Page 1
21.3-inch (54 cm) Color LCD Monitor CL-S200 Installation Manual for system manufacturer Model: CL-S200Bx0xA Important Please read this installation manual carefully before use. Please make sure to read the safety precautions first and keep this manual in a safe and easily accessible place for future reference.
Package contents Please check if the following items are included in this product package. If any item is missing, contact your dealer for replacement. Note Don’t use the power cord packaged with this product for other electronics. □ Monitor □ Power cord (Either one of the followings) 3P: For the U.S.
Connecting the cables Turn off the computer. DVI-D connection : Connect the supplied DVI cable between the display's DIVI-D and the computer's DVI output connector. DisplayPort connection : Connect the a DisplayPort cable between the display’s DisplayPort_IN and the computer’s DisplayPort connector. The recommended cable length is 2 m.
Page 5
How to setup multiple monitors in daisy-chain Please connect multiple monitors together using DisplayPort as below. In relation to monitors and graphics cards supporting daisy-chain, please see our website for more details. To the DisplayPort output of the computer DisplayPort cable DisplayPort cable DisplayPort cable - 5 -...
Switching monitor orientation CL-S200 can be used in both landscape and portrait orientation by changing the resolution setting in Monitor Properties. Follow the instructions below on how to switch monitor orientation. Note: To use the monitor in portrait orientation, a graphics card that supports vertical timing is required.
Adjusting the monitor angle The monitor turns 45 degrees sideways, 5 degrees downward, and 20 degrees upward. Hold the monitor with both hands and adjust the angle slowly. 20 度 45 度 45 度 5 度 Caution Be careful not to pinch your fingers when adjusting the panel angle. - 7 -...
Display information 1. Operation Button Guide “Operation Button Guide” is not displayed on the screen in the default setting. When pressing any operation buttons but the Standby button, “Operation Button Guide ① ” appears. When no operation is done, it will disappear after 10 seconds. When pressing the Menu button, OSD menu and “Operation Button Guide ②...
2. Information 1) Activate “Operation Button Guide ① ” by pressing any operation buttons but the Standby (Power) button _ 2) Press the “MENU” button. 3) Select “INFORMATION” on “MENU 1/2”, and press the “OK” button. And then, the following OSD “INFORMATION 1/2” appears as below. The OSD page switches between "INFORMATION 1/2"...
Current State 1) Activate “Operation Button Guide ① ” by pressing any operation buttons but the Standby (Power) button _ 2) Press the “MENU” button. 3) Select “CURRENT STATE” on “MENU 1/2”, and press the “OK” button. And then, OSD “CURRENT STATE” appears. Current luminance CURRENT STATE...
5. OSD menu hierarchy The OSD structure and a brief explanation for each menu item as follows. MENU Adjustment / setting menu - INFORMATION ..........Shows display information - CURRENT STATE ........Shows display current state - DICOM CONFORMANCE TEST ....DICOM conformance test -...
Page 12
- USB POWER USB Port Power setting - Auto * ..........ON only when USB UP is connected - Off ...........OFF - FACTORY PRESET FACTORY PRESET の実行 ( 長押し ) - Yes ..........Executes FACTORY PRESET - No ...........Cancels FACTORY PRESET • KEY LOCK_1 (The Power button can be locked by long press on it.) •...
Performing the DICOM conformance test Note This function is available only when the monitor has been calibrated to DICOM GSDF. Performing the DICOM conformance test ensures DICOM conformance of the monitor by measuring and checking the monitor properties. The DICOM conformance test is recommended to be performed after 60 minutes of continuous lighting time of the backlight of the LCD panel.
Page 14
Secondly, select “YES” for “START TESTING” on OSD “DICOM CONFORMANCE TEST”. If “DICOM GSDF” is NOT selected, the message “START TESTING” will not ※ appear. DICOM CONFORMANCE TEST 2 M P D I C O M C O N F O R M A N C E T E S T R E S U L T - - -...
Switching configuration data CL-S200 can save up to three settings of luminance-Gamma-color as configuration data within the monitor and they can be switched by selecting a configuration number appeared on the current operating state OSD screen. The following three configuration settings are saved prior to shipping as factory default.
Page 16
CURRENT STATE 2 M P C U R R E N T S T A T E L U M I N A N C E 4 9 9 A M B L I G H T 1 0 0 . 0 lx C O L O R 8 0 0 0 K ( x = 0 .
Dynamic gamma function CL-S200 comes with our original function, dynamic gamma adjustment. By enabling this function, the images on the screen are automatically identified to distinguish between monochrome and color, and the respective images are displayed with the most suitable grayscale.
Page 18
INFORMATION 1/2 2 M P I N F O R M A T I O N 1 / 2 M O D E L C L - S 2 0 0 S E R I A L * * * * * * * * * * * * V S P U R e v .
