AGF-T-30, K-AGF-PV-30, RAF-PV-30, ADF-PV-30 1 Introduction 1.1 Intended Use The filters were designed specifically for front panel installation in analysers or systems. All types can be used for filtering sample gas. When used in explosive atmosphere areas Under certain conditions the filters and filter elements may further be used in areas with explosive atmosphere. The provisions and specifications in chapter "Use in explosive atmosphere areas"...
AGF-T-30, K-AGF-PV-30, RAF-PV-30, ADF-PV-30 2 Safety instructions 2.1 Important notices Operation of the device is only valid if: – the product is used under the conditions described in the installation- and operation instruction, the intended application according to the type plate and the intended use. In case of unauthorized modifications done by the user Bühler Technolo- gies GmbH can not be held responsible for any damage, –...
Intended Use AGF-PV-30. AGF-PV-S2, AGF-T-30. K-AGF-PV-30 and RAF-PV-30 filters may be used in Zone 1 and 2 areas with explosive atmo- sphere. Explosion classes IIA, IIB and IIC are approved. The filters have no innate ignition source and do not fall into the applica- tion of Directive 2014/34/EU and therefore do not bear the CE mark.
Page 7
AGF-T-30, K-AGF-PV-30, RAF-PV-30, ADF-PV-30 The highest surface temperature is directly affected by the temperature of the medium introduced. The medium introduced must not exceed the maximum temperature of 100 °C. If necessary, a temperature monitor should be installed. When operating the filters with moisture detector and controller, these must be inherently safe according to EN 60079-11.
Page 8
AGF-T-30, K-AGF-PV-30, RAF-PV-30, ADF-PV-30 Operation with moisture detector DANGER Explosion hazard To operate the filter in an explosive atmosphere with moisture detector, the moisture detector and controller must be ignition protection class intrinsically safe “i”. The use of non-intrinsically safe components may result in ignition in explosive atmospheres.
AGF-T-30, K-AGF-PV-30, RAF-PV-30, ADF-PV-30 3 Transport and storage The product should only be transported inside the original packaging or a suitable alternative. When not in use, the equipment must be protected from moisture and heat. They must be stored in a covered, dry and dust-free room at a temperature between -20 °C and 60 °C.
AGF-T-30, K-AGF-PV-30, RAF-PV-30, ADF-PV-30 4 Installation and connection 4.1 Installation site requirements Always use washers on the longitudinal slots when installing the retaining ring flange. Furthermore, verify the filter has been inserted all the way to the end of these longitudinal slots.
AGF-T-30, K-AGF-PV-30, RAF-PV-30, ADF-PV-30 5 Operation and control NOTICE The device must not be operated beyond its specifications. BE410011 ◦ 07/2017 Bühler Technologies GmbH...
– All others: Pull the bracket, holding on to the filter glass – Whilst holding the filter head, move the glass back and forth and carefully remove downward. – AGF-PV-30. AGF-PV-S2, AGF-T-30. RAF-PV-30: Remove the filter element and install a new one – ADF-PV-30: Replace adsorbent –...
Filter* Item no. Model Seal 41 50 699 K-AGF-PV-30-A; G1/4 outlet and inlet Viton 41 50 699I K-AGF-PV-30-A-I; NPT 1/4" outlet and inlet Viton * one filter element is included with delivery. Filter element Item no. Model Filter element Material...
AGF-T-30, K-AGF-PV-30, RAF-PV-30, ADF-PV-30 8 Disposal Dispose of parts so as not to endanger the health or environment. Follow the laws in the country of use for disposing of elec- tronic components and devices during disposal. Bühler Technologies GmbH BE410011 ◦ 07/2017...
G1/4 or NPT 1/4" (see ordering information) Weight 0.24 kg Operating pressure max. 4 bar Operating temperature max. 100 °C Fine mesh filter AGF-PV-30 Material – Filter head PVDF Material – Filter cover Glass Material – Gasket Viton Thread (A) G1/4 or NPT 1/4"...