SC-F7000 Serie/SC-B7000 Series
4.10.7 PG Adjustment
REQUIRED TOOL
Thickness gauge
STANDARD VALUE
2.55 pass
2.65 stop
PROCEDURE
1.
When any paper is loaded, remove it.
2.
Unlock the CR Unit, and move the unit to the center of the platen.
3.
Make sure that the PG is ++ (2.6mm).
4.
Place the thickness gauge on the platen to the right of the CR Unit, and slowly
move the unit over the thickness gauge.
Carefully move the CR Unit so that the thickness gauge does not
C A U T I O N
contact with the nozzle surface of the Print Head.
ADJUSTMENT
(P. 56)
5.
Measure the height of the CR Unit to see if it is within the standards.
(Measure it at both left and right of the CR Unit.)
Within the standards: Cap the Print Head and end the adjustment.
Out of the standards: Go to
CR Related Adjustments
Thickness gauge
Figure 4-4. Placing the Thickness Gauge
Step
6.
Revision A
107