Immunity And Compliance Level - Samsung S-Patch3 User Manual

Wearable health monitoring system
Table of Contents

Advertisement

Rated maximum output
150kHz to 80MHz
power of transmitter
[W]
0.01
0.116
0.1
0.368
1
1.166
10
3.687
100
11.660
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

6.5 Immunity and Compliance Level

Immunity test
IEC 60601 Test Level
Conducted RF
3Vrms
IEC 61000-4-6
150kHz to 80MHz
Radiated RF
3Vrms
IEC 61000-4-3
80MHz to 2.5GHz
Samsung Confidential
Separation distance according to frequency of transmitter
80MHz to
800MHz
E
=3V/m
V
=3Vrms
1
1
0.1166
0.3687
1.1660
3.6872
11.6600
Actual Immunity Level
3Vrms
[m]
800MHz to
2.5GHz
E
=3V/m
1
0.2333
0.7378
2.3333
7.3785
23.333
Compliance Level
3Vrms
3V/m
3V/m
[36]

Advertisement

Table of Contents
loading

Table of Contents