Page 1
Firmware version 3.0.34+ Isothermal Technology Limited, Pine Grove, Southport, PR9 9AG, England Tel: +44 (0)1704 543830 Fax: +44 (0)1704 544799 Internet: www.isotech.co.uk E-mail: info@isotech.co.uk The company is always willing to give technical advice and assistance where appropriate. Equally, because of the programme of continual development and improvement we reserve the right to amend or alter characteristics and design without prior notice.
Using 3 or 4 Wire Measurement ............... 22 Using a Thermistor with milliK .................. 24 3.2.1 Configuring milliK to Measure Temperature with a Thermistor ....... 25 Using a Thermocouple with milliK ................26 3.3.1 Using Internal Reference Junction Compensation ..........27 3.3.2...
Page 3
Connecting Thermometers to milliK ................54 Connecting PRTs/SPRTs to milliK ................54 Connecting Thermistors to milliK ................55 Connecting Thermocouples to milliK ............... 56 Connecting 4-20mA Transmitters to milliK .............. 57 Calibration ........................58 Standards Required for Calibration ................59 5.1.1 Making a 4-Terminal Short-Circuit ..............
(USB, RS232, ethernet) allows the user to access the comprehensive features of the milliK. A USB keyboard and mouse can be plugged into the milliK to make control and data entry with the milliK as simple as using a laptop PC.
Page 7
The milliK is likely to be connected to thermometer sensors in use and the user should take care to ensure that the complete system is safe. For example, metal sheathed thermometers may be connected to the milliK and then placed in a furnace powered from a 230V electrical supply.
Page 10
Plug the DC connector from the power supply into the “DC Power” socket on the rear of the milliK. Press the on/off button on the front panel to turn on your milliK. The milliK will load the operating system and software (25s) and then restore itself to the settings used when it was de-powered.
On the front of your milliK you will find two sets of connectors (Lemo 6-pin circular connectors for SPRTs/PRTs/thermistors and miniature thermocouple connectors) for channels 1 and 2. On the rear of your milliK you will find the 4mm sockets for channel 3 (used exclusively for 4-20mA transmitters).
Page 12
You can also connect a keyboard and/or mouse to the USB port (rear panel) and use these to control the milliK. The function keys below the display are mapped to keys F1- F4 on the keyboard so that you can control the milliK completely using a USB keyboard.
Start to proceed (or connect a USB keyboard or mouse and press F4 or click Start). The milliK software will then start and you will see the main Windows: There are four main windows, which can be selected using the function keys below the display.
Page 16
purposes only and cannot be changed. Pressing the key when the ‘Thermometer’ setting is selected opens a window with a list of all thermometers in the milliK’s database: Use the keys to highlight the required thermometer and press the (F4) key to select it.
The ‘Samples/Reading’ setting is initially set to 1 but can be increased to a maximum of 100. This will cause the milliK to take the specified number of samples for each reading and will reduce the noise at the expense of slower measurement speed.
Press to enter the password. A separate password is provided to protect the calibration of the milliK. Both passwords are initially set to “1234” but should be changed prior to use in order to ensure the security of measurements made using your milliK (see section 3.11). A password recovery process is available if you forget your password, please contact Isothermal Technology or its approved distributor for assistance if required (see section 3.11.1).
3 Detailed Description by Function This section describes all the features available on the milliK by function. If you wish to learn how to do something with your milliK use the table of contents on page 2 onwards to look up the appropriate section.
Higher accuracy measurements can be made using calibrated PRTs/SPRTs whose resistance- temperature characteristics have been determined. Before being used, the thermometer and its calibration details must be entered into the milliK’s thermometer database (see section 3.9.1). The thermometer will then appear in the ‘Thermometer’ list. Use the keys to ...
3.1.2 Determining Self-Heating of a PRT/SPRT The normal sense current used by the milliK for PRTs/SPRTs is 1mA. This current causes a small amount of self-heating in the thermometer (typically 1 to 3mK). Provided that the thermometer is calibrated and used at the same current, this leads to negligible uncertainty in the measurement.
Page 23
4 and 6. Provided the wires are all the same gauge and length (same resistance) the voltage drop down the I+ wire will be the same, so the milliK can subtract this correction from the voltage measured between V+ and V- (pins 1 and 6) to determine the voltage across the PRT element.
