttö- Ja Huolto-opas
KOMPRESSIONSSYSTEM
Betjenings- Og Servicehåndbog
TM
ΣΎΣΤΗΜΑ ΣΥΜΠΊΕΣΗΣ
Εγχειρίδιο Λειτουργίας Και Σέρβις
TLAKOVÝ SYSTÉM
Provozní A Servisní Příručka
KOMPRESSZIÓS RENDSZER
Kezelési És Karbantartási Kézikönyv
СИСТЕМА ПНЕВМАТИЧЕСКОЙ
КОМПРЕССИИ
Руководство По Эксплуатации И
Обслуживанию
SYSTEM UCISKOWY
Instrukcja Obsługi I Serwisowania
KOMPRESYON SISTEMI
Operasyon Ve Servis Kılavuzu

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SCD EXPRESS and is the answer not in the manual?

Questions and answers

Summary of Contents for kENDALL SCD EXPRESS

  • Page 1 SCD EXPRESS Compression System Operation and Service Manual SYSTÈME DE COMPRESSION ΣΎΣΤΗΜΑ ΣΥΜΠΊΕΣΗΣ Manuel D’utilisation Et D’ e ntretien Εγχειρίδιο Λειτουργίας Και Σέρβις KOMPRESSIONSSYSTEM TLAKOVÝ SYSTÉM Betriebs- Und Serviceanleitung Provozní A Servisní Příručka SISTEMA DI COMPRESSIONE KOMPRESSZIÓS RENDSZER Manuale Operativo E Di Manutenzione Kezelési És Karbantartási Kézikönyv...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Indications .......................... 1 Contraindications ........................ 1 Cautions ..........................1 Explanation of Symbols Used ....................2 Front Panel Display ......................2 Section I - General Operating Instructions ................3 Set up .....................................3 Start-up ..................................3 Garment Configuration & Detection ..........................3 Normal Operation ................................4 Pressure Settings ................................4 Vascular Refill Detection ...............................4...
  • Page 3 TABLE OF CONTENTS Test Mode 05 - Self Test ............................. 12 Test Mode 06 - Leak Test ............................12 Test Mode 07 - Performance Test ..........................12 Test Mode 08 - Manufacturing Test ........................... 12 Section VI - General Disassembly / Reassembly ..............13 Battery Pack (Removal / Installation) ........................
  • Page 4: Indications

    WARNING: Do not attempt to repair or replace broken tubing connectors as hazardous inflation of the sleeves may occur. Kendall’s SCD EXPRESS Compression System is protected by one or more of the following U.S. patents: 5,022,387; 5,031,604; 5,478,119; 5,876,359; Des.363,988; Des.364,460; Des.369,664; Des.373,192. Other patents pending. Tyco Healthcare expressly states that no implied license with respect to use of non-Tyco Healthcare components is granted by the sale or lease of this system.
  • Page 5: Explanation Of Symbols Used

    Explanation of Symbols Used Refer to Accompanying Controller Serial Number Documents ��� ������������ Reorder number for the device Type BF Protection Against located on the carton label Electronic Shock ������� ��������� Vascular Refi ll Detection Manufacturing Date Code occurring while illuminated CE Mark Underwriters Laboratories (UL) Classifi...
  • Page 6: Section I - General Operating Instructions

    Section I - General Operating Instructions Set up • Place the Controller on the footboard via the bed hook or place it on a horizontal surface appropriate for the environment, such as on a table, within reasonable proximity to the point of use. Be sure to allow adequate air flow to the vents located on either end of the controller. •...
  • Page 7: Normal Operation

    Portable Controller Configuration • The SCD EXPRESS Compression System Controller can be configured without the bed hook for easier portability. The System can also be used without the cord cover door so that the power cord can be disconnected at the Controller instead of at the AC power wall outlet.
  • Page 8: Battery Indicators Status

    System. The battery pack, or a dummy pack (if so equipped), must be installed in the controller to prevent the user from having access to the battery’s electrical connection. The battery pack for the SCD EXPRESS Compression System is optional. To order a new or replacement battery please contact Kendall Customer Service. Battery Indicator Status:...
  • Page 9: Battery Warnings

