Kikusui PCR500MA Interface Manual

Kikusui PCR500MA Interface Manual

Ac power supply pcr-ma series
Hide thumbs Also See for PCR500MA:

Advertisement

KIKUSUI ELECTRONICS CORP.
1-1-3 Higashiyamata, Tsuzuki-ku, Yokohama,
224-0023, Japan
Tel: +81-45-482-6353
Fax: +81-45-482-6261
http://www.kikusui.co.jp/en
Website
PART NO. IA005402
Apr. 2018
Communication Interface Manual
AC Power Supply PCR-MA Series
PCR500MA
PCR1000MA
PCR2000MA
PCR4000MA

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PCR500MA and is the answer not in the manual?

Questions and answers

Summary of Contents for Kikusui PCR500MA

  • Page 1 PART NO. IA005402 Apr. 2018 Communication Interface Manual AC Power Supply PCR-MA Series PCR500MA PCR1000MA PCR2000MA PCR4000MA KIKUSUI ELECTRONICS CORP. 1-1-3 Higashiyamata, Tsuzuki-ku, Yokohama, 224-0023, Japan Tel: +81-45-482-6353 Fax: +81-45-482-6261 http://www.kikusui.co.jp/en Website...
  • Page 2: Command List

    Turns the identify display on or off. Stores the current panel settings to memory FETC:<meas-item>/ MEAS:<meas-item> *SRE Queries the scalar measurement data specified with <meas-item>. Sets the service request enable register. OUTP Turns the output on and off. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 3 Queries the condition of the OPERation status register. FREQ:TRIG STAT:OPER:ENAB Sets the target frequency when INIT/INIT:TRAN and a software trigger are sent. Sets the enable register of the OPERation status register. VOLT Sets the AC voltage. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 4 SYST:COMM:USB:ADDR Queries the address information of the USB interface. SYST:CONF:TRAC Sets whether to display communication errors by performing a debug trace. SYST:ERR Reads the oldest error information or event information from the error queue. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 5 The specifications of this product and the contents of this manual are subject to ■ Trademarks change without prior notice. Copyright 2018 Kikusui Electronics Corp. Microsoft and Windows are registered trademarks of Microsoft Corporation in the United States and/or other countries.
  • Page 6: Installing The Visa Library

    • Keysight VISA by Keysight Technologies (Keysight IO Libraries Suite16.0 or later) • KI-VISA Ver. 5.0.4 or later - Note - If your VISA library is an older version than that specified, you may not be able to use it depending on the interface. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 7 VID (Vender ID) 0x0B3E ■ USB connection PID (Product ID) Use a standard USB cable to connect the PCR to the computer. PCR500MA: 0x1050 PCR1000MA: 0x1051 ■ USB setting PCR2000MA: 0x1052 The factory default USB setting is “USB enabled.” PCR4000MA: 0x1053 For datails of CONFIG settings, see the PCR-MA user's manual.
  • Page 8 - Note - The LAN interface should be shifted remotely by the command ( SYST:COM- M:RLST). Be sure to include this command at the start of the program when you are performing remote programming. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 9 DHCP:ON, Auto-IP:ON, Static:OFF DNS Server Assignment 0.0.0.0 WINS Server Assignment 0.0.0.0 Enable Dynamic DNS Enable Enable mDNS Enable Enable NetBIOS Over TCP/IP Enable To also reset the Hostname and Description, use the Web browser interface. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 10: Welcome Page

    PCR-MA. The instrument information, network information, and VISA resource (I/O resource) information appear on the display. Click items in the navigation menu to move to the other pages. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 11 <Model name> - <Last 5 digits of serial Apply: Applies the edited contents. number> Reset: Resets the network settings. Desired Description KIKUSUI <Model name> AC power supply Default: Returns the network settings to the factory default settings. - <Serial number> Enable Dynamic DNS Enable Back to View Mode: Goes to the network setting item viewing screen (View Mode).
  • Page 12 If you forget the password If you forget the password, reset the LAN interface setting in the CONFIG settings or initialize the PCR-MA to its factory default settings. For datails of CONFIG settings, see the PCR-MA user's manual. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 13: Gpib (Optional)

    Talker Function available Listener Function available Service request Full capability Remote local Full capability Parallel polling No capability Device clear Full capability Device trigger Full capability Controller No capability Electrical interface Open collector driver KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 14: Command Syntax