There are, therefore, no nominal or generic conversion algorithms. The milliK can be used to monitor the resistance of a thermometer declared to be a Default Thermistor, but in order to measure temperature, the thermistor and its calibration details must be entered into the milliK’s thermometer database.
(since it is connected in parallel with the Lemo connector). Your milliK is supplied with two Lemo connectors for you to use with thermistors. The part number for this connector is: Lemo part number: FGG.1B.306.CLAD62Z Additional connectors are readily available internationally from Farnell Electronics and can be ordered via their website www.farnell.com...
The resistance of a thermistor can be measured without entering any calibration information by declaring the ‘Thermometer’ as a “Default Thermistor”. This can be useful when calibrating thermistors against other thermometers: Once you have configured the milliK to work with your thermistor, press either the Graph (F1) or...
3.3.3). 3.3.3 Using an Ice-Point for the Reference Junction The best measurement uncertainty is generally achieved by immersing the reference junction in an ice-point and measuring the EMF it generates (Isotech can provide ice-point reference systems for this application): Reference...
EMFs. 3.3.4 Configuring milliK to Measure a Thermocouple The milliK can be used with both calibrated and uncalibrated thermocouples. Press the Settings (F3) key to access the Settings window, which opens with the ‘Channel’ parameter ...
Page 30
(see section 3.3.3). A typical configuration for a calibrated gold-platinum thermocouple entered into the database as “AuPt thermocouple” and using an ice-point reference would be: Once you have configured the milliK to work with your thermocouple, press either the Graph (F1) or...
The milliK can be used with 4-20mA temperature transmitters by using the 4mm sockets on the rear of the milliK (designated Channel 3). The centre 4mm socket is the current sense input to the milliK, the other 4mm sockets are 0V and a +24V supply. Connect the...
3.4.1 Configuring milliK to Measure a 4-20mA Transmitter The milliK can measure the current output from a 4-20mA transmitter, or it can convert this current to temperature units by defining the conversion in the milliK’s thermometer database (see section 3.9.1).
3.5 Measuring the Difference Between Two Channels The milliK can display the difference between any channel and channel 1 (provided that the units are the same - otherwise the value would be meaningless). This feature is available in...
‘Automatic/Manual’ button and pressing key. 3.7 Using Rolling Statistics The milliK reports the mean and standard deviation of the most recent readings. The number of readings in these rolling statistics can be set independently for each channel in the Settings ...
3.8 Logging Results to a File The milliK can store results to a file on an external USB drive (connected to the USB port on the rear of the instrument) or to internal flash memory. Files stored to internal memory can be transferred to an external USB drive at a later date.
To transfer data logged to the milliK’s internal memory to a PC, it must first be transferred to a USB drive. Plug the USB drive into the USB port on the rear of the milliK. Press the ...
3.8.2 File Format for Logged Data The first 27 rows of the spreadsheet contain information about the configuration of the milliK during logging. Rows 28 onwards contain the elapsed time (in seconds) and logged data for all channels that are enabled:...
Data: Press the key to clear the memory. The milliK will then display the number of files you are about to remove and will ask you to provide the password (see section 3.11) in order to proceed. 3.9 Using the Thermometer Database The milliK includes a database for thermometers you wish to use regularly.
(not all coefficients are required), set the value to zero. Select ‘Eqn Form’ and press key to toggle between the a,b,c and α,β,δ versions of the equation. The milliK will automatically convert the coefficients as you change the equation form selected.
EMF to a temperature. Thermocouples can be calibrated individually to achieve better measurement uncertainty. The calibration is presented as coefficients for a deviation function polynomial or as data pairs. The milliK supports a third- order polynomial correction and data pairs (linear interpolation between the data pairs).
The milliK supports this full third order conversion. Use the keys in the ‘Thermometer’ window to select ‘Conversion’ and then use the keys to select “Polynomial”. Use the ...
The software in your milliK can easily be updated by plugging a USB drive with the updated files (MUST be in root directory) into the USB port on the rear of the milliK and selecting the ...
F2 key (not labelled on screen), the milliK will then generate the recovery key: Email this recovery key to Isotech and we will send you a temporary password to enable you to change the password to a new value. Enter this temporary password as the ‘Old Password’...
