RUN CALIBRATION

 Enter the default Supervisor ID of 1234.
 Remove Calibration Cassette from the pouch.
 Press
to advance to the next screen.
Following the prompts, insert the Calibration Cassette into Sofia 2 and gently close the drawer. Sofia 2
will perform the Calibration Check automatically.
The Sofia 2 display will indicate when the Calibration Check is completed.
NOTE: If Calibration Check cannot be completed successfully, notify the on-site Supervisor or contact
Quidel Technical Support for assistance.
Once calibration is complete, remove and return Calibration Cassette to the storage pouch.
Important: Store the Calibration Cassette in the provided storage pouch between uses. Avoid touching
the read window on the Calibration Cassette.
Run QC according to assay specific Package Insert.