    The available terminators (reception and transmission) is LF (line feed, ASCII 0x0A). Either one can be used as a terminator. When a command string is terminated, the path is reset to the root level. - Note - CR (ASCII 0x0D) is not a terminator. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 15 In the example below, the current limit is set to the minimum value. SOURce:CURRent MINimum Queries can be used to inquire the minimum and maximum values for most parame- ters. SOURce:CURRent? MAX SOURce:CURRent? MIN KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 16: Ieee488.2 Common Commands

    The IEEE standard uses uppercase characters. SCPI commands are not case sensitive. • Commands are accepted whether or not measurement units are specified. • To enter “µ” in the data, use “U” instead. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 17 (Example) When *ESE 16 is transmitted, bit 4 of the event status enable register is set. Each time the execution error bit (bit 4) of the event status register is set, the summary bit (ESB) of the status byte is set. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 18 Returns the command for reproducing the panel settings as an ASCII character (Example) For a PCR1000MA with serial number AB123456 and firmware version string (500 bytes max.). Commands are separated by a semicolon (;). VER1.00, this returns: KIKUSUI,PCR1000MA,AB123456,1.00 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 19 Returns 0 if there is no option installed. "IB22" IB22 GPIB interface board Response "EX08" EX08-PCR-MA External signal interface board Returns 1 when all the commands in standby have been completed. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 20 Memory C OFF(0) The *ESE and *SRE settings are not cleared when the POW- ER switch is turned on. 3 to 10 Memory No. (Example) (Example) *RCL 1 *PSC 0 Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 21 [SOURce:]VOLTage:LIMit:UPPer 315.0 [SOURce:]VOLTage:MODE FIXed [SOURce:]VOLTage:TRIGgered 0.0[V] [SOURce:]VOLTage:OFFSet 0.0[V] [SOURce:]VOLTage:OFFSet:LIMit:LOWer 0 [SOURce:]VOLTage:OFFSet:LIMit:UPPer 445.0 [SOURce:]VOLTage:OFFSet:MODE FIXed [SOURce:]VOLTage:OFFSet:TRIGgered 0.0[V] [SOURce:]VOLTage:RANGe [SOURce:]VOLTage:RANGe:AUTO INITiate:CONTinuous:ACQuire TRIGger :TRANsient:SOURce TRIGger :SYNChronize:SOURce IMMediate TRIGger :SYNChronize:PHASe:ON 0[deg] TRIGger :ACQuire:SOURce DISPlay:AMMeter SENSe:AVERage SENSe:VOLTage:EQUalizer Command *RST KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 22 Sending *SRE 8 sets bit 3 of the service request enable register. Each time the summary bit (bit 3) of the QUEStionable status register in the status byte is set, a service request message is generated. Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 23 If the PCR-MA is not in a condition to accept triggers, an SCPI error (-211,"Trigger *TST? ignored") occurs. See section 10.37 in IEEE 488.2-1992. Response Returns 0 if there are no errors. Returns the error code, if there are errors. Command *TRG KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 24: Abort Command

    If the ABOR command is sent when the ACQuire subsystem is not initiated and the measured data that is held is valid, the measured data is not discarded. A specific trigger subsystem cannot be specified with the ABOR command. It is al- ways interpreted as ALL. Command ABORt[:ALL] KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 25 ABORt:TRANsient If the ABOR:ACQ command is sent when the ACQuire subsystem is not initiated and the measured data that is held is valid, the measured data is not discarded. Command ABORt:ACQuire KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 26: Display Command

    Value: Current (default) Related Command Moving average of the current INIT:CONT:ACQ PEAK Peak current TRIG:ACQ WATTage Wattage TRIG:ACQ:SOUR Settings are reset to default when the *RST command is sent. (Example) DISP:AMM WATT Response: character KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 27 TRIG:TRAN:SOUR Command INITiate:CONTinuous:ACQuire <boolean> INITiate:CONTinuous:ACQuire? Parameter Value: ON(1) Auto continue mode on OFF(0) Auto continue mode off (default) (Example) INIT:CONT:ACQ ON Response: NR1 Related Command INIT:ACQ TRIG:ACQ TRIG:ACQ:SOUR KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 28: Measure/Fetch Command

    <meas-item> is the last node of the command header. -> Tutorial “Measurement (ACQuire)” Command LXI:IDENtify [:STATe] <boolean> LXI:IDENtify [:STATe]? Parameter Value: ON(1) Identify display on OFF(0) Identify display off (default) (Example) LXI:IDEN ON Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 29: Output Command