Page 51
The milliK will then add buttons at the bottom of the ‘Settings’ window to allow you to select a temperature profile: The ‘Setpoint Temp’ button allows you to select a temperature profile (or disable this feature by selecting “None”).
3.12.1 Creating a Temperature Profile The milliK allows you to define temperature profiles and store these in a database. To create a new profile press the Instrument (F4) key, use the keys to select Setpoint Temperatures and then press the key to open the ‘Temperature Profile Menu’...
4 Connecting Thermometers to milliK Channels 1 and 2 on the front of the milliK are intended for connection to either resistance thermometers (PRTs/SPRTs or thermistors) or thermocouples. Each channel is equipped with a Lemo 6-pin circular connector (for resistance thermometers) and a miniature thermocouple connector.
3) and a link that carries no current should be made from pin 1 to pin 2 (see section 3.1.3): PRT/SPRT 4.2 Connecting Thermistors to milliK High accuracy thermistor based thermometers are typically supplied with screened 4-wire cables. The relatively high resistance of thermistors makes them more susceptible to external noise, so it is important to connect the cable screen (if provided) to the body of the Lemo connector.
The relatively high resistance of thermistors means that they are used with much lower sense currents (2µA on milliK) than PRTs/SPRT and the effect of the cable resistance is much lower. Thermistors are therefore sometimes supplied with only 2-wire connections, in which...
In order to connect a 4-20mA transmitter to your milliK, fit 4mm plugs to your transmitter. The milliK supports both self-powered and loop-powered transmitters by providing a 24V DC supply. Connect the transmitter to the 4mm sockets on the rear panel of the milliK using the required configuration:...
5 Calibration The calibration of your milliK can be checked by applying standards to the inputs and noting the indicated readings. The milliK includes calibration software that allows you to adjust its calibration. The process involves attaching suitable standards to the inputs and entering the values of these standards, the calibration software then monitors the measured values and applies any required corrections.
A 0mV source is essentially a piece of copper wire that can be connected across the thermocouples inputs of the milliK. However, you need to take care in making this short circuit in order to avoid creating thermal EMFs that are significant for an instrument as sensitive as the milliK.
Apply a 4-terminal short–circuit (see section 5.1.1) to channel 1 and set the milliK to read its resistance on the 115 range at 1mA by changing the parameters in the...
Page 62
1, repeat the measurement on channel 2. There is no adjustment possible for zero resistance readings, so if the zero readings fall outside the limits you should return your milliK to Isotech for investigation. 5.2.1.2 SPRT Resistance Range (115) – Span Check Connect a calibrated (nominally) 100...
5.2.2 Thermistor Resistance Range (500k) – Span Check Connect a calibrated (nominally) 500k resistor to channel 1 and set the milliK to read its resistance on the 500k range by changing the parameters in the ‘Settings’ window: The high resistance used makes the input prone to picking up electrical interference, so the connection to the resistor should ideally use screened cable (see section 4.2).
(F2) key, clear the statistics and wait for 100 measurements to be made (when the display shows “Mean of 100”). The mean value indicated by the milliK is the zero reading for voltage measurement on channel 1. Repeat the procedure on Channel 2.
10Ω). Repeat the procedure on Channel 2. 5.2.5 4-20mA Transmitter Range – Zero and Span Check Set the milliK to read the current on channel 3 by changing the parameters in the ‘Settings’ window: Press the...
The frequency of calibration is for the user to determine, but Isotech recommend that you calibrate your milliK on an annual basis (the specification for the milliK is given in terms of an initial calibration and additional uncertainty over a 1-year period). The values used to adjust...
Press the (F4) key to start the calibration adjustment. The milliK will then take a number of readings of the attached resistance standard, calculate and then apply an appropriate adjustment. A progress bar appears showing the status of the calibration adjustment: Once started, the calibration adjustment process cannot be aborted.
5.3.7 Adjusting 4-20mA Transmitter Range Connect a calibrated (nominally) 10mA current source to the 4mm sockets designated channel 3 on the rear of the milliK: 10mA Calibrated Current Source In the ‘Calibration Menu’ window, use the ...
(CSV) form can be exported to a USB drive for transfer to a PC. The file can be imported into Excel™ or other suitable applications for viewing. To export the file, connect a USB drive to the USB port on the rear of the milliK, go to the ‘Calibration Menu’ window and press the...