    Battery Warnings The SCD EXPRESS Compression System battery pack contains Lithium Ion (Li-Ion) battery cells and must be used properly for safety and to maintain optimal performance. • Store only between –20° C (-4° F) and 60° C (140° F).
  • Page 10 Alarm Code Fault Type Description Troubleshooting Leg Sleeve pressure is not between 35 and - Check garment application System Pressure 55 mmHg for 12 consecutive cycles; Foot Cuff (too loose or tight). pressure is not between 110 and 150 mmHg - Turn the Controller off and restart.
  • Page 11: Section Iv - Service And Maintenance

    Service technicians should be familiar with the operator’s portion of this manual and the operating principles of the SCD EXPRESS Compression System. If a Controller is to be returned to Kendall for service, a description of the operating conditions and the fault code displayed should accompany the unit.
  • Page 12: Fan Screen And Ventilation

    Controller should be returned to a service center for testing and repair * The SCD EXPRESS was designed to provide sufficient isolation inside the controller between all exposed metal components and the electrical supply mains. As a result there is no externally accessible ground on the device. The recommended electrical safely test for proper grounding is a resistance check of the power cord ground.
  • Page 13: Cleaning

    Section V - Test Methods and Calibration The SCD EXPRESS Compression System has various test modes that can be accessed by the service technician. They are intended for use by qualified personnel. To activate the test modes follow these steps for entering “Test Access Mode”.
  • Page 14: Test Mode Look Up Chart

    Test Mode 03 - Pressure Transducer Calibration NOTE: The transducer used in the SCD EXPRESS Compression System is a state-of -the-art, highly precise and virtually drift free device. Factory calibration certification is void if the case is opened. Recalibration is rarely required and should be done only when necessary.
  • Page 15: Test Mode 05 - Self Test

    5. Test Mode 05 indicator is lit red to indicate that the unit is ready for the 0 mmHg calibration verification. With no pressure applied to the transducer, press the Port B button to complete the verification for this range. 6.
  • Page 16: Section Vi - General Disassembly / Reassembly

    Muffler (Removal / Installation) • The muffler is a custom plastic part used to keep the Kendall SCD EXPRESS Compression System running quietly. • To remove the muffler detach the compressor intake tubing and slide the muffler out from under the power board.
  • Page 17: Power Supply Board (Removal / Installation)

    • Observe the angle and orientation of the fan and screen prior to removal to ease reassembly. (Diagram 04) • Slide the fan and screen out of the case by pulling straight out. • Installation is the reverse of removal. For optimum cooling and quietness, use only Kendall replacement fans. Main Controller Board (Removal/Installation) CAUTION: Use a grounded strap when handling any electronic components.
  • Page 18: Section Vii - Specifications

    Section VII - Specifi cations SCD EXPRESS Compression System Safety Standards Built to UL60601-1, CSA-C22.2 No. 601.1-M90, EN60601-1, and IEC 60601-1-2 Standards UL Classified File # E189131 UL Device Classification Class I Equipment Internally Powered, Portable Type BF Applied Parts...
  • Page 19: Section Viii - Schematics

    Section VIII - Schematics Figure 1 - Parts Assembly Diagram - Exploded View Controller Parts List 10. Fan 1. Front Membrane Panel 11. Fan Screen 2. Front Enclosure 12. Rear Enclosure 3. Main Controller Board 13. Rubber Feet (Set of 2) 4.
  • Page 20 Section VIII - Schematics Figure 1 - Parts Assembly Diagram - Exploded View EN-17...
  • Page 21: Figure 2 - Pneumatic & Electrical Schematic

    Section VIII - Schematics Figure 2 - Pneumatic & Electrical Schematic EN-18...
  • Page 22: Diagram 04

    Section IX - Pump Diagrams Diagrams 01,02, 03 & 04 Transducer Tube Compressor Mounting Bracket Screws Diagram 01 Valve Manifold / Bed Hook Screws Diagram 03 EN-19...
  • Page 23: Section Ix - Pump Diagrams

    Section IX - Pump Diagrams Diagrams 01,02, 03 & 04 Compressor Output Tube Tube Compressor Intake Tube Muffler Intake Tube Diagram 02 Diagram 04 EN-20...
  • Page 24: Diagram 02

    The SCD EXPRESS Compression System has been built and tested according to UL60601-1, E181931 CSA-C22.2 No.601.1- M90, and EN60601-1-2 Standards. Emissions Test issions Test issions Test Compliance Compliance Compliance Electromagnetic Environment - Guidance Electromagnetic Environment - Guidance Electromagnetic Environment - Guidance...