Chatillon DFX II Series User Manual page 17

Digital force gauge
Table of Contents

Advertisement

Find Quality Products Online at:
Procedure
The Chatillon DFX II Series is supplied with a Cer-
tificate of Conformance. The gauge was tested and
found to have a measurement accuracy of 0.5% of
full scale or better.
Calibration Procedure
The DFX II Series calibration procedure is contained
in the gauge firmware. The calibration is a four
point calibration: full scale compression, zero com-
pression, full scale tension, zero tension.
To access the calibration routine, begin with the
gauge power OFF.
Press the ON key. While the "Chatillon Logo" is
displayed, press PEAK then UNITS.
Model DFX II-100
Rev. No. V1.00
Rev. Date
dd/mm/yy
Chatillon
T
Peak
Units
Gauge Service
PEAK to calibrate
UNITS to diagnostics
INFO to exit
T
Peak
Select Gauge Model
DFX II-100
UNITS to change
INFO to proceed
PEAK to exit
T
Peak
GlobalTestSupply
www.
Select CALIB Units
UNITS to change
INFO to proceed
PEAK to go back
T
Note: The units of measure selected must correspond to
the weights that are being used for the calibration.
Place gauge in
horizontal position
INFO to proceed
PEAK to go back
C
T
Note: Placing the gauge horizontally means to lay the
gauge perfectly flat on its backside making sure no load
is being applied to the loadcell shaft.
Place gauge in
compression mode and
exercise 3 times
C
INFO to proceed
PEAK to go back
T
Note: You must use a hanger when performing the
compression calibration routine to eliminate side load
C
effects on the loadcell.
Selecting PEAK during the calibration procedure
will take you back to the last step in the procedure.