Miniature thermocouple plugs are of variable quality and lower quality versions can cause wear to the contacts in your milliK. If you feel that the thermocouple sockets on your milliK are becoming stiff, you can apply a small amount of contact lubricant (Electrolube CG53A is suitable: available internationally from Farnell Electronics via their website www.farnell.com...
7 Interfacing to milliK The milliK is equipped with two RS232 ports and an ethernet port that allow you to control its operation remotely. The USB host port can be used to attach devices such as a mouse, keyboard or keypad to provide alternative means of driving the milliK using its user interface.
Page 77
Hyperterminal™ from the internet. Start Hyperterminal™ to set up a new connection. Type in a name for the connection (such as “milliK on COM1”) and select an icon. In the next window, select the PC’s COM port to Connect using which you have connected your milliK from the drop-down list and click OK.
(though operation will be slowed for both). However, if you wish to use the ethernet or COM ports to control the milliK using SCPI commands, we recommend that you send the MILLik:REMote command (see section 8.6.7) to stop the front panel interface.
Page 80
RJ45 cross-over adaptor with an ethernet patch lead. The milliK can be used with a fixed IP address or acquire its IP address from a DHCP server on the network. To select whether a static (fixed) or dynamic (DHPC) address is used and ...
You can send and receive SCPI commands via the ethernet interface in the same way as with the RS232 interfaces on your milliK. The milliK treats the ethernet port, the two COM ports and the user interface) as separate input streams so that they do not interact with each other.
Page 83
Click on the required device (“Triton00” in the above example) and then click OK, cerhost will then establish a connection and display the screen as it appears on your milliK: Click Tools|Config and select the “Whole Screen” radio button in the “Update” area to ensure that the screen remains synchronised with your milliK.
8 SCPI Command Set The command format and protocol used by the milliK is based on the SCPI (Standard Commands for Programmable Instruments) standard. This was developed to provide a consistent command language for all types of programmable instruments. The objective was...
8.5 Units Numeric parameters may optionally include units (if you omit units, the milliK assumes appropriate default units). The only units required for milliK are ohms (which are indicated by ‘R’, ‘r’, or OHM) and volts (indicated by ‘V’ or ‘v’).
To do this, send the SCPI command “MILLik:REMote” on any port. 8.6.1 Measuring Resistance using SCPI Commands In order to measure the resistance of a PRT, you need to: select the measurement channel set the milliK to measure resistance ...
Page 88
‘remote’ mode (see section 8.6.7) and you need to: ensure the milliK is in ‘remote’ mode select the measurement channel set the milliK to measure voltage specify the thermocouple type ...
MEASure:CURRent? 8.6.4 Measuring Temperature using SCPI Commands In order to measure temperature, the milliK must be in ‘remote’ mode (see section 8.6.7) and you need to specify how the base measurements (resistance for PRTs/SPRTs/thermistors or voltage for thermocouples) are converted to temperature units. You need to specify a number of parameters, which depend on the sensor used.
Page 90
IEC60751. An index system is used to specify which internal probe is to be used. You can interrogate the milliK to find out which probe index to use as follows:...
Page 92
MILLik:REM;MEASure:TEMPerature1? IEC60751(4-WIRE),C,400,NORMal 8.6.4.3 Measuring Temperature using SCPI Commands with a Thermocouple in the database In order to measure temperature using a thermocouple defined in the milliK’s database, you need to: ensure the milliK is in ‘remote’ mode select the measurement channel ...
Page 93
INIT FETCh? The milliK will then report the measured temperature, for example “21.23456”. The commands INIT and FETCh? can be replaced by the single command READ?. To repeat the measurement using the same settings, send READ?. To make 10 measurements with the...
Page 94
MILLik:REMote;MEASure:TEMPerature2? TYPe K,C,NONe 8.6.4.5 Measuring Temperature using SCPI Commands with a Thermistor in the database In order to measure temperature using a thermistor defined in the milliK’s database, you need to: ensure the milliK is in ‘remote’ mode select the measurement channel ...
8.6.4.6 Measuring Temperature using SCPI Commands with a 4-20mA Transmitter in the database In order to measure temperature using a 4-20mA transmitter defined in the milliK’s database, you need to: ensure the milliK is in ‘remote’ mode select the measurement channel (this must be channel 3) ...
