Contents PREFACE INTENDED USE GENERAL DESCRIPTION OF MAJOR COMPONENTS WARRANTY INFORMATION SPECIFICATIONS TROUBLESHOOTING SERVICE INSTRUCTIONS WIRING DIAGRAM SPARE PARTS LIST DRUCKER DIAGNOSTICS SM024 HORIZON 6 SERVICE MANUAL REV A...
Failure to follow the procedures in this document may result in personal injury or instrument damage. Drucker Diagnostics will not be held liable for any injury or damage because of improper servicing.
Clicking noise during braking Rotor is loose Tighten rotor nut PROBLEM POSSIBLE CAUSE SOLUTION Debris in air intake / exhaust ports Remove power before clearing debris. Whistling noise while running Damaged gasket Requires service DRUCKER DIAGNOSTICS SM024 HORIZON 6 SERVICE MANUAL REV A...
• The rotor should be checked periodically for signs of wear. • Remove the rotor from service if any of the following are found: cracks, deep scratches, corrosion or discoloring. DRUCKER DIAGNOSTICS SM024 HORIZON 6 SERVICE MANUAL REV A...
Page 6
To install, line up the holes on the lid tray assembly to the screw holes on the cabinet and install the four screws. • Complete the installation by plugging the wire harness into the control board header ‘J2’. DRUCKER DIAGNOSTICS SM024 HORIZON 6 SERVICE MANUAL REV A...
Page 7
Flip the guard bowl. • The motor is held in place with four #8 Nylok nuts. Remove them with an 11/32” nut driver. • Slide the motor out of the guard bowl’s motor well. DRUCKER DIAGNOSTICS SM024 HORIZON 6 SERVICE MANUAL REV A...
Page 8
Carefully place the cabinet onto the base taking care not to pinch any wires between the two. • Complete the assembly by replacing the six screws in the bottom and three in the rear. DRUCKER DIAGNOSTICS SM024 HORIZON 6 SERVICE MANUAL REV A...