Philips SEM 525 Instructions page 3

Advertisement

The "X" and "Y" controls are approximately centered at 500 X 500 with TILT
at 0 degrees.
Do not adjust the TILT less than 0 degrees nor greater than 90 degrees.
Turning ON the electron
Turn ON the high tension.
gauge.
Adjust the current to saturation (usually 5-6 clicks).
saturate the filament.
Aligning the electron
With a spot size of 500 and minimum magnification, turn up the filament
current until the current no longer rises on the gauge.
gain to provide a circular illumination.
using the "gun shift" then use the "gun tilt" controls to give the brightest
illumination.
Changing or inserting
Be sure the filament current and high tension is off and the filament has
cooled for at least 1 minute.
Press the OFF button on the vacuum control panel.
Press the air button.
Wait for the specimen chamber door to release and then carefully pull the
stage assembly out.
Use the 1.5 mm Allen wrench to secure
the specimen stub to the holder.
over-tighten the set screw.
to be firm.
Push the stage assembly into the
chamber being absolutely sure that t h e
specimen will not hit the EDAX detector.
beam:
There should be a "blip" in the filament current
gun (if necessary):
Center this circle on the screen
a new specimen:
Do not
It only needs
Adjust the detector
Do not handle specimen stubs
with your bare fingers.
Contaminates from your fingers
will volatilize in the high
vacuum and contaminate the
final beam aperture and
detectors which in turn will
significantly degrade the image.
Do not over-