N.B. The optional keyword [MILLik:] is included in the command for use in systems in which millisKanners are used with a milliK. In such systems, the PC may be connected to the end of the “daisy-chain” of millisKanners (leaving the other RS232 port on the milliK free for connection to an Isotech temperature source).
N.B. The optional keyword [MILLik:] is included in the command for use in systems in which millisKanners are used with a milliK. In such systems, the PC may be connected to the end of the “daisy-chain” of millisKanners (leaving the other RS232 port on the milliK free for connection to an Isotech temperature source).
PRT that has a resistance of 390Ω at its highest operating temperature. The milliK will then set the range to 0-460 Ω since this is the most sensitive range that can accommodate the requested maximum resistance.
Format: SENSe:RESistance:WIRes? Reports whether 3 or 4-wire measurement is selected for the next resistance measurement on a PRT/SPRT. Example: a milliK that has been set to make a 4-wire resistance measurement will respond to the command SENS:RES:WIR? by returning “4”.
Format: SENSe:CURRent? Reports the current that will be used for the next resistance measurement. Example: a milliK that has been set to normal current and one of the PRT ranges (0-115 Ω or 0-460Ω) will respond to the command SENS:CURR? with: “1.00000E-003”...
Reports the thermometer probe to be used in the next a measurement of temperature (see sections 8.6.4.1 to 8.6.4.6). this will be either an index to a thermometer in the milliK’s database or a standard thermometer type (see section 8.6.19).
8.6.26 Command FETCh[:SCALar]? Format: FETCh[:SCALar]? Reports the result of the last measurement made. Example: a milliK that has made a measurement on channel 1 using the commands shown in the example in section 8.6.25 will respond to FETC? with a value in the form: “0.00154514”.
Example: a milliK that has been configured to make a voltage measurement using SENSe commands (such as SENS:CHAN 1 ; SENS:FUNC volt ; SENS:RJC None ; SENS:PROB NONe) will respond to READ? with a value in the form: “0.00154514”. 8.6.28 Command MEASure[:SCALar]:TEMPerature<channel#>? Format: MEASure[:SCALar]:TEMPerature [:DC]<channel#>?
NORMal or ROOT2. Example: for a milliK that has been found to require a resistance gain of 1.001234 for the resistance range 0-460Ω with normal (1mA) sense current by using a resistance standard with serial number xyz1956, the gain can be set using: CAL:RES:GAIN 460,NORM,1.001234,SN_xyz1956...
Format: CALibrate:VOLTage:OFFSet <offset> <channel> Reports the added to voltage measurements made on in volts. Example: a milliK with a +1µV offset added to voltage measurement made on channel 1 will respond to CAL:VOLT:OFFS1? with: “0.00000100” 8.6.71 Command CALibrate:VOLTage:GAIN <gain>[,<standard>][,<additional info>] Format: CALibrate:VOLTage:GAIN <gain>...
Example: for a milliK that has been found to require a voltage gain of 1.001234 by using a standard resistor with serial number xyz1956 and a current source with serial number abc9960, set the gain using: CAL:VOLT:GAIN 1.001234,SN_xyz1956,SN_abc9960 CAL:VOLT:GAIN 1.001234 8.6.72 Command CALibrate:VOLTage:GAIN?
Format: CALibrate:IDENtification <name>,<company>,<location>,<temperature> Sets identification information used to “stamp” any calibrations that are made after this command (until the milliK is restarted or the command is resent). This information would typically be used to store information on the calibration technician, the calibration organisation and environment.
Changes the password used to lock (protect) the calibration of the milliK’s measurement system. This is initially set to “1234”, but should be changed before using the milliK for any critical or traceable measurement or calibration work. The new password must be at least 4 characters in length and must be typed in identically twice in order to effect the change.
Page 124
The curves below show how the temperature accuracy varies with measured temperature (both initial and 1- year) for all the thermocouple types supported by the milliK when it is used with either an ice-point reference (for highest accuracy) or internal reference junction compensation (the least accurate, but most convenient method).
10 Approvals The milliK has been verified as complying with the regulatory requirements of the EU and FCC for electromagnetic compatibility and safety (EU only). 10.1 CE Declaration European Community Electromagnetic Compatibility Directive (89/336) European Community Low Voltage Directive (93/68)
Need help?
Do you have a question about the milliK and is the answer not in the manual?
Questions and answers