    VOLTage:ACDC AC+DC voltage Vrms Response Returns the measured value in NR3 format in response to FETC:<meas-item>?/ MEAS:<meas-item>?. Returns the measured values in the <measitem> list order in comma-separated NR3 format in response to FETC:ALL?/MEAS:ALL?. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 30 The output is set to off. EXT-AC (when the analog interface board is installed only) EXT-DC (when the analog interface board is installed only) (Example) (Example) OUTP:PON:STAT RCL0 OUTP:COUP DC Response: character Response: character KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 31: Sense Command

    Sets the averaging period of the measured values (excluding the peak current). OUTPut:PROTection:CLEar Command SENSe:AVERage <NRf> SENSe:AVERage? Parameter Value: 1, 2, 4, 8, 16 (1 by default) Settings are reset to default when the *RST command is sent. (Example) SENSe:AVER 16 Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 32 Parameter Value: SHORt Update every approximately 0.3 seconds (default) LONG Hold the peak value for approximately 5 seconds Settings are reset to default when the *RST command is sent. (Example) SENS:CURR:HOLD LONG Response: character KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 33: [Source:]Current Command

    Settings are reset to default when the *RST command is sent. PCR4000MA 0.8 to 42.0 (42.0 by default) (Example) Unit: CURR:OFFS 3.5 Settings are reset to default when the *RST command is sent. (Example) Response: NR3 CURR 2.5 Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 34: [Source:]Frequency Command

    Parameter <lim_upper_numerc> Upper limit value of frequency Value: 40.0 to 500.0 (500 by default) Unit: Hz (all parameters) Settings are reset to default when the *RST command is sent. (Example) FREQ 400 FREQ 60,55,65 Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 35 Unit: Unit: Settings are reset to default when the *RST command is sent. Settings are reset to default when the *RST command is sent. (Example) (Example) FREQ:LIM:LOW 55 FREQ:LIM:UPP 65 Response: NR3 Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 36 Settings are reset to default when the *RST command is sent. When you send the FREQ command, this setting is set to the FREQ value. FREQ:MODE STEP (Example) Response: character FREQ:TRIG 400 Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 37: [Source:]Voltage Command

    Parameter <lim_upper_numeric> Upper limit value of AC value Value: 0 to 315.0 (315 by default) Unit: V (all parameters) Settings are reset to default when the *RST command is sent. (Example) VOLT 120 VOLT 100,90,110 Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 38 Enables the trigger function Unit: Settings are reset to default when the *RST command is sent. Settings are reset to default when the *RST command is sent. (Example) (Example) VOLT:MODE STEP VOLT:LIM:UPP 110 Response: character Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 39 -222.5 to 445.0 (0 by default) 310V range -445.0 to 445.0 (0 by default) Unit: V (all parameters) Settings are reset to default when the *RST command is sent. (Example) VOLT:OFFS -10.5 VOLT:OFFS 48.0, 45.0, 51.0 Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 40 -445.0 to 445.0 (445.0 by default) Unit: Settings are reset to default when the *RST command is sent. (Example) Settings are reset to default when the *RST command is sent. (Example) VOLT:OFFS:MODE STEP VOLT:OFFS:LIM:UPP 190 Response: character Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 41 VOLT:RANG 310 Settings are reset to default when the *RST command is sent. Response: NR3 When you send the VOLT:OFFS command, this setting is set to the VOLT:OFFS val- (Example) VOLT:OFFS:TRIG -10.5 Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 42 VOLT:RANG:AUTO ON Settings are reset to default when the *RST command is sent. Response: NR1 When you send the VOLT command, this setting is set to the VOLT value. (Example) VOLT:TRIG 120 Response: NR3 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 43: Status Command

    NOT USED NOT USED If both filters are cleared, event reporting is disabled. NOT USED NOT USED Operation Complete Request Control Query Error RQS/MSS Device Dependent Error Execution Error Command Error Reserved Power On KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 44: Status Byte Register

    User Request(URQ) Not used OPERation status enable register is set. Power On(PON) Set when the power is turned on. 8-15 Not Used 8-15 Reserved Not used KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 45: Operation Status Register

