Figure 5-14 Adjusting The Lift-Off Noise (Second Alternative) - Olympus nortec 600 User Manual

Eddy current flaw detector
Hide thumbs Also See for nortec 600:
Table of Contents

Advertisement

DMTA-10040-01EN, Rev. E, February 2018
— Second alternative:
If the surface of your sample is made of the same material as the material of
the hole to be inspected, this alternative is usually more practical and easier.
While the scanner is rotating the probe, simply gently touch the surface of the
sample with the head of the probe, and adjust the signal angle so that the tip
of the signal extends to the right of the impedance plane (see Figure 5-14 on
page 130).
Figure 5‑14 Adjusting the lift‑off noise (second alternative)
4.
Press the GAIN key (
keep the entire crack signal within the screen area.
A good practice is to set the signal's maximum deflection to 10 % of the screen's
horizontal deflection (see Figure 5-15 on page 131).
130
Chapter 5
), and then decrease the gain with the knob in order to