Switching input source Please operate input source change following the procedure mentioned below. Setting method 1 You can change input source by OSD. 1) Activate “Operation Button Guide ① ” by pressing any operation buttons but the Standby (Power) button _ 2) Press the “MENU”...
Auto text mode function CL-S200 comes with the function called "Auto text mode", which automatically detects a text window (white window) used for showing information such as report and patient list, and reduces its luminance depending on the size of the display area of the window.
Page 21
By doing so, you can easily turning on and off this function by using a button. The following steps tell you how to functionalize the “F3” button. 1) Activate “Operation Button Guide ① ” by pressing any operation buttons but the Standby (Power) button _...
Human presence sensor function Human presence sensor is mounted on this product, and it can detect human motion. When leaving your seat, “Standby” mode is automatically activated. It helps to reduce power consumption. The default is “Disable”, but OSD operation can change it to “Enable.”...
To fix the monitor orientation CL-S200 can be used both Portrait and Landscape orientation rotating the panel. If the user would like to use with proper screen even if changes the panel orientation, need to change EDID data which is operated in the OSD (SWITCHING EDID). It is set to Auto Selection as default.
How to display the Test pattern To display the Test pattern to check the image quality prior to the use of the monitor, press the button located in the front of the monitor as follows. Select test pattern 1) Activate “Operation Button Guide ① ” by pressing any operation buttons but the Standby (Power) button _...
Page 25
Secondly, select “FUNCTION1” on OSD “FUNCTION”, and press the “OK” button. 4) Select “FUNCTION1” on OSD “TEST PATTERN”, and press the “OK” button. Display method 1) Activate “Operation Button Guide ① ” by pressing any operation buttons but the Standby (Power) button. 2) Display a test pattern by pressing the “F1”...
DisplayPort setting You can reduce power consumption by selecting "Enable" for DP OUT on INFORMATION 2/2 when using only DVI (DisplayPort isn’t used at all). With this setup, DisplayPort OUT connector doesn’t work. Setting method 1) Activate “Operation Button Guide ① ” by pressing any operation buttons but the Standby (Power) button _...
When using a USB hub... CL-S200 has a USB 2.0 compliant hub. By connecting the supplied USB cable to a USB computer or another USB hub, USB devices can be connected to the monitor directly. The default is “Auto”, but OSD operation can change it to “Off.”...
Page 28
The supplied USB cable can also be used for the followings: • To calibrate the monitor (An optional calibration kit is required.) • To calibrate multiple monitors in succession To calibrate the monitor ● Connect the supplied USB cable between one of the monitor's UP's and the computer's USB down port.
Restore factory preset Please follow the steps as below when restoring your monitor to “Factory Preset.” Restore method 1) Activate “Operation Button Guide ① ” by pressing any operation buttons but the Standby (Power) button _ 2) Press the “MENU” button. 3) Select “FACTORY PRESET”...
Standby button Lock function CL-S200 has the function to lock the Standby button. You can make the Standby button unavailable in front of the monitor to avoid turning off the monitor accidentally. Standby button 1) How to lock Long press the Standby (Power) button for more than four seconds without any OSD on the screen.
Operation button Lock Function CL-S200 has the function to lock operation buttons . You can make the operation buttons unavailable in front of the monitor to avoid changing your setting accidentally. Operation button (4 pieces) 1) How to lock Long press the “▲" button or the “▼" button for more than four seconds without any OSD on the screen.
Attaching an arm stand The tilt stand is interchangeable with commercially available arms that: • Come with a mount plate with 4 slotted screw holes at 100 mm x 100 mm, and • Bear the total weight of the LCD panel (without the stand) and the cables connected to it.
Antitheft security slot The monitor is equipped with the security slot supported by MicroSaver Security Lock System by ACCO Brands Corporation. The MicroSaver locks are available at PC supply stores. * Kensington is the registered trademark of ACCO Brands Corporation. - 33 -...
How to change the fuse Note for AC INLET with fuse AC Inlet has a built-in fuse. When replacing the fuse, be sure to use only a correctly rated approved type, re-fit the fuse cover. How to replace the fuse Open the fuse compartment with the blade screwdriver, and replace the fuse.
When a failure is suspected... Symptom Cause and Solution • Make sure the monitor is plugged ( P4 ). The monitor won't turn on! (The power indicator is off.) • Check if the monitor is on. • Check if cables are connected correctly ( P4 ).
Page 39
CL-S200 Installation Manual October 2017 Edition PZZ11-2532 Notes for the Instruction Manual • No part of this manual, whether partly or wholly, may be reproduced or copied without authorization. • The content of this manual is subject to change without notice.
Need help?
Do you have a question about the CL-S200 and is the answer not in the manual?
Questions and answers