    Indicates whether the PCR-MA is waiting for a setting change trigger (*TRG/TRIG:TRAN). NOT USED CV output NOT USED 1024 NOT USED 2048 NOT USED 4096 NOT USED 8192 NOT USED 16384 NOT USED 32768 NOT USED Always zero KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 46 Sets the enable register of the OPERation status register. A query does not clear the contents of the register. Command Command STATus:OPERation:ENABle <NRf> STATus:OPERation:CONDition? STATus:OPERation:ENABle? Parameter Response: NR1 Value: 0 to 65535 (0 by default) Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 47: Stat:oper:ntr

    Sets the positive transition of the OPERation status register. Command Command STATus:OPERation:NTRansition <NRf> STATus:OPERation:PTRansition <NRf> STATus:OPERation:NTRansition? STATus:OPERation:PTRansition? Parameter Parameter Value: 0 to 65535 (0 by default) Value: 0 to 65535 (0 by default) Response: NR1 Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 48: Questionable Status Register

    Overpower overload has been activated 4096 CL-RMS(Current Limit on Overcurrent (peak) overload has been activated RMS) 8192 SF(Remote sensing failer) Sensing error has occurred. 16384 MEAS-OVLD(Measurement Overload has been detected. overload detected) 32768 Not Used Always zero KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 49 Sets the enable register of the QUEStionable status register. A query does not clear the contents of the register. Command Command STATus:QUEStionable:ENABle <NRf> STATus:QUEStionable:CONDition? STATus:QUEStionable:ENABle? Parameter Response: NR1 Value:0 to 65535 (0 by default) Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 50: Stat:ques:ntr

    Sets the positive transition of the QUEStionable status register. Command Command STATus:QUEStionable:NTRansition <NRf> STATus:QUEStionable:PTRansition <NRf> STATus:QUEStionable:NTRansition? STATus:QUEStionable:PTRansition? Parameter Parameter Value:0 to 65535 (0 by default) Value:0 to 65535 (0 by default) Response: NR1 Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 51: Preset Status

    (including sub registers) to their default values Command Default values: STATus:ENABle = 0x0000 SYSTem:COMMunicate:GPIB:ADDRess <NR1> STATus:PTRansition = 0x7FFF SYSTem:COMMunicate:GPIB:ADDRess? STATus:NTRansition = 0x0000 Parameter Command Value: 1 to 30 STATus:PRESet (Example) SYST:COMM:GPIB:ADDR 23 Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 52 RWLock Sets the PCR-MA operation to remote mode. All panel keys are locked (the LOCAL key is also locked). This is a substitute command for the IEEE488.1 llo (Local Lock Out) command. (Example) SYST:COMM:RLST REM Response: character KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 53 Response mat. Returns +5025. VID (vendor ID) <NR1>, vendor name <character>, PID (product ID)<NR1>, prod- uct name <character>, serial number <character> Response example: For a PCR500MA with a serial number AB123456 2878,KIKUSUI,4176,PCR500MA,AB123456 is returned. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 54 (Example) When there are no errors or events OFF(0) Not display communication errors (default) +0"No error" (Example) When a command that cannot be executed under the current operating (Example) conditions has been received SYST:CONF:TRAC ON -221,"Settings conflict" Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 55 Sets and releases the panel operation lock (keylock). Command Command SYSTem:ERRor:COUNt? SYSTem:KLOCk <boolean> SYSTem:KLOCk? Response Parameter Value: ON(1) Key lock is set. Returns +0 if there are no errors. OFF(0) Key lock is released. (Example) SYSTem:KLOC ON Response: NR1 KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 56: Syst:loc/ Syst:rem/ Syst:rwl

    Returns the optional interface board that is installed in the PCR-MA in comma-sep- arated string format. Returns 0 if there is no option installed. “IB22” IB22 GPIB interface board “EX08” EX08-PCR-MA External signal interface board KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 57: Trigger Command

    TRIGger Command TRIGger Command SYST:VERS Queries the version of the SCPI specifications to which the PCR-MA conforms. TRIG:ACQ Command SYSTem:VERSion? Executes a software trigger for a ACQuire trigger subsystem. Response Command Returns 1999.0. TRIGger:ACQuire[:IMMediate] KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 58 Start the measurement immediately (default) Wait for a software trigger (*TRG, TRIG:ACQ, or IEEE488.1 get (Group Execute Trigger)) to start the measurement Settings are reset to default when the *RST command is sent. (Example) TRIG:ACQ:SOUR BUS Response: character KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 59 (use the *TRG, TRIG:TRAN, or IEEE488.1 get (Group Exe- (Example) cute Trigger) command to change the setting; default). TRIG:SYNC:PHAS 270 Settings are reset to default when the *RST command is sent. Response: NR3 (Example) TRIG:TRAN:SOUR BUS Response: character KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 60: A List Of Errors

