Solution:
Stop using the recorder. If the error repeats, call Mindray Customer Service
Department or the distributor.
7.2.17 Press Bar Up
Possible causes: The press bar of the recorder is up.
Solution:
Push it back If the error still remains, call Mindray Customer Service
Department or the distributor.
7.2.18 Lyse Out
Possible causes: insufficient lyse or wrong lyse volume setting.
Solution:
Check if there is sufficient lyse left. If so, access Setup
Reagents and adjust the remaining lyse volume; if not, change a new container
of lyse.
7.2.19 Diluent Expired
The diluent has expired or its expiration date is not correctly set.
Check whether the diluent has expired. If so, change a new container of diluent;
if not, access Setup
7.2.20 Rinse Expired
Possible causes: rinse has expired or its expiration date is not correctly set.
Solution:
Check the rinse. If it is expired, change a new container of rinse; if not, access
Setup
→
7.2.21 Lyse Expired
The lyse has expired or its expiration date is not correctly set.
Check whether the lyse has expired. If so, change a new container of lyse; if not,
access Setup
7.2.22 Filter Error
Something is wrong with the filter of the vacuum chamber.
Solution:
Access Service
if the result is normal. If not, call Mindray Customer Service Department or the
distributor.
7.2.23 Real-Time Clock Error
Something is wrong with the clock.
Access Setup
restart the analyzer to enable the new settings. If the error still remains, call
→
Settings
Settings
→
Reagents and adjust the expiration date.
→
Settings
→
→
Self-test
→
→
Settings
→
BC-2800 Auto Hematology Analyzer (V1.0)
→
Reagents and adjust the expiration date.
Reagents and adjust the expiration date.
Tubing and test the filter. The error is cleared
Date & Time and set the time and date and
Chapter7
Troubleshooting
→
Settings
→
7-7
Need help?
Do you have a question about the BC-2800 and is the answer not in the manual?
Questions and answers
A/d interrupt abnormal
Abnormal background display on the screen
An abnormal background error on the Mindray BC-2800 can be caused by several factors:
1. Dirty or Contaminated System – Cleaning tests should be performed to remove any buildup or debris.
2. Electrical Issues – Poor grounding or electrical noise can interfere with the system.
3. Reagent Problems – Ensure the use of genuine Lyse reagent.
4. Electrode Issues – A greenish color on the electrode inside the counting bath may indicate a soldering problem between the silver plate and the copper wire. Resoldering or replacing the bath may be necessary.
5. Faulty Tubing or Connections – Loose tubing connections can contribute to the error.
6. Startup Procedure Needed – Running the startup procedure by pressing [F3] at the Count screen may resolve the issue.
If the issue persists after these steps, contacting Mindray Customer Service or a distributor is recommended.
This answer is automatically generated