Press Bar Up; Lyse Out; Diluent Expired; Rinse Expired - Mindray BC-2800 Service Manual

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

Solution:
Stop using the recorder. If the error repeats, call Mindray Customer Service
Department or the distributor.

7.2.17 Press Bar Up

Possible causes: The press bar of the recorder is up.
Solution:
Push it back If the error still remains, call Mindray Customer Service
Department or the distributor.

7.2.18 Lyse Out

Possible causes: insufficient lyse or wrong lyse volume setting.
Solution:
Check if there is sufficient lyse left. If so, access Setup
Reagents and adjust the remaining lyse volume; if not, change a new container
of lyse.

7.2.19 Diluent Expired

The diluent has expired or its expiration date is not correctly set.
Check whether the diluent has expired. If so, change a new container of diluent;
if not, access Setup

7.2.20 Rinse Expired

Possible causes: rinse has expired or its expiration date is not correctly set.
Solution:
Check the rinse. If it is expired, change a new container of rinse; if not, access
Setup
7.2.21 Lyse Expired
The lyse has expired or its expiration date is not correctly set.
Check whether the lyse has expired. If so, change a new container of lyse; if not,
access Setup
7.2.22 Filter Error
Something is wrong with the filter of the vacuum chamber.
Solution:
Access Service
if the result is normal. If not, call Mindray Customer Service Department or the
distributor.
7.2.23 Real-Time Clock Error
Something is wrong with the clock.
Access Setup
restart the analyzer to enable the new settings. If the error still remains, call
Settings
Settings
Reagents and adjust the expiration date.
Settings
Self-test
Settings
BC-2800 Auto Hematology Analyzer (V1.0)
Reagents and adjust the expiration date.
Reagents and adjust the expiration date.
Tubing and test the filter. The error is cleared
Date & Time and set the time and date and
Chapter7
Troubleshooting
Settings
7-7

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BC-2800 and is the answer not in the manual?

Questions and answers

Pradeep
January 21, 2025

A/d interrupt abnormal

User image 678f2af3a4580
Bett
February 13, 2025

Abnormal background display on the screen

1 comments:
Mr. Anderson
February 13, 2025

An abnormal background error on the Mindray BC-2800 can be caused by several factors:

1. Dirty or Contaminated System – Cleaning tests should be performed to remove any buildup or debris.
2. Electrical Issues – Poor grounding or electrical noise can interfere with the system.
3. Reagent Problems – Ensure the use of genuine Lyse reagent.
4. Electrode Issues – A greenish color on the electrode inside the counting bath may indicate a soldering problem between the silver plate and the copper wire. Resoldering or replacing the bath may be necessary.
5. Faulty Tubing or Connections – Loose tubing connections can contribute to the error.
6. Startup Procedure Needed – Running the startup procedure by pressing [F3] at the Count screen may resolve the issue.

If the issue persists after these steps, contacting Mindray Customer Service or a distributor is recommended.