    The suffix contained more than 12 characters. -138 Suffix not allowed A suffix was encountered after a numeric element which does not allow suffixes. -140 Character data error This error is generated when parsing a character data ele- ment. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 61: Device-Specific Errors

    Indicates that a legal program data element was parsed but -363 Input buffer overrun Buffer overrun error. This error applies when the RS232C could not be executed due to the current device state. interface is used. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 62: Query Errors

    12.5.2. This event also sets the operation complete bit (bit 0) of the Standard Event Status Register. Error code Error message description -800 Operation complete The instrument has completed all selected pending opera- tions in accordance with the IEEE 488.2, 12.5.2 synchroniza- tion protocol. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 63: Processing Time Of Commands

    When the VOLTage:RANGe is set to 155 V, VOLTage setting cannot exceed 157.5 V. ■ Query To query the maximum and minimum allowed values for the VOLTage and FREQuen- cy commands, use the query with the MINimum and/or MAXimum parameters. VOLTage? MINimum KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 64 It is set to 0 for operational safety. ■ Query To query the maximum and minimum allowed values for the VOLTage:OFFSet com- mand, use the query with the MINimum and/or MAXimum parameters. VOLTage:OFFSet? MINimum VOLTage:OFFSet? MAXimum KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 65 If you repeatedly change the output, a trigger error (-210) may occur. By using the *OPC command, you can prevent this error. -> Tutorial “Waiting for Operation Com- plete” TRIGger;*OPC ‘Apply a trigger and wait until the output change is complete. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 66 Setting items that are programmable in the TRANsient subsystem are VOLTage, VOLTage:OFFSet, and FREQuency. CURRent limit settings are not associated with the trigger subsystem. The TRIGger:TRANsient:IMMediate command applies a software trigger to the KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 67 Available measurement items vary depending on the selected output mode. For ex- ample, DC measurement cannot be performed while AC mode is selected. Peak current measurement and peak hold measurement can be performed in either KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 68 If you want the ACQuire subsystem to start the measurement on a software trigger, change the trigger source to BUS. TRIGger:ACQuire:SOURce BUS ‘Selects trigger source to BUS INITiate:ACQuire ‘Initiates ACQuire subsystem TRIGger:ACQuire ‘Applies a software trigger to ACQuire subsystem KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 69 The output-on phase control is not a function that synchronizes the output frequency tion does not automatically continue to the new change or measurement. to the input commercial AC line. The SYNChronize subsystem does not have INITiate commands. It is always initiated and cannot be aborted. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 70: Waiting For Operation Complete

    NR1 format ) in the output queue. 'Check whether the OV bit is set STATus:QUEStionable? INITiate:IMMediate:ACQuire;*OPC? <Read the response> At power-on or after the IEEE488 sdc/dcl, *RST command is sent, the PCR-MA is in the OCIS and OQIS state KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 71: Error Checking

    Sending the *CLS command clears the communication error display. If in local mode, the debug trace function is temporarily disabled. The communication error display can be enabled or disabled with the SYSTem:CON- Figure:TRACe command. SYSTem:CONFigure:TRACe {ON|OFF} KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 72 DNS hostname that is managed by an external DNS server (a full-qualified domain name— FQDN). If you are using an mDNS hostname, Apple Bonjour (alternatively, iTunes or Safari) must be installed on your PC. For VISA, the alias can be used for the I/O resource. KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 73 End Sub 'Set the voltage and frequency Private Sub Button2_Click(ByVal sender System.Object, ByVal System.EventArgs) Handles Button2.Click msg.WriteString("OUTP 0" & vbLf) msg.WriteString("VOLT 110" & vbLf) msg.WriteString("FREQ 60" & vbLf) msg.WriteString("OUTP 1" & vbLf) End Sub KIKUSUI Electronics Corp. PCR-MA Interface Manual...
  • Page 74 'Querys measurement DC voltage Private Sub Button3_Click(ByVal sender System.Object, ByVal System.EventArgs) Handles Button3.Click msg.WriteString("MEAS:VOLT:DC?" & vbLf) TextBox1.Text = msg.ReadString(256) End Sub Private Sub Form1_Disposed(ByVal sender Object, ByVal System.EventArgs) Handles Me.Disposed msg.Close() End Sub End Class KIKUSUI Electronics Corp. PCR-MA Interface Manual...