SCD EXPRESS Compression System Operation and Service Manual SYSTÈME DE COMPRESSION ΣΎΣΤΗΜΑ ΣΥΜΠΊΕΣΗΣ Manuel D’utilisation Et D’ e ntretien Εγχειρίδιο Λειτουργίας Και Σέρβις KOMPRESSIONSSYSTEM TLAKOVÝ SYSTÉM Betriebs- Und Serviceanleitung Provozní A Servisní Příručka SISTEMA DI COMPRESSIONE KOMPRESSZIÓS RENDSZER Manuale Operativo E Di Manutenzione Kezelési És Karbantartási Kézikönyv...
TABLE OF CONTENTS Indications .......................... 1 Contraindications ........................ 1 Cautions ..........................1 Explanation of Symbols Used ....................2 Front Panel Display ......................2 Section I - General Operating Instructions ................3 Set up .....................................3 Start-up ..................................3 Garment Configuration & Detection ..........................3 Normal Operation ................................4 Pressure Settings ................................4 Vascular Refill Detection ...............................4...
Page 3
TABLE OF CONTENTS Test Mode 05 - Self Test ............................. 12 Test Mode 06 - Leak Test ............................12 Test Mode 07 - Performance Test ..........................12 Test Mode 08 - Manufacturing Test ........................... 12 Section VI - General Disassembly / Reassembly ..............13 Battery Pack (Removal / Installation) ........................
WARNING: Do not attempt to repair or replace broken tubing connectors as hazardous inflation of the sleeves may occur. Kendall’s SCD EXPRESS Compression System is protected by one or more of the following U.S. patents: 5,022,387; 5,031,604; 5,478,119; 5,876,359; Des.363,988; Des.364,460; Des.369,664; Des.373,192. Other patents pending. Tyco Healthcare expressly states that no implied license with respect to use of non-Tyco Healthcare components is granted by the sale or lease of this system.
Explanation of Symbols Used Refer to Accompanying Controller Serial Number Documents ��� ������������ Reorder number for the device Type BF Protection Against located on the carton label Electronic Shock ������� ��������� Vascular Refi ll Detection Manufacturing Date Code occurring while illuminated CE Mark Underwriters Laboratories (UL) Classifi...
Section I - General Operating Instructions Set up • Place the Controller on the footboard via the bed hook or place it on a horizontal surface appropriate for the environment, such as on a table, within reasonable proximity to the point of use. Be sure to allow adequate air flow to the vents located on either end of the controller. •...
Portable Controller Configuration • The SCD EXPRESS Compression System Controller can be configured without the bed hook for easier portability. The System can also be used without the cord cover door so that the power cord can be disconnected at the Controller instead of at the AC power wall outlet.
System. The battery pack, or a dummy pack (if so equipped), must be installed in the controller to prevent the user from having access to the battery’s electrical connection. The battery pack for the SCD EXPRESS Compression System is optional. To order a new or replacement battery please contact Kendall Customer Service. Battery Indicator Status:...
Battery Warnings The SCD EXPRESS Compression System battery pack contains Lithium Ion (Li-Ion) battery cells and must be used properly for safety and to maintain optimal performance. • Store only between –20° C (-4° F) and 60° C (140° F).
Page 10
Alarm Code Fault Type Description Troubleshooting Leg Sleeve pressure is not between 35 and - Check garment application System Pressure 55 mmHg for 12 consecutive cycles; Foot Cuff (too loose or tight). pressure is not between 110 and 150 mmHg - Turn the Controller off and restart.
Service technicians should be familiar with the operator’s portion of this manual and the operating principles of the SCD EXPRESS Compression System. If a Controller is to be returned to Kendall for service, a description of the operating conditions and the fault code displayed should accompany the unit.
Controller should be returned to a service center for testing and repair * The SCD EXPRESS was designed to provide sufficient isolation inside the controller between all exposed metal components and the electrical supply mains. As a result there is no externally accessible ground on the device. The recommended electrical safely test for proper grounding is a resistance check of the power cord ground.
Section V - Test Methods and Calibration The SCD EXPRESS Compression System has various test modes that can be accessed by the service technician. They are intended for use by qualified personnel. To activate the test modes follow these steps for entering “Test Access Mode”.
Test Mode 03 - Pressure Transducer Calibration NOTE: The transducer used in the SCD EXPRESS Compression System is a state-of -the-art, highly precise and virtually drift free device. Factory calibration certification is void if the case is opened. Recalibration is rarely required and should be done only when necessary.
5. Test Mode 05 indicator is lit red to indicate that the unit is ready for the 0 mmHg calibration verification. With no pressure applied to the transducer, press the Port B button to complete the verification for this range. 6.
Muffler (Removal / Installation) • The muffler is a custom plastic part used to keep the Kendall SCD EXPRESS Compression System running quietly. • To remove the muffler detach the compressor intake tubing and slide the muffler out from under the power board.
• Observe the angle and orientation of the fan and screen prior to removal to ease reassembly. (Diagram 04) • Slide the fan and screen out of the case by pulling straight out. • Installation is the reverse of removal. For optimum cooling and quietness, use only Kendall replacement fans. Main Controller Board (Removal/Installation) CAUTION: Use a grounded strap when handling any electronic components.
Section VII - Specifi cations SCD EXPRESS Compression System Safety Standards Built to UL60601-1, CSA-C22.2 No. 601.1-M90, EN60601-1, and IEC 60601-1-2 Standards UL Classified File # E189131 UL Device Classification Class I Equipment Internally Powered, Portable Type BF Applied Parts...
Section VIII - Schematics Figure 1 - Parts Assembly Diagram - Exploded View Controller Parts List 10. Fan 1. Front Membrane Panel 11. Fan Screen 2. Front Enclosure 12. Rear Enclosure 3. Main Controller Board 13. Rubber Feet (Set of 2) 4.
Page 20
Section VIII - Schematics Figure 1 - Parts Assembly Diagram - Exploded View EN-17...
The SCD EXPRESS Compression System has been built and tested according to UL60601-1, E181931 CSA-C22.2 No.601.1- M90, and EN60601-1-2 Standards. Emissions Test issions Test issions Test Compliance Compliance Compliance Electromagnetic Environment - Guidance Electromagnetic Environment - Guidance Electromagnetic Environment - Guidance...